Edgetech Instruments DPS Specifications

ISO 9001:2008 Registered
ISO/IEC 17025:2005 Accredited
The Edgetech Instruments, Inc. DPS Series
consists of heated, primary method, chilled mirror hygrometer systems that continuously measure the dew point of high dew/frost point gases from high temperature environments.
liquid water to condense in the apparatus. Under these
conditions, reliable dew/frost point measurement us not possible.
The DPS Series heated hygrometers solve this problem by heating all the components of the sensing system to a temperature above the dew/frost point temperature of the sampled gas, preventing condensation in the measurement system. The entire sampling system within the DPS, including the chilled mirror sensor, is heated. The heated sensor and system allow reliable, accurate measurement of high dew/frost point temperatures that are above the ambient temperature.
The DPS heated chilled mirror hygrometer extracts a gas sample from a high temperature, high dew point process, transports it through a heated hose and measures its dew/ frost point over the range of -35°C (-31°F) to as high as +95°C (203°F) dew/frost point temperature. The three
models are the DPS3 standard unit, the DPS4 featuring the
ability to measure and automatically compensate for pressure and the DPS5 deluxe unit that offers an advanced display, operational software and interface options. For details see the DPS Series application note on the Edgetech Instruments website.
All Edgetech Instruments hygrometers are manufactured, calibrated and supported in the USA in a modern, ISO 9001:2008 registered facility containing Edgetech InstrumentsISO/IEC 17025:2005 accredited calibration laboratory. All certifications and calibrations are traceable to NIST.
Benefits:
Allows measurement of dew/frost point temperature that is above ambient temperature
Primary method sensor provides absolute knowledge of dew/
frost point temperature
Prevents measurement system failures
Improves process control with accurate knowledge of dew/frost
point temperature
Reduces down-time and maintenance frequency
Heated Dew/Frost Point Hygrometer System
Features:
Automatic Balance Control mode provides automatic correction for optics contaminants
User programmable system temperature
controller
Current or voltage outputs, RS-232 available
User programmable alarm functions
Applications:
Wood veneer drying
Combustion processes
On-engine test
Leakage testing of nuclear reactors
Fluidized bed dryers
Product drying chambers
Bakeries
Exothermic process chambers
Fuel cell testing
Furnace applications
Pharmaceutical powder drying
Elevated temp process chambers
Plastics molding processes
Chemical reactors
Heat treating and annealing ovens
Nuclear process monitoring
Toll Free: 800-279-3729 Ph: 508-263-5900 Fax: 508-486-9348
E-mail: h2o@edgetechinstruments.com
Web: www.edgetechinstruments.com
Technical Specifications
ISO 9001:2008 Registered
ISO/IEC 17025:2005 Accredited
Specifications:
Primary method chilled mirror sensors:
DPS3 available with the HTS2 two stage chilled
mirror or HTS3 three stage chilled mirror
DPS4 and DPS5 available with HTS2 two stage
chilled mirror or HTS3 three stage chilled mirror
with standard aluminum body or the HTX3
stainless steel body
Heated system temperature range:
20 to 99°C (68 to 210°F)
Dew/frost point measurement range:
-35 to 95°C (-31 to 203°F)
Measurement accuracy:
±0.2°C (±0.4°F)
Repeatability:
±0.05°C (0.9°F)
User programmable temperature controller:
Standard
Display parameters:
Dew/frost point temperature, system set point
temperature, system actual temperature
Two alarm relays:
Standard
Analog Outputs:
Selectable 4 to 20mA, 0 to 5 V dc or 0 to 10 V dc
Serial RS232 output: Standard Sample Connection:
¼ inch compression fitting
Sample Pressure:
0 to 200 psig
Heated sample hose:
3 ft. included
AC power:
100 or 240 V ac, 50/60Hz, 250 Watts, specify
voltage
Mirror material:
Standard: chrome Optional: stainless steel or platinum
Sample Flow:
0.5 to 5.0 SCFH (0.25 to 2.5 l/min) Automatic Balance Control standard Integral adjustable sample flowmeter
Options:
Pressure transducer kit, live pressure for ppmv
output
Sample flow detection, sample flow switch with
SPST dry contacts
Sample flow detection, sample flow switch triggers
red lamp alarm
Vacuum pump to extract sample flow from process
to the sensor
Cart mount, specify model analyzer Extended length sample hose, 10 ft.