The EE300Ex intrinsically safe sensor reliably measures relative humidity (RH) and temperature (T) in explosion hazard areas. It com-
plies with the classications for Europe (ATEX), International (IECEx),
USA / Canada (FM) and Korea (KCs) for ammable gas and dust
applications.
The entire device can be placed in the explosion endangered area.
The remote sensing probe allows for classication up to T6.
Measurement performance
The well proven E+E humidity sensors and competence in calibration
allow for highly accurate and long term stable measurement over the
full range 0…100 % RH and -40…180 °C (-40...356 °F), with pressure
rating up to 20 bar (300 psi).
Besides the RH and T measurement, the EE300Ex calculates all
humidity related parameters such as dew point temperature (Td),
frost point temperature (Tf), absolute humidity (dv) or mixing ratio (r).
Moisture in oil measurement
The EE300Ex with ATEX, IECEx and KCs approval is suitable also
for measuring water content (X) in ppm and water activity (aw) in
isolation, lubrication and hydraulic oils. Typical applications include oil
puriers and online monitoring of lubrication and hydraulic oils.
Supply and outputs
The device can be powered by any intrinsically safe supply unit or via
Zener barriers. The measured or calculated data is available on two
4…20 mA, 2-wire outputs and on the LCD display.
Robust, functional design
The stainless steel enclosure and sensing probe are suitable for harsh environment in challenging industrial applications.
The EE300Ex design facilitates the installation as well as the replacement of the measuring section (electronics and
probe) without time consuming wiring.
Easy Conguration and Adjustment
The setup of the analogue outputs as well as the adjustment of the RH and T reading can be easily performed with the
optional EE-PCA Product Conguration Adapter and the free EE-PCS Product Conguration Software.
Typical Applications
Chemical process control
Pharmaceutical applications
Explosive / hazardous storage rooms
Flour mills
Oil purifiers
18
Features
Gas and dust in zone 0 / 20 and Div. 1
Stainless steel enclosure and probe
Best accuracy up to 180 °C (356 °F)
Pressure tight up to 20 bar (300 psi)
Inspection certicate according to DIN EN 10204-3.1
EE300Ex-M1
Page 2
EE300Ex-M1
v2.8 / Modification rights reserved
Protective sensor coating
The E+E proprietary sensor coating is a hygroscopic layer applied to the sensing elements, their leads and soldering
points. The coating substantially extends sensor life-time and ensures optimal measurement performance in corrosive
environments (salts, o-shore applications). Additionally, it improves the long term stability of E+E sensors in dusty, dirty
or oily applications by preventing stray impedance caused by deposits on the active sensor surface or on the electrical
connections.
Types
TypePressure rangeWorking rangeProbe Ø mm (inch)
T1 Wall mount-40...60 °C
(-40...140°F)12 (0.47)
T7 Remote probe with cut-in tting, pressure tight0.1...20 bar (1.5...300 psi)-40...180 °C (-40...356°F)12 (0.47)
Remote probe with sliding tting for assembly /
T10
