EKOM DK50 4x2VT/M, DK50 6x2VT/M User Manual

EN
D
K
D
K
User ma
50
50
nual
x2
V
x2
V
T/
M
T/
M
November 2018
Z_2364/2018 Chap. 7 -pgs. 8,9,10
COMPRESSOR
DK50 4x2VT/M
DK50 6x2VT/M
EKOM spol. s r. o.
Priemyselná 5031/18 SK-921 01 Piešťany Slovak Republic tel.: +421 33 7967255 fax: +421 33 7967223
www.ekom.sk email: ekom@ekom.sk
DATE OF LAST REVISION
11/2018
NP-DK50-Nx2VT-M-EN-3_11-
2018
112000381-0001
DK50 4x2VT/M DK50 6x2VT/M
NP-DK50-Nx2VT-M-EN-3_11-2018 - 4 - 11/2018
CONTENTS
IMPORTANT INFORMATION .......................................................................................................... 5
1. CONFORMITY WITH REQUIREMENTS IN EUROPEAN UNION DIRECTIVES ................ 5
2. INTENDED USE ................................................................................................................... 5
3. CONTRAINDICATIONS AND SIDE-EFFECTS .................................................................... 5
4. WARNINGS AND SYMBOLS ............................................................................................... 5
5. WARNINGS .......................................................................................................................... 6
6. STORAGE AND TRANSPORT ............................................................................................ 7
7. TECHNICAL DATA ............................................................................................................... 8
8. PRODUCT DESCRIPTION ................................................................................................ 10
9. PRODUCT FUNCTIONALITY ............................................................................................ 12
10. PNEUMATIC DIAGRAM .................................................................................................... 14
INSTALLATION .............................................................................................................................. 15
11. CONDITIONS FOR USE .................................................................................................... 15
12. PLACEMENT OF THE COMPRESSOR ............................................................................ 16
13. PNEUMATIC CONNECTIONS ........................................................................................... 17
14. ELECTRICAL CONNECTIONS .......................................................................................... 18
OBSLUHA ...................................................................................................................................... 19
15. COMMISSIONING .............................................................................................................. 19
16. SWITCHING THE COMPRESSOR ON ............................................................................. 19
17. COMPRESSOR SHUT-DOWN .......................................................................................... 22
MAINTENANCE ............................................................................................................................. 22
18. DEVICE MAINTENANCE ................................................................................................... 22
TROUBLESHOOTING AND REMEDYING PROBLEMS ............................................................... 29
19. SOLVING COMMON PROBLEMS ..................................................................................... 30
20. REPAIR SERVICE ............................................................................................................. 32
21. DISPOSAL OF THE DEVICE ............................................................................................. 32
22. INSTALLATION RECORD ................................................................................................. 32
DK50 4x2VT/M DK50 6x2VT/M
11/2018 - 5 - NP-DK50-Nx2VT-M-EN-3_11-2018
IMPORTANT INFORMATION
1. CONFORMITY WITH REQUIREMENTS IN EUROPEAN UNION DIRECTIVES
This product conforms to the requirements of the MDD93/42/EEC and 2006/42/EEC
directives and is safe for its intended use if all safety instructions are followed.
2. INTENDED USE
The compressor is used as a central source of clean, oil-free compressed air to power individual devices and equipment with compressed air used in large dental clinics and laboratories, in hospital wards and in other settings where the parameters and properties of the compressed air are suitable for the specific application.
The compressed air supplied by the compressor is unsuitable for use with artificial lung ventilation devices without further filtration.
Any use of the product outside the framework of its intended use is considered improper use. The manufacturer is not liable for any damages or injury resulting from misuse.
3. CONTRAINDICATIONS AND SIDE-EFFECTS
There are no contraindications or side-effects known.
4. WARNINGS AND SYMBOLS
The following symbols are used in the user manual, device and its packaging to denote important details and information:
General warnings
Danger, electric shock hazard
Read the user manual!
CE-marking
Compressor is controlled automatically and may start without warning
Caution! Hot surface
Ground connection
Terminal for ground connection
Alternating current
Handling mark on package – FRAGILE
Handling mark on package – THIS SIDE UP
Handling mark on package – KEEP DRY
Handling mark on package – TEMPERATURE LIMITATIONS
Handling mark on package – LIMITED STACKING
Mark on package – RECYCLABLE MATERIAL
Manufacturer
DK50 4x2VT/M DK50 6x2VT/M
NP-DK50-Nx2VT-M-EN-3_11-2018 - 6 - 11/2018
5. WARNINGS
The product is designed and manufactured to be safe for the user and its surrounding environment when used in the defined manner. Keep the following warnings in mind. This keeps risks to a minimum.
