Edaphic Scientific ESRAD-200 Operating Instructions Manual

A
Y
NOTE: Remove Air Intake Plug (N) before
Remote CO2 Storage Safety Dual Alarm with Oxygen Monitor
Operating Instructions
M
odel: ESRAD-200
Table of Contents
2. Package Content Check & Main Unit View
3. LCD Display Symbol
4. SEU (Sensor Unit) Function Instructions
5. RDU (Remote Display Unit) Function Instructions
6. Safety Notes
7. Caring For Product
8. Installation Instructions
9. Customizing Settings
10. CO2 & O2 & Temperature Specifications
11.Relay Outputs
12. Weight & Dimensions
13. Fault Codes & Troubleshooting Guide
13. Support & Warranty
1. Product Instructions
T
hank you for selecting the ESRAD-200 CO2 and O2 Monitor. It is designed to detect the presence of carbon dioxide and oxygen in the ambient air to protect people in confined spaces. High concentrations of CO2 or low concentrations of oxygen in confined spaces are dangerous, and may lead to health problems ranging from headaches and fatigue to asphyxiation and death. The RAD-0200 is widely used to protect workers and employees in CO2 storage areas, breweries, wineries, cellars, beverage dispensing areas, and fast food outlets. It has both audible and visual alarms which activate when CO2 or O2 concentrations reach user-defined levels. Detection of high levels of CO2 will also activate 2 relays that could be used for a fan or air-conditioning system to ventilate the confined space, and reduce CO2 concentration and add O2 concentration in the area.
The ESRAD-200 CO2 & O2 Monitor is cost-effective and has many features:
1. Dual Beam NDIR (Non-Dispersive Infrared) technology is used to measure CO2 electrochemical technology to measure O2 concentration up to 30%.
2. With the SEU (Sensor Unit) and RDU (Remote Display Unit), it can connect multiple RDUs for safety notices.
3. Large LCD display clearly indicates ambient CO2& O2 concentrations.
4. Relay outputs can control a fan to ventilate confined spaces.
5. Audible and visual alarm indications.
6. IP65 Water Proof Protection of SEU (Sensor Unit).
2. Package Content Check & Main Unit View
he ESRAD-200 package comprises the following parts:
T
concentration up to 50,000ppm (parts per million) and
powering SEU. Use Plug when cleaning SEU.
SEU (Sensor Unit)
B
O
N
| Z
A. LCD display B. Buzzer
C. Green LED (P D. Red 1 LED(
E. Red 2 LED( AL2 ) M. Power Supply F. Yellow LED( Fault indication) N. CO2 entry G. Reset Button O. O2 entry H. Enter Button Z. Gas Calibration Port
ower indication)
AL1 ) L. Communication Cable to RDU
I. Mode Button J. Relay 2 output
Blue & white: NC) K. Relay 1 output Blue & white: NC)
JKLM
(red & white: NO,
(red & white: NO,
RDU (Remote Display Unit)
QRS
P
T
U V
W
C
D
E F G H I
Main Unit:
SEU( Sensor Unit) RDU (Remote Display Unit) Panel holders
User manual 8 meters (26 feet) communication cable
Accessories:
Plug lock 1 pcs Screw 10pcs Nail cable clip 10 pcs Warning signs 2 pcs
Expansion plug
10 pcs
Y
X
P. Green LED (Power indication) Q. Red 1 LED (AL1 ) V.Enter Button
R. Red 2 LED ( AL2) W. Buzzer S. Yellow LED ( Fault indication) X. RJ45 Plug for next RDU (Output)
T. LCD display
U. Mode Button
. RJ45 Plug for SEU (Input)
3. LCD Display Symbol
Symbol Meaning Description
CO2 Concentration ppm
Ambient CO2 concentration
(Parts Per Million) O2 Concentratio
Alarm Alarm
Diagnostics
n Ambient O2 concentration
icon
T
est communications between
the SEU and RDU Relay 1 will be triggered when the
CO2 concentration exceeds the
CO2/
O2 First
Alarm level
first alarm level or O2 concentration drops below the first alarm level, the Red 1 LED will flash, and buzzer will sound. Relay 2 will be triggered when CO2 concentration exceeds the
CO2/
O2 Second
Alarm level
second alarm level or O2 concentration drops below the second alarm level. The Red 1 & Red 2 LEDs will flash, and the buzzer will sound. Calibrate the CO2 / O2 sensors
Calibration
when the accuracy deviates from the actual concentration.
