E Instruments E4500, BTU4500, E6000 Quick Manual

OK
Compliant with: EN 50379-1, EN 50379-2
Features
Combustion, Flue gas & Emissions analysis Calculating of stack heat loss a nd efficiency CO environment measuremen t Measuring differential pressure Draft measurement Pressure measurement in the gas supply pipe Store Smoke values, calculati ng mean value
Date: 7/08/14 Time: 10.10
Fuel: Natural gas Altitude: 0 m R.H. air: 50 %
O2 4.2 ٪ CO2 9.3 ٪ Ex. Air 20 ٪ T flue 374.2 °F T air 60.4 °F ΔT 313.8 °F Loss tot 8.6 ٪ Eff. Tot 91.4 ٪ CO 148 ppm NO 40 ppm NOX/NO: 1.03 NOX 41 ppm CO amb 0 ppm
Draft: 0.05 inH2O Smoke: 3 1 2
Aver n. : 2
7/08/14
10:00
Eff. tot
%
91.4
Loss tot
%
8.6
ΔT
°F
74.7
T air
°F
15.4
T flue
°F
190.1
Ex. Air
1.25
CO2
%
9.3
O2
%
4.2
Memory Average analysis
7/08/14
10:00
Print Report
Model
partial
Analysis
average
OK
Memory
1
7/08/14
10:00
Print Report
Model
partial
Analysis
In average
F1
Memory
1
In case the automatic print mode is set then the average an alysis printing start automatically. Conversely, after the sampling of the third analysis, the a verage analysis is shown, so that it can be sent to the printer as explained in the following.
OK
Memory Save
7/08/14
10:00
OK
Memory
1
Mode
data logger
Samples
10
Interval
s
60
Automatically saves the first
sample when the
set interval
expires, and so on
until the third
sample.
7/08/14
10:00
Eff. tot
%
91.4
Loss tot
%
8.6
ΔT
°F
74.7
T air
°F
15.4
T flue
°F
190.1
Ex. Air
1.25
CO2
%
9.3
O2
%
4.2
Combustion analysis
7/08/14
10:00
Eff. tot
%
91.4
Loss tot
%
8.6
ΔT
°F
74.7
T air
°F
15.4
T flue
°F
190.1
Ex. Air
1.25
CO2
%
9.3
O2
%
4.2
Combustion analysis
1
120
KEYBOARD FUNCTIONS
KEYS FUNCTION
Activate the co ntext keys s hown on the display
Enters the Memory menu
Enters the Print menu
Enters the Configuration menu
Previes the combustion analysis
Enters the measurements menu
Turns the instrument on/off
Quits the current screen
Selects and/or modifies
Confirms data
CONTEXT KEYS
CONTEXT KEY FUNCTION
ZERO the pressure sensor
Updates the measure
Saves the current measurement or analysis in the memory location selected from the ‘Select Memory’ menu
Starts the report ticket
Enters the Print menu
Stops the flue gas analysis
Enters the modify mode for the selected parameter
Confirms the settings
Cancels the change/operation and returns to the previous screen
Zooms the screen
Shows the details for the s elected parameter
‘Find’ function; sta rts a quick search for
the analysis to recall
OK
O
Esc
OK
Configuration Fuel
7/08/14
10:00
Biogas
Propane-Air
Fuel oil
Diesel oil
Butane
L.P.G.
Propane
Natural gas
OK
7/08/14
10:00
Analysis
Operator
Information
Language
Instrument
Alarms
Diagnostic
Configuration
Restore
7/08/14
10:00
Fuel
O2 reference
Measure units
Measures list
Condensation
NOx/NO ratio
Autozero
Configuration Analysis
WARNING
F1: stop
Printing.
Please wait...
E Instruments respects the nature and the en vironment, therefore provides this quick user guide t o avoid any unwanted waste of paper. However the complete manual for use and m aintenance of t he instr ument i s alrea dy a vailable in the USB stick (included),. Respect your environment: think before printing the full m anual on paper.
The magnets in the back of the instrument can damage credit cards, hard drives, mechanical watches, pacemakers, defibrillators and other devices proven sensitive to magnetic fields. It is recommended to keep the instrument at a distance of at least 10” away from any of these devices.
402 Middletown Blvd, Suite 216, Langhorne, PA 19047 USA
Tel: (215) 750-1212; Fax: (215) 750-1399 info@E-Inst.com - www.E-Inst.com
To view different test measurements, use ' ' key to scroll through the vario us test measurement screens.
WARNING! - Before proceeding with the combustion analysis remember to s et the proper fuel.
Xair = Excess Air ΔT = Differential Temperature (NET Temperature)
Tg = Flue Gas Stack Temperature Eff. tot = Total Combustion Efficiency
Ta = Ambient or Incoming Air Temperature Loss tot = Stack Losses
K000000000EJ 027994 010217
COMPLETE MANUAL
ON USB FLASH DRIVE
7/08/14
10:00
Measurements Ambient CO
CO
ppm
0
CO max
ppm
0
It is mandatory to perform the instrument autozero in fresh, clean outdoor air.
7/08/14
10:00
Measurements Ambient CO
CO
ppm
0
CO max
ppm
0
Connect the gas probe to the instrument and perform
the measurement.
Wait 5 minutes.
7/08/14
10:00
Measurements Draft
P diff
inH2O
0.000
O
7/08/14
10:00
Measurements Draft
P diff
inH2O
- 0.020
O
Once the
pressure zeroing
is completed
insert the probe in
the chimney and
measure the draft.
AA SF--A
AAC TA03A
AAC KP01
AA SA08
7/08/14
10:00
Measurements Draft
P diff
inH2O
0.001
O
Before zeroing the
pressure it is mandatory to REMOVE the
probe from the
chimney.
View & Save the
DRAFT
measurement.
ZERO Draft
Sensor
O
Save the
measurement
Draft
Ambient CO
7/08/14
10:00
Eff. tot
%
91.4
Loss tot
%
8.6
ΔT
°F
74.7
T air
°F
15.4
T flue
°F
190.1
Ex. Air
1.25
CO2
%
9.3
O2
%
4.2
Combustion analysis UNI 10389
KEEP
PRESSED
FOR 2
SECONDS
UNTIL IT
BEEPS
Insert the gas probe in the chimney. The gas autozero w i ll b e a u to ma ti ca ll y performed thanks to the internal diverting electrovalve.
Memory Select
7/08/14
10:00
Customer
E Instruments
Memory
1
Address
402 Middletown
Ste 216
Phone
215-750-1212
Langhorne, PA 19047
Boiler
xxxx
Date
7/08/14
In "Select" can be in ser te d al l the customer data.
Memory Data logger
7/08/14
10:00
Samples
10
Mode
Data logger
Interval
ppm
60
Print
auto
Memory
auto
In "Data logger" , the user can define the an aly sis , mem ory selection and printing modes.
7/08/14
10:00
Save
Select
Delete
Average
Data logger
Usage %
Memory
OK OK
Data logger
Select
WARNING!
Be sure all connections are tight to assure accurate sampling. It is important that water trap & filter be installed vertically to
prevent damage from moisture & particles to sensors.
When conducting measurements, the water trap/filter
assembly MUST be in a VERTICAL position.
When testing is completed, always drain the water
trap with any condensation (after EVERY test!).
AA AL05
K000000000EJ 027994 010217