EDTM GLASS-CHEK+ GC2001 User Manual

GLASS-CHEK+
Glass and Air Space Thickness Meters
GENERAL DESCRIPTION:
The GLASS-CHEK+ measures the thick­ness of both lites of glass, as well as the airspace of an insulating glass window as­sembly. All thickness measurements can be obtained from a single side. Simply press the GLASS-CHEK+ meter against a single lite of glass, or a double pane window, push the power button and hold it until the measurements are digitally displayed. The thickness of each lite of glass, the air space, and the overall thickness of the IG unit will simultaneously appear on the 2-line LCD display. On a separate results screen the meter will also indicate the presence and location of any Low-E coatings on the rst pane of glass.
The GLASS-CHEK+ contains a single pane Low-E coating detector. The single lite Low­E
detector will test the nearest lite of glass in an IG unit. To fully analyze an IG unit garding Low-E coatings, the window should be tested on both sides to conrm the pres­ence or absence of the Low-E coating. If you only have access to one side of the window, we recommend upgrading to our Glass-Chek PRO (Model # GC3000).
The user may choose to have the thick­ness results displayed in one or two of the following options: fractions of an inch, thousandths of an inch, or millimeters. The results screen will ash between the differ­ent thickness options selected by the user.
Figure 1: Recommended placement of meter.
MODEL# GC2001
re-
GLASS THICKNESS & AIR SPACE MEASUREMENT
Simply press the GLASS-CHEK+ meter against the single or dual lite window as­sembly and push the button. Continue to hold the button and the measurement re­sults will appear within two to ve seconds depending on the application. Tinted glass applications will typically take longer to mea­sure,
while standard clear windows usually result in the fastest measurement. Hold the meter steady during the measurement. The results will be displayed on the screen and will continually ash back and forth between thickness data and the Low-E location for the rst pane of glass. For best accuracy please see the “FOUR OPTIMAL TEST LOCATIONS FOR MOST ACCURATE RESULTS” section below. Windows often deect inward or outward due to varying barometric pressures. The window will hold closest to its original designed specication along the edges and at the corners where the spacer bars hold the glass xed. The thickness sensing area of the GLASS­CHEK+ meter is located along the bottom edge of the meter. Therefore the four test­ing locations shown in Figure 1 below will yield the best results. We recommend tak-
ing multiple readings to verify your re­sults.
Whenever possible, ensure that the glass sample under test is clean and that the exit location (aperture) of the laser is clear of lint, dirt or other contaminants. The laser aperture is on the back side of the GLASS-CHEK+ enclosure.
WE STRONGLY RECOMMEND ONLY US­ING
COMPRESSED AIR TO CLEAN THE LASER APERTURE. However, if there are ngerprints or other smudge marks on the laser aperture window that cannot be removed in any other way, a lint-free cloth can be used to wipe off the smudges.
The GC2001 instrument can test the thick­ness of installed windows from either indoors or outdoors. In most all applications it does not matter what is behind the window, as long as there is nothing touching the far side surface of the window. For testing windows in the factory or warehouse, make sure you are only testing one window at a time. If testing a stack of IG’s, you must move the unit you are testing away from the stack so the additional panes of glass are not added to the measurement viewing area. If testing in a lab, be sure to move the glass off the table surface before testing.
FOUR OPTIMAL TEST LOCATIONS FOR MOST ACCURATE RESULTS
Hold the meter steady AND FLAT against the glass during measurement. The meter should be centered on one of the four edges of the window with the bot-
tom edge of the meter placed parallel to the edge of the window (please see the image to the left). This will minimize the effect of a concave window. As you move away from the edge of a window, many windows will bow inward (concave). Be­cause the sensor runs parallel to the bot­tom edge of the meter, it is important to place the bottom edge of the meter along one of the four edge locations as shown.
GLASS CHEK+ LOW-E DETECTOR OPERATION
The GLASS-CHEK+ will automatically test each measured window for the presence and location of Low-E coatings on the rst pane of glass. The meter will differentiate if the coating is on Surface 1 or Surface 2. It will also indicate if the rst pane of glass is clear glass, meaning there are no Low-E coatings present on the glass. Remember that Low-E coatings are typically placed on surface 2 or 3 of a window. If you are not sure which side of the window was intended to have the Low-E coating, you may want to test both sides to ensure you have fully tested for the coating. If testing both sides of the window is not possible, you should consider our Glass-Chek PRO (Model# GC
3000) that will test both sides of a double
pane window from a single side for Low-E coatings.
