E+E EE35 Specifications

Page 1
EE35
Industrial Transmitter for
Dew Point Measurement
Exact dew point monitoring is increasingly playing a more impor­tant role in many industrial applications, such as drying process­es, air pressure pipelines, etc. For these purposes the multifunc­tional EE35 Series offers the ideal features.
The EE35 Series is based on a functional, user-friendly housing concept and on the proven polymer humidity sensors of the HC Series. A specially developed autocalibration process enables measure­ments in a measurement range of -60...60°C Td (-76...140°F Td), with a Td measurement accuracy of ±2°C (±3.6°F). Two freely configurable and scalable analogue outputs are available for the two measurement values (Td, T). An optional hygrostat output, which can be set by means of a potentiometer, provides an alarm signal in a simple way when a threshold of the permitted dew point is exceeded. An optional display for the measurement values and the associated MIN/MAX values allows a quick overview of the current situation.
Autocalibration
Dew points in the range of -60...-20°C (-76...-4°F) at room temperatures correspond to relative humidity values of 0.08...5.37% RH. The measurement of such low humidity values is not possible with conventional capacitive measurement methods. For the EE35 Series, a special autocalibration process is used to compen­sate for the usual drift effects and thus to achieve high accuracy measurements also at -60°C Td
(-76°F Td).
Installation
In addition to the direct mounting of the dew point probe, a ball valve installation enables the mounting and removal of the probe without having to interrupt the running process.
Alarm Output
An optional alarm module with one relay output is available for control and alarm purposes. The setting of the Td threshold can be easily done with the potentiometer on the printed circuit board.
Integrated power supply
A power supply, integrated in the back module of the housing, can be ordered optionally (100...240V AC, 50/60Hz; ordering code V01). The power supply V01 is available for both polycarbonate and metal housing and comes standard with two plugs for supply and outputs to allow an easy connection.
Typical Applications
Features
industrial processes monitoring of air pressure pipelines warehouses drying processes paper industries chemical industries
104
measuring range -60...60°C Td (-76...140°F Td)
accuracy of measurement ±2°C Td (±3.6°F Td)
traceable calibration
alarm output for dew point
autocalibration
V2.7 / Modification rights reserved
EE35
Page 2
Housing Dimensions (mm)
Installation Example
Housing:
polycarbonate housing metal housing
135 (5.3”)
163 (6.4”)
For use in harsh industrial environ­ments the EE35 series is available in a robust metal housing.
90 (3.5”)
135 (5.3”)
66.5 (2.6”)
90 (3.5”)
Model:
32
13 (0.51”)
(1.3”)
code “probe length”
12 (0.47”)
code “cable length”
15 (0.59”)
EE35-xEx
Remote probe for T up to 60°C (140°F) and pressure-tight up to 20bar (290psi)
Probe material: stainless steel
66.5 (2.6”)
ball valve installation
1/2” ISO or 1/2” NPT
1/2” ISO or 1/2” NPT
(pressure-tight up to
20bar/290psi)
bore diameter: >13.1 (0.5”)
23 (1”)
64 (2.5”) resp. 164 (6.5”)
Connection Versions
Standard
2x M16x1.5
minimum installation depth
maximum installation depth
Plug Option C03
Lumberg
RKC 5/7
fixed installation
(pressure-tight up to 20bar/290psi)
1/2” ISO or 1/2” NPT
bore diameter:
(0.5”)
>13.1
Plug Option C06
Lumberg
RSC 5/7
M16x1.5
EE35
V2.7 / Modification rights reserved
Power supply + Analogue output
105
Page 3
Technical Data
80
60
40
20
0
-20
-40
-60
-80
0
10
20 30
40
50 60
limit of measuring range
uncertainty of measurement ±2 degC
Analog Ausgang / analogue output
OUT 2
8
Versorgung / supply
OUT 1
GND
GND
V+
V
mA
V
mA
7
6
5
4
3
2
1
8-35V DC 12-30V AC
Serielle Schnittstelle für Konfiguration / serial connection for configuration
GND Ser
TxD/A+
RxD/B-
Measuring Quantities
Dew point
Humidity sensor HC1000-400 Measuring range standard calibration: -40...60°C (below 0°C / Accuracy ±2°C ( ± 3.6°F) Traceable to intern. standards,
administrated by NIST, PTB, BEV...
