Humidity / Temperature / Dew Point / Absolute Humidity...
The precise and reliable measurement of humidity in industrial
processes is gaining more and more importance. The multifunctional
transmitters series EE31 offer the ideal solution.
The result of many years of experience in humidity measurement technology for industrial applications, the EE31 series builds on the E+E highquality HC series capacitive humidity sensor elements.
The optimal hardware structure for varying applications is achieved by
combining various standard mechanical and electronic modules. User
friendly MS Windows software tools simplify the configuration of the
transmitter, the data recording, visualization and processing.
The measured values are available on two freely configurable and
scalable analogue outputs and on the serial RS232 interface. With an
optional RS485 module up to 32 transmitters can be connected to a network and one single PC interface allowing easy remote monitoring.
Two freely configurable optional alarm outputs can be set by software.
The measured data and the corresponding MIN/MAX values can be
viewed on the optional LC display.
Other features especially tailored for harsh industrial applications are the
new housing concept consisting of three modules, the easy on-site
adjustment and calibration, and the pluggable sensor option.
These features allow for very fast and easy servicing of the transmitter.
Model A
Model B
Model D / E
By selecting a suitable housing version the EE31 series can be used for
the entire range of humidity measurement applications:
- Model A for wall mounting
- Model B for duct mounting
- Model D with remote sensing probe for measurements in the extended
temperature range -40...180°C (-40...356°F).
- Model E with remote sensing probe for pressure tight applications between 0.01...20 bar (0.15...300psi).
Network with up to 32 transmitters
Up to 32 transmitters can be connected in a RS-485 bus system to a single
PC interface.
The measured and calculated data is stored in a PC database which is available for further processing.
The data base can also be stored in ASCII format or in a database with
ODBC interface.
Product Configuration Software (EE-PCS)
The Configuration Software is used for:
- flexible, easy and fast setup of the analogue and alarm outputs.
- adjustment of the humidity and temperature outputs.
- exchange of the sensing probe or of the sensors.
12
v1.7 / Modification rights reserved
EE31
EE31
Page 2
Easy calibration and adjustment of the transmitter
The modular housing of the EE31 enables a fast and easy on-site adjustment and calibration. Using the
optional extension cable one can adjust or calibrate the entire measurement loop without interrupting the
measurement. No need for time-consuming dismounting and wiring of the instrument.
This feature makes the EE31 series suitable for use in regulatory environments (e.g. FDA, GAMP).
The adjustment of humidity and temperature (2 points or 1 point) is performed either with a simple routine
using two push buttons on the printed circuit board or with the configuration software.
2 Status LEDs
Two status LEDs on the printed circuit board indicate the transmitter status and eventual errors,
especially useful during installation or service operations.
Sensor Coating
Operation in heavily polluted and/or corrosive environments is typical for many industrial processes and can
lead to drift or damage of the humidity sensor and thus to false measured values. The unique protective
coating developed by E+E for the sensing probe brings a significant improvement on the long-term stability
of the transmitter in very dirty and aggressive environments. (ordering code: HC01)
Integrated Display
The actual measured and calculated values as well as the corresponding Min/Max
values can be indicated on an optional display. The physical quantity to be displayed is chosen with the push buttons on the housing. (ordering code: D05)
Pluggable sensing probe
The pluggable sensing probe with plug connection can be easily exchanged in the
versions D and E. The installation of the probe cable (up to 20m / 65ft) is
significantly simplified and can be installed prior to fitting the transmitter.
(ordering code: P01)
Alarm outputs
An optional alarm module with 2 relay outputs is available for control and alarm purposes. The selection of
the physical quantity for the relay outputs and the setting of threshold and hysteresis can be
easily made with the configuration software included in the standard scope of supply.
Integrated power supply
A power supply, integrated in the back module of the housing, can be ordered
optionally (100...240V AC, 50/60Hz; ordering code V01). The power supply V01 is
available for both polycarbonate and metal housing and comes standard with two
plugs for supply and outputs to allow an easy connection.
EE31
v1.7 / Modification rights reserved
13
Page 3
Connection versions
standard
2x M16x1.5
plug option C03
Lumberg
RKC 5/7
power supply+
analogue output
plug option C06
Lumberg
RSC 5/7
RS232
Dimensions in mm
Housing:
polycarbonate housing metal housing
135 (5.3”)
66.5
(2.6”)
M16x1.5
plug option C08
Y splitter*
Lumberg
RSC 5/7
RS485 network
* Siemens 6ES7 194-1KA01-0XA0
135 (5.3”)
Lumberg
RKC 5/7
power supply+
analogue output
66.5 (2.6”)
90 (3.5”)
Models:
50 (2”)
*)
L
∅12 (0.5”)
EE31-xAx
Wall mounting
Probe material: PC
code “probe length”
90 (3.5”)
163 (6.4”)
For use in harsh industrial environments all models of the EE31 are
available in a robust metal housing.