Low-E Coated
Surface
('
PANE 1
PANE 2
PANE 1
PANE 2
AIR SPACE
LOW-E ON
SURFACE 1
AIR SPACE
LOW-E ON
SURFACE 2
DISPLAY OPTIONS
The display of the GLASS-CHEK+ meter is pre-programmed at the factory to offer six different display formats for the results screen. The user can choose to have the thickness measurements displayed in:
Fractions of an inch … with 1/32” resolution
1/8 1 /2 1/8 IN TOTAL IG: 3/4 IN
Decimal inches … with 0.001” resolution
0.125 0.500 0.125 IN TOTAL IG: 0.750 IN
Millimeters … with 0.1 mm resolution
The user may choose to have any one of these units of measure displayed, or a com­bination
3.0 12.5 3.0 MM TOTAL IG: 18.5 MM
of two of them. If a single unit of measure is selected, the results screen will alternate back and forth between the thick­ness results and the Low-E test results. If the user selects two units of measure, the results screen will alternate between the rst unit of measure for thickness, the Low-E results screen and then the second unit of measure selected.
To change the display format, take a mea­surement and continue to hold the button while the results are displayed multiple times. After approximately 40 seconds, the screen will prompt you to enter set-up mode.
745 Capital Commons Drive
Toledo, Ohio 43615 USA
PHONE: (419) 861-1030
FAX: (419) 861-1031
www.EDTM.com
1
Email: sales@edtm.com
© Copyright 2012 Electronic Design to Market, Inc. All rights reserved.
PANE 1
AIR SPACE
CLEAR GLASS
Follow the instructions on the screen to se­lect the display option of choice. When the display option of choice appears, simply re­lease the button. Your meter is now equipped to read in the display format you selected.
APPLICATIONS
WILL TEST …
The GC2001 instrument is calibrated and designed for testing at glass applications. It will test both single pane and double pane windows. The glass can be clear glass or coated with Low-E coatings, self clean coat­ings or other specialty coatings as long as the glass/coating is transparent. The GC 2001 will also measure tinted glass as long as it has a visible light transmission value of approximately 35% or above.
WON’T TEST …
The GC2001 is not capable of measuring triple pane windows or glass that is coated with reective (mirror-like) coatings. The GLASS-CHEK+ is also not capable of test­ing suspended lm applications such as Heat Mirror. Laminated glass with Low-E coatings positioned against the laminate can also NOT be measured. For most of
PANE 2
these applications we recommend our Glass-Chek PRO (Model# GC3000).
LAMINATED GLASS:
The GC2001 will test regular laminated glass assemblies. It will measure the over-
all thickness of the laminated assembly, as
long as there is no Low-E coating placed against the laminate inner-layer. The lami­nate assembly can be a single piece or com­bined
into an insulating unit. The GC2001 will NOT measure the PVB inner-layer or the individual pieces of glass making up the laminated glass, it will only measure the overall thickness of the laminated glass.
TRIPLE PANE & SUSPENDED FILM
The GC2001 will not attempt to measure triple pane windows. The user must pur­chase the Glass-Chek PRO (#GC3000) for this application.
HIGH/LOW RESOLUTION … SPEED OF MEASUREMENTS
The GC2001 has been equipped with a High Resolution and Low Resolution Mode. meter is shipped in Low Resolution this allows for the quickest measurements to be taken. The quicker measurements come at the expense of a little accuracy. If extreme accuracy is not important, then the low resolution mode is perfect for get­ting the quickest test results possible.
If you are testing glass for an application that requires higher accuracy, you can switch the meter into High Resolution Mode. This will improve the meters accuracy as stated in the specication section of this manual. The improved accuracy will cause the meter take measurements more slowly. To change the resolution/speed of the meter, take a
KEEP THE COMPETITIVE
EDGE WITH PRODUCTS
FROM EDTM, INC.
glass & air space laser meters, tempered glass detectors, SHGC, solar, visible, & uv meters Low­E type detectors, 4 point sheet resistance meters, tin side detec­tors,
self-clean coating detectors,
sales kits, temperature guns & sales kit accessories.