5.1. General warnings
READ THE USER MANUAL CAREFULLY BEFORE USING THE DEVICE AND KEEP IT FOR FUTURE USE!
The user manual aids in the correct
installation, operation and maintenance of the product. It is included with the product and must be kept close to it at all times. Careful review of this manual will provide the information necessary for the intended use and correct operation of the product.
Only the original packaging ensures the
protection of the equipment during transport. Save this packaging should you ever have to return the equipment. The manufacturer is not liable for damages caused by faulty packaging when returning a product for transport during the warranty period.
Use a fork lift truck or similar hoisting
equipment for any movement or handling of the product.
This warranty does not cover damages
originating from the use of accessories or consumables other than those specified or suggested by the manufacturer.
The manufacturer only guarantees the
safety, reliability and function of the equipment if:
installation, new settings, changes,
modifications and repairs are performed by the manufacturer or its representative, or a service provider authorized by the manufacturer.
the product is used pursuant to the user
manual.
The user manual corresponds to the
configuration of the product and its compliance with the applicable safety and technical standards at the time of its printing. The manufacturer reserves all
rights for the protection of its configuration, methods and names.
The translation of the user manual is
performed in accordance with the best available knowledge. The Slovak version is to be used in the event of any uncertainties.
5.2. General safety warnings
The manufacturer designed and manufactured the product to mitigate all risks when used correctly for the intended use. The manufacturer considers it its duty to describe the following general safety measures:
The use and operation of the product must
comply with all local codes and regulations. The operator and user are responsible for following all appropriate regulations in the interests of performing work safely.
Only the use of original parts guarantees
the safety of the operating personnel and the reliable operation of the product itself. Only use accessories and parts mentioned in the technical documentation or expressly approved by the manufacturer.
The manufacturer assumes no liability for
any damages or other risks if any accessories or parts other than those mentioned in the technical documentation or expressly approved by the manufacturer are used. This warranty does not cover damages originating from the use of accessories or consumables other than those specified or suggested by the manufacturer.
The user must make sure that the
equipment is functioning correctly and safely every time it is used.
The user/operator must be capable of
safely using and properly operating the product. The user must be trained to operate the product and must be experienced.
Create operating regulations for the
persons operating the product
DK50 4x2VT/M DK50 6x2VT/M
11/2018 - 7 - NP-DK50-Nx2VT-M-EN-3_11-2018
Wear hearing protection when starting the
product, during operation and any time it is in operation.
Operating the product in operating
premises that may contain mixtures of flammable gases or in areas that may contain mixtures of particulate, such as coal dust, is prohibited.
Attention, explosion hazard!
The use of the product in wet or damp
environments is prohibited.
The user must inform the supplier
immediately if any adverse event occurs during the use of the equipment.
5.3. Safety warnings on protection from electric current
The equipment may only be connected to a
properly installed socket connected to the earth (grounded).
Before the product is plugged in, make
sure that the mains voltage and frequency
stated on the product are the same as the power mains.
Check for possible damage to the product
and the connected air distribution system before use. Replace damaged pneumatic and electrical conductors immediately.
Immediately disconnect the product from
the mains in hazardous situations or when a technical malfunction occurs.
During repairs and maintenance, ensure
that:
the product is disconnected from the
mains
the pressure is released from all
lines
Only the manufacturer, or a qualified
specialist trained by the manufacturer, may install, modify or upgrade the product itself.
Only a qualified electrician may install
electrical components!
6. STORAGE AND TRANSPORT
The compressor is shipped from the manufacturer in transport packaging. This secures the equipment from damage during transport.
The original compressor packaging must be used for transport whenever possible. The compressor is shipped in a vertical position and must be secured using transport straps.
Protect the compressor from humid and dirty environments and extreme temperatures during transport and storage. Compressors in the original packaging may be stored in a warm, dry and dust-free area. Do not store near any chemical substances.
Keep the packaging material, if possible. If not, please dispose of the packaging material in an
environmentally-friendly way. Packaging cardboard can be recycled with old paper.
The compressor must be transported only when all air has been vented. Before moving or transporting the compressor, release all the air pressure from the tank and hoses and drain the condensed water.
Ambient conditions for storage and transport
Products may only be stored and transported in vehicles that are free of any traces of volatile chemicals under the following conditions:
Temperature : +0°C to +50°C, Max. relative humidity.: 90%,
DK50 4x2VT/M DK50 6x2VT/M
NP-DK50-Nx2VT-M-EN-3_11-2018 - 8 - 11/2018
7. TECHNICAL DATA
The compressors are designed for dry and well-ventilated indoor environments with the following conditions:
Temperature: +5°C to +40°C, Max. relative humidity.: 70%, Max. absolute humidity.: 15 g/m
3
.
