System Setup
1
VolumeView sensor
2
VolumeView femoral arterial catheter
3
VolumeView thermistor manifold
4
Edwards Oximetry Central Venous Catheter
5
TruWave pressure transducer Oximetry cable
6
4
2
EV1000 Clinical Platform Startup with the VolumeView Set
Before startup, please ensure monitor, cables and patient cables are all connected.
1. Turn on monitor and databox. System self-test starts.
2. Enter new patient data OR continue with existing patient.
3. Set up and prime VolumeView sensor and TruWave CVP transducer and connect to databox.
4. Ensure the sensor and transducer are level to patient’s phlebostatic axis and zero transducers on monitor using Clinical Actions button.
5. Choose continuous parameters, and defi ne alarms/target range.
6. Start transpulmonary thermodilution ­inject bolus and review and edit boluses results.
7. Monitor patient in real time with one of the six available
screens.
Thermodilution Bolus - VolumeView Set
1. Insert the VolumeView femoral arterial catheter
according to the directions for use.
2. Insert a Central Venous Catheter (CVC) (if using the Edwards oxymetry CVC , refer to the Edwards oxymetry CVC setup guide).
3. Ensure the TruWave central venous pressure
transducer and VolumeView sensor are primed and then connected to the EV1000 databox.
4. Ensure the sensor and transducer
are level to patient’s phlebostatic axis and zero transducers on monitor using Clinical Actions button.
5. Connect the VolumeView sensor
to the VolumeView femoral arterial catheter and confi rm the appropriate waveforms and pressures are displayed.
6. Connect the databox thermistor cable to the VolumeView femoral arterial thermistor and confi rm appropriate temperature on the EV1000 monitor.
7. Prime the VolumeView set manifold, removing all air. Connect the TruWave central venous pressure transducer to the VolumeView manifold and then connect it to the central venous catheter and confi rm the appropriate waveform
and pressure are displayed.
8. Connect the in-line injectate thermistor probe to the VolumeView
manifold.
9. Connect a pre-chilled syringe with (15 or 20 ml) normal saline to the VolumeView manifold valve/port.
10. Touch the Clinical Actions button and then touch the Thermodilution
button to start the TPTD procedure.
11. Select the desired injectate volume on
the keypad (prefer 15 or 20 ml).
12. Select Lung Resection, if appropriate,
as well as the descriptor (for example RUL = right upper lobectomy).
13. Select Parameters, to choose indexed or non-indexed parameter.
14. Touch Start Set.
15. Wait is highlighted until the thermal baseline is reached.
16. When Inject is highlighted, use a rapid, smooth, continuous
method to inject the cold solution with the volume amount previously selected.
17. Observe the calculation.
18. Repeat steps 14 and 15.
19. Touch review to edit the series of boluses.
20. Touch Accept.