AUTO - SCANNING WITH DIGITAL CONTROL
LCD COLOR MEDICAL MONITOR
RadiForce EX210 / EX240W
Operation Manual
For future reference, record the serial number of your display
monitor in the space below :
Serial Number
The serial number is located on the back of the monitor
back of the monitor
The title «WARNING» is used to inform the users of possible causes that could inflict the injury, death, or property
damage to the patients.
The title «CAUTION» is used to inform the users of possible causes that could inflict the patients although it might
not severe enough to cause deaths.
EX210/EX240W 1
ENGLISH
WARNING
CAUTION
NOTE
The title «NOTE» is used to inform the users of items that
are of importance in terms of installation, operation, or
maintenance of the Equipment although the failure does
not inflict the bodily harm to the patients.
EIZO Nanao Corporation.2010
All other trademarks are the property of their reference owners.
This document is subject to change without notice.
EIZO provides this information as reference only. Reference to other vendor’s
products does not imply any recommendation or endorsement.
Revision Control
DateDescription
REV.F 091310Document number
2 User’s Guide
TABLE OF CONTENTS
Product Description and Intended Use ......................................................
Warnings and Cautions .............................................................................
Symbol Explanations .................................................................................
EU Declaration of Conformity for Medical Applications .............................
Please check the following items are present when you unpack the box, and save
the packing materials in case you will need to ship or transport the monitor in future.
•EX210/EX240W LCD Monitor and two video cables (1) HD15 VGA
cable* (1) DVI-D cable*
ENGLISH
*
•AC Power cord**
•AC-Adapter
•Composite Video BNC Jack Cable and Super Video Cable
CAUTION
Model No : JMW1150KA2400F04
Part No : PS-52142415001A
*
•User Manual and 4pcs VESA screws
*Might be optional item, check with local representative
**Might vary pending on region stardard
4 User’s Guide
Warnings and Cautions
Please read this manual and follow its instructions carefully. The words warning,
caution, and note carry special meanings and should be carefully reviewed:
Warning The personal safety of the patient or physician may be
involved. Disregarding this information could result in
injury to the patient or physician.
Caution Special service procedures or precautions must be
followed to avoid damaging the instrument.
Note Special information to make maintenance easier or important
information more clear.
An exclamation mark within a triangle is intended to alert the
user to the presence of important operating and maintenance
instructions in the literature accompanying the product.
A lightning bolt within a triangle is intended to warn of the
presence of hazardous voltage. Refer all service to authorized
personnel.
Warning TO AVOID POTENTIAL SERIOUS INJURY TO THE USER
AND THE PATIENT AND/OR DAMAGE TO THIS DEVICE,
THE USER MUST :
Warranty is void if any of these warnings are disregarded.
EIZO Nanao Corporation accepts full responsibility for the effects on safety,
reliability, and performance of the equipment only if:
• Re-adjustments, modifications, and/or repairs are carried out exclusively
by EIZO Nanao Corporation.
• The electrical installation of the relevant operating room complies with the
applicable IEC and CE requirements.
Warning Federal law (United States of America) restricts this
device to use by, or on order of a physician.
The EIZO Nanao Corporation EX210/EX240W monitor has been tested under UL
60601-1 standard and UL listed for Medical application.
EIZO Nanao Corporation reserves the right to make improvements in the product(s)
described herein. Product(s), therefore, may not agree in detail to the published
design or specifications. All specifications are subject to change without notice.
Please contact EIZO Nanao Corporation directly or phone your local EIZO Nanao
Corporation sales representative or agent for information on changes and new
products.
EX210/EX240W 5
Warnings
1. Read the operating manual thoroughly and be familiar with its contents prior to using
this equipment.
2. Carefully unpack the unit and check if any damage occurred during shipment.
3. Should any solid object or liquid fall into the panel, unplug the unit and have it checked
by qualified personnel before operating it any further.
4. Uplug the unit if it is not to be used for an extended period of time. To disconnect the
cord, pull it out by the plug. Never pull the cord itself.
5. Be a qualified physician, having complete knowledge of the use of this equipment.
6. Test this equipment prior to a medical procedure. This monitor was fully tested at the
factory before shipment.
7. Avoid removing covers on control unit to avoid electric shock.
8. Attempt no internal repairs or adjustments not specifically detailed in this operating
manual.