Tab.1
Type
DK50 4x2VT/M DK50 6x2VT/M
Compressor output at 6 bar with dryer l/min 440 660
Voltage / frequency
V/Hz
3 x 400 / 50 3 x 400 / 50
Power (max.) A 14.5 22
Main circuit protection device rating A
50 50
Main electrical feeder mm2
10 10
Enclosure IP10 IP10
Air tank capacity l 220 330
Working pressure bar
6 8 6 8
Safety valve
bar 12 12
Noise level dB 73 74
Weight kg 330* 420*
Weight with dryer kg 360 450 Dimensions
(W x L x H)
mm 2315* x 1740 x 500 2315* x 1740 x 500
Dimensions with dryer (W x L x H)
mm 2475** x 1740 x 500 2475** x 1740 x 500
Operating mode
S1 – 100% S1 – 100%
Dryer performance with refrigerated dryer (M5) (PDP***)
+3°C +3°C
Time to fill air tank from 0 to 7 bar
s 50 50
Required cooling air changes in space
m
3
/h 1000 1500
Electrical class Class I.
Class I.
(*) - without the dryer (**) - dimensions without accessories (***) - apply the correction factor for the M5 dryer
DK50 4x2VT/M DK50 6x2VT/M
11/2018 - 9 - NP-DK50-Nx2VT-M-EN-3_11-2018
Tab.2
Type
DK50 4x2VT/M DK50 6x2VT/M
Compressor output at 8 bar with dryer l/min 320 480
Voltage / frequency
V/Hz
3 x 400 / 50 3 x 400 / 50
Power (max.) A 15.5 25
Main circuit protection device rating A
50 50
Main electrical feeder mm2
10 10
Enclosure IP10 IP10
Air tank capacity l 220 330
Working pressure bar
810 810
Safety valve
bar 12 12
Noise level dB 73 74
Weight kg 330* 420*
Weight with dryer kg 360 450 Dimensions
(W x L x H)
mm 2315* x 1740 x 500 2315* x 1740 x 500
Dimensions with dryer (W x L x H)
mm 2475** x 1740 x 500 2475** x 1740 x 500
Operating mode
S1 – 100% S1 – 100%
Dryer performance with refrigerated dryer (M5) (PDP***)
+3°C +3°C
Time to fill air tank from 0 to 7 bar
s 50 50
Required cooling air changes in space
m
3
/h 1000 1500
Electrical class Class I.
Class I.
(*) - without the dryer (**) - dimensions without accessories (***) - apply the correction factor for the M5 dryer
7.1. Free air delivery (FAD) correction due to elevation
FAD correction table
Elevation [MASL] 0 - 1500 1501 - 2500 2501 - 3500 3501 - 4500 FAD [l/min] FAD x 1 FAD x 0,8 FAD x 0,71 FAD x 0,60
FAD reference conditions: Elevation: 0 MASL
Temperature: 20°C Atmospheric pressure: 101,325 Pa Relative humidity: 0%
DK50 4x2VT/M DK50 6x2VT/M
NP-DK50-Nx2VT-M-EN-3_11-2018 - 10 - 11/2018
8. PRODUCT DESCRIPTION
8.1. Variants
(Fig.1a, Fig.1b)
The DK50 4x2VT/M compressor (Fig. 1a) is composed of:
- Compressor section - 4 x 2V air pumps
- Air tank section - 2 x 110 l
- Adsorption dryer
The DK50 6x2VT/M compressor (Fig. 1b) is composed of:
- Compressor section - 6 x 2V air pumps
- Air tank section - 3 x 110 l
- Adsorption dryer
DK50 4x2VT/M DK50 6x2VT/M
11/2018 - 11 - NP-DK50-Nx2VT-M-EN-3_11-2018
Fig. 1a DK50 4x2VT/M compressor Fig. 1b DK50 6x2VT/M compressor
8.2. Accessories
Every compressor may be added-on/extended directly by the manufacturer or by placing an order for air regulation and filtration accessories (specified in an order) at a later period of time.
Accessories not included in the standard order must be ordered separately!
Accessories included with the product (filter/regulator, filter and micro filter) are mounted to the compressor frame using the bracket on the air outlet side, downstream of the drain/blow down valve on the air tanks.
8.2.1. Set of compressed air outlet filters
The compressor may be equipped with a set of filters, if specified. The set of filters may be equipped with a pressure regulator. Sets of filters are suitable accessories for all of the compressors specified above.
If a different level of air filtration is required, this specification must be agreed upon with the supplier and specified in the order.
1. Compressor air pumps
2. Air tank
3. Dryer
Loading...
+ 23 hidden pages