Calibration and Measurement
Note: Refer to Appendix 2 on concentration units before proceeding
1. Up to 5 calibration points may be entered. Standards should be prepared such that
their concentrations lie either side of the expected sample range. If 2 standards are to be
used then ideally they should be 10fold apart in concentration. E.g. for an expected
sample concentration of 5ppm, standards of 1 and 10ppm should be prepared.
2. Connect the Ion selective electrode to the BNC input at the back of the meter. If using a
separate reference electrode, connect to the 4mm input. Switch on using the black
ON/OFF
3. In order to work in Concentration mode, any existing pH calibration data MUST be
cleared. Select pH mode. If the CAL flag is displayed, press the Clear key until it
disappears.
4. Press the mode key until Conc is displayed.
5. Place electrodes in the first standard and wait for a stable reading (the value displayed
is an arbitrary number at this stage).
6. Press the x 10 key (if necessary) to adjust the value to the correct decade, then use the
ñò keys to adjust to the exact value of the standard. (The CAL flag will flash during this
procedure.). Press Enter. (The CAL flag will stop flashing).
7. Place electrodes in the next standard and repeat and
8. If further standards are to be used repeat from for each standard.
9. Ion concentration measurements may now be made by immersing the electrodes in the
prepared sample and recording the displayed reading.
Note: The performance of the electrodes can be assessed by pressing the Slope key. This
will display the mV/decade slope which can then be compared with the theoretical value
Section 4 - mV Calibration and Measurement
The mV mode has two ranges. The meter will automatically select the most appropriate
range as follows:
0 to ±400mV with resolution 0.1mV
±400mV to ±2000mV with resolution 1mV
Absolute mV
1. Connect the electrodes
2. Press the Mode key to select mV function
3. Clear any existing data. (Press Clear for 5 seconds)
4. Absolute mV readings may now be made by immersing the electrodes in the sample
and recording the reading.