The very smooth surface and the rounded outlines allow for the use in
clean rooms as well.
*)
L
EE31-xDx
∅12 (0.5”)
Remote probe for T
code “probe length”
up to 180°C
(356°F)
code “cable length”
15 (0.6”)
Probe material: stainless steel
*)
L
*)
L
code “cable length”
200
15 (0.6”)
14
EE31-xBx
Duct mounting
Probe material: stainless steel
*) L = Filter length: refer to data sheet "Accessories"
-40...180°C
Temperature dependence of electronics typ. ± 0.01% RH/°C (0.0055% RH/°F)
Response time with metal grid filter at 20°C / t
Temperature
Temperature sensor element Pt1000 (Tolerance class A, DIN EN 60751)
Working range sensing head EE31-xAx: -40...60°C
EE31-xBx: -40...80°C (-40...176°F) EE31-xEx: -40...180°C (-40...356°F)
Accuracy
1)
1)
0...100% RH
HC1000-400
(including hysteresis, non-linearity and repeatability, traceable to intern. standards, administrated by NIST, PTB, BEV...)
Temperature dependence of electronics typ. ± 0.005°C/°C
Outputs
2)
Two freely selectable and scaleable analogue outputs 0 - 5V -1mA < IL < 1mA
0...100% RH / xx...yy°C respectively 0 - 10V -1mA < I
4 - 20mA R
0 - 20mA R
L
L
< 1mA
L
< 500 Ohm
< 500 Ohm
Serial interface RS232C
RS485 optional
Max. adjustable measurement range
2)3)
from up to units
EE31-A EE31-B EE31-D,EHumidity RH 0 100 100 100 % RH
Temperature T -40
(-40)60 (140)80 (176)180 (356) °C (°F)
Dew-point temperature Td -40 (-40)60 (140)80 (176)100 (212)°C (°F)Frost-point temperature Tf -40 (-40)0 (32)0 (32)0 (32)°C (°F)Wet-bulb temperature Tw 0 (32)60 (140)80 (176)100 (212)°C (°F)Water vapour partial pressure e 0 (0)200 (3)500 (7.5) 1100 (15) mbar (psi)Mixture ratio r 0 (0)425 (2900)999 (9999) 999 (9999)g/kg (gr/lb)Absolute humidity dv 0 (0)150 (60) 300 (120) 700 (300)g/m
3
3
(gr/f
Specific enthalpy h 0 (0)400 (50000) 1000 (375000) 2800 (999999)kJ/kg (Btu/lb)
General
Supply voltage 8...35V DC
12...30V AC (optional 100...240V AC, 50/60Hz)
Current consumption - 2x voltage output for 24V DC/AC: typ. 40mA
- 2x current output typ. 80mA
Pressure range for pressure tight probe 0.01...20bar
System requirements for software WINDOWS 2000 or later; serial interface
Housing / protection class PC or Al Si 9 Cu 3 / IP65; NEMA 4
Cable gland M16 x 1.5 cable Ø 4.5 - 10 mm
Electrical connection screw terminals up to max. 1.5mm² (AWG 16)Working and storage temperature range of electronics -40...60°C (-40...140°F)
-20...50°C (-4...122°F) - housing with display
Electromagnetic compatibility according to EN61326-1 EN61326-2-3 ICES-003 ClassB
Industrial Environment FCC Part15 ClassB
1) Refer to the working range of the humidity sensor. 2) Can be easily changed by software.
3) Refer to accuracies of calculated values (www.epluse.com/feuchtemessung).
*) The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation).
The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement).
(0.15...300psi)
(0.18 - 0.39”)
)
)
EE31
v1.7 / Modification rights reserved
15
Page 5
Technical Data for Options
Display graphical LC display (128x32 pixels), with integrated push-buttons
for selecting parameters and MIN/MAX function
Alarm outputs 2 x 1 switch contact
250V AC / 6A 28V DC / 6A
Threshold + hysteresis can be adjusted with configuration software
Switching parameters freely selectable between:
RH Relative humidity T Temperature
Td Dew-point temperature
Tf Frost-point temperature Tw Wet-bulb temperature
e Water vapour partial pressure
r Mixture ratio
dv Absolute humidity
h Specific enthalpy
Working range humidity sensor
The grey area shows the allowed measurement range for
the humidity sensor.
