Eickemeyer LifeVet CP User Manual

Product Information
• Product Model: LifeVet CP
• Product Name: Capnograph and Pulse Oximeter
• Manufacturer: EICKEMEYER KG
• After-service Contact Information:
Address: Eltastraße 8, 78532 Tuttlingen, Germany Phone: +49 7461 96 580 0 Fax: +49 7461 96 580 90 Email: info@eickemeyer.de
Revision History
This manual has a revision number. This revision number changes whenever the manual is updated due to software or technical specification change. Contents of this manual are subject to change without prior notice.
• Revision number: V1.0
• Release time: Oct. 2016
II
Statement
Manufacturer holds the copyright of this manual, and we are also entitled to deal with this manual as confidential files. This manual is only used for operation, maintenance and service of product, someone else can not publish the manual.
This manual contains exclusive information protected by copyright laws and we reserve its copyright. Without written approval of manufacturer no parts of this manual shall be photocopied, xeroxed or translated into other languages.
The contents contained in this manual are subject to amendments without notification.
Manufacturer's Responsibility
Only under the following circumstances will manufacturer be responsible for the safety, reliability and performance of the instrument:
• All the installation, expansion, readjustment,
renovation or repairs are conducted by the personnel certified by manufacturer.
• The storage condition, operation condition and
electrical status of the instrument conforms to the product specification.
• The instrument is used in accordance with the users
manual.
3
About this manual
This manual contains the instructions necessary to operate the product safely and in accordance with its function and intended use. Observance of this manual is a prerequisite for proper product performance and correct operation and ensures patient and operator safety.
This manual is based on the maximum configuration and therefore some contents may not apply to your product. If you have any question, please contact us.
This manual is an integral part of the product. It should always be kept close to the equipment so that it can be obtained conveniently when needed.
All illustrations in this manual serve as examples only. They may not necessarily reflect the setup or data displayed on your product.
Conventions:
• Bold Italic text is used in this manual to quote
the referenced chapter or sections.
• 【】is used to enclose screen texts.
• → is used to indicate operational procedures.
4
Signs in this manual:
Warning:
Indicates a potential hazard or unsafe practice that, if not avoided, will result in death or serious injury.
Caution: Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage.
Note: Provides application tips or other useful information to ensure that you get the most from your product.
Contents
Chapter 1 General Introduction ......................................... 1-1
Intended Use ........................................................... 1-1
Main Unit ................................................................ 1-1
Display Views ......................................................... 1-6
Chapter 2 Safety .................................................................. 2-1
Safety Information .................................................. 2-1
Explanation of Symbols .......................................... 2-5
Chapter 3 Basic Operations ................................................ 3-1
Unpacking and Checking ........................................ 3-1
Getting Started ........................................................ 3-1
Starting the Monitor ................................................ 3-2
Screen Brightness Setting ........................................ 3-3
Auto-Rotate Setting ................................................ 3-3
Date & Time Setting ............................................... 3-3
Patient Information Setting ...................................... 3-4
Demo Mode Setting ................................................ 3-5
Language Setting .................................................... 3-5
Checking the Version .............................................. 3-6
Restoring the Factory Configuration ........................ 3-6
Shutting off the Monitor .......................................... 3-6
Chapter 4 Alarm .................................................................. 4-1
Alarm Categories .................................................... 4-1
Alarm Levels ........................................................... 4-2
Alarm Indicators ..................................................... 4-3
Alarm Status Symbol .............................................. 4-6
Alarm Tone Configuration ....................................... 4-6
Pausing Alarms ....................................................... 4-7
Shutting off the Alarm Volume ................................ 4-8
Alarm Reset ............................................................ 4-9
When an Alarm Occurs ........................................... 4-9
Chapter 5 Measuring CO2 .................................................. 5-1
Introduction ............................................................ 5-1
Safety Information .................................................. 5-1
Monitoring Procedure ............................................. 5-2
CO2 Display ........................................................... 5-6
Respiratory Rate ...................................................... 5-7
Setting CO2 ............................................................. 5-7
CO2 Zero .............................................................. 5-10
Calibration ............................................................ 5-11
Removing Exhaust Gases from the System ........... 5-12
Chapter 6 Measuring SpO2 ................................................ 6-1
Introduction ............................................................ 6-1
Safety Information .................................................. 6-2
Monitoring Procedure ............................................. 6-4
SpO2 Display .......................................................... 6-5
PR Display .............................................................. 6-6
Setting SpO2 ........................................................... 6-7
Setting Desat Limit ................................................. 6-8
Chapter 7 Trend Review ..................................................... 7-1
Introduction ............................................................ 7-1
Review Interface ..................................................... 7-1
Review Setup .......................................................... 7-2
Chapter 8 Battery ................................................................ 8-1
Introduction ............................................................ 8-1
Charging the Battery ............................................... 8-2
Optimizing Battery Performance ............................. 8-2
Checking the Lithium Battery ................................. 8-3
Disposing of the Batteries ........................................ 8-4
Chapter 9 Maintenance and Cleaning ................................ 9-1
Introduction ............................................................ 9-1
Seasonal Safety Checking ....................................... 9-2
Cleaning the Monitor .............................................. 9-4
Cleaning SpO2 Sensor ............................................ 9-5
Disposal .................................................................. 9-5
Chapter 10 Accessories ..................................................... 10-1
VIII
Appendix A Product Specifications ........................................ 1
Safety Specifications .................................................... 1
Physical Specifications ................................................ 1
Environmental Specifications....................................... 2
Charging Specifications ............................................... 2
Hardware Specifications .............................................. 3
Data Storage ................................................................ 4
Measurement Specifications ........................................ 5
Appendix B Factory Defaults ................................................. 9
Alarm Setup ................................................................. 9
System Setup ................................................................ 9
CO2 Setup .................................................................. 10
SpO2 Setup ................................................................ 10
Trend Setup ................................................................ 10
Appendix C Alarm Message .................................................. 11
Physiological Alarm .................................................... 11
Technical Alarm ................................ ......................... 12
Prompt Message ......................................................... 14
Appendix D EMC ................................................................. 15
Appendix E Warranty Registration Card ........................... 16
LifeVet CP User Manual
1-1
Chapter 1 General Introduction
Intended Use
LifeVet CP patient monitor is intended for continuously monitoring or spot checking of CO2, RR, SpO2 and PR signals in animals.
This device can be used in institutions or units with health care capability. For instance, outpatient departments, emergency rooms and departments of internal medicine in hospitals, and ordinary departments in clinics, nursing hospitals and medical institutions for communities.
Main Unit
Front View
LifeVet CP User Manual
1-2
6 1
2
5
3
4
Fig 1-1 Front View of the Monitor
1. Alarm indicating lamp
When an alarm occurs, this lamp will light up as defined
below:
• High level alarm: the lamp quickly flashes red.
• Medium level alarm: the lamp slowly flashes yellow.