9. Pay close attention to the care, cleaning instructions in this manual. A deviation may
cause damage (refer to the Cleaning section).
10. DO NOT STERILIZE MONITOR.
11. Read the entire instruction manual before assembling or connecting the camera.
12. Do not place the monitor or any other heavy object on the power cord. Damage to the
cable can cause fire or electirc shock.
13. Monitor with power supply is suitable for use in patient environment.
14. DO NOT stack more than 8 boxes high
This equipment has been tested and found to comply with the limits for medical devices in
IEC 601-1-2:2003. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to other devices, which can be
determined by turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment into an outlet on a circuit different from that to which the other
device(s) are connected.
- Consult the manufacturer or field service technician for help.
NOTICES TO USER
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions:
(1) this device may not cause harmful interference, and (2) this device must accept any inter-
ference received, including interference that may cause undesired operation.
FCC WARNING
This equipement generates or uses radio frequency energy. Changes or modifications not
expressly approved by the party responsible for compliance could void the user’s authority
to operate the equipment. The user could lose the authority to operate this equipment if an
unauthorized change or modification is made.
ENGLISH
6 User’s Guide
Cautions
1. The AC Adapter must be plugged into a Grounded power outlet.
2. Use only the proprietary EX210/EX240W power supply for the
EX210/EX240W monitor. Make a proper connection by ensuring that the
shrink tubing completely secures the connection between the DC power cord and the
extension cord.
3. Turn power off when unit is not in use.
4. Never operate the unit right after having transported from a cold location directly to a
warm location.
5. Do not expose the monitor to moisture or directly apply liquid cleaners directly to the
screen. Spray the cleaning solution into a soft cloth and clean gently.
6. Handle the monitor with care. Do not strike or scratch the screen.
7. Do not block the monitor cooling vents. The monitor is cooled by natural convection
and has no fan.
8. Do not force the monitor past 28 degrees of vertical when adjusting the screen
position. (For monitors equipped with stands only.)
9. Remove the power module and connection when transporting the unit.
10. Save the original carton and associated packing material. They will be useful should
you have to transport or ship the unit.
11. Allow adequate air circulation to prevent internal heat buildup.
12. Do not place the unit on surfaces (rugs, blankets, etc.) or near materials (curtains,
draperies) that may block the ventilation slots.
13. Do not install the unit near sunlight, excessive dust, mechanical vibration or shock.
14. The unit is designed for operation in a horizontal position. Never operate the unit in
a vertical position.
15. Keep the unit away from equipment with strong magnets (i.e. a large loudspeaker.)
16. Do not expose the monitor to moisture or excessive dust.
17. Equipment with SIP/SOP connectors should either comply with IEC 60601-1 and/or
IEC 60601-1-1 harmonized national standard or the combination should be evaluated.
Do not touch the patient with signal input or output connectors.
18. Use only a hospital grade power supply cord.
19. This equipment has been tested and found to comply with the limit for a Class B digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable generate uses and can radiate radio frequency energy, if not installed and
used in accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not occur in
a particular installation, which can be detemined by turing the equipment off and on,
the user is encourage to try to correct the interference by one or more of the following
measures:
• Reorient or relocate the receiving device.
• Increase the separation distance between the equipment.
• Connect the equipment to an outlet on a circuit different from that to which the
other device(s) are connected.
• Consult the manufacturer or field service technican for help.
20. Grounding reliability can only be achieved when the equipment is connected to an
equipment receptacle labeled “Hospital Only” or “Hospital Grade.”
NoteTo connect to an international power supply, use a an attachment plug
appropriate for the power outlet.
NoteRefer to the “Electromagnetic Compatibility” (EMC) section of this manual to
ensure EMC. The EX210/EX240W must be installed and operated
according to the EMC information provided in this manual.
Symbol Definitions
Dangerous: High Voltage
Direct Current
Serial Number
Top – Bottom
Fragile
Do not get wet
EX210/EX240W 7
ENGLISH
51LJ
Medical Equipment
E215822
Consult accompanying documents.
Indicates proof of conformity to applicable European Economic
Community Council directives and to harmonized standards
published in the official journal of the European Communities.
Authorized Representative in Europe
Indicates protective earth ground.
For indoor use only.
Medical Equipment is in accordance with UL 60601-1 and CAN/
CSA C22.2 No. 601.1 in regards to electric shock, fire hazards,
and mechanical hazards.