disassembly under pressure, pressure tight
0.1...20 bar (1.5...300 psi)-40...180 °C (-40...356°F)13 (0.51)
Dimensions in mm (inch)
179
Types: T1 / T7 / T10
Enclosure
(7)
60 (2.4)151 (5.9)
98 (3.9)
L
Ø13
(0.51)
adjustable min. 23 (0.9) /
max. 164 (6.5) / 364 (14.3)
Ø12
(0.47)
code „probe length “
code „cable length“
15 (0.6)
G1/2" ISO or NPT
Type: T10
Remote probe 20 bar (300 psi) with sliding fitting for assembly /
disassembly under pressure
53 (2.1)
code „cable length“
L
Type: T1
Wall mount
Ø 12 (0.47)
Ø12
(0.47)
L
code „probe length “
G1/2" ISO or 1/2" NPT
15 (0.6)
Types: T7
Remote probe 20 bar (300 psi)
L - length of filtermm (inch)
Stainless steel sintered filter 33 (1.3)
PTFE, H2O2 filter33 (1.3)
Stainless steel grid filter39 (1.5)
Oil filter32 (1.26)
19
Page 3
v2.8 / Modification rights reserved
EE300Ex-M1
∆T [°C]
∆T [°C]
Technical Data EE300Ex
Measurands
Relative humidity
Measuring range 0...100 % RH
Accuracy1)
(including hysteresis, non-linearity and repeatability,-15...40 °C (5...104 °F)≤90 % RH ± (1.3 + 0.3%*mv) % RH
traceable to international standards, administrated by NIST, PTB, BEV...)-15...40 °C (5...104 °F)>90 % RH± 2.3 % RH
Freely selectable and scalable outputs 2 x 4-20 mA (2-wire) galvanically isolated RL= (Vcc-9V)/20mA
General
Supply voltage V
Current consumption Max. 20 mA per channel
Protection class of enclosure IP65 / NEMA 4
Cable gland M16 for cable diameter 5 - 10 mm (0.2" - 0.4")
Working temperature range Probe according measuring range
Storage temperature range Electronics and probe -20...60 °C (22...140 °F)
Electromagnetic compatibility EN 61326-1 EN 61326-2-3 ICES-003 ClassB
Material Enclosure stainless steel 1.4404
1) The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was calculated in
accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement).
-25...70 °C (-13...158 °F)± (1.4 + 1%*mv) % RH
Remote probe: -40...180 °C (-40...356 °F)
130
-400-10-20-30105040302060100908070140
110150 170160
120
Wall mountRemote probe
(140)100 (212)°C (°F)
Output 1 must be connected!
= (9+RL*0.02) V DC V
cc min
= 28 V DC RL = load resistor
cc max
M20 for cable diameter 10 - 14 mm (0.4" - 0.6")
Electronics without display -40...60 °C (-40...140 °F)
Electronics with display -20...60 °C
(-4...140 °F)
Industrial Environment FCC Part15 ClassB
Probe cable PTFE
Probe (without filter) stainless steel 1.4301
T [°C]
T [°C]
180
20
Page 4
EE300Ex-M1
v2.8 / Modification rights reserved
Ex - Classications
Europe (ATEX)
Certicate: TPS 13 ATEX 38892 003 X by TÜV SÜD Product Service GmbH
Safety factors: Ui = 28V; Ii = 100mA; Pi = 700mW; Ci = 2.2nF; Li ≈ 0mH
Ex-Designation:
Transmitter without display II 1 G Ex ia IIC T4 Ga / II 1 D Ex ia IIIC T80°C Da
Transmitter with display II 2 G Ex ia IIC T4 Gb / II 1 G Ex ia IIB T4 Ga
Remote probe II 1 G Ex ia IIC T6-T1 Ga / II 1 D Ex ia IIIC T80°C...220°C Da
International (IECEx)
Certicate: IECEx FMG 14.0017 X by FM Approvals
Safety factors: 6.4 Vdc ≤ Ui ≤ 28Vdc; Ii = 100mA; Pi = 700mW; Ci = 2.2nF; Li = 0mH
Ex-Designation:
Transmitter without display Ex ia IIC T4 Ta = -40°C to 60°C Ga / Ex ia IIIC T131°C Da
