E&J 6110.4ZL User Manual

Page 1
OPERATING MANUAL
FOR

AGRI-THERM II HANDHELD DIFFERENTIAL THERMOMETER

MODEL 6110.4ZL
TEMPERATURE RANGE:
-40°C to 100°C or -40°F to 212°F
Physical Address: 1891 North Oracle Tucson, AZ 85705-6444
Mailing Address: P. O. Box 5276 Tucson, AZ 85703-0276
Telephone: (520) 792-4545 Fax (520) 792-4546 Toll Free: (800) 422-4342
E-mail: info@EVERESTInterscience.com
Website:
www.EVERESTinterscience.com
Page 2
TABLE OF CONTENTS
CHAPTER INFORMATION PAGE
1 INTRODUCTION 3
1.1 Unpacking Your Infrared Thermometer 4
2 OPERATING INSTRUCTIONS
2.1 Battery Charging Procedure 5
2.2 Abbreviations on Rear Panel 5
2.3 Turning the AGRI-THERM II On 6
2.4 Emissivity Settings 6
2.5 Surface Temperature Measurement 6
2.6 Temperature Differential Measurement 6
2.7 Ambient Temperature Measurement 6
2.8 Signal Output 7
3 THROUGH-THE-LENS INTRA-OPTICAL LIGHT
SIGHTING
3.1 TTL Intra-Optical Light Sighting 8
3.2 Focusing with the VARIO-ZOOOM 8
Chapter 4 AGRI-THERM II SPECIFICATIONS
4.1 Model 6110.4ZL Specifications 10
Chapter 5 MAINTENANCE
5.1 General 11
5.2 Cleaning 11
Chapter 6 LIMITED WARRANTY SERVICE INSTRUCTIONS 12
2
Page 3

CHAPTER 1

INTRODUCTION

verest Interscience, Inc., offers the most sophisticated non-contact infrared thermometers on the market today. The new AGRI-THERM II Handheld
E
thirty-eight years of research and development by our founder and President. These are revolutionary handheld infrared thermometers, as they can be focused in the field or lab using our Vario-Zooom (Patent Pending) variable focusing system. And they incorporate our TTL (Through-The-Lens) Intra-Optical Light Sighting System (Patent No. 4,494,881).
An infrared thermometer measures the temperature of a target surface by focusing a spot on the target through an optical field-of-view (telescope) onto a temperature-sensitive detector. Infrared radiation is transmitted through the optics and impinges on the infrared detector, heating it slightly.
A major shortcoming of infrared thermometers is that the infrared “light” (radiation) is invisible to the human eye. Therefore, the position and physical characteristics of the spot being measured, such as its location, shape, size, surface texture, etc., cannot be ascertained by the operator. It is very important for the operator to know where all parts of the target spot are located, because, if any part of the spot is off the edge of the target surface, serious reading errors will occur. The Everest TTL (Through-The-Lens) Intra­Optical Light Sighting System (Patent No. 4,494,881) solves this problem as explained in Chapter 3 of this manual.
The Model 6110.4ZL Portable Infrared Thermometers takes readings in a fraction of a second over a temperature range of -40ºC to 100ºC or -40 to 200ºF with 0.1° resolution.
The AGRI-THERM II Differential (Patent No. 4,301,682) Handheld Infrared Thermometer is light-weight, weighing less than two pounds. It is powered by rechargeable nickel-cadmium batteries and has a standard output of 10.0mV/ºC or °F.
We are pleased that you chose an Everest infrared surface thermometer. We are positive that you will be pleased with the versatility, dependability and reliability that have been engineered into these highly sophisticated infrared thermometers.
If you have questions about a particular application, or if you have difficulty with any aspect of the operation or functioning of your new infrared thermometer, please don’t hesitate to contact us. Our telephone numbers and addresses are listed on the front cover of this manual.
Differential Infrared Thermometer (Model 6110.4ZL) is the culmination of over
3
Page 4