• Low level alarm: the lamp lights yellow without
flashing.
2. Display screen
LifeVet CP User Manual
1-3
The device uses resistive touchscreen, using stylus or fingernail will improve sensitivity.
3. Alarm pause button
• It can pause the alarm for 120s when alarm volume is
on.
• Pressing it can change the alarm message to prompt
message when “Sensor off” alarm happens.
4. Main interface button
• Press this button to return to main interface when it is
on menu setting.
• Press this button to shift between different display
modes when it is in main interface.
5. Menu
• Press this button to enter into menu interface when it is
on main interface.
• Press this button to return to main interface when it is
on menu setting interface.
6. Battery charging indicating lamp
• It is orange when the device is being charged.
• It turns off when the battery is full or device isnt being
charged
Rear View
LifeVet CP User Manual
1-4
Fig 1-2 Rear View of the Monitor
Side View
Topside:
1 2
LifeVet CP User Manual
1-5
Downside:
3
Rightside:
4
5
Fig 1-3 Side View of the Monitor
1. CO2 connector
2. SpO2 probe connector
3. Micro USB connector
• Connect with power adapter.
4. Shortcut key
Press this button to start or pause the CO2 measurement.
5. Power buttom
• Press it about two seconds to turn on when the monitor is
on the condition of shutdown.
LifeVet CP User Manual
1-6
• Press it about two seconds to turn off when the monitor is
on the condition of working.
• Calibration of touch screen
Press shortcut key firstly and press power button and immediately loose shortcut key, click the center of appearing point on screen. If the calibration passes, it will enter the normall interface, if not, a red fork will appear on screen and continue to calibrate.
Display Views
This device has a function of automatic display rotation (Gravity Activated) which provides for vertical and horizontal positioning to maximizing space utilization and visibility.
LifeVet CP User Manual
1-7
Multi-Parameter Display Mode
6
8
7
5
4
3
2
1
9
10
Fig 1-4 Multif-parameter Display Mode
1. Patient ID No.: Click and set patient information, its range
from 1 to 96.
2. SpO2 parameter area: The current SpO2 and its higher
and lower alarm limits are shown in the area.
3. SpO2 waveform area: The waveform shown in the area is
current SpO2 volume curve.
LifeVet CP User Manual
1-8
4. CO2 parameter area: The current CO2 and its higher and
lower alarm limits are shown in the area
5. CO2 waveform area: CO2 waveform is shown in the area.
6. Physiological alarm area: Current physiological alarm
information is shown in the area.
7. Technical alarm and prompt information area: Current
technical alarm and prompt information are shown in the area.
8. Alarm status area: Alarm status symbols are shown in the
area.
9. System time: Current time is shown in the area.
10. Battery symbol: The symbol indicates the current quantity
of electricity of batteries and whether the device is connecting power source, the alternating-current symbol is above battery symbol when the device is connecting power source.
LifeVet CP User Manual
1-9
SpO2 Display mode
1
2
3
Fig 1-5 SpO2 Display Mode
1. SpO2 waveform area: The waveform shown in the area is
current SpO2 volume curve.
2. SpO2 parameter area: The values shown in the area are
current SpO2 value and its higher and lower alarm limits.
3. PR parameter area: The values shown in the area are
current PR value and its upper and lower alarm limits.
LifeVet CP User Manual
1-10
CO2 Waveform Display Mode
1
2 3
Fig 1-6 CO2 Display Mode
1. CO2 waveform display area: Waveform shown in the area
is current CO2 waveform
2. CO2 parameter area: The values shown in the area are
current HR value and its higher and lower alarm limits.
3. RR parameter area: The values shown in the area are
current RR value and its higher and lower alarm limits.
LifeVet CP User Manual
2-1
Chapter 2 Safety
Safety Information
:
Warning:
• Explosion hazard: Do not use the monitor in the
presence of flammable anesthetics mixture with air, oxygen, or hydrogen.
• When the monitor is in use, there should not be any
great power appliances as high voltage cables, X-ray machine, ultrasound equipment and electrizer in use nearby.
• Do not open the monitor housings; electric shock
hazard may exist. All servicing and future upgrades must be carried out by the personnel trained and authorized by manufacturer only.
• When the monitor is connecting with high-frequency
devices, sensors and cables should avoid touching high-frequency devices, in order to leakage current burns patient.
• Keep the monitor away from dust, vibration,
corrosive substances, explosive materials, high temperature and moisture.
LifeVet CP User Manual
2-2
• Do not come into contact with the patient during
defibrillation. Otherwise serious injury or death could result.
• When the monitor is connecting with high-frequency
devices, sensors and cables should avoid touching high-frequency devices, in order to leakage current burns patient.
• The monitor is not designed for the sterilized room.
• The monitor should be handled with care so as to
avoid shocks and falls.
• When the monitor is in use, it must be ensured the
batteries have sufficient capacity; otherwise there might be such phenomena as starting-up abnormalities or inaccurate measurement data, etc.
• The use of accessories, sensors, and cables other than
those specified may result in increased emission, low anti-disturbance and/or create invalid readings of the monitor. It is advised to check it at least once a month.
• The physiological data and alarm messages displayed
on the monitor are for reference only and cannot be directly used for diagnostic interpretation.
• Disposable devices are intended for single use only.
They should not be reused as performance could
LifeVet CP User Manual
2-3
degrade or contamination could occur.
• At the end of its service life, the product described
in this manual, as well as its accessories, must be disposed of in compliance with the guidelines regulating the disposal of such products. If you have questions concerning disposal of products, please contact manufacturer or its representatives.
• To avoid inadvertent disconnection, route all cables in
a way to prevent a stumbling hazard. Wrap and secure excess cabling to avoid risk of entanglement or strangulation by patient or personnel
Caution :
• The monitor does not contain any parts for self-repair
by users. The repair of the instrument must be conducted by the technical personnel authorized by manufacturer.
• To ensure patient safety, use only parts and
accessories specified in this manual
• When the monitor is connected to AC power, the
battery is in a state of being recharged. When it is
LifeVet CP User Manual
2-4
unable to be connected to the AC power, the battery can be used to supply power, and at this time it is unnecessary to use the electrical wires, and the instrument can be switched on directly.
• The monitor can only monitor one patient at a time.
• When a defibrillator is applied on a patient, the
monitor may have transient disorders in the display of waveforms. If the electrodes are used and placed properly, the display of the monitor will be restored within 10s. During defibrillation, please note to remove the electrode of limb lead to the side of the limb.
• In order to have more accurate measurements results,
the monitor should be used in quiet and comfortable environment.
• To guarantee the normal and safe operation of the
monitor, a preventive check and maintenance should be conducted for the monitor and its parts every 6 to 12 months (including performance check and safety check) to verify the instrument can work in a safe and proper condition and it is safe to the medical personnel and the patient and has met the accuracy required by clinical use.