The
Mode as
to
measurement and continue to hold the but­ton
while the results are displayed multiple times. After approximately 40 seconds, the screen will prompt you to enter set-up mode. Follow the instructions on the screen. After going through all of the display options discussed previously, the meter will then show the High Resolution screen and the Low Resolution screen. When the resolu­tion option you desire appears on the dis­play, simply release the button. Your meter is now equipped to operate in the resolu­tion mode you selected. PLEASE NOTE: changing the resolution mode WILL NOT affect the display option you were previously operating in. For instance if you were in milli­meter mode before changing the resolution, the meter will remain in millimeter mode.
IMPROPER GLASS THICKNESS AND AIR SPACE MEASUREMENT
If the window area un­der test is excessively dirty or contaminated, the GLASS-CHEK+ may give improper re­sults or issue a warning message that states:
MEASUREMENT ERROR: CLEAN, MOVE & RETEST
The window area under test must be REL­ATIVELY CLEAN. Clean the test area or move the GLASS-CHEK+ to a clean loca­tion on the window.
The GLASS-CHEK+ uses a light/laser beam to mea­sure
the thickness of the glass and air space. The unit may be affected by extremely bright lights or facing the sun directly. The screen will display the error message “EXCESSIVE LIGHT” if this con­dition
occurs. Simply move to a different lo­cation or go to the other side of the window or shade the laser aperture. If the window is operable you may be able to open the window and reach through the opening to block the incident light from the back side of the meter.
The GLASS-CHEK+ must be used with the surface opposite the Power button against the window. Using the small end surface of the unit or tilting the unit away
from the surface will give improper results.
The GLASS-CHEK fam­ily of glass thickness and air space meters use a Class 2M laser to take its measurements. DO NOT push the power button and point the instrument laser at a persons’ eyes. Always check the other side of the window being tested to ensure that no one will be looking directly into the laser.
GLASS-CHEK+ IMPROPER LOW-E OPERATING CONDITIONS
Placing your hand or a metallic object against the single lite of glass you are testing may cause the GLASS-CHEK+ to yield im­proper
Low-E results. Re-
move your hand or object.
PROBLEM
GLASS-CHEK+ displays error message: “Clean, Move & Retest” OR a partial reading occurs.
TROUBLE-SHOOTING GUIDE
CORRECTIVE ACTION
1. Ensure that the GLASS-CHEK+ is placed at against the
2. Clean the glass surface or move to a cleaner area of
3. Remove dust particles from the laser aperture area on
If the window area under test is excessively dirty or contaminated, the GLASS-CHEK+ may give improper Low-E results. The window area under
4. Check that the two window surfaces are relatively par-
5. Verify the operating temperature range.
6. If only a partial reading occurs, move to another loca-
test must be RELATIVE­LY CLEAN. Clean the test area or move the GLASS-CHEK+ to a clean location on the window.
The temperature where I need to take readings is outside the GLASS-CHEK+
for cold temperatures:
1. Attempt to keep the unit in a warmer environment until
operating range.
BATTERY & REPLACEMENT
The GC2001 meter is powered by a 9-volt alkaline battery. When the battery is start­ing to get low, a warning message will ap­pear during your results screen that rec­ommends you replace the battery soon. This gives the user plenty of warning that a
2. Carry the GLASS-CHEK+ in your pocket or in some
3. Operate the unit several times repeatedly in an attempt
battery replacement is needed in the near future. The instrument will still function dur­ing this period. This is the initial warning that a battery replacement is encouraged.
Once the battery voltage gets below a dan-
for hot temperatures:
1. Attempt to keep the unit in a cooler environment until
gerous level that could begin impacting the performance of the instrument, an error message will appear on the display that tells you to replace the battery immediately. The instrument will NOT be able to take mea­surements once it reaches this level.
