EKF Diagnostics Stanbio HemoPoint H2 User Manual

HemoPoint® H2
Hemoglobin Analyzer
User Manual
Diagnostics for life
Hemoglobin Analyzer
HemoPoint
User Manual
®
H2
The copyright of this document is protected! Copying and
distribution to third parties only by written permission of
EKF Diagnostics Inc.
EKF Diagnostics Inc. 1261 North Main Street Boerne, Texas 78006 U.S.A. Tel: (830) 249-0772 Fax: (830) 249-0851 e-mail: stanbiolab@ekfdiagnostics.com
HemoPoint® H2 Analyzer manufactured in Germany for: EKF Diagnostics Inc., Boerne, Texas U.S.A.
Revised: 02. June 2016 Software-Version: From 2.07.0
Subject to modifications.
Table of contents
0. Table of contents
0. TABLE OF CONTENTS .................................................. 1
1. IMPORTANT INFORMATION ......................................... 5
1.1 WELCOME! .................................................................... 5
1.2 EXPLANATION OF THE GRAPHIC SYMBOLS ........................ 5
1.3 SAFETY NOTES .............................................................. 5
2. INTENDED USE .............................................................. 8
3. SET UP ............................................................................ 9
3.1 UNPACKING THE ANALYZER ............................................ 9
3.2 SETTING UP THE ANALYZER .......................................... 10
3.3 SWITCHING THE ANALYZER ON AND OFF ......................... 11
3.3.1Using for the first time ........................................... 11
3.3.2To operate using the power adaptor ..................... 12
3.3.3Battery operation ................................................... 14
3.3.4Stand-by mode ...................................................... 14
3.3.5Displays and symbols in the ready mode ............. 15
4. SAMPLING AND TESTING PROCEDURE .................. 16
4.1 TAKING A SAMPLE ........................................................ 16
4.1.1Notes on using the microcuvette ........................... 16
4.1.2Taking a sample of capillary blood ........................ 17
4.1.3Taking a sample of venous or arterial blood ......... 21
4.2 PERFORMING THE TEST ................................................ 23
4.3 QUALITY CONTROL ....................................................... 27
4.3.1AutoCheck ............................................................. 27
4.3.2Blank reading ........................................................ 27
4.3.3Control cuvette ...................................................... 28
4.3.4External quality control .......................................... 30
5. FURTHER FUNCTIONS ................................................ 31
5.1 TOUCHSCREEN ............................................................ 31
5.1.1General .................................................................. 31
5.1.2Meaning of the buttons / Navigation ..................... 32
5.2 MEMORY (DATA STORAGE) ........................................... 32
5.2.1Displaying results from memory ............................ 32
5.2.2Printing out results ................................................ 34
HemoPoint
®
H2 User Manual
1
Table of contents
5.3 MENU FUNCTIONS ........................................................ 35
5.4 STANDARD PRINT MODE ................................................ 42
5.5 EXPANDED PRINT MODE ................................................ 43
5.6 CONNECTING ACCESSORIES ......................................... 45
5.2.3Deleting the stored test data ................................. 34
5.3.1Information about the analyzer .............................. 36
5.3.2Connect ................................................................. 37
5.3.3Setting the date ..................................................... 37
5.3.4Setting the time ..................................................... 38
5.3.5Setting the units ..................................................... 38
5.3.6Hgb mode and limits.............................................. 39
5.3.7Setting the display contrast ................................... 39
5.3.8Settings ................................................................. 40
5.3.8.1 Setting the language ............................................ 40
5.3.8.2 Setting date format ............................................... 40
5.3.8.3 Setting time format ............................................... 41
5.3.8.4 Setting the power frequency ................................. 41
5.3.9Menu options ......................................................... 41
5.3.9.1 Hematocrit mode .................................................. 41
5.3.9.2 Setting the print mode .......................................... 42
5.4.1Header (Example) ................................................. 42