LifeVet CP User Manual
2-5
• This manual describes all features and options. Your
monitor may not have all of them.
Explanation of Symbols
Symbol
Symbol Note
Type CF applied part, defibrillation protected The unit displaying this symbol contains an F-Type
isolated (floating) applied part providing a high degree of protection against shock, and is defibrillator-proof.
Refer to User Manual.
Alternating current
IPX1
Degree of protection against ingress of liquid
Alarm volume off
Alarm paused
LifeVet CP User Manual
2-6
Alarm reset
QRS volume off
Date of manufacture
Manufacturer
Serial number
Power button
CO2
Short for “Carbon dioxide”
RR
Short for “Respiratory Rate
SpO
2
Short for “Pulse Oxygen Saturation”
Symbol for the marking of electrical and electronics devices according to Directive 2002/96/EC.
LifeVet CP User Manual
3-1
Parts
Standard
Optional
Quantity
CO2 nose sampling tube
1
CO2 filter
1
SpO2 probes
1
User Manual
this
manual
QC certificate
1
Packing list
1
Power adapter
1
USB data cable
1
Carrying case
1
Suction mount
1
Chapter 3 Basic Operations
Unpacking and Checking
Open the package. Parts are as follows in the package .Take
out the monitor and its accessories.
Getting Started
Before you start to make measurements, carry out the
LifeVet CP User Manual
3-2
following checks on the monitor including all connected modules.
——Check for any mechanical damage; ——Check for any incorrect connection of all the external
cables and accessories
:
Warning:
• If the monitor is mechanically damaged, or if it is not
working properly, do not use it for any monitoring procedure on a patient. Contact your service personnel.
• To avoid explosion hazard, do not use the monitor in
the presence of flammable anesthetics, vapors or liquids.
Starting the Monitor
Press the button about two seconds to turn on the monitor. The alarm indicating lamp flashes, and then goes out. The system gives a beep and enters the main screen.
LifeVet CP User Manual
3-3
Screen Brightness Setting
MenuSystem】,click the right of Brightness】,
you can set the screen brightness to a value between 1to 5, choose the low level brightness to save power.
Caution: If the monitor is used outdoors or the ambient
light is strong, set the screen brightness to a higher level.
Auto-Rotate Setting
Menu】→【System】,click the right of Auto-rotate
to select On or Off. If you select On, the screen can react to the gravity. When the monitor rotates, the screen will rotate the display direction automatically.
Date & Time Setting
After starting up, you need to set date and time of this
monitor. Operations are as follows:
1. SelectMenu】→【Systemto enter the System menu
shown as follows:
LifeVet CP User Manual
3-4
2. Select Date Format , it can be set to
Year/Month/Day 】 、 【 Month/Day/Year 】 or Day/Month/Year】.
3. SelectTime Format】, it can be set to24h】or12h】.
4. Set the current date and time and selectOKto confirm
it.
Patient Information Setting
Please select patient information correctly before measuring,
Click IDon the left bottom of main screen to enter into
Patient Info.. You also can select 【Menu】→【System】 →【Patient Info.. Setting shown as follow:
1. Click the right of IDto set it values.
2. SetType】to【Horse】,【Dog】orCat.
Caution: The alarm limits of different parameters
depend on the patient type. If you set patient type incorrectly, the monitor will judge patient condition by current setting, which might be wrong for your patient.
LifeVet CP User Manual
3-5
Demo Mode Setting
To enter the demo mode:
Select Menu】→【System】→【Maintenance】→enter
the required password. Click the right of Demoto turn on.
To exit the demo mode:
Select Menu】→【System】→【Maintenance】→enter
the required password. Click the right of Demoto turn off.
Caution: The Demo mode is for demonstration purpose
only. To avoid that the simulated data are mistaken for the monitored patients data, you should not enter the Demo mode during a patient is being monitored. Otherwise, improper patient monitoring and delayed treatment could result.
Language Setting
Select Menu】→【System】→【Maintenance】, enter the required password. On Factory Mainten.interface, you can select Language】and then choose a desired language.
LifeVet CP User Manual
3-6
Checking the Version
SelectMenuSystemto check the version of the
monitor .
Restoring the Factory Configuration
If you have changed the system’s configuration and want to
restore the factory configuration, follow this procedure:
1. SelectMenuSystem.
2. Selec【t Set to Default】, popping up a confirming window,
selectOKto restore the factory configuration.
Shutting off the Monitor
Pressing power button about 2s can turn off the monitor.
1. Confirm that the patient monitoring is finished.
2. Disconnect all sensors and cables form the monitor.
3. Press the power button and hold it for 2s to turn off the monitor If the monitor cant be switched off normally, forced close the monitor by pressing and holding the power switch more than 5s. This may cause some damages to the device. The device will turn off automatically if any operation or
measurement is going on. Auto power-off setting: Menu】→
LifeVet CP User Manual
3-7
System】→【Maintenance, enter the required password, click the right of Auto power-off setting, you can select
off, 10min”, 30min”.
LifeVet CP User Manual
4-1
Chapter 4 Alarm
Alarm refers to a prompt that is given by the monitor for
medical personnel through visual, audible and other means when a vital sign appears abnormal or the monitor occurs technical problem.
Q?
Note: The monitor generates all the audible and visual
alarms through speaker, alarm lamp and screen. When the monitor powers on, the alarm lamp will be lighted in red and yellow one time and the speaker will give a beep voice, which indicates the alarm system of the monitor is working normally.
Alarm Categories
By nature, the monitors alarms can be classified into three
categories:
1. Physiological alarms Physiological alarms are triggered by a monitored parameter
value that violates set alarm limits or an abnormal patient condition. Physiological alarm message are displayed in the physiological alarm area.
LifeVet CP User Manual
4-2
2. Technical alarms Technical alarms are triggered by a device malfunction or a
patient data distortion due to improper operation or system problems. Technical alarm messages are displayed in the technical alarm area.
3. Prompt messages As a matter of fact, prompt messages are not alarm messages.
Apart from the physiological and technical alarm messages, the pulse monitor will show some messages telling the system status. Prompt messages are displayed in the technical alarm area.
Alarm Levels
1. By severity, the monitor’s physiological alarms can be classified into three categories: high level alarms, medium level alarms and low level alarms.
• High level alarms
Indicate that the patient is in a life threatening situation and an emergency treatment is demanded.
• Medium level alarms
Indicate that the patients vital signs appear abnormal and an immediate treatment may be required.
• Low level alarms
LifeVet CP User Manual
4-3
Indicate that the patents vital signs appear abnormal and an immediate treatment may be required.
2. By severity, the monitors technical alarms can be classified into four categories: high level, medium level alarms, low level alarms and prompt message.
Caution:
• The levels of technical alarms are predefined before
the monitor leaves the factory and cannot be changed by users.
• The level of technical alarm can’t be changed by the
user.
Alarm Indicators
When an alarm occurs, the monitor will raise user’s attention
by the following indications:
• Alarm tone: According to alarm level, speaker in the
monitor gives alarm sound in different tone.
