E+E EE061 Specifications

EE061
The EE061 probe is ideal for cost-effective, accurate and reliable measurement of relative humidity (RH) and temperature (T) in OEM applications. The measured RH data is available as 4…20 mA, 2-wire output. The device is available also with an additional 4-wire passive T output.
The EE061 features high quality E+E RH sensing elements, which stand for outstanding measurement performance and high resistance to chemicals such as chlorine and ammonia.
The combination of IP65 protection class and E+E proprietary sensor protection lead to outstanding long term stability even in polluted environment.
Protective Sensor Coating
The E+E proprietary sensor coating is a hygroscopic layer applied to the active surface of the RH sensing element. The coating extends substantially the life-time and the measurement performance of the E+E sensor in corro-
sive environment (salts, o󰀨-shore applications). Additionally, it improves the
sensor’s long term stability in dusty, dirty or oily applications by preventing stray impedances caused by deposits on the active sensor surface.
Humidity / Temperature Probe
with Current Output
sensor coating
Typical Applications Features
stables green houses humidifiers and dehumidifiers monitoring of storage rooms
excellent price / performance ratio
very good long term stability
easy installation
compact design
Technical Data
Measurands
Relative humidity
Working range 0...100 % RH Analogue output 0...100 % RH 4...20 mA (2-wire) RL<500 Ohm Accuracy at 20 °C ±5 % RH (0...10 % RH and 90...100 % RH) Temperature dependence typ. ±0.03 % RH/°C
(68 °F), 12 V DC
Temperature (passive)
Output resistive, 4-wire Choice of T-sensor according to ordering guide
General
Supply voltage 9 V DC - 28 V DC Current consumption typ. 1.5 mA Electrical connection cable PVC 0.5 m model M2: 2 x 0.50 mm² model M6: 8 x 0.14 mm² Enclosure material Polycarbonate Protection class IP65 Electromagnetic compatibility EN61326-1 EN61326-2-3 Working temperature range -40...+60 °C Storage temperature range -40...+60 °C (-40...140 °F)
1) Traceable to intern. standards, administrated by NIST, PTB, BEV,... The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement).
90
1)
±3 % RH (10...90 % RH)
(1.6 ft) / 3 m (9.8 ft) / 10 m (32.8 ft), with wire ferrules
(-40...140 °F)
v3.3 / Modification rights reserved
EE061
EE061
Dimensions in mm (inch)
Ordering Guide
Model
T-sensor passive
(see www.epluse.com/R-T_Characteristics)
Filter
Coating
Hardware Configuration
Cable length
Order Example
EE061-M6TP1F3C1KL300
131 (5.2")
34 (1.3")
12
(0.5")
EE061-
humidity M2 humidity with temperature passive M6 Pt100 DIN A TP1 Pt1000 DIN A TP3 NTC 10k ±1 %, B25/100 = 3950 k TP5 membrane F2 metal grid F3 without coating no code with coating C1
0.5 m
(1.6 ft) no code
3 m
(9.8 ft) KL300
10 m
(32.8 ft) KL1000
Model: humidity with temperature passive T-sensor passive: Pt 100 DIN A Filter: metal grid Coating: with coating Cable length: 3 m
(9.8 ft)
Accessories
Plastic mounting flange Ø12 mm (0.47“) black HA010214 Wall mounting clip Ø12 mm Protection cap for Ø12 mm
(0.47“) HA010211
(0.47“) probe HA010783
EE061
v3.3 / Modification rights reserved
91