Edif Echo PC Service manual

E
I1122A – Echo PC Maintenance manual – Rel. 1.1
EI0002A
MAINTENANCE MANUAL
Edif Instruments s.r.l. – Via Ardeatina, 132 – 00179 Roma – ITALY
Tel.+39 065127161 – Fax +39 065127550
Web www.edif.it
– email edif@edif.it
1
E
I1122A – Echo PC Maintenance manual – Rel. 1.1
Model
: Echo PC
Version :
Serial Number :
Installation Date :
Warranty : YES NO
Servicing Agreement : YES NO
Technical Service Address :
Edif Instruments s.r.l. – Via Ardeatina, 132 – 00179 Roma – ITALY
Tel.+39 065127161 – Fax +39 065127550
Web www.edif.it
– email edif@edif.it
2
E
I1122A – Echo PC Maintenance manual – Rel. 1.1
SUMMARY
Page
ECHO PC ACCESSORY BOX MINIMUM QUANTITY OF SPARE PARTS & CONSUMABLES 4
SPARE PARTS & CONSUMABLES LISTS 5
ERROR AND WARNING MESSAGES 7
TROUBLE SHOOTING 9
MAINTENANCE 11
MAINTENANCE DATA SHEET N#1 12
Peristaltic pump activation / deactivation
MAINTENANCE DATA SHEET N#2 13 Flow cell aspiration level check and adjust
MAINTENANCE DATA SHEET N#3 14
Washing procedure
MAINTENANCE DATA SHEET N#4 15
Flow cell washing procedure
MAINTENANCE DATA SHEET N#5 16
Syringe exchange
MAINTENANCE DATA SHEET N#6 17
Lamp exchange
MAINTENANCE DATA SHEET N#7 18
Paper change
MAINTENANCE DATA SHEET N#8 19
Sample circuit change
MAINTENANCE TABLE 20
Edif Instruments s.r.l. – Via Ardeatina, 132 – 00179 Roma – ITALY
Tel.+39 065127161 – Fax +39 065127550
Web www.edif.it
– email edif@edif.it
3
E
I1122A – Echo PC Maintenance manual – Rel. 1.1
EDIF CODE OLD CODE
EI0703CV1 P10 0000 0023 Halogen lamp (MATE-N-LOCK connector)
1 1 EI0801B 1 EI0802A 1 EI0710A 1 EI0711A 1 EI0714A 1 EI0715A 1 EI0716A 1 EI0717A 1 EI0702B 1 EI0721A 1 EI0722A 1 EE0203A 1 EI0702B 1 EI1126A 1 EI0732A 1 EE0202B
P10 0000 0008 Tubes kit P10 0000 0010 Sampling and Aspiration Needles kit P10 1000 0001 Reagent bottles 25 ml (27 pos.) pack (30 pcs) P10 1000 0002 Reagent bottle caps 25 ml (27pos.) pack (30 pcs) P10 0000 0003 Needle cleaning filters pack P10 0000 0004 Reaction segments pack (100 pcs) P10 0000 0005 Serum cups pack (1000 pcs) P10 0000 0006 Thermal paper pack (3 pcs) P10 0000 0012 Fuses kit (230V) P10 0000 0017 Water charge bottle P10 0000 0018 Waste bottle S10 2100 0009 Null modem host cable P10 0000 0012 Fuses kit 230V Brief user manual Documentation CD Power cord
ECHO PC ACCESSORY BOX
MINIMUM QUANTITY OF SPARE PARTS & CONSUMABLES
EDIF CODE OLD CODE
EI0703CV1 P10 0000 0023 Halogen lamp (MATE-N-LOCK connector)
1
EI0703C P10 0000 0011 Halogen lamp (old MX396 connector)
1 EI0801B
P10 0000 0008 Tubes kit P10 0000 0007 Tubes kit (old cover with vertical diluter)
1 EI0802A
1 EI0719A
P10 0000 0010 Sampling and Aspiration Needles kit
P10 0000 0013 Diluter Syringe
Edif Instruments s.r.l. – Via Ardeatina, 132 – 00179 Roma – ITALY
Tel.+39 065127161 – Fax +39 065127550
Web www.edif.it
– email edif@edif.it
4
E
I1122A – Echo PC Maintenance manual – Rel. 1.1
ECHO PC CONSUMABLES & SPARE PARTS
ECHO PC CONSUMABLES
EDIF
CODE
OLD CODE
DESCRIPTION
PACK
EI0710A EI0711A EI0712A EI0713A EI0714A EI0715A EI0716A EI0717A
EI0801B EI0822A EI0802A
EI0703CV1
EI0703C
EI0701B EI0702B EI0719A EI0832A EI0102A EI0721A EI0722A
P10 1000 0001
Reagent bottles 25 ml (27 pos.) pack 30 P10 1000 0002 Reagent bottle caps 25 ml (27pos.) pack 30 P10 0000 0001 Reagent bottles 40 ml (18 pos.) pack 20 P10 0000 0002 Reagent bottle caps 40 ml (18 pos.) pack 25 P10 0000 0003 Needle cleaning filters pack 25 P10 0000 0004 Reaction segments pack 100 P10 0000 0005 Serum cups pack 1000 P10 0000 0006 Thermal paper pack 3
P10 0000 0007 Tubes kit (old over with vertical diluter) 1 P10 0000 0008 Tubes kit 1 P10 2000 0003 Flow cell (new model, 10 x 10 mm) 1 P10 0000 0010 Sampling and Aspiration needles kit 1
Halogen calibrated lamp with support (MATE-N-