• Alarm lamp: According to alarm level, alarm lamp on
monitor flashes in different color speed.
• Alarm message: Alarm message are displayed on the
LifeVet CP User Manual
4-4
screen.
• Flashing numeric: The numeric of parameter in alarm
flashes.
Caution: For different alarm levels, the alarm lamp,
alarm tone and alarm messages presented are different.
Alarm Tone
The different level alarms are indicated by the system in
following different audio ways:
Alarm level
Audible prompt
High “DO-DO-DO------DO-DO,DO-DO-DO------DO-DO
Medium
DO-DO-DO
Low “DO-”
Alarm Lamp
When an alarm occurs, the alarm levels are indicated in the
LifeVet CP User Manual
4-5
following different visual ways:
Alarm level
Visual prompt
High Alarm lamp flashes in red with 2 Hz.
Medium
Alarm lamp flashes in yellow with 0.5 Hz.
Low
Alarm lamp lights on in yellow without flashing.
Caution:
• When multiple alarms of different levels occur at the
same time, the monitor will select the alarm of highest level give visual and alarm indications.
• When multiple alarms occur at the same time, the
alarm message will be displayed in the alarm area in turn.
Alarm Message
When an alarm occurs, the alarm message will be displayed
in the alarm area:
◆ The system uses the following symbols to match the
alarm level of physiological alarm messages:
High level alarms: ***
LifeVet CP User Manual
4-6
Medium level alarms: ** Low level alarms: *
◆ The system uses different background colors for the alarm
message to match the alarm level:
High level alarms: re d Medium level alarms: yellow Low level alarms: yellow Prompt message: blue
Flashing Numeric
When a physiological alarm occurs, the numeric of
parameter will flash.
Alarm Status Symbol
To identify the control for alarm paused or to
indicate that the alarm system is in the alarm system is in the alarm paused state.
indicates the alarm sound is turned off.
To identify the control for alarm reset.
Alarm Tone Configuration
1. The minimum alarm volume setting.
LifeVet CP User Manual
4-7
Select Menu】→【System】→【Maintenance】, enter
the required passwo d, select Machine Mainten.】→【Alarm
Setup】→【Min.Alm.Vol.】, you can select Off, High, Mid, Low”.
2. Alarm volume setting
Select Menu】→【System】→【Alarm Volume】, you
can select from X to 4. X indicates the minimum alarm volume.
Pausing Ala rms
Press the button
on the front panel of monitor, you can
suspend all alarm indicators of the monitor:
• The visual alarm and audible alarm are all suspended.
• The parameters of physiological alarm stop flashing.
• The alarm message in the physiological alarm area will not
be displayed.
• The remaining time and the icon will be shown in the
physiological alarm area.
After the alarm paused time, the monitor will automatically cancel the alarm pausing. Press again the button pausing can be cancelled by manual operation.
, the alarm
LifeVet CP User Manual
4-8
Shutting off the Alarm Volume
Set the Min.Alm.Vol.】andAlarm Volume】to off
to shut off the alarm volume. Then there will be a symbol shown in the alarm status area. The alarm lamp and alarm
messages are still active after the alarm volume is off. The audible alarm is reactivated automatically when:
• The factory configuration is finished;
• Set the alarm volume to a non-off value.
When a factory configuration is selected, the alarm volume of the monitor may be lower than the minimum alarm volume. In this case the
alarm volume is automatically adjusted
according to the minimum alarm volume.
:
Warning:
• When the alarm sound is switched off, the monitor will
give no audible alarm tones even if a new alarm occurs. Therefore the user should be very carefully about whether to switch off the alarm sound or not.
• Don’t rely exclusively on the audible alarm system for
patient monitoring. Adjusting alarm volume to a low level may result in a hazard to the patient. Always keep the patient under close surveillance.
LifeVet CP User Manual
4-9
Alarm Reset
Select Menu】→【System】→【Maintenance】→enter the required passwo d, select Machine Mainten.】→【Alarm Setup】. You can turn on alarm reset. Alarm reset will be
displayed on the interface of system.
Select Menu】→【System】→【Alarm reset】. After licking alarm reset :
• You can reset current alarm system, it will exit alarm pause
if it is on the condition of alarm pause.
• It only turns off audible alarm, the visual is going on for the
existing alarm.
When an Alarm Occurs
Q?
Note: When an alarm occurs, you should always check
the patient’s condition first.
Check the alarm message appeared on screen. It is needed to identify the alarm and action appropriately, according to the cause of the alarm.
1. Check the patient’s condition.
LifeVet CP User Manual
4-10
2. Identify alarming parameter and alarm category.
3. Identify the cause of the alarm.
4. Silence the alarm, if necessary.
5. When cause of alarm has been over, check that the alarm system is working properly.
You will find the alarm message for the individual parameter
in Appendix C Alarm message.
LifeVet CP User Manual
5-1
Chapter 5 Measuring CO2
Introduction
The monitor adopts infrared absorption technology to measure the carbon dioxide (CO2) concentration in the breathing airway of patient. Because CO2 molecule can absorb infrared light of special wavelength, and the amount of absorbed infrared light directly relates to the concentration of CO
2,
therefore while the infrared light radiated from the infrared light source passing through the gas sample containing CO2, part of energy will be absorbed by CO2 in the gas. At another side of infrared light source, a photodetector is used to measure the remaining infrared energy and convert it to electric signal, which will be compared with the energy of infrared light source and adjusted so as to correctly reflect the CO2 concentration in the gas sample.
Sidestream: Takes a sample of the respiratory gas with a constant sample flow from the patient’s airway and analyzes it with the CO2 sensor.
Safety Information
:
Warning:
• Do not position the sensor cables or tubing in any
LifeVet CP User Manual
5-2
manner that may cause entanglement or strangulation.
• Performance is not guaranteed if an item labeled as
single patient use is reused.
• Monitor the CO2 waveform (Capnogram). If you see
changes or abnormal appearance check the sampling tube. Replace it if needed.
• Monitor the CO2 waveform (Capnogram) for elevated
baseline. Elevated baseline can be caused by sensor or patient problems.
• Do not operate the CO2 module when it is wet or has
exterior condensation.
• Do not use device on patients that cannot tolerate the
withdrawal of 50 ml/min±10 ml/min from the airway or patients that cannot tolerate the added dead space to the airway.
• Do not connect the exhaust tube to the ventilator
circuit.
Monitoring Procedure
1. The measurement value will be more accurate if the
monitor has 2 minutes’ warming-up time. To connect the
LifeVet CP User Manual
5-3
CO2 filter with 3-way joint of the loop of anesthetic machine by sampling tube, or to directly connect patients nose by the hose.
Fig 5-1-1 Connection of Sampling and Filter
LifeVet CP User Manual
5-4
Fig 5-1-2 Connection of Sampling with 3-way joint or
patient’s nose
Q?
