Edgetech OxyTrans Operator's Manual

OxyTrans
Loop Powered Two Wire Oxygen Transmitter
OPERATORS MANUAL
19 Brigham Street Unit 8 Marlborough, MA 01752 508-263-5900 Fax: 508-486-9348 www.edgetechinstruments.com h2o@edgetechinstruments.com
Rev. A April 2009
OxyTrans
QUICK STARTUP GUIDE
STARTUP
1. Using the supplied Mounting Kit, mount the OxyTrans in position to measure the desired gas.
Caution: The cell must be mounted vertically with Digital
Display at the top.
2. Connect input and output tubing. See Manual for proper tubing
materials. Shutoff valves should be installed on both sides of the sensor.
3. Connect loop power supply, using cable supplied.
4. Connect data acquisition system or other series load if
required.
5. Allow sufficient equilibration time before recording data.
TABLE OF CONTENTS Page
1.0 Quick Startup Guide
2.0 Commitment to Quality 3
3.0 Introduction 4
3.1 General Description 4
3.2 Micro Fuel Sensor 4
3.3 The Effect of Pressure 4
3.4 Calibration Characteristics 5
3.5 Standard Configurations 5
3.6 Options and Accessories 6
4.0 Installation 8
4.1 Mounting 8
4.2 Plumbing 8
4.3 Electrical Connections 9
5.0 Operation 11
5.1 Digital Display 11
5.2 Analog Output 11
6.0 Maintenance
6.1 Calibration 12
6.2 Out of Service Procedure 13
6.3 Troubleshooting 13
6.4 Cell Replacement 14
7.0 Specifications 18
8.0 Appendix 19
8.1 Warranty 20
8.2 Oxygen Cell Housing Drawing 21
8.3 Intrinsically Safe Barrier (Optional) 22
1
LIST OF ILLUSTRATIONS Page
3-1 Cell I/O Curve 5 3-2 Standard Configurations 6 4-1 Mounting Kit 8 4-2 Mounting the Transmitter 9 4-3 Electrical Connections 10 5-1 Digital Display 11 6-1 Calibration Adjustment 12
LIST OF TABLES Page
3-1 Standard Factory Ranges 6
2
2.0 EDGETECH INSTRUMENTS INC.'S COMMITMENT TO QUALITY
To Our Customers:
Thank you for purchasing one of our products. At Edgetech Instruments Inc., it is our policy to provide cost-effective products and support services that meet or exceed you requirements, to deliver them on time, and to continuously look for ways to improve both. We all take great pride in the products we manufacture.
We want you to be entirely satisfied with your instrument. The information contained in this manual will get you started. It tells you what you need to get your equipment up and running, and introduces its many features.
We always enjoy hearing from the people who use our products. Your experience with our products is an invaluable source of information that we can use to continuously improve what we manufacture. We encourage you to contact or visit us to discuss any issues whatsoever that relate to our products or your application.
The Employees of Edgetech Instruments Inc.
3
3.0 INTRODUCTION
3.1 GENERAL DESCRIPTION
The Edgetech Instruments Inc. OxyTrans 2-wire 4 to 20 mA Loop Powered Oxygen Transmitter consists of two sub-systems:
1. The Analyzer Unit, which is an electro-chemical fuel sensor designed
for measuring oxygen.
2. The Electronics Unit, which provides sensor signal conditioning, digital
display, and electrical analog outputs.
The gas to be measured is transported to the sensor membrane of the Micro Fuel sensor. This sensor is an electro-chemical and galvanic cell that transforms the oxygen in the measured gas into an electrical current.
The Electronics Unit measures this electrical signal, providing visual and electrical information to the operator, directly in either parts-per-million of oxygen (ppm O2), or as a percentage of the total measured gas (%O2).
3.2 MICRO FUEL SENSOR
The oxygen sensor is a sealed plastic disposable electrochemical sensor. The active components are a cathode and an anode, which are immersed in a 15% aqueous KOH electrolyte solution. When oxygen in the sample gas is sensed by the membrane, the cell converts the resulting chemical reaction into an electrical current change. Its action is similar to that of a battery, with one important difference. In a battery, all reactants are stored within the cell. In the oxygen sensor, one of the reactants, oxygen, comes from outside the device as a constituent of the sample gas being analyzed.
3.3 THE EFFECT OF PRESSURE
In order to determine the amount of oxygen present in the gas sample, it is
necessary that the gas diffuse into the cell under constant pressure, by Dalton’s
Law. If the total pressure increases during a measurement, the rate that oxygen reaches the cathode through the diffusing membrane will also increase. The electron transfer, and therefore the resulting measured output current, will increase even though the oxygen concentration in the sample has not changed.
Therefore, it is important that the sample pressure at the fuel cell remain constant during measurements.
4
0
10
20
30
40
50
60
70
80
90
100
0 0,2 0,4 0,6 0,8 1
oxygen concentration %
output zell in µA
3.4 CALIBRATION CHARACTERISTICS
Since the total pressure of the sample gas at the input of the measuring cell is constant, a convenient characteristic of the cell is that the current produced in an external circuit is directly proportional to the rate at which oxygen molecules reach the cathode, and this rate is directly proportional to the concentration of oxygen in the gaseous mixture. In other words it has a linear characteristic curve, as shown in Figure 3-1. An advantage is that measuring circuits do not have to compensate for nonlinearities.
In addition, since there is zero output in the absence of oxygen, the characteristic curve has close to an absolute zero (within ± 1 ppm oxygen). In practical application, zeroing is still used to compensate for the combined zero offsets of the cell and the electronics. The electronics is zeroed automatically when the instrument power is turned on.
Figure 3-1 Characteristic Input/Output Curve for the Oxygen Cell
3.5 STANDARD FACTORY CONFIGURATIONS
The Edgetech Instruments Inc. OxyTrans series of transmitters are available in a number of convenient configurations. These include both vertical and horizontal mounts, so that the digital display may be viewed either from the top or from the front. Intrinsically Safe barriers are also available if required.
5
A number of standard ranges are available. These include:
RANGE
PARTS-PER-MILLION
PERCENT OXYGEN
0 to 10
X
0 to 100
X
0 to 1000
X
0 to 10,000
X
0 to 1
X
0 to 25
X
Table 3-1 Standard Factory Ranges
3.6 OPTIONS AND ACCESSORIES
Other standard configurations include a Panel Mount unit with optional safety barriers. A Portable unit is available in a carrying case with a built-in sampling system, including a flow meter, flow adjustment valve, sintered filter, vacuum pump, and power supply. Also available is an Explosion-Proof (ATEX approved) version. A unit with electrical interface compatible with the Hart Protocol is a standard option. Additional ranges are available on special order. Some ranges provided in the past include 0 to 20,000 ppm, 0 to 50,000 ppm, and 0 to 100% O2. See Figure 3-2 below.
Figure 3-2a Explosion-Proof Unit Figure 3-2b Vertical Mount
Figure 3-2c Horizontal Mount Figure 3-2d Portable Unit
6
Loading...
+ 17 hidden pages