Tested to comply with FCC Class B standards.
DC power control switch
8 User’s Guide
EU Declaration of Conformity for Medical Applications
A Declaration of Conformity has been filed for this product. A sample
of this document may be found in the addendum which accompanied
this manual. For a copy of the Declaration of Conformity document,
please contact EIZO Nanao Corporation. and request for EX210/
EX240W DOC.
Prepare for Unpack
Before you unpack your monitor, prepare a suitable workspace. You
need a stable and level surface near a grounded wall outlet in an area
which is relatively free of glare from sunlight or other sources of bright
light. The monitor is cooled by natural convection (it has no fan).
For optimum performance, do not block the cooling vents.
While unpacking the monitor, inspect it and other package contents
for shipping damage that could cause a fire or shock hazard. Immediately report any shipping damage to the carrier or transportation
company and contact customer service for monitor in the future or in
case of return.
After you unpack the monitor, make sure the following items are included
• Monitor with video cable
• AC adapter with cable
CAUTION: AC Adapter must be plugged into Grounded a power outlet
CAUTION : AC adapter
Model No: JMW1150KA2400F04
Part No: PS-52142415001A
• This operations manual
Note: Your system provider may offer alternative cords or cables
depending on the installation requirement and local geography issues.
EX210/EX240W 9
SAFETY PRECAUTION
· Avoid placing the monitor, or any other heavy object, on the power cord
to prevent fire or electrical shock from damage to the power cord.
· Do not expose the monitor to rain, excessive moisture, or dust to avoid
fire or shock hazard.
· Do not cover the slots or openings of the monitor for proper heat
dissipation. Always put the monitor in a place where there is adequate
ventilation.
· Avoid placing the monitor against a bright background or where sunlight
or other light sources may reflect on the area of the monitor. Place the monitor
just below eye level.
· Handle with care when transporting the monitor.
· Refrain from giving the shock or scratch to the screen, as screen is
fragile.
CLEANING YOUR MONITOR
No specific liquid or chemical necessary when cleaning this LCD monitor
However, we suggest to clean the monitor with non-abrasive cloths and cleaning
solutions used in hospitals to clean similar equipment. We recommend using 70%
Isopropyl alcohol for the screen surface and warm water and a mild detergent for all
other surfaces. Other acceptable cleaning agents are listed below:
ENGLISH
• 70% isopropyl alcohol
• Cidex (2.4% glutaraldehyde solution)
• 0.5% Chlorhexidine in 70% isopropyl alcohol
To clean the screen, do not spray liquid cleaners directly on to the unit. Stand away
Form the monitor and spray cleaning solution onto a cloth. Without applying excessive pressure, clean the screen with the slightly dampened rag.
10 User’s Guide
POWER MANAGEMENT FUNCTION
The monitor is equipped with the power management function which automatically
reduce the power consumption when not in use in three power level modes.
•Stand-by Mode
The monitor goes into stand-by mode when the horizontal sync signal is off for
about 10 seconds. In this mode, the screen goes off and the power LED blinks
for 1 seconds On and 1 second Off. The screen is displayed after the horizontal
sync signal is restored.
•Suspend Mode
The monitor goes into suspend mode when the vertical sync signal is off for
about 10 seconds. The power consumption during this is 27 W. In this mode,
the screen goes off and the power LED blinks for 1 seconds On and 1 second
Off. The screen is displayed after the vertical sync signal is restored.
•Off Mode
The monitor goes into power-off mode when the vertical and horizontal sync
signals are off for about 10 seconds. In this mode, the screen goes off and the
power LED blinks for 1 seconds On and 1 second Off. The screen is displayed
after the vertical and horizontal sync signals are restored.
Power Management System
The EX210/EX240W Medical Monitor power management proposal. Provides four
phases of power-saving modes by detecting the horizontal sync signal as shown in
the table below.
State
Horizontal SyncActiveInactiveActiveInactive
Vertical SyncActiveActiveInactiveInactive
VideoActiveBlankedBlankedBlanked
Power IndicatorGreen
Power
Consumption
Normal
Operation
100W27W27W27W
DPMS
Standby
Green Flashing (1
sec. Interval)
DPMS
Suspend
Green Flashing (1
sec. Interval)
DPMS Off
Green Flashing (1
sec. Interval)
When the monitor is power saving mode or detects an incorrect timing, the screen
will be blank and power LED indicator will blink.
