EDT directION QC355TX User Manual

Page 1
Table of Contents
Section 1 - Introduction and Set Up
Section 2 - pH Calibration & Measurement
Section 3 - Use with a Printer or Computer
Section 4 -
Section 5 -
Appendix 1 - pH Electrodes
Troubleshooting
QC355TX - Routine QC pH Meter - Users Manual
Page 2
SECTION 1 - Introduction
Description
The QC355(Tx) pH meter is a simple-to-operate bench pH meter. The meter utilises automatic calibration to one or two points: automatic temperature compensation is compulsory and the inbuilt calibration clock reminds the user when re-calibration is necessary.
Unpacking
Verify that you have received all equipment. If you have any questions about the shipment, please call EDT Direct ION Ltd. or your agent. When you receive the shipment, inspect the container for any signs of damage. Note any evidence of rough handling in transit. Immediately report any damage to the agent.
Note
The carrier will not honour any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material in the event that re-shipment be necessary. The following items are packed in the box:
QC355TX Meter E8051 Temperature Probe E8061 Electrode Stand AC Adapter 9V Battery pH Buffers 4,7,10 Operator’s Manual
Setting Up - AC Operation
Only use the approved power adaptor supplied Check that the adaptor is the correct voltage for your power supply Plug the adaptor into the power socket at the back of the meter, then connect to the AC supply.
Battery Installation
Approximately 24 hours of continuous use is afforded by the 9V battery. The BAT flag appears on the display to indicate a low battery. To install or replace the battery, slide off the back cover Remove the old battery and insert a new one ensuring that the polarity is correct Replace back cover.
Figure 1. Bottom view of instrument showing power connections
Page 3
Setting Up (Continued)
A Note on Electrodes
pH, Ion Selective and Redox electrodes may all be used with the QC355. These may be combination or half-cell types. Combination or sensing half-cells should have a BNC terminal and be connected to the socket marked pH/ION input at the back of the meter. Reference half-cells should have a 4mm bunched terminal and be connected to the socket marked REF. Always refer to electrode instructions before use. See also, Appendix 1.
Figure 2. Rear view of QC355(Tx)
Page 4
Section 2 - pH Calibration and Measurement
For accurate results, stir all buffers and samples. Always rinse electrodes and blot dry before transferring from one solution to another to prevent contamination. Ensure that any electrode filling holes are left uncovered during use.
Automatic Calibration & Temperature Compensation
1. Connect the pH electrode(s) and temperature probe and switch on using the power
ON/OFF switch on the back panel.
2. Select pH mode by pressing the pH/°C key. Note: A flashing display indicates that the instrument is uncalibrated.
3. Place the pH electrode(s) and the temperature probe in the pH 7 buffer and wait for a stable reading. Indicated by ◄ on the display.
4. Press the CAL key– CAL flag will flash. Press the 7pH key—CAL flag will stop flashing. The first calibration point has now been entered automatically at the correct value for the temperature of measurement.
5. Rinse, blot and place electrode(s) and temperature probe in pH buffer 4 or 10. Wait for a stable reading indicated by ◄on the display.
6. Press the CAL key– CAL flag will flash. Press the 4/10pH key—CAL flag will stop flashing. The second calibration point has now been entered automatically at the correct value for the temperature of measurement.
7. pH measurements may now be made by immersing rinsed electrodes in the sample and recording the stable reading indicated by the ◄ flag.
8. To read the temperature press the pH/°C key at any time.
7pH
4/10pH
Cal
Send
pH/°C
Calibration Keys, automatically selects and enters correct pH value
Starts calibration routine
Selects pH or temperature mode
Transmits data to printer or computer
Figure 3. Front Panel
Page 5
Section 3 - (Tx Version Only) Operation With a Printer or Computer
Printer
1. Connect a printer (set at 1200 Baud) to the meter via the RS232C port on the back panel.
2. Follow the calibration procedure given in Section 2
3. To print out a sample reading, press the Send key. The first time the key is pressed the
following printout is obtained
Computer
Connect a computer using 1200 Baud via the RS232C port at the back of the meter. A computer program is required to receive and send characters from the computer. The current readings can be sent to the computer by pressing the Send key. Each line is terminated with a Carriage Return (CR), Line Feed (LF). All characters are ASCII printable alpha-numeric
Three commands, CA, PR and RD can be sent from the computer:
DATE………………….
