QUICK GUIDE
mA
+
-
10*)...28V DC *)10V+0.02*R
L
Output 1: 4-20 mA
8
4
mA
Output 2: 4-20 mA
5
6
3
2
RS485 A
RS485 B
www.epluse.com
WIRING / ANSCHLUSS
EE364 - Moisture in Oil Transmitter
10*)...28V DC *)10V+0.02*R
mA
Output 1: 4-20 mA
Output 2: 4-20 mA
8
4
5
mA
EE364-T63xx
View on sensor-plug
Blick auf Sensorstecker
6
3
2
RS485 A
RS485 B
Note:
To satisfy the EMC Directive 2004/108/EC a shielded
cable must be used.
Accessory HA0103xx - connecting cable for EE364-T63xx
Zubehör HA0103xx - Anschlusskabel für EE364-T63xx
1 WH NC (not connected)
2 BN RS485 B
3 GN RS485 A
4 YE Out1 4-20mA
5 GY Out2 4-20mA
6 PK GND
7 BU NC (not connected)
8 RD V+
L
+
3
4
5
1
10*)...28V DC *)10V+0.02*R
-
1 2
EE364-Tx3xx
View on sensor-plug
Blick auf Sensorstecker
3
4
5
Anmerkung:
Zur Erfüllung der EMV Richtlinie 2004/108/EG ist ein
geschirmtes Kabel zu verwenden.
Accessory HA0108xx - connecting cable for EE364-Tx3xx
Zubehör HA0108xx - Anschlusskabel für EE364-Tx3xx
1 BN V+
2 WH NC (not connected)
3 BU GND
4 BK RS485 A
5 GY RS485 B
L
+
-
RS485 A
RS485 B
MODBUS MAP
Instructions for Modbus-Protocol-Setup please see
Application Note AN0103 (www.epluse.com/EE364).
• The EE364 can be used in Modbus networks with
• The bus termination shall be realized with 120 Ohm
max. 32 units.
resistor (not included in the scope of supply)
• For proper function the power supply must be strong
enough to ensure supply voltage within the specified
range (see technical data) at any time and at all
devices in the bus. This is particularly relevant when
using long and thin cables which can cause high voltage drop.
Factory setting:
Slave-ID 243; Baudrate 9600, Parity even, Stopbit 1
Grundlagen zur Modbus-Protokoll-Einstellungen finden Sie
in der Application Note AN0103 (www.epluse.com/EE364).
• Der EE364 ist für Modbus Netzwerke mit max. 32
Busteilnehmern geeignet.
• Der Abschluss der Busleitung (bus termination) muss
mit einem 120 Ohm Widerstand (nicht im Lieferumfang
enthalten) realisiert werden.
• Für eine einwandfreie Funktion muss das Netzteil stark
genug sein, um die Versorgungsspannung innerhalb
des angegebenen Bereichs (siehe technische Daten)
jederzeit und auf allen Geräten im Bus zu gewährleisten. Dies gilt insbesondere bei der Verwendung von
langen und dünnen Kabel, die einen hohen
Spannungsabfall verursachen können.
Werkseinstellung:
Slave-ID 243; Baudrate 9600, Parity even, Stopbit 1
32Bit FLOAT: 16Bit Data:
Ø30
(1.2)
Ø12
(0.5)
G1/2“
Coil / Register
Numbers
30052 0x33 water activity Aw
30054 0x35 water content Xk/Xm
30026 0x19 temperature Tx
30224 0xDF A parameter (read)
30226 0xE1 B parameter (read)
60101 0x64 A parameter (write)
60103 0x66 B parameter (write)
DataAddresses
Parameter name
Coil / Register
Numbers
SETUP AND ADJUSTMENT / SETUP UND JUSTAGE
The EE364 transmitter is ready to use and does not require any
confi guration by the user. The factory setup of EE364 corresponds
to the type number ordered. If needed, the user can change
the factory setup or make an adjustment by using the optional
USB-Modbus converter (E+E part no. HA011013) and the E+E
Product Configuration Software (EE-PCS). The E+E Product
Configuration Software (EE-PCS) is free and can be downloaded
from www.epluse.com/configurator.
Der EE364 ist sofort einsatzbereit und erfordert keine
Konfi guration durch den Kunden. Die Werkseinstellung des EE364
entspricht der bestellten Typenbezeichung. Bei Bedarf kann die
Werkseinstellung mit dem optionalen USB-Modbus Konverter
(E+E Artikel Nr. HA011013) und der E+E Product Confi guration
Software (EE-PCS) geändert werden.
Die E+E Product Configuration Software (EE-PCS) steht unter
www.epluse.com/configurator gratis zum Download bereit.
INSTALLATION / EINBAU
DataAddresses
60001 0x00 Slave-ID
60002 0x01 RS485 Setting
Parameter name
OK
DIMENSIONS / ABMESSUNGEN
mm (inch)
ISO NPT
34 (1.36)
Ø12
(0.5)
G1/2“
SCOPE OF SUPPLY / LIEFERUMFANG
- EE364 Transmitter according to Ordering Guide
- M12x1 Attachable mating connector - 8 pin, straight
- User manual - Quick Guide
- Inspection certifi cate according to DIN EN10204 - 3.1
140 (5.5)
AD27
77 (3)
92 (3.6)
Ø30
NOT OK
51 (2)
(1.2)
1/2“NPT
- EE364 Messumformer lt. Bestellinformation
- Konfektionierbarer Gegenstecker M12x1, 8 pol., gerade
- Bedienungsanleitung - Quick Guide
- Abnahmeprüfzeugnis gemäß DIN EN10204-3.1
AD22
89 (3.5)
74 (2.9)
E+E Elektronik Ges.m.b.H.
Langwiesen 7 • A-4209 Engerwitzdorf
Tel: +43 7235 605-0 • Fax: +43 7235 605-8
info@epluse.com • www.epluse.com
LG Linz Fn 165761 t • UID-Nr. ATU44043101
Place of Jurisdiction: A-4020 Linz • DVR0962759
BA_ee364 // v3.3 // Technische Änderungen vorbehalten // 302724