E+E EE08 Specifications

EE08
High-Precision Miniature
Humidity / Temperature Transmitter
Accurate humidity / temperature measurement over a wide working range, fitted in a small-sized housing and high flexibility have been the main goals for the development of the EE08 series. Low power consumption and short start-up time support efficient energy management for battery operated systems. For this application an additional version (V10) with supply voltage 4.5-15 V DC has been developed. Calibration data and other relevant functions like linearization or temperature compensation are stored in the probe. This feature, together with the optional connector, allows for easy replacement of the probe without a need for re-adjust­ment of the reading device (interchangeability). The humidity and temperature measurement are available as analogue outputs (0-1/2.5/5 V) and as a digital interface (E2-interface). Easy implementation and data processing is warranted. Humidity and temperature reading can be re-adjusted using the calibration software; available as an accessory. The configuration equip­ment allows humidity and temperature adjustment of the sensor.
Typical Applications Features
meteorology / weather stations humidity / temperature data logging incubators fermentation chambers green houses snow machines dry storage facilities
with cable
with connector
small dimensions
wide working range, high accuracy
traceable calibration
customer adjustment possible
interchangeable in seconds
low power consumption / short start-up time
analogue outputs / digital interface
Technical Data
Measuring values
Relative Humidity
Sensor HC101 Working range Digital output (2 wire) Analogue output 0...100 % RH 0-1/2.5/5/10 V -0.2 mA < I Accuracy at 20 °C Temperature dependence typ. 0.03 % RH/°C
Temperature
Sensor Pt 1000 (DIN A) Digital output (2 wire) Analogue output 0-1/2.5/5/10 V -0.2mA < I Accuracy at 12/24V DC
General
Supply voltage output 0-1 V / 0-2.5 V 4.5-15 V DC or 7-30 V DC output 0-5 V 7-30 V DC output 0-10 V 12-30 V DC
Current consumption typ. < 1.3 mA Digital interface E2-interface level = 3.3 V / ±0.1 V Housing polycarbonate / IP65 Sensor protection metal grid filter Electromagnetic compatibility EN61326-1 EN61326-2-3 Industrial Environment Temperature ranges working temperature: -40...80 °C storage temperature: -40...80 °C (-40...176 °F)
1) refer to the working range of the humidtiy sensor HC101 2) serial protocol refer to www.epluse.com *) The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation).
The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement).
68
1)
0...100 % RH
2)
output value: 0.00...100.00 % RH
)
(68 °F) and 12 V DC
2)
output value: -40.00...+80.00 °C (-40...176 °F)
*
±2 % RH (0...90 % RH) ±3 % RH (90...100 % RH)
°C
(typ. 0.02 % RH/°F)
< 0.2 mA
L
< 0.2 mA
L
°C
(-40...176 °F)
v2.0 / Modification rights reserved
EE08
EE08
Dimensions (mm)
EE08 with connector (Type D)EE08 with cable (Type E)
M12x161 (2.4”)
83 (3.3”)
M16x1.5
61
(2.4”)
71 (2.8”)
12 (0.5”)
25 (1”)
73 (2.9”)
12 (0.5”)
Connection Diagram
Type E:
Temp. active Temp. passive, 4-wire
T-passive white T-passive blue GND pink pink T-out grey grey RH-out yellow yellow SCL green green
E2-inferface
}
SDA brown brown +UB red red
(not connected) white, black
(not connected) blue, violet
(not connected)
Type D:
1 T-passive 2 SDA 3 SCL 4 RH-out 5 T-out 6 GND 7 T-passive
8 +UB
}
E2-interface
Ordering Guide
HOUSING MODEL OUTPUT SUPPLY
polycarbonate (P) humidity active / temperature active (FT) 0 - 1 V1) (1) 4.5 - 15 V DC (V10) Pt 100 DIN A (A) with connector (D)
humidity active / temperature passive (FP) 0 - 2.5 V
0 - 5 V 0 - 10 V
1)
(7) 7 - 30 V DC (V11) Pt 1000 DIN A (C) with cable (E)
2)
(2)
2)
(3)
T-SENSOR
(passive, 4-wire)
EE08-
FILTER COATING CABLE LENGTH
metal grid filter (6) without coating (no code) 1 m (3.3ft) (01) metric (no code) -40...80 (T22)
with coating (HC01) 2 m
(Type E only)
(6.6ft) (02) non metric (E01) -40...60 (T02) (16.4ft) (05) -30...70 (T08)
5 m
T-UNIT
T-SCALING
-20...80 (T24)
-20...50 (T48) other (Txx)
1) possible with supply 4.5 - 15 V DC (V10) or 7 - 30 V DC (V11)
2) possible with supply 7 - 30 V DC (V11) only
3) T-Sensor details see www.epluse.com/R-T_Characteristics
3)
TYPE
connector M12 (8-pole)
Order Example
EE08-PFT2V11E602T22
housing: polycarbonate model: humidity active / temp. active output: 0 - 5V supply: 7 - 30V DC type: with cable
Scope of Supply
- EE08 Transmitter according to ordering guide
- Inspection certifi cate according to DIN EN10204 - 3.1
Accessories / Replacement Parts
- M12 connection cable for type D, length 1.5 m (5 ft) (HA010322)
- M12 connection cable for type D, length 3 m
- M12 connection cable for type D, length 5 m
- M12 connection cable for type D, length 10 m (32.8 ft) (HA010325)
- Radiation shield for Type E (HA010502)
- Radiation shield for Type D (HA010506)
Configuration equipment: The confi guration equipment allows humidity and
temperature adjustment of the sensor.
- configuration cable (HA011005)
- configuration software: free download under www.epluse.com/EE08
EE08
v2.0 / Modification rights reserved
(10 ft) (HA010323) (16.4 ft) (HA010324)
filter: metal grid filter coating: without cable length: 2m (6.6ft) T-unit: metric T-scaling: -40...80°C
- Protection cap for 12 mm probe (HA010783)
- M12 female socket with wires (HA010703)
- M12 female cable connector assembly possible (HA010704)
- metal grid filter (HA010113)
HA011005
PC
69
Loading...