P10 0000 0023
LOCK connector) 1 Halogen calibrated lamp with support (old MX396
P10 0000 0011
connector) nd Fuses kit (115V) 1 P10 0000 0012 Fuses kit (230V) P10 0000 0013 Diluter Syringe complete of plunger 1 P10 0000 0015 Flow cell quartz + O-ring kit (old flow cell) 2 P10 0000 0016 Heating reel 1 P10 0000 0017 Water charge bottle 1 P10 0000 0018 Waste bottle 1
Edif Instruments s.r.l. – Via Ardeatina, 132 – 00179 Roma – ITALY
Tel.+39 065127161 – Fax +39 065127550
Web www.edif.it
– email edif@edif.it
5
E
I1122A – Echo PC Maintenance manual – Rel. 1.1
ECHO PC SPARE PARTS
EDIF CODE
OLD CODE DESCRIPTION
QTY ON
BOARD
EI0201A EI0202A EI0244A EI0203A EI0204A EI0205A EI0206A EI0207A EI0208A EI0210A EI0211A EI0212A EI0213A
EI0218A
EI0220A EI0222A
EI0223C
EI0223B EI0224B EI0226A EI0301B2 V1 EI0301B2 EI0302B1 V1 EI0302B1 EI0214A EI0307B EI0401A EI0245B EI0109C
S1001000001 Optical limit switch 7 S1001000002 Cooling fan for power board 2 nd Bottom cooling fan 1 S1001000003 Diluter solenoid valve 1 S1001000004 Peltier cell for optics 1 S1001000005 Heating etched foil resistance (outer plate) 1 S1001000006 Heating etched foil resistance (pre-heating reel) 1 S1001000007 Sampling and aspiration needle step motors 2 S1001000008 RP, IP, OP, FS. PM stepper motors 5 S1001000011 Flow cell temperature sensor 1 S1001000010 Outer (reaction) plate temperature sensor 1 S1001000009 Diluter stepper motor 1 S1001000012 Optical pre-amplifier assembly 1 S1001000017
(2)
Boards interconnection flat cable 2
S1001000019 Reagents plate rotating wiring kit 1 S1001000021 Optical signal cable 1
nd
Low-voltage wiring assembly (AMPMODU, MATE-N-
LOCK)
1
S1001000022 Low-voltage wiring assembly (old MX396 connectors) == S1001000023 High - voltage wiring assembly 1 S1001000025 Stepper motors protection diodes modules 8
nd Power board (AMPMODU green connectors) 1
S1001000026 Power board (MX396 old white connectors) ==
S1001000027 CPU board 1
nd CPU board (old version with CRT controller) == S1001000013 Internal printer assembly 1 S1001000028 Power supply transformer assembly 1 S1001000031 Power supply for reagent cooling 1 S1021000013 Level sensor module 1 S10 2100 0012 Internal PC PC104 assembly 1
Edif Instruments s.r.l. – Via Ardeatina, 132 – 00179 Roma – ITALY
Tel.+39 065127161 – Fax +39 065127550
Web www.edif.it
– email edif@edif.it
6
E
I1122A – Echo PC Maintenance manual – Rel. 1.1
ERROR AND WARNING MESSAGES
ERROR ON SAMPLING NEEDLE
ERROR ON ASPIRATION NEEDLE
ERROR ON REAGENT PLATE
ERROR ON INNER PLATE
ERROR ON OUTER PLATE
ERROR ON DILUTER SYRINGE
ERROR ON FILTER WHEEL
Error during Auto-Zero: OVER
The home positioning of the sampling needle failed. Check the motor, the limit switch, the mechanics, the electronic drive. The home positioning of the aspiration needle failed. Check the motor, the limit switch, the mechanics, the electronic drive. The home positioning of the reagent plate failed. Check the motor, the limit switch, the mechanics, the electronic drive. The home positioning of the sample plate failed. Check the motor, the limit switch, the mechanics, the electronic drive. The home positioning of the reaction plate failed. Check the motor, the limit switch, the mechanics, the electronic drive. The home positioning of the diluter syringe failed. Check the motor, the limit switch, the mechanics, the electronic drive. The home positioning of the filter wheel failed. Check the motor, the limit switch, the mechanics, the electronic drive. The autozero value of the black (filter 0) is over 100.0 mV Check the optical preamplifier, try to adjust the offset
Edif Instruments s.r.l. – Via Ardeatina, 132 – 00179 Roma – ITALY
Tel.+39/065127161 – Fax +39/065127550
email edif@edif.it
7
Loading...
+ 14 hidden pages