1.1 UNPACKING YOUR AGRI-THERM II INFRARED THERMOMETER

hen you unpack your AGRI-THERM II Infrared Thermometer, please be sure the following items are enclosed:
W
If any items are missing or appear to have been damaged during shipment, please save the shipping box and notify Everest Interscience immediately at (520) 792-4545, Toll Free at (800) 422-4342 or by E-mail to info@EVERESTInterscience.com.
1.
Your new Model 6110.4ZL AGRI-THERM II Infrared Thermometer
2. A Carrying Case
3. A Battery Charger
4. A Signal Output Cord
5. A Certification of Calibration
6. An Operating Manual
4
Page 5

CHAPTER 2

OPERATING INSTRUCTIONS

2.1 BATTERY CHARGING PROCEDURE

our AGRI-THERM II has been charged at the factory before shipment. However, you may want to charge it again overnight before extended use. To
Y
plug on the end of the six-foot lead of the supplied battery charger into the Rear Panel Jack of the infrared thermometer labeled “BATT.” The “BATT” jack can be found on the left hand side of the Rear Panel. Then, plug the AC adaptor into 90-250 Volt AC 50/60 Hz power source.
Allow ten (10) hours for a complete charge; partial charges are okay. When charging is complete, unplug the AC adaptor from the outlet and then remove the plug from the instrument.
2.2 ABBREVIATIONS ON REAR PANEL
T
recharge the batteries in your AGRI-THERM II Infrared Thermometer, plug the
he abbreviations on the Rear Panel of the AGRI-THERM II stand for the following:
9 SEL = Select: IR, DIFF or AMB
9 IR = Infrared Surface Temperature Reading
This is the current infrared surface temperature reading of the target that the instrument is focused on.
9 DIFF = Temperature Differential
This is the Difference between the Dry Bulb Ambient Air Temperature and the Temperature of the target.
9 AMB = Dry Bulb Ambient Air Temperature
This is the temperature of the ambient air surrounding the Infrared Thermometer.
9 PWR = Power: On/Off
9 LITE = TTL Intra-Optical Light Sighting: On/Off
9 DATA = Plug for outputting the data selected by “SEL” at the rate of
10mV/º.
9 BATT = Place to plug in battery charger.
5
Page 6
2.3 TURNING THE INSTRUMENT ON
imply move the toggle switch on the Rear Panel entitled “PWR” to On. The “PWR” switch is located on the left side of the Rear Panel.
S
2.4 EMISSIVITY SETTINGS
E
(Target Emission Efficiency in %)
missivity Compensation is done by pushing the “plus” (+) or “minus” (-) buttons on the Rear Panel of the infrared thermometer. Press the “+” or “-" to set in the Emissivity you desire.
2.5 SURFACE TEMPERATURE MEASUREMENT
he infrared surface temperature measurement can be made by selecting “IR” by pushing the “SEL” button the number of times necessary until the light comes on
T
the Infrared Surface Temperature.
2.6 TEMPERATURE DIFFERENTIAL
T
the antenna out of the front of the instrument. This removes the sensor on the end of the antenna from the heating and cooling effects of the instrument and the operator’s hand. Now, when the instrument is on, the Temperature Differential will be read out on the Liquid Crystal Display. Let the instrument equilibrate for at least ten (10) minutes before making Differential readings for better accuracy.
2.7 AMBIENT TEMPERATURE MEASUREMENT
T
Dry Bulb Ambient Air Temperature.
above “IR.” Then, when the instrument is on, the numbers on the display indicate
o take the Temperature Differential (Patent No. 4,301,682) between the target temperature and the dry bulb ambient air temperature, first press the “SEL” button the number of times necessary until the light is shown above “DIFF.” Then, pull
he Dry Bulb Ambient Air Temperature can be measured by pushing the “SEL” button the number of times necessary until the light is on above the “AMB” notation. Now, when the infrared thermometer is on, the display is indicating the
6
Page 7
The following photograph shows where the “SEL,” “IR,” “DIFF,” and “AMB” buttons are located:
2.8 “DATA” OUTPUT
He “DATA” output is on the upper right hand side of the face place of the AGRI-
THERM II. The instrument is provided with a signal cord to connect to a data
T
itself. The “DATA” connection provides for the capability of running the AGRI-
THERM II continuously and sending the data to an outside source where the analog
signal can be converted to a digital signal and the data can be logged and saved.
logger through this receptacle. No data is stored in the infrared thermometer
7
Page 8
CHAPTER 3
THROUGH-THE-LENS (TTL) INTRA-OPTICAL LIGHT
SIGHTING (Patent No. 4,494,881)
3.1 TTL SIGHTING
ith the Everest Through-The-Lens (TTL) Intra-Optical Light Sighting System (Patent No. 4,494,881), an illuminated image of the infrared detector is
W
projected directly through the infrared optics of the infrared temperature sensor, illuminating the exact area where a temperature measurement is being taken. Because this light shares the internal optical path, it can never be knocked out of alignment. It will always illuminate the area where the infrared/visible optics are focused.
3.2 FOCUSING WITH THE VARIO-ZOOOM
n order to focus on a smaller or larger area, the front of the Vario-Zooom (Patent Pending) infrared thermometer need only be pushed in or out until the target is
I
completely illuminated. Simply point the instrument at a target in subdued ambient
light and push the front optics in or pull it out in order to focus the infrared thermometer to display the field-of-view (FOV) of the instrument.
A picture of the Front Optics of the AGR-THERM II Infrared Thermometer with the antenna extended for Dry Bulb Ambient Air Temperatures is shown below:
8
Page 9
To focus an instrument that will be used in the field or out-of-doors, first set the instrument up in subdued lighting in a laboratory. There you will be able to see the light being emitted by the Light Emitting Diode (LED) and determine the distance to spot size ratio. Then, when you are in the field, you can determine the area being measured by knowing that the ratio of the distance to spot size will remain constant as it was in the lab.
9
Page 10

