EE071 is optimized for use in demanding OEM applications.
In addition to the precise measurement of humidity (RH) and
temperature (T), the EE071 calculates physical quantities such
as dew point temperature, mixing ratio and absolute humidity.
All measured and calculated values are available on the
RS-485 interface with Modbus RTU protocol.
The RH and T sensor HCT01 is perfectly protected against dust
and dirt by the E+E proprietary coating. Furthermore, all solder
pads are sealed against corrosion. With the appropriate filter
cap the EE071 offers outstanding long term stability even in
harsh environment. The compact design with M12 connector
allows for easy installation and fast replacement of the probe.
With the optional Modbus configuration adapter the user can
perform RH and T adjustment and set the Modbus parameters.
Humidity and Temperature Probe
with Modbus Interface
Metall
Polycarbonate
sealed
solder pads
EE071
Sensor coating
Typical Applications
process and climate technology
agriculture, stables
incubators, hatchers
outdoor measurement
storage rooms, cooling chambers
Technical Data
Measured values
Relative Humidity
Sensor element HCT01-00D
Modbus output range 0.00...100.00 % RH
Accuracy incl. hysteresis and nonlinearity ±2 % RH (0...90 % RH) ±3 % RH (90...100 % RH)
Temperature dependence < (0.025 + 0.0003 x RH) [% RH/°C]
Temperature
Sensor Pt1000
Modbus output range -40.00...+80.00 °C
Accuracy:
Standard
Key Features
highest accuracy
excellent protection against pollution
outstanding long term stability
temperature compensation
low power consumption
calculated physical quantities
(-40...176 °F)
High
0.4
0.3
0.2
0.1
-0.1
-0.2
-0.3
-0.4
-0.48
General
Supply voltage
Current consumption typ. 0.4 mA at a measuring rate of 1 sec.
Current pulse during power-up at UB 7 V: I
(with serial resistance 100 Ohm) at UB 12 V: I
Warmup Time after Power-Up max. 800ms
Interface / Bus RS485 / Modbus in slavemode
Housing / polycarbonate or stainless steel / IP65
Electromagnetic compatibility
FCC Part 15 Class B ICES-003 Issue 5 ClassB
Working and storage temperature -40...80°C
Max. cable length 100m (328.1ft)
1) For bus operation with terminal resistor (120Ω) min. UB: 4,5V DC
2) No terminal, pull-up or pull-down resistor integrated in the probe
3) EE071 is not protected against voltage spikes (surge)
42
1) 2)
4 - 28 V DC
3)
EN613226-1 EN61326-2-3
www.epluse.com/EE071
60 mA; current draw drops below 10 mA within 350 μs
max
110 mA; current draw drops below 10 mA within 400 μs
max
(-40...176°F)
v1.19 / Modification rights reserved
EE071
EE071
Page 2
Dimensions in mm (inch)
polycarbonate housing - EE071-HTPx
(3”)
(0.3”)
8
75
34 (1.34”)
metal housing - EE071-HTMx
(0.3”)
8
L
1
L
2
13
(0.51”)
M12x1
12
(0.47”)
12
(0.47”)
M12x1
FilterL
Stainless steel grid 79.5 mm (3.13“)38.5 mm (1.52“)
H2O
2
1
73.5 mm (2.89“) 33 mm (1.3“)
Connection Diagram
EE071:
1...+UB
2...B-RS485 (= Data-)
3...A-RS485 (= Data+)
4...GND
M12x1 flange (HA010705, Accessories)
brown...+UB
white.....B-RS485 (= Data-)
blue......A-RS485 (= Data+)
black....GND
grey......shielding
M16x1.5
Modbus Map
The measured values are saved as a 32Bit float value from 0x19 to 0x25 and as 16Bit signed integer
between 0x27 and 0x2D.
The factory setting for the Slave-ID is 247 as an integer 16Bit value. This ID can be customised in the
register 0x00 (value margin 1 - 247 permitted).
The serial number as ASCII-code is located at register address 30001-30008.
FLOAT (read register):INTEGER (read register):
Register
address
300260x19Temperature [°C]
300280x1BTemperature [°F]
300300x1DRel Humidity [%]
300320x1FAbs Humidity [g/m³]
300340x21Dew Point [°C]
300360x23Dew Point [°F]
300380x25Mixing ratio [g/kg]
1) Values are stored with a scaling of 1:100 (e.g.: 2550 is equivalent to 25.5°C)
2) Read function code: 0x03 Write function code: 0x10
3) Ambient pressure in mbar, with 2 decimal digits (e.g. 1008.25)
Protocol
address
Parameter
name
Register
address
300400x27Temperature [°C]
300410x28Temperature [°F]
300420x29Rel Humidity [%]
300430x2AAbs Humidity [g/m³]
300440x2BDew Point [°C]
300450x2CDew Point [°F]
300460x2DMixing ratio [g/kg]
Protocol
address
Parameter
name
1)
For Modbus protocol setting please see Application Note (www.epluse.com/EE071).
INTEGER (write register):
Register
address
600010x00Slave-ID
Protocol
address
FLOAT (read & write register):
Register
address
5001
Protocol
address
2)
0x1388Air pressure
L
2
12
(0.47”)
Parameter
name
Parameter
name
3)
Radiation shield
For outdoor applications EE071 must be used with the optional
radiation shield HA010502, which protects the device against rain, snow,
ice and solar radiation.
EE071
v1.19 / Modification rights reserved
www.epluse.com/EE071
EE071 with
radiation shield
HA010502
43
Page 3
E+E Sensor Coating
The E+E proprietary sensor coating is a protective layer applied to the active surface of the HCT01 sensing
element. The coating extends substantially the lifetime and the measurement performance of the E+E sensor
in corrosive environment. Additionally, it improves the sensor’s long term stability in dusty, dirty or oily applications by preventing stray impedances caused by deposits on the active sensor surface.
Humidity and Temperature (HT) polycarbonate (P) membrane (B) Standard (x) 9600 (A) odd (O) 1 stopbit (1)
1)
(M) metal grid (C) High (C) 19200 (B) even (E) 2 stopbits (2)
metal
PTFE (E)38400 (C) no parity (N)
1)
(L)
H
2O2
stainless steel grid
1)
(I)
EE071-
1) The metal housing (M) is only available with stainless steel grid filter and with H2O2 filter (L). The stainless steel grid filter is only available with metal housing (M).