E Instruments 1500-Oil, 1500-1 User manual

Page 1
R
R
3
WARNING! - Be fore proceeding with the combustion analysis remember to set th e p r oper fu el.
Configuration Fuel
7/08/14
1.25
190.1
15.4
74.7
91.4
1
10:00
1
7/08/14
10:00
10:00
4.2
Automatically
9.3
saves the first
sample when the
set interval
expires, and so on
until the third
sample.
8.6
Date: 7/08/14 Time: 10.10
Fuel: Natural gas Altitude: 0 m R.H. air: 50 %
O2 4.2 ٪ CO2 9.3 ٪ Ex. Air 20 ٪ T flue 374.2 °F T air 60.4 °F ΔT 313.8 °F Loss tot 8.6 ٪ Eff. Tot 91.4 ٪ CO 148 ppm CO amb 0 ppm
Draft: 0.05 inH2O Smoke: 3 1 2
Aver n. : 2
—————————————————————— —————————————————————— —————————————————————— —————————————————————— ——————————————————————
7/08/14
10:00
Configuration
Analysis
Operator
Information
Language
Combustion analysis
O2
%
CO2
%
Ex. Air T flue
°F
T air
°F
ΔT
°F
Loss tot
%
Eff. tot
%
Instrument
Alarms
Diagnostic
Restore
7/08/14
10:00
1.25
190.1
15.4
74.7
91.4
4.2
9.3
8.6
OK
Mode Memory Samples
Interval
s
Fuel
reference
O
2
Autozero
Memory Save
OK
In case the automatic print mode is set then the average analysis printing start automatically. Conversely, after the sam pling of t he third a nalysis , the avera ge anal ysis is sh own, so t hat it ca n be se nt to the printer as explained in the following.
Memory Average analysis
O2
%
CO2
%
Ex. Air T flue
°F
T air
°F
ΔT
°F
Loss tot
%
Eff. tot
%
x To view different test measurements, use ' ' key to scroll through the various test measurement screens.
Xair = Excess Air ΔT = Differential Temperature (NET Temperature)
7/08/14
10:00
1.25
190.1
15.4
74.7
91.4
Print
4.2
9.3
Memory Analysis Model
Report
average
8.6
OK
Tg = Flue Gas Stack Temperature Eff. tot = Total Combustion Efficiency
Configuration Analysis
Condensation
Measure units
Measures list
data logger
7/08/14
10:00
1
partial
7/08/14
OK
10:00
10 60
7/08/14
10:00
1
OK
Memory Analysis Model
F1
Natural gas
Propane
OK
L.P.G. Butane Diesel oil Fuel oil Propane-Air Biogas
Combustion analysis
O2
%
CO2
%
Ex. Air T flue
°F
T air
°F
ΔT
°F
Loss tot
%
Eff. tot
%
120
7/08/14
Print Report
In average
WARNING
partial
Printing.
Please wait...
F1: stop
Ta = Ambient or Incoming Air Temperature Loss tot = Stack Losses
1500
KEYBOARD FUNCTIONS
KEYS FUNCTION
Activate the context keys shown on the display
Enters the Memory menu
Enters the Print menu
Enters the Configuration menu
Previes the combustion analysis
Enters the measurements menu
Turns the instrument on/off
Quits the current screen
Selects and/or modifies
Confirms data
E Instruments respects the nature and the environment, therefore provides this quick user guide to avoid any unwanted wast e of pa p er . However the complete manual for use and maintenance of the instrument i s already available in the USB stick (included),. Respect your environment: think before printing the full manual on paper.
Compliant with: EN 50379-1, EN 50 379-2 Features
x Combustion, Flue gas & Emissions analysis x Calculating of stack heat loss and efficiency x CO environment measurement x Measuring differential pressure x Draft measurement x Pressure measurement in the gas supply pipe x Store Smoke values, calculating mean value
The magnets in the back of the instrument can damage credit cards, hard drives, mechanical watches, pacemakers, defibrillators and other devices proven sensitive to magnetic fields. It is recommended to keep the instrument at a distance of at least 10” away from any of thes e dev ices .
CONTEXT KEYS
CONTEXT KEY FUNCTION
O
ZERO the pressure sensor Updates the measure
Saves the current measurement or analysis in the memory locatio n selected from the ‘Select Memory’ menu
Starts the report ticket Enters the Print menu Stops the flue gas analysis
Enters the modify mode for the selected parameter
Confirms the settings
OK
Cancels the change/operation and
Esc
returns to the previous screen Zooms the screen Shows the details for the selected
parameter
‘Find’ function; starts a quick s earch for
the analysis to recall
COMPLETE MANUAL
ON USB FLASH DRIVE
K000000000EJ 027995 010217
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
AA AL05
AAC TA03A
AAC KP01
AA SA08
WARNING!
x Be sure all connections are tight t o assure accurate s ampling. x It is importan t that water trap & filter be installed vertically to
prevent damage from moisture & particles to sensors.
x When conduc ting measureme nts, the water trap/filter
assembly MUST be in a VERTICAL position.
x When testing is completed, always drain the water
trap with any condensation (after EVERY test!).
WARNING! During autozero DO NOT insert the gas probe in the chimney.
KEEP
PRESSED
FOR 2
SECONDS
UNTIL IT
BEEPS
AA SF--A
Combustion analysis UNI 10389
O2
%
CO2
%
Ex. Air T flue
°F
T air
°F
ΔT
°F
Loss tot
%
Eff. tot
%
7/08/14
10:00
1.25
190.1
15.4
74.7
91.4
4.2
9.3
8.6
15/01/1
4
Configuration Air temperature
Probe T
Analysis
°C
25.0
Air T
Air temp.
It saves the value, acquired or entered in the parameter
‘Air T’.
Configuration Air temperature
°C
15/01/1
Probe T
25.0
°C
Air T
-----
°C
While in manual analysis, holding pressed both keys and
makes the instrument switch off the suction fumes pump and blocks the refresh of any current measure. To switch on the suction fumes pump again and reactivate the refresh of the current measure, press again the keys
and .
-----
4
Acquires the temperature value detected from the sampling probe. That value is reported in
the parameter ‘Air T’.
Configuration Air temperature
15/01/1
Probe T
25.0
°C
Air T
25.0
°C
7/08/14
10:00
Measurements Draft
P diff
Draft
inH2O
0.001
Before zeroing the
pressure it is mandatory to REMOVE the
probe from the
chimney.
It is mandatory to perform the instrument autozero in fresh, clean outdoor air.
7/08/14
10:00
Measurements Ambient CO
CO
O
O
Measurements Draft
P diff
0.000
4
inH2O
7/08/14
10:00
pressure zeroing
is completed
insert the probe in
the chimney and
measure the draft.
ZERO Draft
Sensor
Once the
Ambient CO
ppm
CO max
ppm
Connect the gas probe to the instrument and perform
the measurement.
Wait 5 minutes.
0
0
O
7/08/14
10:00
0
Save the
measurement
P diff
inH2O
Measurements Draft
- 0.020
7/08/14
10:00
View & Save the
DRAFT
measurement.
CO
ppm
Measurements Ambient CO
CO max
0
ppm
O
K000000000EJ 027995 010217
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...