5.4.2Hgb limit mode (On), Hematocrit (On) .................. 43
5.5.1Header (Example) ................................................. 43
5.5.2Hgb limit mode (On), Hematocrit (On) .................. 43
5.5.2.1 Setting the tone signal .......................................... 43
5.5.2.2 Setting the backlight mode ................................... 44
5.5.2.3 Setting the stand-by time ...................................... 44
5.5.2.4 Broadcast ............................................................. 45
5.5.2.5 The service menu ................................................. 45
5.5.3Information on contacting us ................................. 45
5.6.1Connecting to a PC ............................................... 46
5.6.2Connecting a printer .............................................. 46
5.6.2.1 Connecting the printer cable ................................ 47
6. MAINTENANCE ............................................................ 48
6.1 CLEANING AND DISINFECTION OF THE INSTRUMENT ........ 48
6.1.1Housing and touchscreen ..................................... 48
6.1.2Cuvette holder ....................................................... 48
6.1.3Optical unit ............................................................ 49
6.1.4Power adaptor ....................................................... 50
6.2 CHARGING AND CARE OF THE BATTERY .......................... 50
6.2.1Charging procedure .............................................. 51
6.2.2Charging time ........................................................ 51
2
HemoPoint
®
H2 User Manual
Table of contents
6.2.3Self discharge ....................................................... 52
6.3 REPAIRS ...................................................................... 52
6.4 PROPER DISPOSAL ....................................................... 52
7. TROUBLESHOOTING .................................................. 53
7.1 PROBLEM SOLVING ....................................................... 53
7.2 RESETTING OF THE ANALYZER ...................................... 58
8. TECHNICAL DATA ....................................................... 59
8.1 HEMOPOINT
8.2 NXT
®
H2 ANALYZER ............................................ 59
®
MICROCUVETTE ................................................ 60
9. REFERENCE RANGE ................................................... 61
9.1 NORMAL RANGE ........................................................... 61
9.2 UNDERSTANDING YOUR RESULT .................................... 61
10. APPENDIX ...................................................................... 62
10.1 REPLACEMENT PARTS AND CONSUMER MATERIALS ........ 62
10.2 FURTHER INFORMATION ................................................ 63
11. INDEX ........................................................................ 64
HemoPoint
®
H2 User Manual
3
Important information
1. Important information
1.1 Welcome!
Congratulations on purchasing the HemoPoint® H2 hemoglobin testing system. Before you begin using your new HemoPoint system we strongly recommend you read this manual completely, with particular attention to the following section with its introductory explanations and safety notes.
1.2 Explanation of the graphic symbols
DANGER!
WARNING!
TIP
The DANGER symbol!
This symbol warns of situations or actions that could lead to serious damage to the health of the user or a patient.
The WARNING symbol!
This symbol warns about incorrect handling that could cause measuring errors or damage to the instrument or any accessories used.
The TIP symbol! Alongside this symbol we provide useful additional information about the current matter.
®
H2
1.3 Safety notes
It is essential that you read the following notes, in order to avoid risks to persons and damage to the analyzer and other equipment. EKF Diagnostics does not accept responsibility for damage arising from non-observance of the following notes.
HemoPoint
®
H2 User Manual
5
Important information
WARNING!
DANGER!
WARNING!
DANGER!
WARNING!
WARNING!
Follow the user manual!
Each time the analyzer is used, precise knowledge and attention to these operating instructions is required. Only use the HemoPoint
®
H2 system for the purpose which will be described in Section 2.
Danger of fatal electric shock!
Under no circumstances should you open the AC power
adaptor. There are no elements inside which require servicing or maintenance.
Never use a mechanically damaged AC power adaptor –
live connections might be exposed.
The AC power adaptor is not waterproof. Therefore, never
let liquids come into contact with it. A lightly dampened cloth, however, can be used to clean it when disconnected. Please see the notes on care in Section 6.