“REPLACE BATTERY”
1. Replace the battery with a 9 volt alkaline battery ONLY
appears on the display.
appears on the display
“EXCESSIVE LIGHT”
1. Attempt to block the source of the bright light. If the light
2. Pointing the GLASS-CHEK+ directly into the sun may
SPECIFICATIONS
POWER SOURCE 9 volt alkaline battery (NEDA 1604A) only PHYSICAL DIMENSIONS 3.5” x 5.5” x 1.70 inch (89 x 140 x 43mm) WEIGHT 0.75 pounds STORAGE TEMPERATURE -10 to +60 degrees Celsius OPERATING TEMPERATURE 0 to +40 degrees Celsius DISPLAY 2 lines X 20 characters ACCURACY:
-High Resolution Mode -Low Resolution Mode (Default) Pane 1 0.004 inch (0.1 mm) Pane 1 0.008 inch (0.2 mm) Air Space 0.012 inch (0.3 mm) Air Space 0.012 inch (0.3 mm) Pane 2 0.006 inch (0.15 mm) Pane 2 0.012 inch (0.3 mm)
MINIMUM GLASS THICKNESS ALLOWED 0.080 inch (2.0 mm) MINIMUM AIR SPACE ALLOWED 0.187 inch (4.75 mm) OVERALL THICKNESS RANGE MAXIMUM Single Pane 2.3 inch (58 mm) Double Pane with 3/32” (2.5 mm) glass 1.5 inch (38 mm) Double Pane with 1/4” (6 mm) glass 1.6 inch (40 mm) Double Pane with 1/2” (12 mm) glass 1.7 inch (44 mm)
GLASS THICKNESS ALLOWED FOR LOW-E READINGS up to 0.5 inch (12.7 mm) MAXIMUM LASER OUTPUT < 1 mW (CLASS 2M laser product) LASER WAVELENGTH 650 - 680 nm
This product complies with IEC 60825-1 and FDA regulation 21 CFR 1040.10.
The thickness tolerance is calibrated to NIST traceable standard No. 821/268634-03.
AMERICAN SOCIETY FOR TESTING AND MATERIALS (ASTM)Tolerance specications for at glass
TRADITIONAL DESIGNATION
single
1/8 in. (double)
5/32 in. 3/16 in.
1/4 in.
5/16 in.
3/8 in. 1/2 in.
DESIGNATION TOLERANCE
mm
inches
2.5 3 4 5 6 8
10 12
0.09
0.12
0.16
0.19
0.23
0.32
0.39
0.50
mm min.
11.91
2.16
2.92
3.78
4.57
5.56
7.42
9.02
mm max.
2.57
3.40
4.19
5.05
6.20
8.43
10.31
13.49
inches min.
0.085
0.115
0.149
0.180
0.219
0.292
0.355
0.469
inches max.
0.101
0.134
0.165
0.199
0.244
0.332
0.406
0.531
glass surface and that you are testing in the optimal test location described on page 1.
the glass.
the back of the GLASS-CHEK+. Use compressed air to remove any dirt or dust particles from the laser aperture area.
allel to each other and to the GLASS-CHEK+ unit.
tion on the glass or take a reading from the other side.
you are ready to take the readings. The GLASS-CHEK+ will record accurate readings up to the point the elec­tronics reach the outside temperature which is below the operating range. Depending on the degree of cold­ness, this will allow for several readings to be taken.
other manner which will keep the unit near a warmer temperature.
to warm up the die of the laser module itself. This will work in temperatures which are not greatly beneath the operating temperature range.
you are ready to take readings. The GLASS-CHEK will record accurate readings up to the point the electron­ics reach the outside temperature which is above the operating range. Depending on the extent of the heat, several accurate readings should be obtained.
(NEDA 1604A).
source is on the other side of the window you are test­ing, move to another location or block the light source.
result in an “excessive light” condition. Try rotating the GLASS-CHEK+ meter to change the angle that the sun is hitting the laser aperture area on the back side of the meter.
WARRANTY
The manufacturer warrants all models of the GC2001 to be free from defects in material and workmanship under normal use and service as specied within the operator’s manual. The manufacturer shall repair or replace the unit within twelve (12) months from the original date of shipment after the unit is returned to the manufacturers factory, prepaid by the user, and the unit is disclosed to the manufacturers satisfac­tion, to be thus defective. This warranty shall not ap­ply to any unit that has been repaired or altered other than by the manufacturer. The aforementioned provi­sions do not extend the original warranty period of the unit which has been repaired or replaced by the manufacturer. Batteries are not covered by warranty.