Recover Setting
Factory
Recover factory default settings and cancel customized settings.
Indicate the CO2 leakage once
ESC
the CO2 level is
above the
second alarm level. The CO2 concentration is above
Hi
5%, or the O2 concentration is lower than 30%.
Fan
If CO2 > O2<O2 Alarm Level, fan will run.
CO2 Alarm Level or
4. SEU (Sensor Unit) Function Instructions
The SEU (Sensor Unit) should be placed in a room where the CO2 is likely to accumulate or O2 reduced, such as a room where CO2 canisters or tanks are stored. The large LCD displays the ambient CO2 & O2 concentration.
The SEU has the “DIAG”, “AL1”, “AL2”“CALI”, “ReFactSet” function. The “DIAG” function executes communication tests between the SEU and RDU. The user can do the calibration under the “CALI” mode when n ecessary. If data-setting is done incorrectly, the user can use the “ReFactSet” back to the original factory setting.
There are two alarm levels: “AL1” and “AL2”. Both alarm levels are adjustable. The first alarm level of CO2 can be set to 5,000ppm, 1%, 1.5%, 2%. The default first alarm level is 1.5%. The first alarm level of O2 can be set to 18%, 18.5%, 19%, 19.5% or 20%. The default first alarm level is 19%.The second alarm level of CO2 can be set to 1.5%, 2%, 2.5%, 3%,
3.5% or 4% CO2. The default second alarm level is 3%. The second alarm level for O2 can be set to 16%, 16.5%, 17%, 17.5% or 18%. The default second alarm level for O2 is 17%.
When the RAD-0200 CO2 &O2 Monitor detects CO2 concentration exceeds the first alarm level or O2 concentration drops the first alarm level, the red 1 LED will blink and the buzzer will sound intermittently, and the relay will be
triggered. When CO2 concentration drops below the first alarm level, or O2 concentration raises to the first alarm level, the red 1 LED will stop blinking and the buzzer will stop beeping.
If the concentration of CO2 continues to rise above the second alarm level, or the concentration of O2 drops below the second alarm level, the red 1 and red 2 LED’s will both flash together, the tempo of the buzzer will increase, and the second relay will be triggered. When the CO2 concentration drops below the first alarm level the red 1 and red 2 LED’s will stop blinking, and the buzzer will stop beeping. However, the yellow FLT LED will stay on to show an event has occurred until the RAD-0200 is restarted with the reset button RAD-0200 or the AC adapter is unplugged and reconnected.
The green LED will light continuously when the power is normally supplied. Warning: If the ambient CO2 concentration reaches the second alarm level
on SEU & RDU, there will be a safety notice “ESC” displayed on the LCD. Careful action, such as ventilating t he space, must be take n before entering the room where the SEU is placed.
If the communication cable between the SEU & RDU isn’t connected properly, the fault LED on the SEU will blink. To fix this, reconnect the cable. If the cable is accidently inserted into the output port on the RDU, after one minute “Er7” will flash on the LCD. Unplugging the cabl e and plugg ing i t into the In put port will normally correct the problem.
5. RDU (Remote Display Unit) Function Instructions
The RDU (Remote Display Unit) should be placed outside the CO2 storage room during use. The RDU is connected to the SEU with a CAT5 cable. An 8 meter (26 ft.) cable is supplied, although up to 3 daisy-chained RDUs over 76 meters (250 ft.) using CAT5 cable has been tested. The RDU should be placed where it can be conveniently observed before entering the room where the SEU is located. The RDU is a repeater, and displays the measurements made by the SEU on an easy-to-read digital LCD along with important safety information.
