E+E HUMOR20 Operating Manual

HUMOR 20 +
Automatic Calibration Module
HUMIDITY CALIBRATOR
BA_Humor20_e // v3.3 // technical data are subject to change // 302252
MANUAL
Hardware + Software
Table of contents
2.1 HUMOR 20 4
2.2 Automatic calibration module (optional) 4
3.1 General 5
3.2 HUMOR 20 5
5.1 HUMOR 20 7
5.2 Automatic calibration module (optional) 7
6.1 Setup 7
6.2 Starting up HUMOR 20 8
6.3 Starting up HUMOR 20 with automatic calibration module 9
6.4 Verifying the 100%RH Point 11
7. CALIBRATION AND ADJUSTMENT PROCEDURE 12
7.1 Cubical transmitters (room design) with HUMOR 20 12
7.2 Transmitter with sensor probe with HUMOR 20 12
7.3 Transmitter EE33 Model J with HUMOR 20 13
7.4 Analogue Output HUMOR 20 14
7.5 Calibration with the automatic calibration module 14
7.6 Power failure during the calibration procedure 14
7.7 Ending the calibration or adjustment procedure 14
8. HUMOR 20 SOFTWARE 15
8.1 Installation the software 15
8.2 Configuration mode control panel 16
8.3 Menu bar 16
8.3.1 Operation 16
8.3.2 Adjustment / Calibration 22
8.3.3 Pressure Compensation 23
8.4 Measurements 24
8.4.1 Start screen 24
8.4.2 "Test unit" configuration field 24
8.4.3 "Measurements" configuration field 26
8.5 Fixed RH Value (only available when using an automatic calibration module) 27
9. MAINTENANCE 27
9.1 Re-fill water (at error message - waterlevel low) 27
9.2 Drain water (at error message - waterlevel high or long idle periods) 27
9.3 Cleaning 27
10.1 Humidity - display flashes 28
10.2 Warning: OUT OF SPEC 28
10.3 Warning: WATERLEVEL HIGH 28
10.4 Warning: WATERLEVEL LOW 29
10.5 Error message: heat defect 29
10.6 Error message: pressure excess 29
10.7 Humidity - display is incorrect 29
10.8 Stabilisation time too long 29
10.9 Leakage in filter set 30
10.10 Electronic defect - replace electronics 30
10.11 Fuse defect - Replacing the fuse (Humor20 + Autokit) 30
12. ACCESSORIES 32
12.1 Oil-free compressor 32
12.2 Covers for measuring chamber 32
12.3 Calibration certificate 32
12.4 Automatic calibration module 32
2
LIMITED LIABILITY
E+E Elektronik® shall not be held liable for any damage or consequential damage (for example, but not restricted to, loss of earnings, interruption of business, loss of information and data or any other financial losses) resulting from the setup, use or impossibility of use of an E+E Elektronik® software product and any associated support services or non-performance of support services.
E+E Elektronik® Ges.m.b.H. is not responsible for the completeness of this publication and shall not be liable for the products described herein in the event of improper handling. This publication may contain technical inaccuracies or typographical errors. The information con­tained herein is revised regularly. These changes will be implemented in later editions. The products described herein may be improved or changed at any time.
© Copyright E+E Elektronik GmbH // All rights reserved // Subject to change without notice
1. FOREWORD
E+E Elektronik® Ges.m.b.H. developed this device to accurately generate the relative humidity (Uw). As such, this professional tool represents the manufacturer's current status of knowledge with regard to development, design and production. The manufacturer is confident that the humidity calibrator HUMOR 20 meets all the requirements and expectations that you, the buyer, demand of a new acquisition. If the device is handled properly and serviced regularly it should work reliably for several years.
This manual forms part of the scope of supply and serves to ensure proper handling and optimal performance of the device. Therefore, make sure you read the manual before you take the device into operation. The manual is also to be brought to the attention of any person involved in transport, setup, operation, mainte­nance or repair.
