Edgetech Instruments 1500 Specifications

ISO/IEC 17025:2005 Accredited
ISO 9001:2008 Cered
A self-contained portable, multi-function hygrometer
that accurately and reliably measures the moisture content in gases, with the battery life and ruggedness required for field use.
The Model 1500 is a drift-free, low maintenance hygrometer featuring high durability, NIST traceable certification and ease of use. Its primary method chilled mirror sensor provides
precision dew/frost point temperature measurement in a
robust, transportable package.
Simply attach the sample line, adjust the flow with the integral flow meter and turn on the power. Within minutes you will be collecting field dew/frost point data with the same chilled mirror accuracy of a laboratory standard hygrometer. With a nominal battery life of 12 hours, you will be able to collect data for the entire shift or spot check your processes for
several days without recharging.
The Model 1500 is highly regarded by calibration technicians and manufacturing personnel for its quick dry down, fast response and long life. It can be fitted with a precision air temperature sensor and pressure transducer, and either locally or remotely mounted Edgetech Instruments D-Probe or S-Series chilled mirror sensors.
calibrated and supported in the USA in a modern, ISO
9001:2008 certified facility and calibrated in an ISO/IEC 17025:2005 accredited calibration laboratory. All calibra­tions are traceable to NIST.
Features:
Primary method sensor provides absolute knowledge of
dew/frost point temperature
Automatic Balance Control mode provides automatic correc-
tion for optics contaminants
Temperature measurement option for live relative humidity
measurement
Pressure measurement option for live parts per million by
volume measurement
Current or voltage outputs, RS-232 optional
User programmable alarm functions
Internal vacuum pump to pull sample
Portable Chilled Mirror Dew/Frost Point Hygrometer
Applications:
Ambient air
Glove boxes
HVAC ducts
Environmental test chambers
Dry air circulation pipes
Refrigerated storage rooms
Engine test rooms
1-29 REV 1
www.edgetechinstruments.com
The Edgetech Instruments Model 1500
Toll Free: 800-279-3729 Ph: 508-263-5900 Fax: 508-486-9348
E-mail: h2o@edgetechinstruments.com
Web: www.edgetechinstruments.com
Technical Specifications
ISO/IEC 17025:2005 Accredited
ISO 9001:2008 Cered
Sensor Type:
Primary method chilled mirror dew/frost point
sensor
Sensor Options:
The remotely mounted D-Probe insertion sensors
are air cooled with two stages.
The S-Series flow through sensors are available in
both locally and remotely mounted configurations with either two or three stages. Note: Units fitted with S3 sensors are AC powered only.
Measurement and Performance:
Dew/Frost Point Measurement Ranges:
Two-Stage DS2: -4 to 122°F (-20 to 50°C) air cooled Two-Stage DX2: -40 to 122°F (-40 to 50°C) air cooled Two-Stage S2SC: -40 to 122°F (-40 to 50°C) fan cooled Three-Stage S3: -76 to122°F (-60 to 50°C) fan cooled Three-Stage S3: -103 to 122°F (-75 to 50°C) liquid cooled
Air/fan cooled sensor performance at 25°C ambient and 1 Atm.
Accuracy:
Dew/Frost Point: +/- 0.36°F (+/- 0.20°C)
Functional:
Power Charger:
115-230 VAC, 50-60 Hz
Operating Temperature:
32 to 122°F (0 to 50°C)
Sample Connection:
¼ inch compression fitting
Sample Flow Rate:
0.5 to 5 SCFH; Integral flow meter and control valve standard with S­Series sensors
Sample Pressure:
0 to 150 psig;
Higher pressures available, consult factory
Output: Digital:
RS-232, Optional Output: Analog: 4 to 20 mA or 0-5 VDC; -50 to 50°C or
-75 to 75°C dew/frost point span
Compliance:
9.0 VDC, 450 ohms
Display:
Red LED, 0.5” tall; 8 digit alphanumeric
Alarm:
Visual: Flashing ALARMmessage
Relay:
1 Form C, non-latching
10A at 240 VAC, 8A at 24 VDC, 1/2 HP at 240 VAC
Dimensions:
14.75” H x 18” W x 7” D
Weight:
24 lbs.
Enclosure:
Plastic carrying case
The Edgetech Instruments Model 1500