E+E EE23 Specifications

Page 1
EE23
The EE23 is optimized for reliable and cost effective use in industrial applications. In addition to highly accurate measurement of relative humidity (RH) and temperature (T), the sensor also calculates the dew point (Td) and the frost point temperature (Tf).
Measurement Performance
The EE23 employs high-end E+E humidity sensing elements manufactu-
red in state-of-the-art thin lm technology, which are the prerequisite for
outstanding accuracy.
Long Term Stability
lter cap, the EE23 tackles even challenging industrial applications.
Humidity / Temperature Sensor for
Industrial Applications
Model T1
Outputs and Power Supply
The measured data is available on two voltage or current outputs as well as on the display. Additional features like alarm (relay) output and integrated supply module 100…240V AC facilitate the use of the EE23 in a wide range of applications.
Easy Installation and Service
The enclosure consists of the back cover with the terminals for wiring,
the pluggable active part with the electronics and the probe, and the front cover. Once installed, the active part of EE23 can be plugged on and off without rewiring. The plastic enclosure is appropriate also for mounting onto DIN rails.
Remote Probe and Accessories
The remote probe with cable length up to 20m (65.6 ft) together with a
wide choice of accessories such as mounting anges or brackets, drip
water protection or radiation shield allow for easy integration of the EE23
into any measurement task.
User Congurable
The user can easily perform a two-point humidity and temperature
adjustment. The analogue and alarm outputs can be freely congured.
Model T2
Model T4/T5
Model T6
Features
• Temperature range -40...180 °C (-40...356 °F)
• Outstanding long term stability
• Calculation of dew point and frost point temperature
• Easy mounting and maintenance
• Alarm output
Inspection certicate according to DIN EN 10204 – 3.1
Protective Sensor Coating
The E+E proprietary sensor coating is a protective layer applied to the active surface and leads of the sensing elements. The coating extends substantially the lifetime and the measurement performance of the E+E sensor in corrosive environment. Additionally, it improves the sensor’s long term stability in dusty, dirty or oily applications by preventing stray impedances caused by deposits on the active sensor surface.
24
v5.1 / Modification rights reserved
EE23
EE23
Page 2
Technical Data
Measurands
Relative Humidity
Working range 0...100% RH
Accuracy1) (including hysteresis, non-linearity and repeatability, traceable to intern. standards, administrated by NIST, PTB, BEV...) EE23-T1/T2/T4/T5 EE23-T6
-15...40°C (5...104°F) 90% RH ± (1.3 + 0.3%*mv) % RH ± (1.8 + 0,3%*mv) % RH
-15...40°C (5...104°F) >90% RH ± 2.3% RH ± 2.8% RH
-25...70°C (-13...158°F) ± (1.4 + 1%*mv) % RH ± (1.9 + 1%*mv) % RH
-40...180°C (-40...356°F) ± (1.5 + 1.5%*mv) % RH ± (2 + 1.5%*mv) % RH Temperature dependence electronics typ. ± 0.015% RH/°C
Response time t
Temperature
Probe working range EE23-T1 -40...60°C (-40...140°F) EE23-T2/T6 -40...80°C (-40...176°F)
EE23-T4 -40...120°C (-40...248°F)
EE23-T5 -40...180°C (-40...356°F)
Accuracy
Temperature dependence of electronics typ. 0.002°C/°C
Output Scale Span
Humidity RH 0 100 100 100 100 % RH Temperature T -40 (-40) 60 Dew point temperature Td -40 (-40) 60 (140) 80 (176) 100 (212) 100 (212) °C (°F) Frost point temperature Tf -40 (-40) 0 (32) 0 (32) 0 (32) 0 (32) °C (°F)
with metal grid filter at 20 °C (68 °F) < 15 sec.
90
∆°C ∆°C
EE23-T1/T2/T4/T5
from up to units
EE23-T1 EE23-T2/T6 EE23-T4 EE23-T5
(140) 80 (176) 120 (248) 180 (356) °C (°F)
EE23-T6
°C°C
Outputs 0 - 1 V -0.5 mA < I
0 - 5 / 0 - 10 V -1 mA < IL < 1 mA
< 0.5 mA
L
0 - 20mA / 4 - 20 mA RL < 470 Ohm
General
Supply voltage for 0 -1 V / 0 - 5 V outputs 10.5 - 35V DC or 12 - 28V AC for 0 - 10 V / 0 - 20 mA / 4-20 mA outputs 15.0 - 35V DC or 15 - 28V AC
100...240V AC, 50/60Hz supply module (optional) Current consumption for voltage output for DC supply 25 mA (with alarm module 35 mA) for AC supply ≤ 45 mA Current consumption for current output for DC supply 55 mA (with alarm module 65 mA) for AC supply ≤ 100 mA Enclosure / protection class PC or Al Si 9 Cu 3 / IP65; NEMA 4
Cable gland M16x1.5 cable Ø 4.5 - 10 mm (0.18 - 0.39”)
Electrical connection screw terminals max. 1.5 mm² (AWG 16)
Working temperature range of electronics -40...60°C Working temperature range with display -30...60°C (-22...140°F)
Storage temperature range -40...60°C (-40...140°F) CE compatibility according EN61326-1 EN61326-2-3 ICES-003 ClassB Industrial Environment FCC Part15 ClassB
Alarm Module
2)
Output SPDT-Switch max. 250V AC/8A or 28V DC/8A
threshold hysteresis
Setting range 10...95% RH 3...15% RH Setting accuracy ± 3% RH
1) The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was calculated in accordance with
EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement).