This manual must not be used for competitive purposes without our written consent and must not be passed on to third parties. Copies for personal use are permitted. All information, technical data and illustrations contained in this manual are based on the informati­on available at the time of publication.
E+E Elektronik® Ges.m.b.H. reserves the right to change the technical data or make other technical modifications at any time and without prior notice without being committed to retrofitting models produced before such a change was made.
General
This manual is part of the scope of supply and serves to ensure the optimum operation and perfor­mance of the device. To ensure correct operation, this manual must be read before the device is taken into operation.
Explanation of symbols
This symbol indicates a safety instruction.
These safety instructions must always be followed. The manufacturer is not liable for any damage resulting from non-compliance. In this case, the user is solely responsible for his actions.
This symbol indicates a note.
These notes should be observed to achieve optimal performance of the equipment.
3
2. SCOPE OF SUPPLY
2.1 HUMOR 20
1
2
3
4
5
6
1 Power supply cable IEC Europe (230V) + power supply cable IEC Northamerica (110V) 2 Water drain pipe with connector 3 Funnel 4 Allen key (10mm / 5 Measuring beaker 6 Measuring and calibration software 7 Face pin wrench 8 Plexiglas cover for room transmitter testing
0.4”)
7
8
9
10
11
12
13
14
9 O-ring for room transmitter 10 Knurled nut 11 Cover for measuring chamber (ordering code HA0202xx) (not inlcuded in the scope of supply HUMOR 20) 12 Fixing bracket for filter set (pre-mounted) 13 Works certificate acc. DIN EN 10204-3.1 14 Filter set with oil separator
2.2 Automatic calibration module (optional)
• Automatic calibration module
• Power supply cable IEC Europe (230V)
• Power supply cable IEC North America (110V)
• RS232 connection cable to HUMOR 20
• Compressed-air connecting hose to HUMOR 20
4
3. SAFETY INSTRUCTIONS
3.1 General
HUMOR 20 is designed based on state-of-the-art technology, which means it will operate reliably if it is kept in good condition and operated and serviced correctly. This device may pose a risk if it is operated improperly by untrained personnel.
Improper use can:
• Injure operating personnel and damage the device itself or other assets belonging to the user.
• Prevent the device from working efficiently and accurately.
Observe the following information to ensure operational safety:
• Only qualified or specially trained staff may operate or perform work on the humidity calibrator. Unauthorised changes or modifications are either not permitted or require the express permission of the manufacturer.
• Read this manual carefully before taking the device into operation. The manual must also be made available to any person involved in transport, setup, operation, mainte­nance or repair (particularly when loaning or selling the device to a third party).
• Only operate the device if it is in good condition. Any fault must be repaired by authori­sed personnel or by an E+E Elektronik sales partner before it is taken into operation again.
3.2 HUMOR 20
• Check whether the supply voltage data on the nameplate matches the local supply voltage. This also applies to the automatic calibration module.
• Only power the humidity calibrator via a grounded power outlet (safety precaution). This also applies to the automatic calibration module.
• HUMOR20 and automatic calibration module should be placed in a way which allows an as easy as possible disconnection from the power supply.
• Before opening the water intake, make sure that the humidity calibrator is no longer under pressure (controller turned to the left as far as it will go; if HUMOR 20 already in service, display shows RH > 90%).
• HUMOR 20 may only be operated with distilled (deionised) water.
• Before connecting the compressed air or nitrogen supply, make sure that the controller is turned to the left as far as it will go.
• The device may only be operated with filtered compressed air with a contamination level of <0.01mg/m³.
• The media supply may not exceed an absolute pressure of 10 bar (pressure gauge and safety value recommended for use in the supply line).
• If a humidity calibrator is filled with water and ready for operation, it must not be tilted by more than 20° during transportation. If this is not possible, HUMOR 20 must be drained completely, see section "9.2 Drain water".