Note:
• Inserting the sampling tube into the receptacle
automatically starts the sampling pump. Removal of the sampling tube turns the sample pump off.
• To remove the CO2 filter from CO2 connector, press
and rotate anticlockwise the filter, and then pull out filter.
LifeVet CP User Manual
5-5
2. Connect CO2 filter into CO2 connector and rotate CO2
filter closewise.
3. Connect sampling tube into CO2 filter. If the sampling
tube is occluded or damaged, perform a “Check sampling line” on screen..
4. Ensure that the CO2 sensor exhaust tube vents gases away
from the sensor environment.
5.
Using the shortcut key on the right of monitor to start or pause CO2 measurement.
Caution:
• Always disconnect the filter from the CO2 connector
when not in use.
• Do not insert the things other than filter into CO2
connector.
• The sampling tubes are disposable. Please keep the
sampling tube clean, and prevent the tube from clogging by dust. It is advised to replace the sampling tube every 12h (up to 120h of use with filter tip), the sampling tube leaks or has been damaged and contaminate .
LifeVet CP User Manual
5-6
CO2 Display
• CO2 parameter display
1
5
2
4
3
Fig 5-2 CO2 Display
1.CO2 label 2.CO2 high alarm limit
3. CO2 low alarm limit 4. CO2 value
5. CO2 unit
• CO2 waveform display
Fig 5-3 CO2 Waveform Display
LifeVet CP User Manual
5-7
Respiratory Rate
1
5
2
4
3
Fig 5-4 RR Display
1. RR label 2. RR higr alarm limit
3. RR low alarm limit 4. RR value
5. RR unit
Setting CO2
Select Menu】→【CO2 Setup】, enter into CO2 setup
interface.
LifeVet CP User Manual
5-8
Fig 5-5 CO2 Setup Interface
Setting CO2 and RR Alarm
Click the right of Alarm, you can set CO2 and RR
Alarm , you can select Mid, High”.
Setting CO2 and RR Alarm Limits
Click the right of Uplimit】or Downlimit】, you can set
up limit and down limit of CO2 and RR. Attention: The high
LifeVet CP User Manual
5-9
alarm limit should greater than the lower one.
Setting CO2 Scan Speed
Select scan speed of CO2 waveform. Click the right of
Speed, you can select “6.25 mm/s12.5mm/s, 25 mm/s”.
Setting CO2 Unit
Click the right ofCO2 Unit】, you can select mmHg%
kPa”.
Setting Scale
You can adjust the position of wave scale manually, and the
waveform amplitude will vary along with it. Click the right of
Scale, you can select “61mmHg76mmHg91 mmHg 106 mmHg”.
Setting Apnea Alarm
You can select the apnea time as required in the options. The
monitor indicates an alarm when a pre-adjusted time has elapsed since the last detected breath. Click the right Apnea Alm, you can select Off, 5s, 10s, 20s, 40s, 60s, 80s, 100s, 120s”.
LifeVet CP User Manual
5-10
CO2 Zero
While zeroing is recommended the first time the monitor is
connected to the filter, it is only absolutely necessary when the message Zero Required” is displayed. Follow these steps:
1) Ensure that the sampling tube is not connected to the patient or close to any source of CO2 (including the patient's, your own, exhaled breath and ventilator exhaust valves).
2) Select Menu】→【System】→【Maintenance】, enter password.
3) Select Machine Mainten.】→【CO2 Setup】→【CO2
Zero,The unit zeroes the module and displays the Zero In Progress message for approximately 15-20 seconds.
The message disappears upon completion of the zeroing.
Caution:
• Always ensure that the filter is properly connected to
the CO2 connector before zeroing.
• Do not attempt zeroing for 20 seconds after removing
the sampling tube from the patients airway. This time
LifeVet CP User Manual
5-11
allows any CO2 remaining in the sampling tube to dissipate before zeroing.
• Do not attempt to zero the monitor while the sampling
tube is in the patients airway.
• Do not attempt zeroing if the temperature is not stable.
• Zeroing with CO2 in sampling tube can lead to
inaccurate measurements or other error conditions. If you attempt zeroing while CO2 remains in sampling tube, the time required to zero the monitor may be increased.
Calibration
The monitor has already been calibrated before leaving factory. User can directly apply it to measuring in normal conditions, to the exclusion of the below conditions.
For Sidestream CO2 module, please carry out gain calibration and manual offset calibration, when the following conditions happened:
——The module has been used for between half a year and
one year.
——The accuracy of CO
2
reading has been doubted by clinical
physician.
LifeVet CP User Manual
5-12
——After the latest calibration, atmospheric pressure or height
above sea level varies evidently.
Caution: User may only calibrate the device under the instruction of the technical personnel authorized by manufacturer. Moreover, incorrect calibrating procedure may result in incorrect reading.
Removing Exhaust Gases from the
System
:
Warning:
When using the SidestreamCO2 measurement on patients who are receiving or have recently received anesthetics, connect the outlet to a scavenging system, or to the anesthesia machine/ventilator, to avoid exposing medical staff to anesthetics.
This monitor removes exhaust gas to outside directly.
LifeVet CP User Manual
6-1
Chapter 6 Measuring SpO2
Introduction
The measurement of oxygen saturation of arterial blood (also known as pulse oxygen saturation, usually shortened as SpO2) adopts the principles of light spectra and volume tracing. The LED emits lights with two specific wavelengths, which are selectively absorbed by oxygenated hemoglobin and deoxyhemoglobin. The optical receptor measures the changes in the light intensity after the light passes the capillary network and estimates the ratio of oxygenated hemoglobin and the total hemoglobin.
oxygenated hemoglobin
SpO2 % =
x 100%
oxyhemoglobin + deoxyhemoglobin
Wavelengths of the light emitted by the pulse oximeter probe are nominally 660nm for red LED and 940nm for infrared LED.
LifeVet CP User Manual
6-2
Safety Information
:
Warning :
• Use only SpO2 sensors specified in this manual. Follow
the SpO2 sensors instructions for use and adhere to all warnings and cautions.
• When a trend toward patient deoxygenation is
indicated, blood samples should be analyzed by a laboratory co-oximeter to completely understand the patient’s conditions.
• Do not use the monitor and the SpO2 sensor during
magnetic resonance imaging (MRI). Induced current could cause burns.
• Prolonged continuous monitoring may increase the
risk of unexpected changes in skin characteristics, such as irritation, reddening, blistering or burns. Inspect the sensor site every two hours and move the sensor if the skin quality changes. For cats, or patients with poor peripheral blood circulation or sensitive skin, inspect the sensor site more frequently.
• Measurements and pulse signals can be affected by
certain environmental conditions, sensor application
LifeVet CP User Manual
6-3
errors, and certain patient conditions. See the appropriate sections of this manual for specific safety information.
• Check the SpO2 sensor and its package for any sign of
damage before use. Do not use the sensor if any damage is detected.