The following combination options are available to you:
Main PIP
VGAOOOOOOO
DVIOOOOOOOO
SDI1OOOOOOO
SDI2OOOOOOO
C-VideoOOOOOOO
S-VideoOOOOOOOO
Component /
RGBs
SOGOOOOOO
VGADVISDI1SDI2C-VideoS-Video
OOOOOOOO
Component
/ RGBs
ENGLISH
SOG
14 User’s Guide
DDC
To make your installation easier, the monitor is able to Plug and Play with your
system if your system also supports DDC protocol. The DDC (Display Data
Channel) is a communication protocol through which the monitor automatically
informs the host system about its capabilities, for example, supported resolutions
and corresponding timing. The monitor supports DDC1 and DDC2B standard.
INSTALLATION
To install the monitor to your host system, please follow the steps as given below:
Steps
1. Use the supplied video cable (DVI, VGA, S-Video, C-Video) then connect to the
host system accordingly.
2. Connect the DC power to the DC power connector on the monitor.
3. Connect one end of AC power cord into the AC Adapter and the other end to AC
power outlet.
4. Then turn the host system on and then the monitor.
5. If the monitor still does not function properly, please refer to the troubleshooting
section to diagnose the problem.
DC Power
C - Video /
SOG
DVIVGASDI INS - Video
IN
OUT
RGB / ComponentSDI OUT
EX210/EX240W 15
CONNECTING the POWER CORD
* Check first to make sure that the power cord you use is the correct type required
for your area.
* This monitor has an universal AC adapter that allows operation in either
AC 100 - 240 V ac voltage area. No user-adjustment is required.
* Plug one end of the power cord to the AC adapter, plug another end to a proper
AC outlet.
The cord set should have the appropriate safety approvals for the country in which
the equipment will be installed and marked HAR.
For 120 volt Applications, use only UL Listed deachable power cord with NEMA
configuration 5-15P type (parallel blades) plug cap. For 240 volt applications use
only UL Listed Detachable power supply cord with NEMA configuration 6-15P type
(tandem blades) plug cap.
ENGLISH
Hospital-grade
power cord
Power Supply
Model No: JMW1150KA2400F04
Part No: PS-52142415001A
Extension power cord (optional)
Manufacturer: Bridgepower
Part No: CB-47D2004P57MF (15ft) or
CB-47D2022P86MF (75ft)
EX210/
EX240W
16 User’s Guide
User Functions / Messages
Key Name and Function
Power LED : Lights up to indicate the power is turned ON.
Power Switch : To power ON or OFF the monitor.
Rotary switch : With the OSD menu activated, increases the value of the selected
(Ture Right) parameter or moves rightward in OSD menu.
Rotary switch : With the OSD menu activated, increases the value of the selected
(Ture Left) parameter or moves leftward in OSD menu.
PIP Switch : To enable or disable Picture In Picture function.
PBP Switch : To enable or disable Picture By Picture function.
POP Switch : To enable or disable Picture On Picture function.
Adjust Switch : To activate or deactivate image adjustment in sub screen of PIP,
PBP, POP mode.
Locker Switch : To turn ON or OFF the system.
How to access the menu
1. Push the “Knob” button to activate the OSD menu.
2. Able to turn left or right the button to select those icons. The icon will be
highlighted when is selected.
3. Push the “knob” button to go into the another OSD menu.
4. Turn left or right “Knob” button to increase or decrease the number of selected
function.
5. In order to exit from the OSD menu in a different layer, then choose the “Exit
option”. In case of long pushing and holding the button around two or three
seconds, then it will completely out from the menu regardless wherever you are.
6. Turn the “Knob” button in a fast way to the left or right, while the menu is not
activated, then input signal menu will turns up with “V” mark which is chosen
current available input signal. You can switch to another input signal source by
turning the knob slowly then select the target source by pushing the knob.