OPERATOR…………...
SAMPLE………………
pH = 7.00pH T=21.7°C
Command—CA 1 CR LF pH = 7.06 CR LF T = 15.8 °C CR LF mV = -0.2mV CR LF 2 CR LF Second calibration point
Command—RD pH = 7.00pH CR LF
T = 21.5°C CR LF
mV = 0.2mV CR LF
CA—Send Calibration Data
RD –Send Current Readings
Command—PR Sl = 95.3% NERNST CRLF
E° = 5.2mV CR LF
PR –Send Probe Status Data
RS232 Connection details
Rx
Tx
CTS
GND
4. Pressing and releasing the Send key subsequently
will result in a printout of the displayed reading and temperature only.
5. To obtain a printout of other parameters for the same sample, press the Mode key and then the Send key
6. To print a new identifier, press and hold down the Send key.
Page 6
Section 4 - Troubleshooting
Symptom
Probable Cause
No display
Battery is flat or not installed Power supply disconnected
‘BAT’ flag displayed
Battery Low
Wildly erratic readings or display reads -­on left hand side
Electrodes disconnected Electrodes not immersed in solution Reference electrode not filled Reference junction dry
Drifting readings
Inconsistent or lack of stirring Reference filling solution contaminated Buffers contaminated
Erratic/drifting readings or display reads – on left hand side when shorting plug is attached.
Return meter for servicing
Error Codes
PRO Temperature probe malfunctioning or not connected Buf Wrong or Contaminated Buffer SL Poor Electrode Slope caused by faulty electrode or poor
buffers/ standards
E° Faulty or Aged electrode E6, E7 Calibration error. Recalibrate with fresh standards or buffers
Electrodes out of solution.
In the event of a malfunction, it is important to pinpoint the problem to either the meter or the cell. If a spare cell is available, substitute it for the one in use.
There are no user serviceable parts in this instrument. Please ensure that the instrument, together with all accessories, is returned to EDT Direct ION Ltd or the agent with a full description of the symptoms.
No attempt should be made to repair the meter.
Page 7
Section 5 -
Appendix 1
Specifications
pH electrodes
Before Use
Remove the protective cap covering the glass sensing bulb and replace with the protective guard if applicable. Inspect the filling solution for air bubbles and remove by shaking in a downward direction. Soak the electrode in pH storage solution for 30 minutes.
Cleaning
Soak the electrode in 0.1M HCl for 15 minutes followed by soaking in pH storage solution for 30 minutes
Storage.
Put some pH storage solution in the protective cap and place over the glass bulb. Never store the electrode in distilled or de-ionised water. Never allow the electrode to dry out.
pH electrode storage solution is made up by dissolving 1g KCl in 100mL of pH 7 buffer
pH Range 0.00 to 14.00 pH Accuracy ±0.02pH Temperature Compensation 0-100°C Auto Calibration 2 points, 7 or 4/10 °C Range
-30.0 to +130.0°C
°C Resolution
0.1°C
°C Accuracy
±0.3°C
Display Power Instrument Size Instrument Weight
12.7mm LCD 9V Battery or AC 210 x 150 x 88mm 550g
Page 8
For more information on our products
please visit our website www.edt.co.uk
EDT directION
Unit 5 Waldershare Park, Dover. CT15 5DQ.
Telephone: 01304 829960 Email: [email protected]
Company No, 04135318 VAT No. GB 765175410
Loading...