• When disposing the disposable SpO2 probe or useless
SpO2 probe, please observe all local, state, and federal regulations that relate to the disposal of this products or similar products.
Caution: In case it is necessary to add a clip to fix the
fingertip sensor, the cable instead of the sensor itself should be clipped. Please note that the cable of sensor should not be pulled with force.
Q?
Note:
• The pleth wave is not equal to the intensity of PR
signal.
• The monitor does not provide automatic self-
LifeVet CP User Manual
6-4
examination alarm signal and the operator has to use SpO2 simulator for self-examination.
Monitoring Procedure
1. Selecting SpO2 Sensor
Depending on the patient category, weight and application site, you can select the SpO2 sensor as required.
2. Connecting SpO2 Sensor
Plug the SpO2 sensor cable into the SpO2 connector on the measurement module.
3. Applying SpO2 Sensor
Clean the application site, such as colored nail polish, and apply the sensor to the patient
Place the sensor on the tongue or ear of the animal. The preferred sensor application site for dogs, cats and horses is on the tongue with the optical components of the sensor positioned at the center of the tongue. Alternatively, the sensor and the clip may be applied to the lip, toes, ear, foreskin or vulva of the animal.
:
Warning:
• Do not use the SpO2 sensor on a limb where the NIBP
LifeVet CP User Manual
6-5
cuff is applied. This may result in inaccurate SpO2 reading due to blocked blood flow during cuff inflation.
• Do not conduct SpO2 measurement on the finger
smeared with nail polish, otherwise unreliable measurement results might be produced.
• When using finger sensor, make sure the nail faces to
the light window.
SpO2 Display
• Parameter Display
1 5
2
4
3
Fig 6-1 SpO2 Parameter
1. SpO2 label 2. High alarm limit of SpO
2
3. Low alarm limit of SpO
2
4. SpO2 value
5. SpO2 unit
LifeVet CP User Manual
6-6
• Waveform Display
Fig 6-2 SpO2 Volume Curve
PR Display
1
5
2
4
3
Fig 6-3 PR Display
1. PR label 2. High alarm limit of PR
3. Low alarm limit of PR 4. PR value
5. PR unit
LifeVet CP User Manual
6-7
Caution: During the monitoring of HR and PR, displaying of HR has priority. That is PR will be displayed only when there isn’t HR monitoring.
Setting SpO2
Select Menu】→【SpO2 Setup】, enter into SpO2 Setup
interface.
Fig 6-4 SpO2 Setup
LifeVet CP User Manual
6-8
Setting SpO2 Alarm
Click the right of Alarm】, you can set alarm level of SpO
2
and PR, you can select Mid, High
Setting SpO2 Alarm Limits
Click the right of Uplimit】or Downlimit】, you can set
the SpO2 uplimit and downlimit. Attention: The high alarm limit should greater than the lower one.
Setting Scan Speed
Click the right of 【Speed, you can select “6.25 mm/s
12.5mm/s, 25 mm/s”.
Setting Average Time
Click the right ofAvg Time】, you can select 4s, 8s, 16s”.
Setting QRS Volume
Click the right of【QRS Vol., you can select Off, High,
Mid, Low”.
Setting Desat Limit
SpO2 desat means when SpO2 measuring value is lower than
LifeVet CP User Manual
6-9
the desat limit, a high physiological alarm will be trigged. Its setting is as follows. 1 SelectMenu】→【System】→【Maintenance】, enter the
required password. 2 Select Machine Mainten】→【SpO2 Setup】→【Desat
limit】, click the right of Desat limitto set its value.
LifeVet CP User Manual
7-1
Chapter 7 Trend Review
Introduction
Select Menu】→【Trendto enter trend reviewing
window. In the window, you can review CO2, RR, SpO2 and PR data stored before.
Review Interface
Fig 7-1 Review Interface
LifeVet CP User Manual
7-2
If the trend date is not only one page, you can turn pages
by the next/return button.
Review Setup
Click the right of IDto select patient’s ID, you can review patient’s trend review by selecting different ID.
Fig 7-2 ID Review Interface
Click the right of More】on the top of review interface, the
drop-down window shown as following:
LifeVet CP User Manual
7-3
Fig 7-3 The drop-down Window of “More”
You can set Save time, Delete, Delete all,
Transmission】in this interface.
• Save time: To adjust recording time, you can select10s,
30s, 1min, 2 min, 5min, 10min”.
• Delete: To delete trend data of the selected ID No.
• Delete all: To delete trend data of all patients.
LifeVet CP User Manual
8-1
Chapter 8 Battery
Introduction
A rechargeable and maintenance-free battery is designed for Patient Monitor, which enables continuous working when AC power off.
When r a lithium ion battery is used, the battery icon
indicates the battery status as follows:
1. Indicates that the power of the battery is full
2. Indicates that the power of the battery is 3 grids left
3. Indicates that the power of the battery is 2 grids left
4. Indicates that the power of the battery is 1 grid left
5. Indicates that the battery is almost depleted. Battery power supply can only last for a period of time. If
the voltage of batteries is too low, an alarm of Battery Low will be triggered. Please insert the monitor to battery charger to charge the battery. The monitor will be switched off automatically 10 minutes after the first Battery Low alarm is given.
LifeVet CP User Manual
8-2
Charging the Battery
To charge the battery:
1. Connect the Micro USB in power adapter,
2. Connect the other connector of Micro USB in the monitor, and plug the adapter into the AC mains,
3. The indicating lamp on the monitor is on to indicate that the battery is in charge,
4. When the battery charging indicating lamp on the monitor turns off, the battery is fully charged.
Optimizing Battery Performance
A battery needs at least two optimizing cycles when it is put into use for the first time. A battery cycle is one complete, uninterrupted charge of the battery, followed by a complete, uninterrupted discharge of the battery. A battery should be conditioned regularly to maintain its useful life. Condition a battery once when it is used or stored for two months, or when its run time becomes noticeably shorter.
To optimize a battery, follow this procedure:
1. Disconnect the monitor from the patient and stop all
monitoring and measuring procedures.
LifeVet CP User Manual
8-3
2. Place the monitor in the charger stand and connect the AC
mains. Allow the battery to be charged uninterruptedly for
above 4 hours.
3. Remove the AC mains and allow the monitor to run from
the battery until it shuts off.
4. Replace the monitor in the charger stand and connect the
AC mains. Allow the battery to be charged uninterruptedly
for above 4 hours.
5. The optimizing of the battery is over.
Checking the Lithium Battery
The performance of a battery may deteriorate over time. To check the performance of a battery, follow this procedure:
1. Disconnect the monitor from the patient and stop all
monitoring and measuring procedures.
2. Place the monitor in the charger stand and connect the AC
mains. Allow the battery to be charged uninterruptedly for above 4 hours.
3. Disconnect AC mains and allow the monitor to run on the
battery until it shuts off.