Information Adjust Custom Entry, SN, Run Time, Input Format
Exit Exit the menu
18 User’s Guide
OSD Section
Chromatic
Color Temp Change the color temperature - D65, D93, S1, S2
Gamma Change the gamma value - 1.8, 1.9, 1.95, 2.1, 2.1S, 2.2, 2.3,
2.4, 2.4S, S0, Radio graph
Exit Exit the menu
Visual
Brightness Adjust the brightness of panel (Range 0 - 100)
Contrast Adjust the contrast of video (Range 0 - 100)
Phase Adjust the phase of video (Range 0 - 100)
*available for C and S - video
Chroma Adjust the chroma of video (Range 0 - 100)
*available for C and S - video
Sharpness-H Set the sharpness of horizontal image (Range 0 - 100)
Sharpness-V Set the sharpness of vertical image (Range 0 - 100)
Exit Exit the menu
OSD Section
Setting
Scale Mode Change the scale mode - Fill all, One to One,
Vertical - fill, Horizontal - fill, Fill aspect ratio
Color Space Change color space between RGB and YPbPr
*available for Component and RGBS
Mirror Off / On mirror image
Freeze Frame Off / On freeze frame
Zoom / Pan Enable zoom - in and pan function
EX210/EX240W 19
ENGLISH
PIP Enable PIP (Picture In Picture) function
POP Enable POP (Picture On Picture) function
PBP Enable PBP (Picture By Picture) function
Exit Exit the menu
20 User’s Guide
OSD Section
Sub menu of PIP
Mode PIP Mode ON / Off
Source PIP sub screen source - SDI1,2, C-Video, S-Video,
Component, VGA, DVI, SOG
Position PIP sub screen position - Top L, Top R, Bottom L, Bottom R
Size PIP sub screen size - Small, Medium, Large
Blending PIP sub screen blending (Range 0 - 100)
Swap PIP sub screen swap
Exit Exit the menu
Sub menu of PBP
Mode PBP Mode ON / Off
Source PBP sub screen source - SDI1,2, C-Video, S-Video,
Component, VGA, DVI, SOG
Swap PBP sub screen swap
Exit Exit the menu
OSD Section
Sub menu of POP
Mode POP Mode ON / Off
Source POP sub screen source - SDI1,2, C-Video, S-Video,
Component, VGA, DVI, SOG
Position POP sub screen position - Top L, Top R, Bottom L, Bottom R
Size POP sub screen size - Small, Medium, Large, Full
Blending POP sub screen blending (Range 0 - 100)
EX210/EX240W 21
ENGLISH
Swap POP sub screen swap
Exit Exit the menu
22 User’s Guide
OSD Section
EIZOced
OSD Position Change the OSD menu - Position, Background,
Control OSD time out, Language
Screen Control Control and adjust H and V position, Freqency, Phase
Noise Reduction, Motion Offset
DPMS Change the DPMS
Auto Source Adjust Auto Source Select between on and off
Select
Smart Select Enable / Disable smart select sub menu
Restore Factory Changes the all OSD value to factory outgoing status
Key Lock Set to key lock mode
Exit Exit the menu
OSD Section
EX210/EX240W 23
Sub menu of OSD Control
H-position Adjust OSD H - position (Range 0 - 100)
V-position Adjust OSD V - position (Range 0 - 100)
Back ground Adjust transparency of OSD back ground (Range 0 - 100)
OSD Time out Adjust OSD time out - 5s, 10s, 15s, 20s, 1m, 2m
Language OSD language - English, Japanese, Chinese, Korean,
French, German
Exit Exit the menu
Sub menu of Screen Control
H-position Adjust screen H - position (Range 0 - 100)
*available for VGA
ENGLISH
V-position Adjust screen V - position (Range 0 - 100)
*available for VGA
Frequency Adjust frequency (Range 0 - 100)
*available for VGA
Phase Adjust phase (Range 0 - 100)
*available for VGA
Noise Noise reduction (Range 0 - 100)
Reduction *available for Component, C-Video and S-Video
Motion Offset Motion offset (Range 0 - 15)
*available for Component, C-Video and S-Video
Exit Exit the menu
24 User’s Guide
OSD Section
Information
Custom Entry Change the user or monitor’s name
SN Display the serial number
Run Time Display the total run time
Input Format Display the current input resolution and vertical frequency
Exit Exit the menu
EX210/EX240W 25
TROUBLESHOOTING
Before sending your LCD monitor for servicing, please check the troubleshooting
list below to see if you can self-diagnose the problem.