The manufacturer assumes no liability for the conse­quential damages of any kind through the use or mis­use of the GC2001 product by the purchaser or oth­ers. No other obligations or liabilities are expressed or implied. All damage or liability claims will be limited to an amount equal to the sale price of the GC2001, as established by the manufacturer.
2
GLASS-CHEK+
Mesureurs L’Epaisseur du Verre et de la Lame D’Aire
DESCRIPTION GÉNÉRALE:
Le GLASS-CHEK+ mesure l’épaisseur des deux vitrages ainsi que la lame d’air d’une fenêtre à double vitrage. Toutes les mesures d’épaisseur peuvent être obtenues depuis un seul côté. Il suft de poser le GLASS-CHEK+ contre une fenêtre à simple ou double vi­trage, d’appuyer sur le bouton d’alimentation et de le maintenir enfoncé jusqu’à ce que les mesures apparaissent. L’épaisseur de chaque vitrage, la lame d’air et l’épaisseur totale du double vitrage apparaîtront simultanément sur l’écran LCD à deux lignes. Sur un écran de résultat distinct, le mesureur indiquera éga­lement la présence et l’emplacement de tout revêtement à faible émissivité sur le premier vitrage.
Le GLASS-CHEK+ contient un détecteur de revêtement à faible émissivité sur un seul vi­trage. Le détecteur de revêtement à faible émissivité sur un seul vitrage testera le vi­trage le plus proche d’une fenêtre à double vitrage. An d’analyser pleinement les re­vêtements à faible émissivité d’une fenêtre à double vitrage, la fenêtre doit être testée des deux côtés an de conrmer la présence ou l’absence du revêtement à faible émis­sivité.
Si vous avez accès à un seul côté de la fenêtre, nous vous recommandons d’utiliser notre modèle de gamme supérieure, le Glass­Chek PRO (n° GC3000).
L’utilisateur peut choisir d’afcher les résul­tats de l’épaisseur dans une ou deux des op­tions suivantes: fractions de pouce, millième de pouce ou millimètres. L’écran de résul­tat clignotera entre ces différentes options d’épaisseur sélectionnées par l’utilisateur.
Figure 1: Position recommandée du mesureur.
MODÈLE GC2001
MESURE DE L’ÉPAISSEUR DU VERRE ET DE LA LAME D’AIR
Il suft simplement d’appuyer le mesureur GLASS-CHEK+ contre une fenêtre simple ou double vitrage et de pousser sur le bouton. Maintenez le bouton enfoncé et les résultats des mesures apparaîtront en deux et cinq minutes en fonction de l’application. Les mesures avec du verre teinté sont généra­lement plus longues alors que les fenêtres transparentes permettent de réaliser des mesures très rapidement. Maintenez le me­sureur sans le faire bouger pendant la me­sure. Les résultats seront afchés à l’écran et ils clignoteront en continu pour indiquer les données d’épaisseur et l’emplacement du re­vêtement à faible émissivité sur le premier vi­trage.
Pour une meilleure précision, veuillez consulter la section QUATRE EMPLACE­MENTS
DE TEST OPTIMAUX POUR DES RÉSULTATS LES LUS PRÉCIS POSSIBLE en bas de la page. Les fenêtres échissent souvent vers l’intérieur ou l’extérieur en fonc­tion des variations de pression barométrique. La fenêtre sera le plus proche de sa spécica­tion nominale le long des bords et des angles où les entretoises maintiennent le vitrage. zone de détection d’épaisseur du mesureur GLASS-CHEK+ est située le long du inférieur. De ce fait, les quatre zones de illustrées ci-dessous offriront les meilleurs résultats (Figure 1). Nous vous recomman-
dons de prendre plusieurs mesures pour vérier vos résultats. Dans la mesure du
possible, assurez-vous que l’échantillon de verre testé est propre et que la sortie (ou ture) du laser est dépourvue de poussière, saleté ou autres contaminations. L’ouverture du laser se trouve à l’arrière du boîtier du GLASS-CHEK+.