4. The operating time of a battery reflects its performance
directly.
LifeVet CP User Manual
8-4
Disposing of the Batteries
Batteries that are damaged or depleted should be replaced and discarded properly. Dispose of used batteries according to local regulations.
:
Warning: Do not disassemble batteries, or dispose of
them in fire, or cause them to short circuit. They may ignite, explode, or leak, causing personal injury.
LifeVet CP User Manual
9-1
Chapter 9 Maintenance and Cleaning
Introduction
Keep your equipment and accessories free of dust and dirt. To avoid damage to the equipment, follow these rules:
1. Always dilute according the manufacturers instructions
or use lowest possible concentration.
2. Do not immerse part of the equipment in the liquid.
3. Do not pour liquid onto the equipment or accessories.
4. Do not allow liquid to enter the case.
5. Never use abrasive materials (such as steel wool or silver
polish), or erosive cleaners (such as acetone or acetone­based cleaners).
:
Warning:
• Be sure to shut down the system and disconnect all
power cables from the outlets before cleaning the equipment.
• For optimal performance, product service should be
performed only by qualified service personnel
LifeVet CP User Manual
9-2
Caution: If you spill liquid into the equipment of accessories, connect you service personal or us.
Seasonal Safety Checking
Q?
Note: To ensure the performance and safety of
equipment, it must be checked after using 1 year. When check the equipment, please contact professional technology engineers.
Please clean the plug of power cord at least once a year. Too
much dust on plug may cause the fire.
The following safety checks should be performed at least every 12 months by a qualified person who has adequate training, knowledge, and practical experience to perform these tests.
The data should be recorded in an equipment log. If the device is not functioning properly or fails any of the following tests, the device has to be repaired. Inspect the equipment and accessories for mechanical and functional damage.
Inspect the safety relevant labels for legibility. Verify that the device functions properly as described in the
LifeVet CP User Manual
9-3
instructions for use. Test the earth leakage current according IEC 60601-1 Limit:
NC 500μA, SFC: 1000μA. Test the enclosure leakage current according to IEC 60601-1: Limit: NC 100μA, SFC: 500μA.
Test the patient leakage current (normal operation) according IEC 60601-1 Limit: type CF: for a.c.: 10μA, for d.c.: 10μA.
Test the patient leakage current under single fault condition according IEC 60601-1 Limit: type CF: for a.c.: 50μA, for d.c.: 50μA.
Test the patient leakage current Mains voltage on applied part: According IEC 60601-1: Limit: type CF: for a.c.: 50uA.
:
Warning: No use-serviceable parts inside, before
servicing to authorized representative or manufacturer.
LifeVet CP User Manual
9-4
Cleaning the Monitor
1. Common detergent and non-corrosive disinfectant used in
hospital can be applied to clean monitor, however you must be aware that many kinds of detergents must be diluted prior to utilization, and please use it according to the instruction of detergent manufacturer.
2. Avoid the use of alcohols, amino or acetonyl detergent.
3. The enclosure and screen of monitor shall be free of dust, and
they can be wiped with lint-free soft cloth or sponge soaked in detergent. While cleaning, be careful and do not spill liquid onto the instrument and keep any liquid out of it. When wiping the side panel of monitor, you must be especially careful to keep water out of all kinds of cable and outlet on the panel.
4. Do not use abrasive material including wire brush or metal
brightener during cleaning because this material will damage the panel and monitor screen.
5. Do not submerge the monitor in liquid.
6. While cable or plug of attachment accidentally gets wet,
please rinse it with distilled water or deionized water and dry it in a warm environment for several hours until it is completely dry. Never use a wet or humid cable/plug!
LifeVet CP User Manual
9-5
Cleaning SpO2 Sensor
1. The casing of the sensor and light tube can be cleaned with swab or non-velvet soft cloth dipped with medical alcohol.
2. The sensor cable can be cleaned or sterilized with Hydrogen Peroxide 3% or isopropyl alcohol 70%.
3. It is forbidden to put the monitor in high-pressure containers and put the sensor directly in liquid.
:
Warning: Do not reuse or disinfect the disposable
SpO2 sensor.
Disposal
Dispose of the monitor in accordance with local environment and waste disposal laws and regulations. For the disposal of SpO2 sensor and ECG cable, follow local regulations regarding disposal of hospital waste.
LifeVet CP User Manual
10-1
Chapter 10 Accessories
:
Warning:
• Use only accessories specified in this manual. Using
other accessories may cause damage to the monitor.
• Disposable accessories are designed for single-patient
use only. Reuse of them may cause a risk of contamination and affect the measurement accuracy.
• Check the accessories and their packages for any sign
of damage. Do not use them if any damage is detected.
Type
Mode
PN
CO2
CO2 Filter
15-100-0184
CO2 Tube
15-100-0035
3-Way Joint
15-100-0037
Horse/ Dog nose sampling
tube
15-100-0187
SpO2 sensor
usableHorse
15-100-0013
usableDog
15-100-0014
usableCat
15-100-0015
LifeVet CP User Manual
1
Appendix A Product Specifications
Safety Specifications
CFDA classification
II
CE classification
IIb
Type of protection against electric shock
II, with internal power or external power device
Degree of protection against electric shock
CF
Degree of protection against hazards of explosion
Ordinary equipment, without protection against hazards of explosion
Degree of protection against ingress of liquid
IPX1 Equipment type
Handheld
Mode of operation
Continuous
Physical Specifications
Mainframe weight
500g (full configuration, including the batteries)
Mainframe size
142mm(W)×78mm(H)×36mm(D)
LifeVet CP User Manual
2
Environmental Specifications
Temperature Operating: 5 to +40℃;
Storage: -20 to +55℃;
Atmospheric pressure
Operating: 860hPa to 1060hPa
Storage: 500hPa to 1060hPa
Humidity Operating: 15% to 85%(non condensing)
Storage: 10% to 93%(non condensing)
Charging Specifications
A4.1 Charger
Micro USB
Charge
Power adapter
Input: AC 100240 Output: DC 5V/2A
A4.2 Battery Requirement
Type
Built-in lithium battery
Voltage
3.7V
capacity
4800mAH
Charging time
34h
Run time
>18h
LifeVet CP User Manual
3
Hardware Specifications
Display
Size
4.3inch
Resolution
480*272
Touch
Resistive touch
Autorotation Direction
four direction
Indicating Lamp
Alarm indicating
1 (Yellow/Red), on the top of screen
Battery charging indicating lamp
1 (orange When charged, it lights orange. When fully charged or not charged, it doesnt light.
Audio Indicating
Speaker Gives audible alarm, button tone and
QRS tone Supports Pitch Tone and multi-level volume; Alarm tones meet the requirement of IEC
LifeVet CP User Manual
4
Patient ID
196 Display way
Trend tabular
SpO2 trend interval
10s 30s 1min 2min 5min 10min
Storage
Save when power down
60601-1-8.