ProblemsCurrent StatusRemedy
No PictureLED ON· Using OSD, adjust brightness and
LED OFF· Check the Locker switch and Power
LED Blinking· Check if video signal cable is properly
Abnormal Picture Unstable Picture· Check if the specification of graphics
Display is missing, center shift,
or too small or too large in
display size
contrast to maximum or reset to their
default settings.
switch.
· Check if AC power cord is properly
connected to the AC adapter.
connected at the back of monitor.
· Check if the power to computer system
is ON.
adapter and monitor is in compliance
which may be causing the input signal
frequency mismatch.
· Using Auto Setup button, if still display
abnormal picture then adjust CLOCK,
CLOCK-PHASE, H-POSITION and VPOSITION with non-standard signals.
· Using OSD, in case of missing full-screen
image, please select other resolution in
your Operating System (Windows 95/98,
NT) or other vertical refresh timing.
· Must wait for a few seconds after
adjusting the size of the image before
changing or disconnecting the signal or
powering OFF the monitor.
ENGLISH
26 User’s Guide
SPECIFICATION
Model EX210EX240W
LCD Panel21.3 TFT LCD Panel24.0 TFT LCD Panel
TypeActive MatrixActive Martrix
Resolution1600 x 1200 @ 60Hz1920 x 1200 @ 60Hz
Pixel Pitch0.27mm0.27mm
Display Color16.7M colors16.7M colors
Color ToneUp to 256 color toneUp to 256 color tone
Response Time<25ms Typ.<25ms Typ.
** All contents are subject to change without notice.
EX210/EX240W 27
: Direct Current
MEDICAL EQUIPMENT
51LJ
Medical Equipment
E215822
WITH RESPECT TO ELECTRIC SHOCK,
FIRE AND MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH UL 60601-1, AND
CAN/CSA C22.2 NO. 601.1
: UL approval mark according to the
safety standard for Medical equipment
This monitor is intended for use in Health Care Facilities model EX210/EX240W
Equipment is not suitable for use in the presence of flammable anesthetic mixture
with air or with oxygen or nitrous oxide.
No user serviceable parts inside, ask qualified personnel when accessing inside.
For disposal of waste product, follow the requirement of local code.
Electrical input rating: 24V DC 6.25A
Classification
ENGLISH
Type of protection against electric shock: Class I Equipment.
Degree of protection against the ingress of water: IPX1 compliance.
Mode of operation: Continuous
This monitor has been tested to comply with IEC/EN 60601-1, IEC/EN60601-1-2
and is certified by UL to medical standard UL60601-1(UL/cUL Mark).
Because many medical offices are located in residential areas, this monitor, in addition to the medical requirements, has also been tested and found to comply with
the limits for FCC Class B computing devices in a typically configured system. It is
the system integrator or configurer’s responsibility to test and ensure that the entire
system complies with applicable EMC laws.
Environmental conditions for transport and storage:
- Temperature range within -4° to 140° F (-20° to 60°C)
- Relative humidity range within 10% to 85%
- Atmospheric pressure range within 500 to 1060 hPa.
28 User’s Guide
Electromagnetic Compatibility
Like other electrical medical equipment, the EX210/EX240W requires special
precautions to ensure electromagnetic compatibility with other electrical medical devices.
To ensure electromagnetic compatibility (EMC), the EX210/EX240W must be installed and
operated according to the EMC information provided in this manual.
Note The EX210/EX240W has been designed and tested to comply with IEC
60601-1-2:2001 requirements for EMC with other devices.
Caution Portable and mobile RF communications equipment may affect
the normal function of the EX210/ EX240W..
Warning Do not use cables or accessories other than those provided
with the EX210/EX240W, as this may result in increased electromagnetic
emissions or decreased immunity to such emissions.
Warning If the EX210/EX240W is used adjacent to or stacked with
other equipment, observe and verify normal operation of the EX210/EX240W
in the configuration in which it will be used prior to using it in a surgical procedure. Consult the tables below for guidance in placing the EX210/EX240W.
The Model EX210/EX240W uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment
EX210/EX240W is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes
The Model EX210/EX240W is intended for use in the electromagnetic environment specified below.
The customer or the user of EX210/EX240W should assure that it is used in such an environment.
