![](/html/10/1025/10255eedf662aa57496e9ada6e87ea80ea56a8dc15e1430828b69bb7da544394/bg1.png)
EE08
High-Precision Miniature
Humidity / Temperature Transmitter
Accurate humidity / temperature measurement over a wide
working range, fitted in a small-sized housing and high
flexibility have been the main goals for the development of the
EE08 series.
Low power consumption and short start-up time support
efficient energy management for battery operated systems.
For this application an additional version (V10) with supply
voltage 4.5-15 V DC has been developed.
Calibration data and other relevant functions like linearization
or temperature compensation are stored in the probe.
This feature, together with the optional connector, allows for
easy replacement of the probe without a need for re-adjustment of the reading device (interchangeability).
The humidity and temperature measurement are available as analogue outputs (0-1/2.5/5 V) and as a digital
interface (E2-interface). Easy implementation and data processing is warranted. Humidity and temperature
reading can be re-adjusted using the calibration software; available as an accessory. The configuration equipment allows humidity and temperature adjustment of the sensor.
Typical Applications Features
meteorology / weather stations
humidity / temperature data logging
incubators
fermentation chambers
green houses
snow machines
dry storage facilities
with cable
with connector
small dimensions
wide working range, high accuracy
traceable calibration
customer adjustment possible
interchangeable in seconds
low power consumption / short start-up time
analogue outputs / digital interface
Technical Data
Measuring values
Relative Humidity
Sensor HC101
Working range
Digital output (2 wire)
Analogue output 0...100 % RH 0-1/2.5/5/10 V -0.2 mA < I
Accuracy at 20 °C
Temperature dependence typ. 0.03 % RH/°C
Temperature
Sensor Pt 1000 (DIN A)
Digital output (2 wire)
Analogue output 0-1/2.5/5/10 V -0.2mA < I
Accuracy at 12/24V DC
General
Supply voltage output 0-1 V / 0-2.5 V 4.5-15 V DC or 7-30 V DC
output 0-5 V 7-30 V DC
output 0-10 V 12-30 V DC
Current consumption typ. < 1.3 mA
Digital interface E2-interface level = 3.3 V / ±0.1 V
Housing polycarbonate / IP65
Sensor protection metal grid filter
Electromagnetic compatibility EN61326-1 EN61326-2-3
Industrial Environment
Temperature ranges working temperature: -40...80 °C
storage temperature: -40...80 °C (-40...176 °F)
1) refer to the working range of the humidtiy sensor HC101 2) serial protocol refer to www.epluse.com
*) The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation).
The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement).
68
1)
0...100 % RH
2)
output value: 0.00...100.00 % RH
)
(68 °F) and 12 V DC
2)
output value: -40.00...+80.00 °C (-40...176 °F)
*
±2 % RH (0...90 % RH) ±3 % RH (90...100 % RH)
∆°C
(typ. 0.02 % RH/°F)
< 0.2 mA
L
< 0.2 mA
L
°C
(-40...176 °F)
v2.0 / Modification rights reserved
EE08
EE08
![](/html/10/1025/10255eedf662aa57496e9ada6e87ea80ea56a8dc15e1430828b69bb7da544394/bg2.png)
Dimensions (mm)
EE08 with connector (Type D)EE08 with cable (Type E)
M12x161 (2.4”)
83 (3.3”)
M16x1.5
61
(2.4”)
71 (2.8”)
∅12 (0.5”)
25 (1”)
73 (2.9”)
∅12 (0.5”)
Connection Diagram
Type E:
Temp. active Temp. passive, 4-wire
T-passive white
T-passive blue
GND pink pink
T-out grey grey
RH-out yellow yellow
SCL green green
E2-inferface
}
SDA brown brown
+UB red red
(not connected) white, black
(not connected) blue, violet
(not connected)
Type D:
1 T-passive
2 SDA
3 SCL
4 RH-out
5 T-out
6 GND
7 T-passive
8 +UB
}
E2-interface
Ordering Guide
HOUSING MODEL OUTPUT SUPPLY
polycarbonate (P) humidity active / temperature active (FT) 0 - 1 V1) (1) 4.5 - 15 V DC (V10) Pt 100 DIN A (A) with connector (D)
humidity active / temperature passive (FP) 0 - 2.5 V
0 - 5 V
0 - 10 V
1)
(7) 7 - 30 V DC (V11) Pt 1000 DIN A (C) with cable (E)
2)
(2)
2)
(3)
T-SENSOR
(passive, 4-wire)
EE08-
FILTER COATING CABLE LENGTH
metal grid filter (6) without coating (no code) 1 m (3.3ft) (01) metric (no code) -40...80 (T22)
with coating (HC01) 2 m
(Type E only)
(6.6ft) (02) non metric (E01) -40...60 (T02)
(16.4ft) (05) -30...70 (T08)
5 m
T-UNIT
T-SCALING
-20...80 (T24)
-20...50 (T48)
other (Txx)
1) possible with supply 4.5 - 15 V DC (V10) or
7 - 30 V DC (V11)
2) possible with supply 7 - 30 V DC (V11) only
3) T-Sensor details see
www.epluse.com/R-T_Characteristics
3)
TYPE
connector M12
(8-pole)
Order Example
EE08-PFT2V11E602T22
housing: polycarbonate
model: humidity active / temp. active
output: 0 - 5V
supply: 7 - 30V DC
type: with cable
Scope of Supply
- EE08 Transmitter according to ordering guide
- Inspection certifi cate according to DIN EN10204 - 3.1
Accessories / Replacement Parts
- M12 connection cable for type D, length 1.5 m (5 ft) (HA010322)
- M12 connection cable for type D, length 3 m
- M12 connection cable for type D, length 5 m
- M12 connection cable for type D, length 10 m (32.8 ft) (HA010325)
- Radiation shield for Type E (HA010502)
- Radiation shield for Type D (HA010506)
Configuration equipment: The confi guration equipment allows humidity and
temperature adjustment of the sensor.
- configuration cable (HA011005)
- configuration software: free download under www.epluse.com/EE08
EE08
v2.0 / Modification rights reserved
(10 ft) (HA010323)
(16.4 ft) (HA010324)
filter: metal grid filter
coating: without
cable length: 2m (6.6ft)
T-unit: metric
T-scaling: -40...80°C
- Protection cap for 12 mm probe (HA010783)
- M12 female socket with wires (HA010703)
- M12 female cable connector
assembly possible (HA010704)
- metal grid filter (HA010113)
HA011005
PC