Alarm pressure
45 dB to 85 dB, Testing place is 1 meter from the tone.
Buttons
Power button
Turn on/off
Shortcut key
Start/Pause CO2 measurement Short press to achieve the above function
long press + power button to achieve calibration of LCD
Data Storage
The changing trends of physiological parameters will be shown in the monitor, you can select optionally PC software, to upload trend review to computer by USB.
LifeVet CP User Manual
5
Capacity
500 groups/patient can be stored (only data, no waveform).
Measurement Specifications
CO2 Specifications
CO2 (Sidestream)
Measureme nt Way
Infrared spectrum
Measureme nt Range
019.7% (0-150mmHg or 0-20.0kPa )
CO2 Accuracy
0%-5.3% (0mmHg-40mmHg),±0.3% (±2mm Hg)
5.4%-9.2% (41mmHg-70mmHg), ±5% of reading
9.3%-13.2% (71mmHg-100mmHg);±8% of reading
13.3%-19.7% (101mmHg-150mmHg), ±10% of reading.
CO2 Resolution
0.1mmHg
Gas Flow Rate
60~80ml/min
LifeVet CP User Manual
6
Unit
%, mmHg, kPa
Measureme nt Range of awRR
3~150 rpm
Measuring accuracy of RR
±1% or ± 1 rpm, whichever is greater Response Time
3s
Calibration
Offset calibration, auto/ manual; gain calibration
CO2 Alarm Range
019.7%( 0-150mmHg or 0-20.0kPa ) high/low limit can be adjusted continuously
Alarm Indication
Blinking display of the data and parameters, text prompts, Three levels of alarming: sound-light alarming, alarming with blinked data and parameters, and that with text prompts.
SpO2
SpO
2
Measurement
0~100%
LifeVet CP User Manual
7
range Resolution
1%
Accuracy
70~100%± 2%
Alarm
Select the high and low alarm limit of SpO
2
Sensor
Pulse oximetry sensors contain LEDs that emit red light at a wavelength of approximately 660 nm and infrared light at a wavelength of approximately 905 nm. The total optical output power of the sensor LEDs is less than 15 mW. This information may be useful to clinicians, such as those performing photodynamic therapy.
Data update period
13s
Anti-interference
Anti-motion interference Anti-electrotome interference
Resisting low perfusion ability
With powerful ability of resisting low perfusion, PR amplitude can reach to
0.2% with value of SpO2 displaying.
LifeVet CP User Manual
8
Pitch Tone
with
PR
Measuremen t range
25 bpm ~250 bpm resolution
1 bpm
accuracy
±1% or ± 1 bpm, whichever is greater
Alarm
Select the high and low alarm limit of PR
LifeVet CP User Manual
9
Appendix B Factory Defaults
This section lists the most important factory default settings. These settings can be adjusted and you can load the factory defaults if you need.
Alarm Setup
Alarm Setup
Factory Default
Alarm volume
Medium
Minimum alarm volume
Low SpO2 Alarm Level
Medium
CO2 Alarm Level
Medium
System Setup
System setup
Factory Default
QRS volume
medium
Brightness
3
Scan speed
12.5mm/s
LifeVet CP User Manual
10
CO2 Setup
CO2 setup
Horse
Dog
Cat
EtCO2 High Limit
62 mmHg
62 mmHg
62 mmHg
EtCO2 Low Limit
21 mmHg
21 mmHg
21 mmHg
RR
Horse
Dog
Cat
RR High Limit
35 rpm
40 rpm
40 rpm
RR Low Limit
5 rpm
8 rpm
8 rpm
SpO2 Setup
SpO2 Setup
Horse
Dog
Cat
SpO2 High Limit
100 %
100 %
100 %
SpO2 Low Limit
90 %
90 %
90 %
PR Setup
Horse
Dog
Cat
PR High Limit
50 bpm
160 bmp
200 bpm
PR Low Limit
30 bpm
70 bpm
90 bpm
Trend Setup
Trend Setup
Factory Default
Interval
30s
LifeVet CP User Manual
11
Appendix C Alarm Message
This section lists some important alarm message. In the tables below, “*” means the alarm level is user-adjustable.
Physiological Alarm
SpO2 Alarm Messages
Cause
Level
SpO2 Too High *
A measurement has risen above the high alarm limit or fallen below the low alarm limit
High、
Medium
SpO2 Too Low *
High
Medium
PR Too High *
High
Medium
PR Too Low *
High
Medium
SpO2 Desat
SpO2 measurement has fallen below the SpO2 desat limit.
High
No Pulse
The pulse signal was too
High
LifeVet CP User Manual
12
weak to be analyzed.
CO2 Alarm Messages
Cause Level
EtCO2 Too High *
A measurement has risen above the high alarm limit or fallen below the low alarm limit.
High
Medium
EtCO2 Too Low *
High
Medium
RR Too High
High
Medium
RR Too Low
High
Medium
Apnea
Resp cant be detected on preset-time.
High
Technical Alarm
Message
Cause
Level
SpO2 Sensor Off
The SpO2 sensor detached the patient or the monitor.
Low
LifeVet CP User Manual
13
CO2 Sensor Off
The CO2 sensor detached the patient or the monitor.
Low
Communication Error
Communication error or test model error.
Low
Battery Low
The battery power is low.
Medium
SpO2 Low Perf
The signal detected is weak.
Medium
CO2 measurement over range
CO2 measurement Over range, need ve rify zero
Low
CO2 sensor error
CO2 sensor error.
Low
CO2 sensor Over Temp
Temperature of the sensor is over the normal working temperature.
High
Check CO
2
Sampling Line
Sampling line is occluded or damaged. Sampling tube is kinked or pinched. Exhaust tube is blocked.
Low
Zero Required
The sensor needs verify zero.
High
CO2 sensor no initialized
CO2 sensor no initialized
Low
LifeVet CP User Manual
14
Prompt Message
Message
Cause
Level
searching
Searching pulse
Prompt Message
SpO2 sensor off
SpO2 sensor may be disconnected from the patient or the monitor.
Zero in Progress
Zeroing is in progress.
CO2 sensor off
Sensor dose not connect with monitor when the monitor is running. After confirming alarm message of sensor off, the alarm message will become prompt message.
LifeVet CP User Manual
15
Appendix D EMC
LifeVet CP User Manual
16
Appendix E Warranty Registration Card
Thank you for purchasing EICKEMEYER® products!
Please complete this card and mail back to EICKEMEYER KG within one week. If you need any support or the defects occur, please feel free to contact us by telephone or fax. Warranty will apply with no charge in the warranty period (exclude accident, misuse, abuse or misapplication). You are also and always welcome to our service center, when you need any special service after warranty. Do not repair the product by any person who is not authorized or trained by EICKEMEYER®.
Product
Model
Serial No.
Contract
Date Installed
Warranty
Name
Address
Contact Person
Tele/fax
Loading...