2) only for models T1, T2, T4 and T6.
EE23
v5.1 / Modification rights reserved
(with alarm module 70 mA
eff
(with alarm module 120 mA
eff
(-40...140°F)
)
eff
)
eff
25
Page 3
Dimensions in mm (inch)
135
66.5
Enclosure:
Polycarbonate (PC) Metal
(5.3”)
(2.6”)
135
(5.3”)
66.5
(2.6”)
90 (3.5”)
Probes:
50 (2")
39 (1.54")
12 (0.5")
EE23-T1
Probe material: PC
code “probe length”
EE23-T2
Probe material: stainless steel
32 (1.3")
90 (3.5”)
163 (6.4”)
5 (0.2")
code “cable length”
20
(0.8")
(0.6")
code “cable length”
40
(1.6")
15
32
(1.3")
EE23-T4 / EE23-T5
12 (0.5")
Probe material: stainless steel
code “probe length”
EE23-T6
12 (0.5")
Probe material: stainless steel
Accessories (Additional information see data sheet “Accessories”)
- Mounting flange (HA010201)
- Mounting flange 5mm for model T6 only (HA010208)
- Bracket for installation onto mounting rails* (HA010203)
- Drip water protection (HA010503)
- Radiation shield (HA010502)
- Calibration set
- Stainless steel wall mounting clip Ø12 mm
*Note: Only for plastic enclosure
26
(see data sheet „Calibration Kit“) (HA0104xx)
(0.5”) (HA010225)
v5.1 / Modification rights reserved
EE23
Page 4
Ordering Guide
1)
Model
Enclosure
Filter
Cable length
(incl. probe length)
Probe length
Hardware Configuration
Electrical connection
Optional features
Output Signal
1)
Output 1
Scaling 1 low
Scaling 1 high
Output 2
Scaling 2 low value SBLvalue
Setup - Analogue outputs
Scaling 2 high value SBHvalue
Display mode
Measurand Code
relative humidity % 10
temperature
1) For T1, T2 T4 and T6 adjustment changes on the electronics board- see operation manual For T5 adjustment and configuration changes by E+E PCS Software only - see operation manual
2) Alarm output only available with cable glands (plug option is not possible) / combination alarm output and integrated power supply is not possible
3) Integrated power supply includes 2 plugs for power supply and outputs / combination alarm output and integrated power supply is not possible
4) Measurand on display can be selected with push buttons
polycarbonate metal (Al Si 9 Cu 3)
plastic - metal grid (up to 120 °C / 248 °F) F3 F3 F3 F3
stainless steel sintered no code no code no code no code PTFE F5 F5 F5 F5
stainless steel grid (up to 180 °C / 356 °F) F9
H
2O2
stainless steel membrane Ø 5 mm F17 2 m (6.6 ft) K2 K2 K2 5 m (16.4 ft) K5 K5 K5
10 m (32.8 ft) K10 K10 K10
20 m (65.6 ft) K20 K20
40 mm (1.57ʺ) L40 65 mm (2.55ʺ) L65 L65 L65 200 mm (7.87ʺ) no code no code no code 400 mm (15.75ʺ) L400 L400 L400
cable glands no code 1 plug for power supply and outputs
2) 3)
LC Display D1 D1 D1 D2 E+E sensor coating C1 C1 C1 C1 alarm outputs for RH integrated power supply 100...240 V AC, 50/60 Hz
2)
3)
0-1 V GA1 0-5 V GA2 0-10 V GA3 0-20 mA GA5 4-20 mA GA6 relative humidity RH [%] no code other measurand (xx see measurand code below) MAxx 0 no code value SALvalue 100 no code value SAHvalue temperature T [°C] no code temperature T [°F] MB2 other measurand (xx see measurand code below) MBxx
measurand output 1+2 alternating DT2 DT2 DT2 DT2 measurand output 1 DT3 DT3 DT3 DT3 measurand output 2 DT4 DT4 DT4 DT4
xx
dew point Td
°C 1 °F 2
frost point Tf
T1
wall mount
T2
duct mount
T4
remote probe up to
120 °C (248 °F)
T5
remote probe up
to 180 °C (356 °F)
miniature probe
no code
HS3
F12 F12 F12 F12
EE23
E4
4)
AM2 AM2 AM2 AM2 AM3 AM3 AM3 AM3 AM3
xx
°C 52 °F 53 °C 65 °F 66
T6
D1
Order Example
EE23-T4HS3F3K2D1GA2SBL0SBH50DT2
Type: remote probe up to 120 °C (248 °F)
Enclosure: metal (Al Si 9 Cu 3) Filter: plastic - metal grid Cable length: 2 m (6.6 ft)
Probe length: 200 mm (7.87ʺ)
Electrical connection: cable glands Optional feature: LC Display
EE23
v5.1 / Modification rights reserved
Output Signal: 0-5 V
Output 1 relative humidity [%]
Scaling 1 low: 0 Scaling 1 high: 100
Output 2: temperature [°C]
Scaling 2 low: 0 Scaling 2 high: 50 Display mode: measurand output 1+2 alternating
27
Loading...