32°F the transmitter outputs frostpoint) special calibration: -60...60°C (-76...140°F)
(-40...140°F)
Response time t
dew point temperature (°C)
90
80 sec. -20°C -40°C (-4°F → -40°F) 10 sec. -40°C -20°C (-40°F -4°F)
Process temperature (°C)
Temperature
Sensor Pt1000 DIN A
Measuring range 0...60°C
Accuracy of temperature measurement at 20°C (68°F) ±0.2°C (±0.36°F) Sensitivity error at full scale ±0.1°C (±0.18°F) Temperature dependence of electronics < 0.005°C/°C
Outputs
Two freely selectable and scaleable analogue outputs 0 - 10V -1mA < I
0 - 5V -1mA < I
xx...yy°C T, Td/Tf / xx...yy°C respectively 4 - 20mA RL < 500 Ohm
0 - 20mA RL < 500 Ohm
General
Supply voltage 8...35V DC
12...30V AC (optional 100...240V AC, 50/60Hz)
Current consumption - voltage output typ. 40mA, with autocalibration: 100mA
- current output typ. 80mA, with autocalibration: 140mA
Pressure range 0...20bar
Housing / protection class PC or Al Si 9 Cu 3 / IP65;
Cable gland M16 x 1.5 (option: plug) cable Ø 4.5 - 10 mm (0.18 - 0.39”)
Electrical connection screw terminals up to max. 1.5mm² (AWG 16)
Sensor protection stainless steel sintered filter
Working temperature range probe: -40...60°C
electronic: -40...60°C
with LC display: -20...50°C (-4...122°F)
with alarm module: -40...60°C
Storage temperature range -40...60°C (-40...140°F)
Electromagnetic compatibility according to EN 61326-1 EN61326-2-3 ICES-003 ClassB
Industrial Environment FCC Part15 ClassB
(32...140°F)
(0...300psi)
< 1mA
L
< 1mA
L
Nema 4
(-40...140°F) (-40...140°F)
(-40...140°F)
Technical Data for Options
Connection Diagram
Display graphical LC display (128x32 pixels), with integrated push-
buttons for selecting parameters Td or T and MIN/MAX functions
Alarm output for Td/Tf - range: -60...40°C Td
(-60...40°F Td)
adjustable with the
potentiometer
on the printed circuit board
- 1 switch contact
- 250V AC/6A or 28V DC/6A
Terminal configuration - Alarm output
106
serial connection for configuration
analogue output
supply
REL 1
V2.7 / Modification rights reserved
NC
C
NO
11
12
13
EE35
Page 4
Ordering Guide EE35
EE35-
Hardware Configuration
Housing metal housing M
Type pressure tight E Cable length 1m
(incl. probe length)
Probe length 100mm
Pressure tight 1/2" male thread HA03 feedthrough 1/2” NPT thread HA07 Display without display
Alarm output
Plug cable glands
Probe fixed
Td-Calibration standard -40...60°C
Supply voltage 8...35V DC / 12...30V AC
1)
Software Configuration
Physical parametres temperature T [°C/°F] output 1 B of the outputs dew point temperature Td [°C/°F] output 2 C
Type of 0-5V 2 output signals 0-10V 3
Measured value unit metric [°C]
Scaling of T-output -40...60 (T02) -60...20 (T65) -40...100 (T79) output T Select accorcding to
Scaling of Td/Tf-output -40...60 (T02) 0...60 (T07) -60...60 (T64) output Td resp.Tf Select accorcding to
1) Combination alarm output and plugs is not possible (with cable glands only) / combination alarm output and integrated power supply is not possible
2) Integrated power supply includes 2 plugs for power supply and outputs / further plug options are not possible
polycarbonate housing P
(3.3ft) 01
2m (6.6ft) 02
(16.4ft) 05
5m
(3.9”) 3 (7.9”) 5
200mm
with display D05 without relay with relay SW
1 plug for power supply and outputs C03 1 cable thread / 1 plug for RS232 C06
pluggable P01
special calibration -60...60°C (-76...140°F) CA02
integrated power supply 100...240V AC, 50/60Hz
frost point temperature Tf [°C/°F] D
0-20mA 5 4-20mA 6
non metric [°F] E01
-50...50 (T27) -50...100 (T66) -40...140 (T83) ordering guide (Txx)
-80...20 (T63) -20...70 (T73) -60...120 (T97)
-60...60 (T64) 20...140 (T77)
-10...50 (T03) 0...80 (T21) 32...120 (T90) ordering guide
0...50 (T04) -40...80 (T22) 32...140 (T91) (Tdxx resp. Tfxx)
0...100 (T05) -20...80 (T24) 32...132 (T96)
(-40...140°F)
2)
Other T-scaling refer
to data sheet „„T-Scalings““
Other Td/Tf-scaling refer
to data sheet „„T-Scalings““
V01
Accessories
- Ball valve set 1/2” ISO (HA050101) - Interface cable for PCB („HA010304“)
- Ball valve set 1/2” NPT (HA050104) - Interface cable for plug C06 („HA010311“)
- Display + housing cover in metal („D05M“) - Bracket for installation onto mounting rails* („HA010203“)
- Display + housing cover in polycarbonate („D05P“) - Sealing element (HA050308)
- Stainless steel sintered filter („HA010103“)
*Note: Only for plastichousing, not for metalhousing
Order Example
EE35-ME025HA03D05P01/BC5-T02-Td02
Housing: metal housing Output 1: T Type: pressure tight Output 2: Td Cable length: 2m Probe length: 200mm Pressure tight feedthrough: 1/2" male thread Scaling of T-output: -40...60°C Display: with display Scaling of Td-output: -40...60°C Alarm output: without relay Plug: cable glands Sensing probe: pluggable Td Calibration: standard Supply voltage: 8...35V DC / 12...30V AC
EE35
V2.7 / Modification rights reserved
(6.6ft) Output signal: 0-20mA
(7.9”) Measured value unit: metric
107