Transmitter with display Ex ia IIC T4 Ta = -40°C to 60°C Gb / Ex ia IIB T4 Ta = -40°C to 60°C Ga
Remote probe Ex ia IIC T6-T1 Ta = -70°C to 200°C Ga / Ex ia IIIC T80°C Da
Korea (KCs)
Certicate gas:
Remote probe: 20-AV4BO-0253X
Transmitter without display: 20-AV4BO-0254X
Transmitter with display: 20-AV4B0-0257X (EPL Ga - Zone 0)
20-AV4BO-0258X (EPL Gb - Zone 1)
Certicate dust:
Remote probe: 20-AV4BO-0256X
Transmitter without display: 20-AV4BO-0255X
Safety factors: 6.4 V DC ≤ Ui ≤ 28 V DC; Ii = 100 mA; Pi = 700 mW; Ci = 2.2 nF; Li = 0 mH
Ex-Designation:
Transmitter (without display): Ex ia IIC T4 -40°C ≤ Tamb ≤ +60°C
Ex iaD 20 IP6X T131°C -40°C ≤ Tamb ≤ +60°C
Transmitter (with display): Ex ia IIC T4 -40°C ≤ Tamb ≤ +60°C (up to Zone 1)
Ex ia IIB T4 -40°C ≤ Tamb ≤ +60°C (up to Zone 0)
Remote probe: Ex ia IIC T6-T1 / Ex iaD 20 IP6X T80°C -40°C ≤ Tamb ≤ +60°C
21
Page 5
v2.8 / Modification rights reserved
EE300Ex-M1
USA (FM)
Certicate: No. FM17US0302X by FM Approvals
Safety factors: 6.4 Vdc ≤ Vmax (or Ui) ≤ 28Vdc; Imax (or Ii) = 100mA; Pi = 700mW; Ci = 2.2nF; Li = 0mH
Ex-Designation:
Equipment Group I: EE300Ex without display
Class I, II, III, Division 1, Groups A, B, C, D, E, F and G; T4 Ta = -40°C to +60°C; Entity – M1_139080; IP65
Class I, II, III, Division 2, Groups A, B, C, D, E, F and G; T4 Ta = -40°C to +60°C
Class I, Zone 0, AEx ia IIC T4 Ta = -40°C to +60°C Ga; Entity – M1_139080; IP65
Zone 20, AEx ia IIIC T131°C Ta = -40°C to +60°C Da; Entity – M1_139080; IP65
Remote Probe:
Class I, II, III, Division 1, Groups A, B, C, D, E, F and G; T6…T1; Entity – M1_139080; IP65
Class I, II, III, Division 2, Groups A, B, C, D, E, F and G; T6…T1
Class I, Zone 0, AEx ia IIC T6…T1 Ga; Entity – M1_139080; IP65
Zone 20, AEx ia IIIC T80°C Da; Entity – M1_139080; IP65
Equipment Group II: EE300Ex with display
Class I, Division 1, Groups C, and D; T4 Ta = -40°C to +60°C; Entity – M1_139080
Class I, Division 2, Groups A, B, C and D; T4 Ta = -40°C to +60°C; Entity – M1_139080
Class I, Zone 0, AEx ia IIB T4 Ta = -40°C to +60°C Ga; Entity – M1_139080
Class I, Zone 1, AEx ia IIC T4°C Ta = -40°C to +60°C Gb; Entity – M1_139080
Remote Probe:
Class I, II, III, Division 1, Groups A, B, C, D, E, F and G; T6…T1; Entity – M1_139080; IP65
Class I, II, III, Division 2, Groups A, B, C, D, E, F and G; T6…T1
Class I, Zone 0, AEx ia IIC T6…T1 Ga; Entity – M1_139080; IP65
Zone 20, AEx ia IIIC T80°C Da; Entity – M1_139080; IP65
CANADA (FM)
Certicate: No. FM17CA0154X by FM Approvals
Safety factors: 6.4 Vdc ≤ Vmax (or Ui) ≤ 28Vdc; Imax (or Ii) = 100mA; Pi = 700mW; Ci = 2.2nF; Li = 0mH
Ex-Designation:
Equipment Group I: EE300Ex without display
Class I, II, III, Division 1, Groups A, B, C, D, E, F and G; T4 Ta = -40°C to +60°C; Entity – M1_139080; IP65
Class I, II, III, Division 2, Groups A, B, C, D, E, F and G; T4 Ta = -40°C to +60°C
Zone 0, Ex ia IIC T4 Ta = -40°C to +60°C Ga; Entity – M1_139080; IP65
Zone 20, Ex ia IIIC T131°C Ta = -40°C to +60°C Da; Entity – M1_139080; IP65
Remote Probe:
Class I, II, III, Division 1, Groups A, B, C, D, E, F and G; T6…T1; Entity – M1_139080; IP65