CHAPTER 4

SPECIFICATIONS
4.1 AGRI-THERM II (Model 6110.4 ZL) SPECIFICATIONS
TEMPERATURE MEASUREMENT
Scale Range: -40˚C to 100˚C or -40˚F to 212˚F Resolution: 0.1˚C or 0.1˚F Accuracy: ±0.5˚C or ±1.0˚F Repeatability: ±0.1˚C or ±0.2˚F Temperature: All Functions in ˚C or ˚F, corresponding to voltage out Noise Effective Temperature: <0.1˚C or ºF Measurement Modes: Surface Temperature, Dry Bulb Temperature and
Temperature Differential (Patent No. 4,301,682)
OPTICAL CONSIDERATIONS
Optical Configuration: Robust, aerospace-quality, double-coated Zinc Selenide
Optics per Military Specification MIL-C-13508 Spectral Pass Band: 8 < Wavelengths < 14 Microns Sighting: Intra-Optical Light Sighting (Patent No. 4,494,881):
Visible Light Illuminates Field of View (FOV) Illumination Source: Light Emitting Diode -- SAFE – NO LASERS! Focus: Variable from 2º to 20º with Vario-Zooom (Patent
Pending) Display: Numeric on Liquid Crystal Display on Rear Panel
OPERATING CONDITIONS
Operating Environment: -20˚C to 65˚C or +14˚F to 150˚F, up to 99% Relative
Humidity, Non-Condensing Storage Temperature: Same as Operating Environment Temperature
GENERAL
Response Time: <1 Second Emissivity Compensation: Settable from 10% to 99% on Rear Panel Operating Distance: 2 cm to Infinity
ELECTRICAL INTERFACE
Power Source: Rechargeable Nickel Cadmium Rechargeable Batteries
200 Hours Continuous Use on Full Battery Charge Output Signal: -400 mV to 1000 mV at 10 mV/ºC Battery Charger: 90-240 Volt, 48-62 Hz
WARRANTY
Warranty: One Year-Limited Warranty of Parts & Labor
Everest Interscience, Inc. has a policy of continued design upgrade. Therefore, we reserve the right to change specifications without notice
10
Page 11

