The robust multifunctional hand-held meets the
highest requirements and comes with a wide range
of accurate probes that fit various applications.
Use touch-screen navigation to show, up to three
measurement values simultaneously on the capacitive
TFT display.
A total of 23 measurands (vary according to probe) is
available, including:
• relative humidity RH
• temperature T
• dew point temperature Td
• absolute humidity dv
• mixing ratio r
• air velocity v
• volumetric flow
• air pressure p
• CO
Store measurement values in the internal memory of the device and use the free SmartGraph3 software to
manage data on your PC. The optional carrying case accommodates the hand-held and probes, as well as
accessories.
ppm
2
OMNIPORT 30
with CO2 probewith RH/T probe
Data-Logging & Data-Management
The Omniport 30 offers both continuous and single-point datalogging. The measurement values of up to three channels are
stored simultaneously, together with a time and date stamp, in
the internal memory of the device.
Stored measurement data, as well as the minimum, maximum,
average and standard deviation values can be shown directly on
the display. Scroll through a recorded measurement and switch
between graphs by using the control cross.
Data management is easy and intuitive by using the free SmartGraph3 software to create graphs that contain
measurement channels of interest for better data analysis. Data can be exported in .csv format and then
imported into EXCEL for further processing.
Features
Data logging
Internal memory for 2 million measured values
23 physical quantities
Capacitive TFT touch screen
Displays measurands simultaneously
Real-time HOLD / MIN / MAX / AVG readout
Integrated air pressure sensor
User friendly operation
Free data management software
62
v1.5 / Modification rights reserved
OMNIPORT 30
OMNIPORT 30
Page 2
Technical Data
General
Power supply 4 x Alkaline LR6 AA batteries, 1.5 V (not in the scope of supply)Optional power supply 5V DC via USB (cable included)
Temperature range operating: handheld and handle of sensing probe: 0...50°C
storage: -20...60°C (-4...140°F)
Internal memory for approx. 2 million measured values
Housing / protection class ABS / IP40
Dimensions (HxWxD) 170 x 62 x 34 mm
(6.69 x 2.44 x 1.34”)
Weight ca. 205g (0.45 lbs)
Display TFT display, 54 x 41 mm (2.13 x 1.61”), illuminated
CE compatibility Hand-held: EN61000-6-2:2005 EN61000-6-3:2007
Logprobe: EN61326-1:2013 EN61326-2-3:2013
Integrated air pressure sensor
Measuring range 800 to 1100 mbar (complete accuracy)
Accuracy max. ± 0.5 mbar (at 25 °C, 1013.25 mbar)
Long term stability typ. -1 mbar/year
Working range: 0.2...20m/s (40...4000ft/min) / 0...50°C (32...122°F)
(0.7”)
Accuracy:
± (0.2m/s / 39ft/min + 3% of m.v.)
± 1°C
(0.5”)
∅12
Response time τ
Order code: LOGPROBE65
: ≤ 1.5 sec.
90
CO2 PROBE
LOGPROBE 802/805/810 - CO2 probe
96
(3.8”)
18.5
(0.7“)
Working range: 0...2000 / 5000 / 10000ppm
Technology: dual wave NDIR with autocalibration
Accuracy
and 1013mbar0...5000ppm: < ± (50ppm +3% of measured value)
(77°F and 14.69psi) 0...10000ppm: < ± (100ppm +5% of measured value)
at 25°C 0...2000ppm: < ± (50ppm +2% of measured value)
Warm-up time: 3 min.
Temperature dependency: typ. 1ppm CO
Order code: 0...2000ppm LOGPROBE802
0...5000ppm LOGPROBE805
0...10000ppm LOGPROBE810
Carrying Case
The optional carrying case protects the handheld, probes and accessories
during transport and storage. It helps to avoid mechanical damage, as well
as the contamination of the sensors, which is essential for the optimal long
term performance of the instrument.
/ ±1.8°F (0...50°C / 32...122°F)
/°C (-20...45°C) (-4...113°F)
2
Protective Cover
The optional cover protects the handheld device during usage in dirty or
oily environments. The integrated magnetic plate allows an easy temporary
fixing of the handheld onto a metallic surface on site, while the hang-on
straps helps keep your hands free while taking measurements.
64
v1.5 / Modification rights reserved
HA040906
HA040907
OMNIPORT 30
Page 4
Ordering Guide
POSITION 1 BASIC DEVICE
POSITION 2 CABLE
POSITION 3 PROBES
Humidity / Temperature
Air Velocity
CO
POSITION 4 CARRYING CASE
1) Directly connected to device, cable is optional
2) Cable is necessary
Order Example
2m (6.6ft)HA010813
LOGPROBE 20 - Response time ≤30 sec. LOGPROBE 20 - Response time ≤10 sec. LOGPROBE 16 - HVAC probeLOGPROBE 31 - high temperature probeLOGPROBE 30 - confined space probe