Only use the AC power adaptor in a socket that has been
properly installed.
Check that the AC voltage and frequency printed on the AC power adaptor label match your electrical socket and whether the shape and configuration of the plug contacts are compatible for this connection.
Not to be used in areas where there is a risk of explosion!
The instrument is not approved for use in areas where there is a risk of explosion.
Keep the analyzer away from liquids!
The analyzer is not waterproof. Fluids entering the instrument could destroy the electrical and optical components in the analyzer. A lightly dampened cloth, however, can be used for cleaning. Please see the notes on care in Section 6.
Allow the instrument to reach room temperature!
Particularly on changing from a cold into a warm environment (i.e. after storage or transport) condensation can form inside and on the outside of the instrument. Wait an appropriate time (approx. 1 hour), before you connect to the main power supply or switch on the instrument.
6
HemoPoint
®
H2 User Manual
Important information
Only use original equipment!
Only attach equipment that is expressly approved for use with the
HemoPoint
®
H2. EKF Diagnostics does not guarantee the function
of the instrument when other equipment is used.
Never open the analyzer!
There are no parts in the analyzer that require customer maintenance. Repairs can only be carried out by EKF Diagnostics personnel. Further notes on maintenance can be found in Section
6.
Do not force the cuvette holder!
The cuvette holder is very important for the quality of testing. Therefore mechanical force on the holder should be avoided and only the designated microcuvettes should be used. For cleaning purposes the cuvette holder can be removed from the analyzer after a mechanical locking device is released. Notes on this can be found in Section 6.
HemoPoint
®
H2 User Manual
7
Intended use
2. Intended use
The HemoPoint determination of hemoglobin (Hgb) in whole blood of adults, infants, and children in a professional point-of-care setting. It consists of a dedicated photometer and individual, single-use microcuvettes filled with reagents. Using the microcuvette, a small amount of arterial, venous or capillary blood is taken up by capillary action. The filled cuvette is inserted into the HemoPoint the chemical reaction in the cuvette is measured, and the Hgb value is displayed. The microcuvettes are intended to be used only once and must be disposed after use as potentially infectious waste, in accordance with the current regulations applicable to your establishment. The HemoPoint practices and in clinical laboratories to assist in medical diagnostic investigations. In addition it can be used in emergency and intensive care units and in medical facilities such as blood banks. This test system has been approved for use in laboratories with Waived status as defined in 452 CFR
493.15(c)(a).
EKF DIAGNOSTICS INC. DISCLAIMS ALL EXPRESS AND IMPLIED WARRANTIES OF THE MERCHANTABILITY AND FITNESS PERTAINING TO THIS PRODUCT WHICH ARE NOT EXPRESSLY DETAILED IN THIS PACKAGING INFORMATION OR A WRITTEN AGREEMENT BETWEEN THE BUYER AND SELLER OF THIS PRODUCT.
EKF DIAGNOSTICS INC. MAINTAINS THAT THIS PRODUCT CONFORMS TO THE INFORMATION CONTAINED IN THIS INSERT. PURCHASER MUST DETERMINE THE SUITABILITY OF THE PRODUCT FOR THEIR PARTICULAR USE. USE ONLY IN ACCORDANCE WITH LABELING INSTRUCTIONS.
®
H2 system is intended for the quantitative
®
H2 photometer, the color produced by
®
H2 system is designed for use in medical
8
HemoPoint
®
H2 User Manual
3. Set up
3.1 Unpacking the analyzer
Before you set up the HemoPoint® H2 analyzer, first check that all the items are present and undamaged (see Illustration 1).
Set up
Illustration 1: (1) HemoPoint® H2 analyzer, (1) AC power adaptor and (1) control cuvette.
Materials provided
• One (1) HemoPoint
• One (1) AC power adaptor
• One (1) HemoPoint
®
H2 Analyzer
®
H2 Control Cuvette
• One (1) User Manual
• One (1) Quick Reference Guide
• One (1) Warranty Registration Card
HemoPoint
®
H2 User Manual
9
Set up
WARNING!