The RDU has “DIAG” function. The “DIAG” can test the communication between the SEU and RDU. Resetting the RAD-0200 CO2 and O2 Monitor is only available from SEU.
6. Safety Notes
Warning: Yo ur safety is very important to us. To ensure to use the product correctly and safety, please read these warnings and the entire User Manual before using the product. Otherwise, the protection provided by the equipment may be impaired. These warnings provide i mportant safety information and should be observed at all times.
1. Handle the device carefully; do not subject the product to impact or shock. Otherwise, this may cause the accuracy to drift.
2. Do not place the unit or the adaptor near a heat source. Heat can cause distortion of the unit, which may result in an explosion or fire.
3. Do not touch the exposed electronic circuitry of the device under any circumstances, as there is danger of electric shocks.
4. Use only the included power adaptor. Improper power adaptors or power sources can cause serious damage to the product, or result in injury or death to the user.
5. Use the “DIAG” function to verify the communication between SEU and RDU works normally.
6. Make sure that the power adaptor is secured tightly by a plug lock so the power adapter cannot be disconnected accidently or by hand.
7. Do not enter into the room directly if there has safety notice “ESC” displayed on the LCD of SEU & RDU. Careful and protective action must be taken before entering the room where the SEU is installed.
8. Take care of cable connection between SEU and RDU. Make sure the cable from SEU is connected into the INPUT port of RDU.
9. Ensure the external power supply to a ventilat ion fan is tested while the relay is working. If there has no power to the fan, the relay will not turn it on, which may result in potentially dangerous high CO2 concentration or low O2 concentration in a confined space.
Power Supply for fan
Fan for Ventilation
Power Adapter
SEU (Sensor unit)
8m(26ft) comm. cable
RDU
Display
Door
7. Caring For the Product
To receive the maximum benefit from using this product, please observe the follow guidelines.
1. Repair - Do not attempt to repair the product or modify the circuitry by yourself. Please contact your local dealer or a qualified repairman if the product needs servicing, including the replacement or calibration of the sensors.
2. Note -- The oxygen sensor must be replaced every 3 years.
3. Cleaning - Disconnect the power before cleaning. Clean with a damp cloth only. Do not use liquid cleaning agents such as benzene, thinner or aerosols, as these will damage the device.
4. Maintenance – We recommend testing the communication between the SEU and RDU under the ‘DIAG” function to verify the SEU and RDU are working properly. If the four LED’s blink and the buzzers sound simultaneously, it indicates that SEU and RDU work normally .
When the LCD displays a safety icon “ESC”, please take immediate protective action to check if a CO2 leak has occurred. We suggest a calibration and thorough function chec k at least yearly to make sure that the RAD-0200 CO2 and O2 Monitor is working properly.
8. Installation Instructions
Please carefully take out the SEU (Sensor Unit), RDU (Remote Display Unit), panel holders, network cable connector, communication cable, user manual, plug lock, screws, expansion plugs, nail cable clips, and warning poster from the package. Step-by-Step Installation:
1. Choose a suitable location to install the SEU. Fix a panel holder on the wall with the four screws (included); the recommend instal latio n heigh t is abo ut
0.45 meters (1.5 feet) from the floor and as c lose to the manifolds and valves as possible.
2. Put the SEU on the panel holder. Remove the Air Entry Plug from the air entry port (N). Save plug for use when cleaning SEU.
3. Fix another panel holder in a suitable l oc ati on outs id e the monitored space with screws (included). Push the RDU onto the panel holder, and stick the warning poster next to RDU.
4. The communicatio n cable is pre-wired to the SEU. Route the cable to the RDU and fixed the communication cable to the wall by nail cable clips, and then plug the cable into the input port on the RDU. Communication is now ready between the SEU and RDU.