• Once HUMOR 20 has been switched off, it must also be disconnected from the compressed air (because of possible condensation in the lines as the heating is turned off).
5
4. OPERATING PRINCIPLE
Dry air
Air or nitrogen under pressure p1 is fed into chamber 1 and humidified. The developing saturation vapour pressure ews is a direct result of the temperature of the chamber respec­tively the water. Next, the humidified air is expanded to pressure p and
Pressure [mbar]
Dry air
fed into chamber 2. During expansi­on, the vapour pressure ews is reduced in the same ratio as the overall pressure of the air. Therefore,
Vapour pressure
Vapour pressure
the water vapour pressure in chamber 2 is:
e = ews * p/p1
Thanks to the special design of HUMOR 20 with its high thermal conductivity, chamber 1 always has the same temperature as chamber 2.
Consequently, if the temperature is the same, the saturation vapour pressure ews of chamber 1 is identical to that of chamber 2. As a result of the expansion of the humidified air, chamber 2 contains the water vapour partial pressure e. From the definition for relative humidity follows for chamber 2:
rh = e/ews = ews * (p/p1) / ews = p/p1
Hence, the relative humidity in chamber 2 is a direct result of the ratio of the pressures in chambers 1 and 2.
Air p
Pressure regulator
1
Reducing valve
In addition, the relative humidity generated in HUMOR 20 is practi­cally independent of the ambient temperature and only dependent
p , T
e
e
ws
p1 --> p
on the pressures p and p1. The only precondition is that the tem­perature of chambers 1 and 2 and the water in chamber 1 is the same.
Measuring chamber
The relative humidity in chamber 2 can be varied by adjusting the inlet
Water T, p
1
Saturation chamber
pressure p1. During operation, the pressures in chambers 1 and 2 are measured and the actual relative humidity is calculated with the
Schematic diagram of a two-pressure reactor
above equation and shown in the display of the calibrator.
6
5. OPERATING COMPONENTS
5.1 HUMOR 20
Water intake
Measuring chamber
Humidity controller
Carry handle
HUMOR 20 - Back Side
power cable + switch + fuse
Supply voltage + measuring inputs for test units
Pressure supply (compressed air / N2)
output signal test unit
Switch 2: for test unit 1-4 and
measuring chamber temperature
Bracket for filter set
serial interface
socket "analogue output HUMOR 20"
water out
(pre-mounted)
Display
Switch 1:
Filter set
5.2 Automatic calibration module (optional)
Compressed-air connecting hose
Pressure supply (compressed air / N2)
(*) Lights up when power is supplied (**) - Flashes: Adjusted to setpoint
- Lights up: Setpoint reached
6. GENERAL USER INSTRUCTIONS
Green "power" LED (*) Orange "auto" LED (**)
filter set
Bracket for filter set (pre-mounted)
6.1 Setup
Operating range: 10...40°C / 10...80% rH. Do not perform measurements in direct sunlight or while exposed to other external sources of heat. Do not operate HUMOR 20 in potentially-explosive areas or expose it to strong mechanical vibra­tions. If HUMOR 20 is transported when it is ready for operation (filled with distilled water), do not tilt the housing by more than 20°.
7
6.2 Starting up HUMOR 20
HUMOR 20 must acclimatise at least 6 hours on-site at the calibrati­on location.
Apply voltage supply. Plug the power supply cable into the device and a grounded power outlet.
Activate the main switch.
Turn the humidity controller to the left (anti-clockwise) as far as it will go.
Open the water tank. Ignore the warm-up time. Use the Allen key supplied to open the screw plug.
Before opening the water tank, make sure that the humidity calibrator is no longer under pressure. Turn the humidity controller to the left (anti-clockwise) as far as it will go and wait until "RH > 90%" is shown in the display of HUMOR 20.
