Technical Specifications
ISO/IEC 17025:2005 Accredited
ISO 9001:2008 Certified
Sensor Type:
Primary method chilled mirror dew/frost point
Dew/Frost Point Range (at 25°C ambient):
S Series sensors:
S2 two-Stage: -40 to 122°F (-40 to 50°C)
S3 three stage: -103 to 122°F (-75 to 50°C)
Liquid cooling is required to measure frost points
below -60°C with S Series sensors
X3 Series sensors:
X3 Standard: -50 to 165°F (-45 to 75°C)
X3F Fan Cooled: -85 to 165°F (-65 to 75°C)
X3SF High Efficiency: -95 to 165°F (-70 to 75°C)
X3LC Liquid Cooled: -130 to 165°F (-90 to 75°C)
Liquid cooling is required to measure frost points
below -70°C with X3 Series sensors
Accuracy, Dew/Frost Point:
S Series: ± 0.5°F (± 0.28°C)
X3 Series: ± 0.36°F (± 0.20°C)
General:
Display: Red LED, 0.5” (1.27mm) tall; 8 digit
alphanumeric
Operating temperature: 32° to 122°F (0° to 50°C)
Sample connection: 1/4” compression fitting,
standard model configured for positive pressure
Sample flow rate: 0.5 - 5 SCFH; integral flow
meter
Pressure: 0 - 150 psig (10.34 bar). Higher
pressures available, consult factory
Power:
90-230 VAC ±10%, 50-400Hz, cord and ON-OFF
switch provided
Outputs:
Analog: 0-5 VDC or 4-20 mA scaled for sensor
Compliance: 9.0VDC, 450 ohms
Serial: RS-232
Alarms:
Audible: >98 dba at 2 ft.
Relay: 1 Form C, non-latching 10 A at 240 VAC,
8A at 24 VDC, 1/2 HP at 240 VAC
Visual: Flashing “ALARM” message
Weights & Dimensions:
Weight: 6 lbs. (2.7 kg)
Enclosure Dimensions:
10.7” (27.3cm) H; 14.0” (35.6cm) W; 6”(15.2cm) D
Environmental:
Enamel Coated Steel NEMA-12 (IP52)
Options:
Vacuum pump for sample extraction
Extended range sensors
RS232 digital output
Pressure transducer (for gr/lb. or ppmV
measurement)
In-line particulate/coalescing filter
Remote sensor mounting kit
Strobe alarm light
High pressure sensor
Platinum or stainless steel mirror
Initiating Automatic Balance Control