EFO MFT4 User Manual

MFT4 Multifunction Installation
Tester
- 1 -
User manual
Safety information and explanation of symbols used .............................. 3
Features of the MFT4 ................................................................................... 5
Special Polarity Test Function. .................................................................. 6
Audible tones ............................................................................................... 7
Overview of the switches and LCD ......................................................... 8
Test lead inputs ............................................................................................ 9
Continuity Test Function ........................................................................... 10
Continuity Test Procedure ....................................................................... 10
Lead Nulling ............................................................................................ 10
Hands Free Continuity Testing ................................................................ 11
Insulation Test Function............................................................................ 12
Insulation Test Procedure ........................................................................ 12
Hands Free Insulation Testing ................................................................. 13
Loop Test functions ................................................................................... 14
High Current mode .................................................................................. 14
No Trip Mode ........................................................................................... 14
PFC/PSC................................................................................................. 15
Test lead configuration ............................................................................ 15
Lead configuration for No-Trip testing ..................................................... 16
Lead configuration for High current 2-wire testing ................................... 16
Loop Test Procedures ............................................................................. 17
No Trip Loop test (Zs) ............................................................................. 17
High current test (Ze) .............................................................................. 18
Hands Free Loop testing ......................................................................... 18
RCD Test Function ................................................................ ..................... 19
Ramp test ................................................................................................ 20
Sinusoidal polarity (the 0° or 180° test) ................................................... 20
RCD test procedure ................................................................................ 21
User selected test ................................................................................... 21
30mA Automatic test ............................................................................... 22
Ramp test ................................................................................................ 22
Specifications and tolerances .................................................................. 23
Continuity Test Range Accuracy .............................................................. 23
Insulation Test Range Accuracy .............................................................. 23
Insulation Output Voltage ........................................................................ 23
Loop Test Range Accuracy ...................................................................... 23
RCD Test Range Accuracy ...................................................................... 24
- 2 -
Safety information and explanation of symbols used
Because the MFT4 is a multi function tester used for testing both live and dead circuits there are different safety issues that apply to the individual functions. Before using your MFT4 please read these instructions paying particular attention to the general safety warnings below and those at the start of each section.
Before using the tester check the case and the test leads for damage.
If any damage is noticed the unit should be withdrawn from service and returned to place of purchase for repair.
It is important for safety that only one set of leads can be fitted at a time. In the unlikely event that the interlock cover is damaged the tester should be withdrawn from service and returned to place of purchase for repair.
Caution read this manual for safety information
The Continuity and Insulation functions are rated at 500V Category III
The Loop and RCD functions are rated at 300V Category IV When installing batteries observe correct polarity do not mix old and new
batteries - Dispose of used batteries in accordance with local regulations. Never incinerate batteries.
Do not use this tester in a manner other than that described in this booklet.
To clean the tester wipe with a damp cloth with a mild soap solution taking care not to allow water ingress into the input terminals. Do not use solvents and do not immerse. Allow the tester to fully dry before use.
The MFT4 is fuse protected against damage by accidental connection to an over-voltage supply. The fuse is located inside the battery compartment and can be accessed by removing the two small battery cover retaining screws on the back of the case. Always ensure that test leads are disconnected before removing the battery cover.
- 3 -
The broken fuse indicator on the LCD will flag if the fuse has blown. It must be
MEASUREMENT RANGE
OPERATING RANGE PER EN61557
OTHER CONTINUITY
0.00 Ω - 19.99 kΩ.
0.1 Ω - 9.99kΩ.
IN>200mA Uq < 7V
INSULATION
0.00 MΩ - 1999 MΩ
.
0.1 MΩ - 1990 MΩ.
IN = 1mA
LOOP HI-I
0.01 Ω - 500 Ω
1.04 Ω - 500Ω.
230V 50Hz
LOOP NO-TRIP
0.01 Ω - 500 Ω.
1.04 Ω - 500Ω.
230V 50Hz
RCD TRIP TIME
5 ms – 1999 ms
38 ms – 1999ms.
replaced with the correct type: Fuse type: F 500mA fast blow ceramic 600V.
