E+E EE576 Operating Manual

16
(0.63")
<2.8
(0.1")
>6.2
(0.24")
60
(2.4")
60
(2.4")
16
(0.63")
Air
Velocity
USER GUIDE
EE576 - Very Low Air Velocity Probe
CAUTION
Accurate measurement results are conditioned by the correct positioning of the sensing probe in the air stream.
Best accuracy is achieved in laminar flow.
Avoid mechanical stress onto the probe and mainly onto the sensing head.
Observe the humidity working range 10…95% RH, non-condensing.
Avoid installation in corrosive environment, as this may lead to sensor destruction.
SCOPE OF SUPPLY
EE576 according to ordering guide
User Guide
WIRING
Cable PVC 3 x 0,25mm² with cable end sleeves.
1 white V+ 2 brown GND 3 green output signal
PLEASE NOTE:
The EE576 is not reverse voltage- and short-circuit proof.
DIMENSIONS IN MM (INCH)
Probe:
150 (6”)
120 (4.7”)
80
(3.1”)
60 (2.4”)
alignment strip
24.5 (0.9”)
Detail A
optional (laser cutting):
hole in the wall of the duct:
Ø12 (0.5”)
Mounting Flange:
6 (0.2”)
44 (1.7”)
Bore hole for mounting:
drilling in the wall of the duct:
16
(0.6”)
Front sensor head:
13.4
Detail A:
Recess for
2.5 (0.1”) alignment
strip
By leaving a key notch in the hole in the wall
of the duct, the ange
can be mounted in the correct direction of the air stream.
(0.5”)
1.5 (0.1”)
alignment strip
The arrow indicates the air ow
direction during factory adjustment.
INSTALLATION
The alignment strip along the probe’s tube and the matching mounting flange determine the orientation of the sensor probe. The arrow on the tip of the sensor probe and on the mounting flange marks the direction of the air stream. Install the mounting flange in such a way that the alignment is parallel to the air stream.
CORRECT
Air
Velocity
INCORRECT
The mounting ange allows for an innitely variation of the immersion depth of the sensor probe. Attention should be
paid to that both sensor slots are in the air stream.
CORRECT
Air
Velocity
INCORRECT
If the sensor probe is installed without a mounting ange make sure the air velocity sensor is aligned in parallel with the
air stream.
CORRECT
Air
Velocity
INCORRECT
USA FCC NOTICE:
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the installation manual, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a parti­cular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which thereceiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
CANADIAN ICES-003 ISSUE 5:
CAN ICES-3 B / NMB-3 B
INFORMATION
Langwiesen 7 • A-4209 Engerwitzdorf Tel: +43 7235 605-0 • Fax: +43 7235 605-8 info@epluse.com • www.epluse.com
LG Linz Fn 165761 t • UID-Nr. ATU44043101 Place of Jurisdiction: A-4020 Linz • DVR0962759
+43 7235 605 0 / info@epluse.com
BA_EE576_e // v1.2 // technical data are subject to change // 302443
Loading...