Additional materials required but not provided
• Latex Gloves
• Sampling devices for capillary blood collection
• Biohazardous waste container
• Alcohol swabs and gauze for cleaning puncture site
HemoPoint
HemoPoint
HemoPoint
HemoPoint
®
H2 n•x•t® Microcuvettes
®
H2 Hemoglobin Controls (optional)
®
H2 Optics Cleaner
®
H2 Training Program (optional) – access online at
www.hemopoint.com
Why Register your HemoPoint
®
H2?
Returning the completed Warranty Registration Card will activate your warranty and allow EKF Diagnostics to provide you with the latest product updates and improvements.
Even easier – Go to www.stanbio.com/warranty.php and register your HemoPoint
®
H2 online!
What is your voltage?
Check that the AC voltage and frequency printed on the AC adaptor label match your electrical socket and whether the shape of the plug is compatible for this connection. If not, do not connect the analyzer to the electrical socket under any circumstances. Contact EKF Diagnostics in this case.
3.2 Setting up the analyzer
Select a suitable place for setting up the analyzer, according to the following criteria:
Avoid direct sunlight
Avoid strong electromagnetic fields
Avoid direct influence from ionizing radiation
Avoid rapid temperature fluctuations (keep away from
heaters, open windows, ventilators, fans or air­conditioning, etc.)
Avoid wet areas (i.e. wash basins)
10
HemoPoint
®
H2 User Manual
Place the analyzer on a level counter adjacent to a power socket. Make sure there is enough room for the cuvette holder to be freely accessible.
Illustration 2: Setting up the analyzer.
Set up
3.3 Switching the analyzer on and off
3.3.1 Using for the first time
Allow the analyzer to reach room temperature!
Changing from a cold to a warm environment (i.e. after storage or transport), condensation can form both on the inside and the outside of the analyzer. Wait at least 1 hour, before you connect the analyzer to a power supply.
As the built-in battery has not yet been charged, the analyzer must first be connected to the power adaptor (supplied) and
HemoPoint
®
H2 User Manual
WARNING!
11
Set up
TIP
plugged into an electrical socket until the battery is completely charged.
This procedure is necessary in order for the capacity of the battery to be checked and the charge indicator to function correctly. Further information on this can be found in Section 6.2 “Charging and care of the battery”.
3.3.2 To operate using the power adaptor
Use the power adaptor to operate the analyzer, noting the following steps when connecting the power adaptor:
1. Insert the small round plug on the connecting cable of the power adaptor into the appropriate socket (6VDC) at the back of the analyzer.
Illustration 3: Connecting the power adaptor.
2. Next, insert the power adaptor plug into an appropriate electrical socket. The analyzer itself does not have a power On/Off switch.
12
HemoPoint
®
H2 User Manual
Please be patient after plugging the analyzer into the electrical socket. It may be 4 or 5 seconds before anything appears on the display. This does not indicate a fault but is related to various initialization steps within the analyzer. Following this initialization phase, the display illuminates and a welcome screen appears briefly, after which the instrument is ready to operate.
Set up
TIP
Illustration 4: Welcome display screen.
Illustration 5: Display in ready mode (Open Holder).
The individual symbols and their meanings are explained in Section 3.3.5 “Displays and symbols in the ready mode“.
Now you can either start your testing, view results in the memory, or carry out adjustments in the Menu.
HemoPoint
®
H2 User Manual
TIP
13
Set up
TIP
TIP
WARNING!
TIP
You can find tips on this in Sections 4 “Sampling and Testing Procedure”, 5.2 “Memory (data storage) and 5.3 “Menu functions”.
If the analyzer is not in use for some time, it switches into the energy-saving, Stand-by mode (see Section 3.3.4 “Stand-by mode”).