NOUS VOUS RECOMMANDONS VIVEMENT D’UTILISER UNIQUEMENT DE L’AIR COM­PRIMÉ
POUR NETTOYER L’OUVERTURE DU LASER. Cependant, en cas de traces de doigt ou de saleté qu’il est impossible de retirer de la lentille, nettoyez à l’aide d’un chiffon non pelucheux.
Le GC2001 peut tester l’épaisseur des fe­nêtres installées depuis l’intérieur ou l’ex­térieur. Dans la plupart des applications, ce qui se trouve derrière la fenêtre importe peu, tant que rien ne touche la surface éloignée de la fenêtre. Pour tester les fenêtres en usine ou en entrepôt, assurez-vous que vous tes­tez bien une fenêtre à la fois. Si vous testez une pile de fenêtres double vitrage, vous de­vez déplacez chaque unité que vous testez an que les autres vitrages ne soient pas ajoutés à la mesure. En cas de test en la­boratoire, assurez-vous de déplacer le verre à l’écart de la surface de la table avant de réaliser le test.
QUATRE EMPLACEMENTS DE TEST OPTI­MAUX POUR DES RÉSULTATS LES PLUS PRÉCIS POSSIBLE
Pendant la mesure, maintenez le mesureur
sans bouger, À PLAT contre le vitrage.
Le mesureur devrait être centré par rapport à
l’un des quatre bords de la fenêtre, avec le bord inférieur du mesureur parallèle au bord de la fenêtre (voir l’image de gauche). Ceci réduira les effets d’une fenêtre concave. Au fur et à mesure que vous éloignez des bords d’une fenêtre, de nombreuses fenêtres se creusent (concave). Comme le capteur est parallèle au bord inférieur du mesureur, il est important de placer le bord inférieur du me­sureur le long de l’un des quatre bords de la fenêtre, comme cela est illustré.
EXPLOITATION DU GLASS CHEK+ POUR LA DÉTECTION DE REVÊTEMENT À FAIBLE ÉMISSIVITÉ
Le GLASS-CHEK+ testera automatiquement chaque fenêtre mesurée an de vérier la présence et la position des revêtements à faible émissivité sur le premier vitrage. Le mesureur spéciera si le revêtement est situé sur la Surface 1 ou la Surface 2. Il indiquera également si le premier vitrage est transpa­rent, ce qui signie qu’il est dépourvu de re­vêtement à faible émissivité. N’oubliez pas que les revêtements à faible émissivité sont généralement placés sur les surfaces 2 ou 3 d’une fenêtre. Si vous n’êtes pas certain du côté sur lequel le revêtement à faible émis­sivité devait être posé sur la fenêtre, testez les deux côtés an de vous assurer d’avoir convenablement recherché le revêtement.
S’il n’est pas possible de tester les deux cô
tés de la fenêtre, vous devriez envisager d’utiliser notre Glass-Chek PRO (modèle n° GC3000) qui testera les deux côtés d’une fe-
nêtre à double vitrage depuis un seul côté an de trouver le revêtement à faible émissivité.
Surface avec revêtement
à faible émissivité
'
(
La
VITRAGE 1
VITRAGE 2
bord test
ver-
de
LAME D’AIR
REVÊTEMENT
À FAIBLE
ÉMISSIVITÉ SUR
LA SURFACE 1
OPTIONS D’AFFICHAGE
L’écran du mesureur GLASS-CHEK+ est pré
VITRAGE 1
LAME D’AIR
REVÊTEMENT
À FAIBLE
ÉMISSIVITÉ SUR
LA SURFACE 2
programmé en usine pour proposer six for­mats d’afchage différents à l’écran de ré­sultat. L’utilisateur peut choisir d’afcher les mesures d’épaisseur en:
Fractions de pouce … avec une résolution de 1/32»
1/8 1 /2 1/8 PO TOTAL DV: 3/4 PO
Pouces décimaux … avec une résolution de 0,001»
0,125 0,500 0,125 PO TOTAL DV: 0,750 PO
Millimètres … avec une résolution de 0,1 mm
3,0 12,5 3,0 MM TOTAL DV: 18,5 MM
L’utilisateur peut choisir de visualiser l’une de ces unités de mesure ou une combinaison de deux. Si une seule unité de mesure est sélectionnée, l’écran de résultats alternera entre les résultats d’épaisseur et les résul­tats du test de revêtement à faible émissivi­té. Si l’utilisateur choisit deux unités de me­sure, l’écran de résultats alternera entre la première unité de mesure de l’épaisseur, les résultats du revêtement à faible émissivité et la seconde unité de mesure sélectionnée.