CHAPTER 5

MAINTENANCE

5.1 GENERAL

his AGRI-THERM II will be virtually maintenance free. It is recommended, however, that the instrument be returned to the factory after 24 months for a
T
It is important that you return the original battery charger with the instrument and to use this battery charger only when recharging your infrared thermometer.
When returning the AGRI-THERM II for repair or recalibration, please call Toll Free at
1-800-422-4342 or 1-520-792-4545 to receive a Return Material Authorization (RMA)
No. You can also request the RMA by E-Mail to info@EVERESTInterscience.com or an RMA form can be obtained on our website at www.EVERESTInterscience.com . After receiving the RMA No., please ship the instrument insured, freight prepaid, to:
Please include an explanation of the problem or problems you are having or the service that you want performed.
calibration check and recalibration, if necessary.
EVEREST INTERSCIENCE
1891 North Oracle Road Tucson, AZ 85705-6444
USA

5.2 CLEANING

eriodically the infrared thermometer may need to be cleaned. Use a damp cloth to remove any debris on the outside of the instrument.
P
To clean the optical lens, use denatured alcohol and a Q-Tip to wipe off any residue on the front optics. Do not use anything other than denatured alcohol to clean the optics.
11
Page 12

CHAPTER 6

ONE-YEAR LIMITED WARRANTY SERVICE
INSTRUCTIONS
verest Interscience warrants each instrument to be free of defects in material or workmanship for a period of one year from the date of shipment to the original
E
When examination of the instrument reveals that the fault has not occurred because of misuse or abnormal conditions of operation, Everest Interscience will service, replace or adjust any defective part or parts when the instrument is returned freight prepaid. OEM
clients must show that recurring errors are not a part of the customer’s assembly integration process. Upon the advice of Everest Interscience, certain malfunctions will be considered to be due to these process techniques and this warranty will be considered null and void for any instruments subjected to those methods.
Instruments repaired when abnormal usage has occurred or beyond the effective date of the warranty will be charged at applicable rates. Everest Interscience will submit a quotation for the estimated charges before commencing repair. The customer must provide a Purchase Order or credit card information in the amount of the repair and return shipping before work will commence. After Everest has sent the repair quotation, the customer has up to one year to provide a Purchase Order or credit card information to Everest for either the repair and return or the return only of the instrument. If the customer does not provide this information within one year, Everest will recycle the instrument.
Instruments whose warranty expiration dates exceed five years from the current date may be subject to denial for service and repair. In such instances, please contact the Service & Repair Division at (520) 792-4545.
This warranty includes recalibration and battery replacement during the warranty period, if required.
are present, etc.
If shipping from within the United States, please call 1-800-422-4342 for a Return Materials Authorization (RMA) number. Outside of the US, please call 1-520-792-4545. Then for the United States and all other countries, please ship the instrument insured, freight prepaid to:
purchaser.
NOTE: THE WARRANTY IS NULL AND VOID IF:
The instrument has been dismantled, or The instrument case has been deformed such as dented, gouged, tool marks
12
Page 13
EVEREST INTERSCIENCE, INC.
1891 NORTH ORACLE ROAD
TUCSON, AZ 85705-6444 USA
Please include a letter explaining the difficulty that you are having with the instrument, or complete the RMA Form provided to you, when returning the instrument for repair. An RMA Form can be downloaded from our website: www.EVERESTInterscience.com. It is necessary to send a check in the amount of the Evaluation Analysis Fee along with the instrument. This charge can be obtained on our RMA form on the website or by calling (800) 422-4342 or (520) 792-4545.
We apologize in advance if you do have a problem with one of our instruments. We will call you when we receive the instrument and before commencing any work on the instrument if repairs are necessary. Also, we will do all we can to repair it and return it to you as quickly as possible.
13