E+E EE99-1 Specifications

EE99-1
OEM - Humidity / Temperature Modules
The EE99-1 OEM - RH/T modules are designed to meet the specific requirements of RH/T monitoring in climate chambers. High-end E+E humidity sensor elements of the HC series and accurate temperature compensation of the humidity reading result in an excellent accuracy over a broad measurement range. The analogue output for relative humidity is 4 - 20mA / 3-wire. The passive temperature output can be connected via 3-wire to an external readout. Easy mounting and service is possible with a plug-in screw terminals block and by push buttons for field calibration.
Operation in heavily polluted and/or corrosive environ­ments is typical for many industrial processes and can lead to drift or damage of the humidity sensor and there-
EE99-1
fore to incorrect measurements. The unique protective coating developed by E+E for the sensing probe means a significant improvement of the long-term stability of the transmitter in very dirty and aggressive environments.
Typical Applications
climate chambers drying chambers
remote sensing probe up to 10m (32.8ft)
accuracy ±2% RH
Features
traceable calibration
working range humidity 0...100% RH
working range temperature -50...180°C
(-58...356°F) / up to 200°C (392°F)
short term passive 3-wire temperature output
easy field calibration
Technical Data
Measured quantities
Relative humidity
Humidity sensor 1) HC1000-400 Working range 0…100% RH
Accuracy
-15...40°C (5...104°F) 90% RH ± (1.3 + 0.3%*mv) % RH
-15...40°C
-25...70°C
-50...180°C Output signal 4 - 20mA (3-wire)
Response time with filter at 20°C
Temperature
Temperature sensor element Working range -50...180°C
General Data
Supply voltage 10 - 35V DC or 10 - 28V AC Load resistor for 4 - 20 mA output 10 - 35V DC R
10 - 28V AC R Current consumption for DC supply < 32mA for AC supply < 60mA Working temperature range electronics -40…60°C (-40...140°F) Storage temperature range -40…60°C (-40...140°F) Electrical connection pluggable screw terminals up to max. 1.5mm² (AWG 16) Sensor protection stainless steel grid filter
Electromagnetic compatibility Designed for installment in and with other equipment (OEM) Measurements according to EN61000-4-3 and EN61000-4-6
FCC Part15 ClassB ICES-003 ClassB
1) Refer to the working range of the humidity sensor 3) max. power dissipation 1mW
2) The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement).
50
2)
(including hysteresis, non-linearity and repeatability, traceable to intern. standards, administrated by NIST, PTB, BEV...)
(5...104°F) >90% RH ± 2.3% RH (-13...158°F) ± (1.4 + 1%*mv) % RH
(-40...356°F) ± (1.5 + 1.5%*mv) % RH
(68°F) / t90 < 15 sec.
3)
Pt100 resp. Pt1000 (class A, DIN EN 60751) see Ordering Guide
(-58...356°F) / up to 200°C (392°F) short term
Uv - 5V
< [] (max. 350 Ω)
L
0.02 A
< 350
L
eff
EE99-1
v1.7 / Modification rights reserved
EE99-1
Mounting Dimensions (mm)
(2.8”)
71.8
(2.3”)
58.3
3.5
Connection Diagram
supply voltage
10...35V DC
10...28V AC
89.7
(3.5”)
81.4
(3.2”)
(0.1”)
push buttons for 1 or 2 point
humidity calibration
15
(0.6”)
39
(1.5”)
(7.9”)
200
Probe material: high-tech plastic PPS-GF40
T-sensor passive
3-wire
(2.6”)
/ 65
Ordering Guide
MODEL OUTPUT T-SENSOR VERSION FILTER CABLE LENGTH
Humidity + Temperature
passive (FP)
EE99-1-
PROBE LENGTH
200mm (7.9”) (5) 65mm
(2.6”) (2)
Accessories
Metal grid lter (HA010108)
4 - 20 mA (6) Pt100 DIN A
Pt1000 DIN A
(A) remote sensing probe (D) stainless steel grid filter (8) 2m (6.6ft) (02)
(C) 5m (16.4ft) (05)
10m
(32.8ft) (10)
Order Example
EE99-1-FP6AD8025
Model: Humidity + Temperature passive Output: 4 - 20mA T-Sensor: Pt100 DIN A Version: remote sensing probe Filter: stainless steel grid filter Cable length: 2m Probe length: 200mm (7.9”)
(6.6ft)
EE99-1
v1.7 / Modification rights reserved
51