745 Capital Commons Drive
Toledo, Ohio 43615 USA
TÉLÉPHONE: (419) 861-1030
TÉLÉCOPIE: (419) 861-1031
www.EDTM.com
Courriel: sales@edtm.com
3
An de modier le format d’afchage, prenez une mesure et maintenez le bouton enfoncé pendant que les résultats sont afchés plu­sieurs fois. Après environ 40 secondes, vous serez invité à entrer dans le mode parame­trage. Suivez les instructions à l’écran pour sélectionner l’option d’afchage de votre choix. Lorsque l’option d’afchage apparaît, relâchez simplement le bouton. Votre mesureur est dé­sormais
prêt à lire dans le format d’afchage
que vous avez sélectionné.
APPLICATIONS
PERMET DE TESTER …
Le GC2001 est étalonné et conçu pour tester du verre plat. Il permet de tester les fenêtres à simple et double vitrage. Le verre peut être transparent ou avec des revêtements à faible émissivité, des revêtements autonettoyants ou d’autres revêtements spéciaux, tant que le verre/revêtement est transparent. Le GC 2001 mesure également le verre teinté, tant qu’il bénécie d’une valeur de transmission de la lumière visible d’environ 35% ou plus.
NE PERMET PAS DE TESTER …
Le GC2001 ne permet pas de mesurer les fenêtres à triple vitrage ou le verre revêtu de revêtements rééchissants (miroir). Le GC 2001 ne permet pas de mesurer les appli­cations de lm suspendu comme le Miroir Chauffant. Il ne permet pas non plus de mesu­rer
VITRAGE 2
TRANSPARENT
VITRAGE 1
LAME D’AIR
VERRE
les revêtements à faible émissivité contre
VITRAGE 2
la feuille d’un verre feuilleté. Pour la plupart
de ces pplications, nous recommandons notre Glass-Chek PRO (modèle n° GC
3000)
.
VERRE FEUILLETÉ:
Le GC2001 permet de tester le verre feuilleté. Il mesurera l’épaisseur totale du verre feuille­té
tant qu’il n’y a pas de revêtement à faible émissivité contre la couche intérieure feuille­tée. L’ensemble feuilleté peut être d’une seule pièce ou combiné dans une unité isolante. Le GC2001 NE MESURE PAS la couche in rieure de PVB ou chaque feuille composant verre feuilleté, il indique simplement l’épais­seur générale de l’ensemble feuilleté.
TRIPLE VITRAGE ET FILM SUSPENDU
Le GC2001 ne permet pas de mesurer les fenêtres à triple vitrage. L’utilisateur doit ache­ter
le Glass-Chek PRO (n° GC3000) pour cette application.
HAUTE/BASSE RÉSOLUTION … VITESSE DES MESURES
Le GC2001 dispose d’un mode haute réso tion et d’un mode basse résolution. Le mesu­reur
est expédié en mode basse résolution, car il permet de réaliser les mesures plus rapi­dement. Des mesures plus rapides se font détriment de la précision. Si une précision extrême n’est pas importante, alors le mode basse résolution est parfait pour obtenir les résultats d’essai les plus rapides possible.
Si vous testez le verre pour une application nécessitant une précision élevée, vous pou­vez commuter le mesureur en mode haute résolution. Ceci améliorera la précision du mesureur, conformément aux indications de
CONSERVEZ UNE LONGUEUR
D’AVANCE SUR LA
CONCURRENCE GRÂCE AUX
PRODUITS D’EDTM, INC.
Mesureurs laser de verre et lame d’air, dé­tecteurs de verre trempé, détecteurs de re­vêtement à faible émissivité avec mesure de fsg, lumière solaire, visible et UV, mesureur de résistivité par méthode 4 pointes, détec­teurs de face étain, détecteurs de revêtement autonettoyant, kits de vente, pistolet de me­sure de température et accessoires de kits de vente.
Copyright 2012 Electronic Design to Market, Inc. Tous droits réservés.
té-
le
lu-
au
Loading...
+ 5 hidden pages