The enclosure is double insulated
Protected against over voltage to 550V
For safety reasons the tester is shipped without batteries fitted. To install batteries remove the two small crosshead screws on the back of the instrument that retain the battery cover and fit four Alkaline batteries type AA / LR6 in accordance with the polarity shown.
The MFT4 complies fully with the requirements of EN61010. The following table details the operating ranges for the individual functions
compliant with the performance requirements of EN61557.
- 4 -
Features of the MFT4
The MFT4 is packed with design features that maximise both convenience and safety. These include:
Large display
To give the clearest results the MFT4 uses a large auto-backlit LCD which makes reading the test results easy even when used in poorly lit areas.
Auto shut down
To preserve battery life when not in use the MFT4 incorporates an Auto-Off function that powers the unit down after three minutes of inactivity. To resume use after an Auto shut down a single press of any of the function buttons will power up the unit.
Battery check
The first position either side of the off position of the rotary selector switch is a battery check function.
Extended battery life
For simplicity the tester is powered by just four conventional AA (LR6) alkaline batteries. The MFT4 has much lower power consumption than most testers and therefore gives excellent battery life.
In addition to the battery status indicator that shows on the LCD, when the battery power is becoming very low the Red warning LED will light to show that imminent replacement is necessary. Always use Alkaline rather than zinc carbon batteries.
Easy to locate
The test lead inputs are located on the top of the case allowing the tester to stand vertically or be laid flat. Alternatively the unit can be carried by the supplied neck strap.
Hands Free
Most of the test functions can utilise the Hands-Free mode in which the tester is primed to automatically start the test as soon as the probes are connected to a circuit, thereby leaving your hands free to hold the test probes.
Socket wiring check
To protect both the user and the instrument against harm caused by accidental connection to an incorrectly wired supply the tester will automatically check the polarity upon connection to a live supply. If the wiring has been incorrectly connected testing will be inhibited an alarm will sound with flashing LED.
- 5 -
Special polarity test function.
It is a little known fact that a system can be reverse wired with Line (Phase) to earth/neutral and earth/neutral to Line (Phase). The sockets will all work and conventional loop testers will show and test that everything is correct despite this very dangerous wiring condition. Although extremely rare, this miss wire condition can exist so if your test shows this fault do not proceed – if in any doubt advise your customer to contact their supply company immediately.
Correct Polarity
Reversed Polarity
- 6 -
Audible tones
Danger A rising siren type
alarm
In the event of a potentially dangerous situation such as connecting to a live supply when configured for insulation testing. Will be accompanied by the Red Voltage/Polarity warning LED flashing.
Warning A continuous 2 tone
alarm
An unsuitable supply configuration such as a mains supply with incorrect polarity or having the leads connected wrongly will be accompanied by the Red Voltage/Polarity warning LED flashing.
Wait-Test in progress A steady beeping
sound
Emitted whilst a measurement is in progress. The same tone is sounded when used in Hands free mode to indicate that continual measurement is being made
Test completed A single beep
Sounded upon completion of a measurement to indicate that the result is being displayed
Alert A short 2 tone alarm
Sounded when a test returns a result that is likely to be regarded as a failure e.g. An insulation test that gives a result of less than 2 MΩ
A simple selection of audible tones is used to supplement the visual display. These help the user by providing intuitive feedback during testing. In addition to warning about dangerous or unstable supply conditions they provide a very quick confirmation that the measurement process is taking place and, upon completion of the test, a warning if the results are likely to be regarded as a failure. The meaning of the tone for each individual function is covered in detail in the relevant section. Generally however there are five types of tone emitted.
- 7 -
Test lead inputs
RCD Muliplier
PFC-PSC / RCD Recall
Polarity test touch pad area
Hands free Button
Test Button
Function Selector
Continuity Null Button
Overview of the switches and LCD
The Primary display of the large LCD shows the result of the test being conducted. At the same time a secondary display area shows supporting information e.g. for an insulation test the main display shows the resistance of the insulation whilst the secondary display confirms the test voltage applied.
- 8 -
Loading...
+ 16 hidden pages