3.3.3 Battery operation
®
The HemoPoint
H2 analyzer automatically switches to battery
operation when you unplug the power adaptor from the electrical socket, or from the instrument (assuming, of course, that the battery is sufficiently charged). When the battery is fully charged it will operate for a period of at least 100 hours.
Tips on charging the battery and battery maintenance can be found in Section 6.2 “Charging and care of the battery”.
To avoid problems while testing during battery operation, also note the tips in Section 3.2 “Setting up the analyzer”.
3.3.4 Stand-by mode
The analyzer does not have a separate On/Off-switch. It is designed so that it automatically switches to the Stand-by mode when you are not operating the instrument for a certain time (approximately 5 minutes). The display will go blank in the Stand­by mode.
This elapsed time can be changed in Section 5.5.2.3 “Setting the Stand-by time”.
The Stand-by mode reduces power usage to a minimum.
There are (3) three ways of switching the instrument ON again:
14
HemoPoint
®
H2 User Manual
Touch the Touchscreen
Open or close the cuvette holder
Plug the analyzer into an electrical power supply
3.3.5 Displays and symbols in the ready mode
Illustration 6: Display in ready mode (Open Holder).
Set up
HemoPoint
®
H2 User Manual
15
Sampling and testing procedure
4. Sampling and testing procedure
4.1 Taking a sample
4.1.1 Notes on using the microcuvette
optical eye
Illustration 7: n•x•t
The microcuvette is the most sensitive component in the
HemoPoint
®
microcuvette
®
H2 system. It is therefore necessary to handle the
cuvette carefully, paying particular attention to the following conditions:
1. Each microcuvette can only be used once!
The microcuvette is designed for single use. The cuvette is coated during manufacturing with all the reagents necessary for determining the hemoglobin concentration in the blood sample. As soon as the blood is taken into the cuvette, a reaction occurs (within 1 minute).
2. The microcuvette is sensitive to moisture!
The microcuvette will absorb moisture therefore the cuvettes are supplied in a special airtight container that contains a drying agent. This drying agent assures that any moisture is readily absorbed during storage.
16
HemoPoint
®
H2 User Manual
Sampling and testing procedure
Pay attention to the following notes, to preserve the integrity of the cuvettes:
Store the cuvettes only in the original container and at
room temperature (59-86°F / 15-30°C).
Only remove one cuvette at a time from the container and
then immediately close the lid. Make sure that the lid is completely closed by pressing it down as far as it will go.
The cuvette shelf life is reduced once the lid is opened. See the
HemoPoint
®
H2 cuvette container label for additional information.
Please make a note of the date of opening on the container label.
3. The microcuvette is a precision optical component!
The microcuvette is analyzed optically in the HemoPoint
®
H2
analyzer. The measured absorption is proportional to the hemoglobin concentration. The light from the light source should penetrate through the sample to the photo detector with the least possible influence from the cuvette. Therefore it is important not to touch the optical eye of the cuvette with fingers or sharp objects (see Illustration 7).
4. The microcuvette is only designed for in-vitro testing!
5. The reagents that coat the inner walls of the cuvette are harmful and must not be swallowed.
WARNING!
4.1.2 Taking a sample of capillary blood
1. Take out a microcuvette from the original container and immediately close the lid.
Please note Section 4.1.1 “Notes on using the microcuvette”.
HemoPoint
®
H2 User Manual
WARNING!
17
Sampling and testing procedure
Illustration 8: Preparing to take a blood sample.
2. Make sure that your patient is sitting comfortably. There should be good blood circulation in the hand from which you wish to take blood, i.e. it should be warm and relaxed.
If in doubt, the hands can be warmed in warm water.
TIP
DANGER!
18
Illustration 9: Stimulating the circulation.
3. Lightly massage the fingers, in order to stimulate the
circulation.
Warning, risk of infection!
Please wear suitable protective gloves!
HemoPoint
®
H2 User Manual
Loading...
+ 49 hidden pages