Class I, II, III, Division 2, Groups A, B, C, D, E, F and G; T6…T1
Zone 0, Ex ia IIC T6…T1 Ga; Entity – M1_139080; IP65
Zone 20, Ex ia IIIC T80°C Da; Entity – M1_139080; IP65
Equipment Group II: EE300Ex with display
Class I, Division 1, Groups C, and D; T4 Ta = -40°C to +60°C; Entity – M1_139080
Class I, Division 2, Groups A, B, C and D; T4 Ta = -40°C to +60°C; Entity – M1_139080
Zone 0, Ex ia IIB T4 Ta = -40°C to +60°C Ga; Entity – M1_139080
Zone 1, Ex ia IIB T4 Ta = -40°C to +60°C Gb; Entity – M1_139080
Remote Probe:
Class I, II, III, Division 1, Groups A, B, C, D, E, F and G; T6…T1; Entity – M1_139080; IP65
Class I, II, III, Division 2, Groups A, B, C, D, E, F and G; T6…T1
Zone 0, Ex ia IIC T6…T1 Ga; Entity – M1_139080; IP65
Zone 20, Ex ia IIIC T80°C Da; Entity – M1_139080; IP65
The USA and Canada approvals are valid for air and gas measurement only.
22
Page 6
EE300Ex-M1
v2.8 / Modification rights reserved
Ordering Guide EE300Ex-M1
Wall mountT1
Type
1)
Display
Electrical Connection
Probe Cable Length
Probe Length
Process Connection
(Zone Feedthrough)
Hardware Configuration
Filter
Sensing Element
Protection
Ex Approval
5)
Output 1
Scaling 1 lowValueSALvalue
Scaling 1 highValueSAHvalue
Output 2Measurand (xx see measurand code below)MBxx
Scaling 2 lowValueSBLvalue
Scaling 2 highValueSBHvalue
Setup - analogue outputs
1) No display possible for environments with combustible dust, bers and yings and in
gases with EPL Ga IIC (EX1/EX2) / Gas Groups A, B for Division 1 (EX3/EX9) / Zone
0 IIC (EX5)
2) Allowed only in combination with PA0
Remote probe with cut-in fitting, pressure tight, 20 bar
Remote probe with sliding fitting, pressure tight, 20 bar
Without displayD0
With displayD1
2 x M16 cable glandE2
1/2" NPT conduitE13
2 x M20 cable glandE15
Wall mountK0
(3.3 ft)K1K1
1 m
(6.6 ft) K2K2
2 m
(16.4 ft) K5K5
5 m
(32.8 ft)K10K10
10 m
Wall mount, 50 mm
(2.56ʺ)
65 mm
(3.95")L100
100 mm
(7.9ʺ)L200L200
200 mm
(15.8")L400L400
400 mm
(1.97ʺ)L50
2)
Without probe fittingPA 0PA0
G1/2" ISO - cut-in fitting, Ø 12 mm
1/2" weld cut-in fitting, Ø 12 mm
1/2" NPT - cut-in fitting, Ø 12 mm
G1/2" ISO - sliding fitting, Ø 13 mm
1/2" NPT - sliding fitting, Ø 13 mm
Stainless steel sinteredF4F4F4
3)
PTFE
Stainless steel grid, stainless steel body, up to 180 °CF9F9F9
Catalytic fo H
sterilisation
2O2
Stainless steel with boreholes Ø 3 mm
Without coatingC0C0
With coating
4)
ATEX (Europe)EX1
IECEx (International)EX2
FM (USA)EX3
KCs (Korea)EX5
FM (Canada) EX9
Measurand (xx see measurand code below)MAxx
EE300Ex-M1A6HS2
(300 psi)T7
(300 psi)T10
L65
(0.47")PA20
(0.47")PA21
(0.47")PA22
(0.51")PA23
(0.51")PA25
F5F5F5
3)
(0.12")F13F13
F12F12F12
C1C1C1
3) May not be used in EPL Ga IIC (EX1/EX2) / Gas Groups A, B for Division 1 (EX3/
EX9) / Zone 0 IIC (EX5)
4) Not appropriate for moisture in oil measurement, obligatory for all other applications,
free of charge
5) Assign the most relevant measurand to output 1
Measurand Code for output 1 and 2 in the ordering guide
Relative humidity%10
Temperature
Dew point Td
Frost point Tf
Mixing ratio r
Absolute humidity dv
1) For approval FM (USA / Canada) not allowed.
