S1001000026 Power board (MX396 old white connectors) ==
S1001000027 CPU board 1
nd CPU board (old version with CRT controller) ==
S1001000013 Internal printer assembly 1
S1001000028 Power supply transformer assembly 1
S1001000031 Power supply for reagent cooling 1
S1021000013 Level sensor module 1
S10 2100 0012 Internal PC PC104 assembly 1
Edif Instruments s.r.l. – Via Ardeatina, 132 – 00179 Roma – ITALY
Tel.+39 065127161 – Fax +39 065127550
Web www.edif.it
– email edif@edif.it
6
E
I1122A – Echo PC Maintenance manual – Rel. 1.1
ERROR AND WARNING MESSAGES
ERROR ON SAMPLING NEEDLE
ERROR ON ASPIRATION NEEDLE
ERROR ON REAGENT PLATE
ERROR ON INNER PLATE
ERROR ON OUTER PLATE
ERROR ON DILUTER SYRINGE
ERROR ON FILTER WHEEL
Error during Auto-Zero: OVER
The home positioning of the sampling needle failed.
Check the motor, the limit switch, the mechanics, the
electronic drive.
The home positioning of the aspiration needle failed.
Check the motor, the limit switch, the mechanics, the
electronic drive.
The home positioning of the reagent plate failed. Check
the motor, the limit switch, the mechanics, the electronic
drive.
The home positioning of the sample plate failed. Check
the motor, the limit switch, the mechanics, the electronic
drive.
The home positioning of the reaction plate failed. Check
the motor, the limit switch, the mechanics, the electronic
drive.
The home positioning of the diluter syringe failed. Check
the motor, the limit switch, the mechanics, the electronic
drive.
The home positioning of the filter wheel failed. Check the
motor, the limit switch, the mechanics, the electronic
drive.
The autozero value of the black (filter 0) is over 100.0 mV
Check the optical preamplifier, try to adjust the offset
Edif Instruments s.r.l. – Via Ardeatina, 132 – 00179 Roma – ITALY
Tel.+39/065127161 – Fax +39/065127550
email edif@edif.it
7