EDT directION QC355TX User Manual

Table of Contents
Section 1 - Introduction and Set Up
Section 2 - pH Calibration & Measurement
Section 3 - Use with a Printer or Computer
Section 4 -
Section 5 -
Appendix 1 - pH Electrodes
Troubleshooting
QC355TX - Routine QC pH Meter - Users Manual
SECTION 1 - Introduction
Description
The QC355(Tx) pH meter is a simple-to-operate bench pH meter. The meter utilises automatic calibration to one or two points: automatic temperature compensation is compulsory and the inbuilt calibration clock reminds the user when re-calibration is necessary.
Unpacking
Verify that you have received all equipment. If you have any questions about the shipment, please call EDT Direct ION Ltd. or your agent. When you receive the shipment, inspect the container for any signs of damage. Note any evidence of rough handling in transit. Immediately report any damage to the agent.
Note
The carrier will not honour any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material in the event that re-shipment be necessary. The following items are packed in the box:
QC355TX Meter E8051 Temperature Probe E8061 Electrode Stand AC Adapter 9V Battery pH Buffers 4,7,10 Operator’s Manual
Setting Up - AC Operation
Only use the approved power adaptor supplied Check that the adaptor is the correct voltage for your power supply Plug the adaptor into the power socket at the back of the meter, then connect to the AC supply.
Battery Installation
Approximately 24 hours of continuous use is afforded by the 9V battery. The BAT flag appears on the display to indicate a low battery. To install or replace the battery, slide off the back cover Remove the old battery and insert a new one ensuring that the polarity is correct Replace back cover.
Figure 1. Bottom view of instrument showing power connections
Setting Up (Continued)
A Note on Electrodes
pH, Ion Selective and Redox electrodes may all be used with the QC355. These may be combination or half-cell types. Combination or sensing half-cells should have a BNC terminal and be connected to the socket marked pH/ION input at the back of the meter. Reference half-cells should have a 4mm bunched terminal and be connected to the socket marked REF. Always refer to electrode instructions before use. See also, Appendix 1.
Figure 2. Rear view of QC355(Tx)
Loading...
+ 5 hidden pages