Immunity TestIEC 60601 Test LevelCompliance Level
Electrostatic discharge
(ESD)
IEC61000-4-2
Electrical fast transient
/burst IEC61000-4-4
Surge
IEC61000-4-5
6 kV contact
8 kV air
2 kV for power supply lines
1 kV for input / output lines
1 kV differential mode
2 kV common mode
6 kV contact
8 kV air
2 kV for power supply lines
1 kV for input / output lines
1 kV differential mode
2 kV common mode
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
EX210/EX240W 29
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC61000-4-11
Note: Ut is the A.C. mains voltage prior to application of the test level.
3.0 A/m3.0 A/m
<5% Ut (>95% dip in Ut)
for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec.
<5% Ut (>95% dip in Ut)
for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec.
Power frequency magnetic fields should be
at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
Mains power quaility
should be that of a typical
commercial or hospital
environment. If the user
of the EX210/ EX240W
image intensifier requires
continued operation during power mains interruptions, it is recommended
that the EX210/ EX240W
image intensifier be
powered from an uninterruptible power supply or
a battery.
The Model EX210/EX240W is intended for use in the electromagnetic environment specified below.
The customer or the user of EX210/EX240W should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic environment
Portable and mobile RF communications
equipment should be used no closer
to any part of the EX210/ EX240W,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended Separation Distance
0.15 MHz to 80 MHz
- guidance
ENGLISH
Conducted RF
IEC61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
(a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast, connot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the EX210/EX240W is used exceeds the applicable RF
compliance level above, the EX210/EX240W should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the EX210/EX240W.
(b) Over the frequency range 150kHz to 80MHz, field strengths should be less than [V1] V/m.
3 Vrms
150 KHz to 80 MHz
3 V/m
80MHz to 2.5 GHz
3 Vrms
150 KHz to 80 MHz
3 V/m
80MHz to 2.5 GHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
where P is the maximum oputput power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as deter-mined by an electromagnetic
site survey (a), should be less than the
compliance level in each frequency range
(b).Interferency may occur in the vicinity
of equipment marked with the follwing
symbol:
30 User’s Guide
Recommended Separation Distances Between Portable and Mobile RF Communications Equi-
pement and the EX210/EX240W System
The EX210/EX240W system is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The user of the EX210/EX240W system can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the EX210/EX240W system as recommended below, according to the
Rated maximum output
power (W) of transmitter
0.010.120.120.23
0.10.370.370.74
11.171.172.33
103.703.707.37
10011.7011.7023.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in me-
ters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
maximum output power of the communications equipment.
Separation distance (m) according to frequency of transmitter
150kHz to 80MHz
d=1.17, p
80MHz to 800MHz
d=1.17, p
800MHz to 2.5GHz
d=2.33, p
Dimension Drawing (mm) of EX210
EX210/EX240W 31
ENGLISH
32 User’s Guide
Dimension Drawing (mm) of EX240W
Connectors
DC Input
Connector Jack Power Input
PinDescriptionPinDescription
1GND4NC
2GND5VCC_24V
3VCC_24V
Video Input
24 pin DVI-I connector.
DDWG or equivalent connector.
PinDescriptionPinDescriptionPinDescription
1T.M.D.S. Data2-11T.M.D.S. Data1 Shield21NC
2T.M.D.S. Data2+12NC22T.M.D.S. Clock Shield
3T.M.D.S. Data2 Shield13NC23T.M.D.S. Clock+
4NC14+5V Power24T.M.D.S. Clock-
5NC15Ground (for +5V)25Analog Red
6DDC Clock16Hot Plug Detect26Analog Green
7DDC Data17T.M.D.S. Data0-27Analog Blue
8Analog V - Sync18T.M.D.S. Data0+28Analog H - Sync
9T.M.D.S. Data1-19T.M.D.S. Data0 Shield29GND(for analog signal)
10T.M.D.S. Data1+20NC
EX210/EX240W 33
ENGLISH
15 pin VGA connector.
PinDescriptionPinDescriptionPinDescription
1GRED6GND11NC
2GGREEN7GND12GSDA
3GBLUE8GND13IHS
4GND9G5V14IVS
5GND10GND15GSCL
S - Video connector.
PinDescriptionPinDescription
1GND3CHROMA
2GND4LUMA
C - Video connector.
PinDescription
1DATA IN / OUT
正規代理店
株式会社 ナナオ
〒924-8566 石川県白山市下柏野町153番地
電話: 076-275-4121
Fax: 076-275-4125
Web サイト:http://www.radiforce.com
Authorized Representative in Europe and Exclusive Distributor