Relative humidity [%RH]
Temperature [°C]
Connection diagram
Scope of Supply
EE31 according to ordering guide
Manual EE31 German/English/French
Inspection certifi cate according to DIN EN 10204 – 3.1
Allen key 3.0only for metal housing
Mating plug for integrated power supplyV01
Mating plug RKC 5/7V01 / C03 / C08
Y-junction for network connectionC08 & N
Mating plug RSC 5/7C06 / C08
M16 Cable gland metalexcept C03, C06, C08, V01
16
Operating points outside of this range do not lead to
destruction of the element, but the specified measurement accuracy cannot be guaranteed.
Terminal configuration - Alarm output
NC
REL1
REL2
Included in all versionsAccording to ordering guide
Pressure tight1/2” male threadHA03
feedthrough1/2” NPT threadHA07
InterfaceRS232
Displaywithout display
Alarm output
Plugcable glands
Sensing probefixed
Coating sensorno
Supply voltage8...35V DC / 12...30V AC
1)
Software Configuration
Physicalrelative humidityRH [%](A)Output 1
parameters ofTemperatureT[°C](B)
outputsDew point temperatureTd[°C](C)Output 2
Type of0-5V(2)
output signals0-10V(3)
Measured value units metric / SI
Scaling of T-output-40...60 (T02) -20...80 (T24)0...350(T89)
Scaling of Td-output0...50 (T04)0...180(T26)32...120 (T90)
in°C or °F0...100 (T05)-40...180 (T52)32...140 (T91)
1) Combination alarm output and plugs is not possible (with cable glands only) / combination alarm output and integrated power supply is not possible
2) Integrated power supply includes 2 plugs for power supply and outputs / further plug options are not possible
polycarbonate housingPPPP
PTFE filter5555
filter8888
H
2O2
stainless steel grid filter (up to 180°C/ 356°F)9999
(6.6ft)0202
5m (16.4ft)0505
(32.8ft)1010
10m
(65.6ft)2020
20m
(2.6”)2
(7.9”)555
200mm
(15.8“)66
400mm
RS485NNNN
with displayD05D05D05D05
without relay
with relaySWSWSWSW
1 plug for power supply and outputsC03C03C03C03
1 cable gland / 1 plug for RS232C06C06C06C06
2 plugs for power supply/outputs and RS485 NetworkC08C08C08C08
pluggableP01P01
yesHC01HC01HC01HC01
3
](H)
2)
Output T
Output Td
integrated power supply 100...240V Ac, 50/60Hz
Frost point temperatureTf[°C](D)
wet bulb temperatureTw [°C](E)
water vapour partial pressure e[mbar](F)
mixture ratior[g/kg](G)
absolute humiditydv[g/m
specific enthalphyh[kJ/kg](J)
0-20mA(5)
4-20mA(6)
non metric / USE01E01E01E01
0...60 (T07)-40...100 (T79)32...180 (T92)
-40...120 (T12)-40...350 (T82)32...250 (T94)
0...120 (T16)-40...140 (T83)32...300 (T95)
0...80(T21)-40...300 (T84)32...132 (T96)
-40...80(T22)0...250 (T88)32...350(T101)
V01V01V01
Select according to Ordering Guide(A-H, J)
Select according to Ordering Guide(A-H, J)
Select according to Ordering Guide(2,3,5,6)
Select according to Ordering Guide (Txx)
Select according to Ordering Guide (Tdxx)
Other T and Td-scaling refer to datasheet
T-Scaling
Order Example
EE31-PFTB55SW/BC2-T07-Td03
Housing: polycarbonate housing
Type: humidity + temperature Output 1: T
Model: duct mounting Output 2: Td
Filter: PTFE Filter Output signal: 0-5V
Probe length: 200mm (7.9”) Scaling of T-output: 0...60°C
Alarm output: with relay Scaling of Td-output: -10...50°C
Accessories / Replacement Parts (For further information, see data sheet "Accessories")
- Filter caps (HA0101xx)- Interface cable for plugs C06 (HA010311)
- Display + housing cover in metal (D05M) - Bracket for installation onto mounting rails* (HA010203)
- Display + housing cover in polycarbonate (D05P) - Drip water protection (HA010503)