E+E EE576 Specifications

EE576
The compact EE576 probe is optimized for low air velocity
measurement in applications like laminar ow control or lter monitoring. It operates on the hot-lm anemometer principle
which stands for high accuracy and fast response time.
Reliability
The the ow sensing element manufactured in state-of-the-art E+E thin-lm technology is highly insensitive to contamination and o󰀨ers excellent long-term stability.
Easy installation
in the air ow. The mounting ange within the scope of supply
enables precise setting of the immersion depth.
Output
The measured data up to 1 m/s (200 ft/min) or 2m/s (400 ft/min) is available on the 0-5V or 0-10V output.
Probe for Very Low Air Velocity
Typical Applications
laminar flow control filter monitoring
outstanding accuracy and long term stability
excellent price/performance ratio
easy and fast mounting
Technical Data
Air Velocity
Working range 0...1 m/s (0...200 ft/min)
0...2 m/s Accuracy1) 0.2...1 m/s (40...200 ft/min): ±(0.05 m/s +2 % of mv)
at 20 °C / 68 °F / 45 % RH and 1013 hPa 0.2...2 m/s (40...400 ft/min): ±(0.08 m/s +4 % of mv) mv=measured value
Response time t90 typ. 4 sec. at 1 m/s (200 ft/min) Output signal 0-5 V 0-10 V
General
Supply voltage (Class III) 10 - 19 V DC or 19 - 29 V DC
Current consumption max. 70 mA at 2 m/s Connection 0.5 m cable, PVC 3x0.25 mm² with cable end sleeves
Electromagnetic compatibility EN61326-1 EN61326-2-3 Probe material / Protection class polycarbonate / IP20 (sensing head); IP40 (probe) Humidity working and storage range 10...95 % RH (non-condensing) Temperature working range -20...60 °C storage range -30...60 °C (-22...140 °F)
1) The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation).
The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement).
(0...400 ft/min)
< 1 mA
I
L
(-4...140 °F)
(400 ft/min)
Features
IL = Load current
196
v1.5 / Modification rights reserved
EE576
Dimensions in mm (inch)
Probe:
12
(0.5”)
Flange
44 (1.7”)
150 (6”)
alignment strip
(included in the scope of supply):
80 (3.1”)
60 (2.4”)
6 (0.2”)
120 (4.7”)
cable length:
(19.7”)
0.5 m
24.5 (0.9”)
bore diameter 16 (0.6”)
Front sensing head:
13.4 (0.5”)
1.5 (0.1”)
alignment strip
The arrow indicates the air flow direction during factory adjustment.
2.5 (0.1”)
Cable Assignment
white V+ brown GND green output signal
Ordering Guide
Output
Air velocity range
Supply
Cable length
1) with 19 - 29 V DC supply only
Order Example
EE576-A2HV23V5KL200
EE576-
0 - 5 V
1)
0 - 10 V
0...1 m/s (0...200 ft/min) HV21
0...2 m/s (0...400 ft/min) HV23
10 - 19 V DC V5
19 - 29 V DC V6
0.5 m (1.64 ft)
2 m (6.56 ft) KL200
A2
A3
KL50
Output: 0 - 5 V Air velocity range: 0...2 m/s Supply: 10 - 19 V DC Cable length: 2 m
EE576
v1.5 / Modification rights reserved
197
Loading...