E+E EE870 Operating Manual

Page 1
EE870
Modular CO2 Transmitter
BA_EE870_e // v1.5 // Technische Änderungen vorbehalten
Manual
Page 2
E+E Elektronik® Ges.m.b.H. doesn't accept warranty and liability claims neither upon this publication nor in case of improper treatment of the described products.
The document may contain technical inaccuracies and typographical errors. The content will be revised on a regular basis. These changes will be implemented in later versions. The described products can be improved and changed at any time without prior notice.
© Copyright E+E Elektronik
Ges.m.b.H.
All rights reserved.
2
Page 3
Table of contents
GENERAL ................................................................................................................................................... 3
IMPORTANT NOTES .................................................................................................................................. 4
CONNECTIONS / MOUNTING ................................................................................................................... 4
TECHNICAL DATA ..................................................................................................................................... 5
DIMENSIONS .............................................................................................................................................. 5
HARDWARE SETTINGS ............................................................................................................................ 5
MODBUS MAP ........................................................................................................................................... 6
OPERATION OUTDOORS ......................................................................................................................... 7
ACCESSORIES .......................................................................................................................................... 7
SCOPE OF SUPPLY ................................................................................................................................... 7
GENERAL
For technical data, options and accessories please see EE870 data sheet and operation manual at www.epluse.com/EE870.
The modular CO
EE871 interchangeable CO2 probe with E2 digital output. For technical data please see EE871 data sheet and operation manual (www.epluse.com/EE871)
For readjustment of EE871 please see data sheet and operation manual EE-PCS Product Configuration Software (www.epluse.com/EE-PCS)
HA011014 converter board: The converter board supplies the EE871 CO digital output signal into analogue outputs and Modbus RTU digital interface. The measured data is available simultaneously on the analogue outputs and on the Modbus interface. Technical data, setup and operations instructions for HA011014 are detailed in this document.
Connecting cable between EE871 CO cables, as ordered:
• HA010809 ......... 1 m length
• HA010810 ........ 2 m length
• HA010811 ........ 5 m length
• HA010812 ........ 10 m length
transmitter EE870 consists of following components:
2
probe and HA011014. EE870 scope of supply includes one of the following
2
.
3
probe and converts its
2
Page 4
IMPORTANT NOTES
Connecting the supply voltage to the modbus terminals may destroy the device.
The HA011014 converter board is supplied as a component and must be installed in accordance with the valid ESD guidelines and properly protected against enviromental influences.
Probe
Sensing head
Filter cap
CO2 Sensing cell
The device shall not be exposed to extreme mechanical stress. The sensing head of the probe and mostly the sensing cell might not be exposed to any mechanical stress.
The probe must be operated with the filter cap on at all times. Do not touch the sensing cell or electronics inside the sensing head.
A long response time indicates a dirty filter cap, as it might happen in polluted applications. Do not attempt to clean the filter cap; it would only cause its clogging. Replace the filter cap by an E+E original one, order no. HA010116.
While replacing the filter cap take utmost care to not touch the sensing cell and the electronics.
This device is not appropriate for safety, emergency stop or other critical applications where device malfunction or failure could cause injury to human beings.
CONNECTIONS / MOUNTING
Modbus Dip-switch
Jumper "Bus termination resistor"
Modbus terminals
Terminals for cable EE871
+5V white
SCL black
SDA blue
GND brown
Shield grey
Jumper "Fast/Slow"
Supply terminal*
LED:
Jumper "Output Range"
Analogue outputs*
On = OK Fast flash = Error
* Very important: for failure-free operation and performance according to the specs the supply GND and the
measurement GND must be wired separately.
4
Page 5
TECHNICAL DATA
6
6
78 (3.08“)
96 (3.78“)
Gewicht: 30g
Digital CO2 Probe EE871
Measuring principle Dual wavelength (non-dispersive infrared technology) NDIR Measurement range / 0...2000 ppm: < ± (50 ppm + 2 % from the measured value) Accuracy at 25 °C and 0...5000 ppm: < ± (50 ppm + 3 % from the measured value) 1013 mbar