MAxx / MBxx
°C1
°F2
°C52
°F53
°C65
°F66
g/kg60
gr/lb61
g/m³56
gr/ft³57
Wet bulb temperature Tw
Water vapour partial pressure e
Specific enthalpy h
Water activity
1)
Water content X in mineral transformer oil
Water content X in customer specific oil
°C54
°F55
mbar50
psi51
kJ/kg62
BTU/lb64
aw67
1)
ppm70
1)
ppm70PPMxxx
MAxx / MBxx
23
Page 7
v2.8 / Modification rights reserved
EE300Ex-M1
Order Example
Example 1:Example 2:
EE300Ex-M1A6HS2T7D1E2K10L200PA20F4C1EX1/
MA1SAL-40SAH180MB10SBL0SBH100
Type: Remote probe up to 20 bar (300 psi)
Display: With display
Electrical Connection: 2 x M16 cable gland
Probe Cable: 10 m
Probe Length: 200 mm (7.9")
Process connection
(Zone Feedthrough): G1/2" ISO - cut-in fitting, Ø 12 mm
Filter: Stainless steel sintered
Sensing Element Protection: With coating
Ex Approval: ATEX (Europe)
Output 1: Temperature [°C]
Scaling Output 1: -40...180 °C
Output 2: Relative humidity [% RH]
Scaling Output 2: 0...100 % RH
(32.8 ft)
(0.47")
EE300Ex-M1A6HS2T1D0E2K0L50PA0F9C1EX3/
MA2SAL-40SAH140MB53SBL-40SBH140
Type: Wall mount
Display: Without display
Electrical Connection: 2 x M16 cable gland
Probe Cable: Wall mount
Probe Length: Wall mount, 50 mm
Process connection:
(Zone Feedthrough): Without probe fitting
Filter: Stainless steel grid
Sensing Element Protection: With coating
Ex Approval: FM (USA)
Output 1: Temperature [°F]
Scaling Output 1: -40...140 °F
Output 2: Dew point temperature [°F]
Scaling Output 2: -40...140 °F
Accessories
Blank cover for enclosure base HA011401
Safety barrier, 1-channel, STAHL 9002/13-280-093-001 HA011410
Intrinsically safe supply unit, 1-channel, STAHL 9160/13-11-11 HA011405
Intrinsically safe supply unit, 2-channel, STAHL 9160/23-11-11 HA011406
Sealing plug for unused M16 cable glands HA011402
Sealing plug for unused M20 cable glands HA011404
Ball valve with 1/2 ISO female thread, ATEX certied HA011403
Product Conguration SoftwareEE-PCS (free download: www.epluse.com/congurator)
Adapter Kit for conguration and adjustment
(must be ordered together, see datasheet EE-PCA):
Pos. 1: Product Conguration Adapter EE-PCA
Pos. 2: Connection cable HA011068
(1.97ʺ)
24
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.