EE36 is dedicated for the accurate and reliable measurement of moisture in lubrication oil in maritime applications and is certified according to the “Germanischer Lloyd (DNV GL)”. The device is specifically approved for use in MAN B&W marine diesel engines for instance.
The outstanding long-term stability and resistance to pollution of the EE36 rests on the high-end E+E capacitive sensing element of the HC series.
Humidity measurement in oil
The monitoring of the moisture in lubrication oil is of paramount importance for the long-term performance and predictive maintenance of the machinery. The moisture in oil is described by either the absolute value “water content” x (ppm) or the relative value “water activity” a
- The water content x (ppm) represents the ratio between the mass of water and the mass of oil.
- The water activity a saturated oil.
a
= 0 corresponds to completely dry oil, while aw = 1 means a fully saturated oil.
w
Humidity / Water in Oil Sensor for Maritime Applications
:
w
is the ratio between the actual moisture content and the maximal moisture content of the
w
EE36 +
Ball valve set
EE36 measures the water activity (a rate calculation of the water content of a certain oil requires a set of oil specific parameters.
The measured and the calculated values are available on two free scalable and configurable analogue outputs as well as on the optional display. Additionally, EE36 can be fitted with a relay module for alarms and process control purposes.
Installation
The sensing probe of EE36 is designed for continuous online monitoring. In addition to the fix installation, the use of a ball valve allows for insertion and removal of the probe without process interruption.
Easy Adjustment
The user can easily readjust the transmitter by using either a simple procedure with two push buttons on the printed circuit board or the free configuration software.
) and the temperature (T) and calculates out of them the water content x. The accu-
w
Product Configuration Software (EE-PCS)
The free configuration software (download at www.epluse.com) allows for easy setup of the analogue and alarm outputs, as well as for the adjustment of the water activity and temperature readings. Furthermore, the EE-PCS facilitate the replacement of the sensing element or of the sensing probe.
Features of EE36
Measurement of aw and T at pressure up to 20 bar (300 psi)  Calculation of water content in ppm  Two free scaleable and configurable analogue outputs  Probe cable length up to 20 m (66 ft)  Easy on site adjustment and calibration of a LED indication for operation and sensing probe status  User configuration of the instrument with PC via RS232 interface  Configuration software  Display of a Two free configurable relays outputs optional Pluggable sensing probe optional Connector for power supply and outputs optional
, T and water content with MIN/MAX function optional
w
and T outputs
w
120
v3.0 / Modification rights reserved
EE36
EE36
Dimensions (mm/inch) Installation Example
NO
NO
3
6
NC
NC
1
4
C
C
2
5
REL1
REL2
Metal housing:
135 (5.3”)
90 (3.5”)
163 (6.4”)
66.5 (2.6”)
(pressure-tight up to 20 bar/290 psi)
Material: Al Si 9 Cu 3 Protection class: IP65 / NEMA 4
1/2” ISO or 1/2” NPT
ball valve installation
Probe:
code “cable length”
32
(1.3”)
13
(0.51”)
23 (1”)
64 (2.5”) resp. 164 (6.5”)
code “probe length”
12
(0.47”)
Electrical connection
Standard
15
(0.59”)
Plug Option C03
EE36-xEx
Remote probe for T -40...180 °C
(-40...356 °F) and pressure-tight
up to 20 bar
(300 psi)
probe material: stainless steel
minimum installation depth
maximum installation depth
Plug Option C06
1/2” ISO or 1/2” NPT
bore diameter
(0.5”)
>13.1
fixed installation
(pressure-tight up to 20 bar/300 psi)
1/2” ISO or 1/2” NPT
bore diameter: >13.1 (0.5”)
Plug Option C07
2x M16x1.5
Connection Diagram
EE36
v3.0 / Modification rights reserved
Lumberg RKC 5/7
power supply + analogue output
serial connection for configuration
analogue output
supply
Lumberg
RSC 5/7
M16x1.5
Lumberg RSC 5/7