Edgetech Instruments PPM1 Specifications

ISO/IEC 17025:2005 Accredited
ISO 9001:2008 Registered
The Edgetech Instruments PPM1 is a trace moisture
analyzer for monitoring dew/frost point and ppmV water vapor in gases.
The PPM1 uses an electrolytic, phosphorus pentoxide sensor
in combination with a proprietary semi permeable diffusion membrane. The principle of operation applies Faradays Law of Electrolysis to determine the water vapor content by measuring the dissociation current needed to electrolyze the water molecules in the sample.
The PPM1 is packaged in an IP66/67, NEMA 4 plastic enclosure and comes complete with two FORM A relay al a r ms, t wo analog current outputs a n d a RS -232
bi-directional serial interface.
All Edgetech Instruments hygrometers are manufactured and supported in the USA in a modern, ISO 9001:2008 registered, ISO/IEC 17025:2005 accredited facility. All calibrations and certifications are traceable to NIST.
Features:
Precise, trace moisture measurement
Range: 0.1 to 1250 ppmV
NIST traceability calibration certificate
Primary method measurement principle
Compact, low cost solution
Remote sensor can be installed at process take-off point
Standard sample ports are 1/4 inch compression fittings
Optional VCR sample ports
May be configured for flow-thru or diffusion sampling
Low Cost, Trace Moisture Analyzer
Applications:
Air separation and cylinder transfill
Glove box/isolation chamber
Semiconductor fabrication
Pharmaceutical storage
Pharmaceutical processes
Inert gas purity
MAP systems
Power and energy systems
The PPM1 Trace Moisture Analyzer
Click here for the PPM1 Application note
Toll Free: 800-279-3729 Ph: 508-263-5900 Fax: 508-486-9348
E-mail: h2o@edgetechinstruments.com
Web: www.edgetechinstruments.com
Technical Specifications
ISO/IEC 17025:2005 Accredited
ISO 9001:2008 Registered
Sensor Type
Electrolytic phosphorous pentoxide sensor (P2O5)
on ceramic wafer
Measuring Range
Trace Moisture: 12500.1 ppmV
Dew/Frost Point: -20°C to –95°C
Accuracy
± 5% of ppmV reading
Response Time
90% of step change, one minute, typical
Hysteresis
None
General
Display: 2 Line LCD backlit Operating temperature: 32° to 122°F (0° to 50°C) Sample connection: 1/4” compression fitting,
standard model configured for positive pressure Sample Flow Rate: 0.5 - 5 SCFH, regulated Sample Pressure: 0 to 200 psig, regulated
Sensor Housing: 316 SS, flow thru head, 1/4 Inch
compression fittings, VCR optional Sampling Options: May be configured for diffusion
sampling for glove box or isolation chamber
Power & Outputs
Power: 18 to 30Vdc, unregulated, 1A max Outputs: Analog: two, 4-20 mA, scalable by user
Alarm: two dry alarm relays, FORM A,
adjustable by user Digital: RS-232 serial, bi-directional
Rev 2
Weights & Dimensions
Weight: 4 lbs. (1.81kg) Environmental: Plastic IP66/67, NEMA
4 case
Dimensions: 5.1” (13cm) L x
3” (7.6cm) W x 5.1” (13cm) D
Options
Vacuum pump for sample extraction VCR male or female ports Sensor mounting bracket for diffusion
sampling