0...3 %:
0...5 %: Response time t 60 s without measured data averaging Temperature dependency 0...2000 ppm: (-20...45 °C)
0...10,000 ppm:
0...3 %:
0...5 %: Housing / Protection class Plastic PC / Housing IP65 Cable length max. 10 m Electromagnetic compatibility EN61326-1 (Industrial enviroment) EN61326-2-3
Conversion Board
Supply voltage 10-35 V DC / 10-28.8 V AC Supply current 120 mA at 24 V DC / 300 mA at 10 V DC
Protection class IP00
Electrical connection screw terminal size: 2.5 mm Analog outputs 0-1 V; 0-5 V; 0-10 V -1 mA < IL < 1 mA selectable by jumpers 0-20 mA; 4-20 mA R Resolution 12 bit
Response time t
Modbus RTU setup with dip-switches (see operation manual) Temperature dependence Voltage: typ. ±0.2 mV/°C (0 – 1V) typ. ±0.5 mV/°C (0 – 5V) typ: ±0.6 mV/°C (0 – 10V)
Current: typ. ±1 μA / °C EE870 Operating conditions -40...60 °C EE870 Storage condition -40...60 °C (-40...140 °F) 0...100 % RH (not condensating) 70...110 kPa (10.15...15.95 psi)
1) For averaging output
1)
(77 °F...14,69 psi) 0...10,000 ppm: < ± (100 ppm + 5 % from the measured value)
< ± (1,5 % from full scale + 2 % from the measured value)
105 s with measured data averaging (smooth output)
63
(-4...113 °F) 0...5000 ppm: typ. ± (1 + CO
typ. -0,3 % from the measured value/°C
(32 ft)
60 s or 105 s selectable by jumpers
90
(-40...140 °F) 0...100 % RH (not condensating) 85...110 kPa (12.33...15.95 psi)
concentration [ppm] / 1000) ppm/°C
2
2
< 500 Ohm
L
DIMENSIONS
M12x1
Weight: 30 g (1.06 oz)
Conversion BoardDigital CO2 Probe EE871
(0.24“)
4.5
(0,18“)
18.5 (0.73“)
4.5
(0,18“)
3.2 (1.13“)
39 (1.54“)
13.8 (0.55“)
66 (2.6“)
(0.24“)
25.8 (1.02“)
5
Page 6
HARDWARE SETTINGS
Analogue-Output
Modbus-Interface
Jumper for setting the analogue output:
Current Voltage
- 0-1 V 0-20 mA 0-5 V 4-20 mA 0-10 V
Jumper for setting the response time:
fast response time slow response time
Example:
Jumper "Output Range" set to 0...10 V. Jumper "Fast/Slow" set to "Fast"
Dip-switch for Modbus settings
S1, S2 baud rate S3 stopbit S4 parity S5 - S10 addresse of the converter
MODBUS MAP
Register address Parameter name Protocol address
0x00 0x01 1 0x02 2 0x03 3 0x04 4 0x05 5 0x06 6 0x07 7
0x08 FW Version 8
0x09 0x0A 10 0x0B 11 0x0C 12 0x0D 13 0x0E 14 0x0F 15 0x10 16
0x11 0x12 18 0x13 19 0x14 20 0x15 21 0x16 22 0x17 23 0x18 24
Serial number
Type number
Serial number
17
Jumper for termination resistor active/inactive
Example:
Dip-switch S1 and S2 "Baud rate" set on 19200 Dip-switch S3 "Stopbit" set on "one" Dip-switch S4 "Parity" set on "even" Dip-switch S5 - S10 "Address" set on "13"
(activate S5/S7/S8)
Jumper "Termination resistor" set on "No"
Register address Parameter name Protocol address
0
ASCII
HIGH BYTE - main version LOW BYTE - sub version
9
ASCII
0x19 0x1A 26 0x1B 0x1C 28 0x1D 0x1E 30 0x1F 0x20 32
0x21 0x22 34 0x23 0x24 36
0x25 0x26 38 0x27 0x28 40
0x29 0x2A 42 0x2B 0x2C 44
0x2D 0x2E 46 0x2F 0x30 48
0x31 0x32 50
-
-
fast
CO
2
slow
CO
2
Gain Voltage
Offset Voltage
Gain Current
Offset Current
Measuring range
Measuring range
Analogue Output
low (V)
Analogue Output
high (V)
Slow or Fast
25
27
29
31
33
35
37
39
41
43
45
47
49
FLOAT
ASCII
6
Page 7
OPERATION OUTDOORS
For outdoor applications, the probe of EE870 must be used with the radiation shield order no. HA010507, which protects the device against rain, snow, ice, and solar radiation. The convertor board must protected IP65 (NEMA4) or better.
Ø105
174
Ø5
20
44
94
W
122
ACCESSORIES
Replacement probe EE871-HRxJ2 see data sheet EE871 Cable M12 - flying leads (1 m (39.37") / 2 m (78.74") / 5 m (196.85") / 10 m (393.70")) HA010809/10/11/12 Mounting flange for probe HA010212 Radiation shield HA010507 PFTE Filter cap HA010116 H
filter cap HA010122
2O2
Protection cap for the M12 cable socket HA010781 Protection cap for the M12 probe plug HA010782
SCOPE OF SUPPLY
- EE871 probe according to ordering guide
- Test report according to DIN EN10204 - 2.2 for EE871
- Conversion board HA011014
- Connecting cable HA0108xx
- Operation manual
- Test report according to DIN EN10204 - 2.2 for conversion board
7
Page 8
E+E ELEKTRONIK HEADQUARTERS
E+E ELEKTRONIK Ges.m.b.H.
Langwiesen 7, A-4209 Engerwitzdorf Austria
Tel: +43 7235 605-0 Fax: +43 7235 605-8 info@epluse.at / www.epluse.com
E+E ELEKTRONIK SUBSIDIARIES
E+E ELEKTRONIK China / Beijing
Tel: +86 10 8499 2361 info@epluse.cn / www.epluse.cn
E+E ELEKTRONIK China / Shanghai
Tel: +86 21 6117 6129 info@epluse.cn / www.epluse.cn
E+E ELEKTRONIK Germany
Tel: +49 6172 13881-0 info@epluse.de / www.epluse.de
E+E ELEKTRONIK France
Tél: +33 4 7472 3582 info@epluse.fr / www.epluse.fr
E+E ELEKTRONIK Italy
Tel: +39 02 2707 8636 info@epluse.it / www.epluse.it
E+E ELEKTRONIK Korea
Tel: +82 31 732 6050 info@epluse.co.kr / www.epluse.co.kr
E+E ELEKTRONIK USA
Tel: +1 847 490 0520 office@epluse.com / www.epluse.com
YOUR PARTNER IN SENSOR TECHNOLOGY
R
Ges.m.b.H.
www.epluse.com
Loading...