Fill HUMOR 20 with distilled water. Fill the humidity calibrator with max. 1,300 ml distilled (deionised) water. If the maximum level is exceeded, the warning WATERLEVEL HIGH is shown in the display. If this happens, drain water until the warning disappears (see section "9.2 Drain water").
During the filling process, make sure that no water enters the measuring chamber.
Close the water tank. Use the Allen key supplied to tighten the screw plug.
Wait until the warm-up phase is completed. The warm-up phase will take 20 minutes (display: WARM UP TIME).
8
Attach the oil separator / filter set to the side of HUMOR 20, on the bracket, and establish a connection to the input "Pressure in".
Connect the pressurised (min. 8 bar) compressed air line to the filter set. A supply pressure less than 8 bar can cause leakage at the filter set. (p 10 bar - see safety instructions)
HUMOR 20 is ready for operation.
6.3 Starting up HUMOR 20 with automatic calibration module
Setting up HUMOR 20 with the automatic calibration module.
HUMOR 20 must acclimatise at least 6 hours on-site at the calibrati­on location.
Establish a serial connection between HUMOR 20 and the automatic calibration module. Connect "RS232-Humor" on the automatic calibra­tion module with "RS232" on HUMOR 20. (Max. cable length allowed is 3m. The length of the supplied cable is
1.8m)
Connect the power cable to a grounded power outlet, but do not switch on the devices yet, using the main switch!
Establish a serial connection between the automatic calibration modu­le and the PC/notebook.
First switch on HUMOR 20 using the main switch and then the auto­matic calibration module.
9
Open the water tank. Ignore the warm-up time. Use the Allen key supplied to open the screw plug.
Before opening the water tank, make sure that the humidity calibrator is no longer under pressure. Turn the humidity controller to the left (anti-clockwise) as far as it will go and wait until "RH > 90%" is shown in the display of HUMOR 20.
Fill HUMOR 20 with distilled water. Fill the humidity calibrator with max. 1,300 ml distilled (deionised) water. If the maximum level is exceeded, the warning WATERLEVEL HIGH is shown in the display. If this happens, drain water until the warning disappears (see section "9.2 Drain water").
During the filling process, make sure that no water enters the measu­ring chamber.
Close the water tank. Use the Allen key supplied to tighten the screw plug.
Wait until the warm-up phase is completed. The warm-up phase will take 20 minutes (display: WARM UP TIME).
Establish a compressed-air connection between HUMOR 20 and the automatic calibration module.
Max. allowed hose length is 1m.
Attach the oil separator to the side of the automatic calibration module, on the bracket, and establish a connection to the input "Pressure in".
10
Connect the pressure supply (oil-free compressed air or N2). A supply pressure less than 8 bar can cause leakage at the filter set. (p 10 bar - see safety instructions)
Turn the humidity controller to the right (clockwise) as far as it will go.
HUMOR 20 with the automatic calibration module is ready for operation.
6.4 Verifying the 100%RH Point
To obtain the highest accuracy of the humidity measurement it is recommended to verify the 100 %RH point every 4 weeks during con­tinues operation or each time before putting the Humor 20 into operation.
The verification has to be executed as follows:
• Turn the humidity controller counterclockwise all the way
• Disconnect the air pressure supply to the Humor
• Display reading > 95 %RH
• Open the water inlet
• Open the measuring chamber (one opening is sufficient, see picture)
• Read the display after approx. 5 minutes:
Display reading Final conclusion on performance Required action
99.2% to 100.8%
97.0% to 99.2% and 100.8% to 103.0%
< 97.0% and >103.0%
HUMOR within specifications A high accurate measurement can be
realized
HUMOR outside specifications Adjustment is necessary to obtain the
highest accuracy (see chapter "8.3.2. Calibration").
HUMOR outside specifications Adjustment is not possible. There is a
problem with the humidity measure­ment / indication and the HUMOR should be returned to E+E Elektronik for calibration / repair.
11
Loading...
+ 25 hidden pages