5. The RAD-0200 CO2 and O2 Monitor has two r elay outpu ts: one is for AL1 and the other is for AL2. The relay cable is pre-wired to the SEU. Relays can control a fan or blower to ve ntilate monitored space whe n necessary and the relay will be triggered when the CO2 concentration exceeds the first alarm level or O2 concentration drops below the first alarm level.
6. After finishing the installation, Connecte d the AC power adapter into the electrical supply outlet. Mount the Plug lock by expansion plugs so that the power adapter cannot be disconnected without use of tools.
gas
tanks
RDU Display
to the 2nd/3rd RDU (optional,daisy chain)
7. When the power has been connected, The SEU and the RDU will begin to work. Please use the “DIAG” function to verify the communication between SEU and RDU, the four LED’s will blink and buzzer will sound on SEU & RDU. Communication is ready when the information displayed is the same on the SEU & RDU.
Mount the plug lock
9. Customizing Settings
Once power has been connected, the SEU and RDU will begin to monitor the CO2 & O2 concentration. In order to the get the timely alarm safety information or to meet personal requirements, you can customize the parameters as necessary.
Temperature ºC/ºF:
1. Press “Enter” to switch between ºC and ºF temperature.
Using the DIAG function:
PRESS
MODE
1. Press Mode until the “DIAG” icon flashes
2. Press Enter, the four LED’s will blink on the SEU and the buzzer will sound
3. The four LED’s will blink and buzzer will sound simultaneously on the RDU
Setting the CO2
First Alarm parameter:
PRESS
MODE
1. Press Mode until the “AL1” icon flashes
2. Press Enter. The “AL1” icon will show on the LCD
3. Press Mode to go through 5,000ppm, 1%, 1.5%, 2% CO2 alarm levels
4. Press Enter again to save the setting
PRESS
ENTER
PRESS
ENTER
Setting the CO2 Second Alarm parameter:
PRESS
MODE
PRESS
ENTER
1. Press Mode until the “AL2” icon flashes
2. Press Enter. The “AL2” icon will show on the LCD
3. Press Mode to go through 1.5%, 2%, 2.5%, 3%, 3.5%, 4% CO2 alarm levels
4. Press Enter again to save the setting
Note: The second alarm level must always be higher than the first alarm level when customizing the alarm level parameters.
Using the O2 CALI functio
n:
“ZEr” Calibration means the SEU is saturated in nitrogen gas. “SPA” Calibr ation means the SEU is saturated in 21% Oxygen (outdoor air).
PRESS
ENTER
PRESS
MODE
MODE 10s
PRESS
PRESS
ENTER
Setting the O2 First
Alarm parameter:
PRESS
MODE
PRESS
ENTER
1. Press Mode until the “AL1” icon flashes
2. Press Enter. The “AL1” icon will show on the LCD
3. Press Mode to go through 18%, 18.5%, 19%, 19.5%, 20% O2 alarm levels
4. Press Enter again to save the setting
Setting the O2 Second
Alarm parameter:
PRESS
MODE
PRESS
ENTER
1. Press Mode until the “ AL2” icon flashes
2. Press Enter, the “AL2” icon shows on LCD
3. Press Mode to go through 16%, 16.5%, 17%, 17.5%, 18% O2 alarm levels
4. Press Enter again to save the setting afte r selection
Note: The second alarm level for CO2 must be HIGHER than the first alarm level. The second alarm level for O2 must be LOWER than the first alarm level. Otherwise, the alarms will not function as expected.
Using the CO2 CALI function:
“ZEro” Calibration means the SEU is saturated in nitrogen (0% CO2) gas. “SPAn” Calib ration means the SEU is saturated in 40,00ppm (4%) CO2.
PRESS
ENTER
PRESS
MODE
MODE 10s
PRESS
PRESS
ENTER
1. Press Mode until the CO2 and “CALI” icon flashes.
2. Press Enter. The “CO2”, “CALI” and “ZEro” icons will show on the LCD. Press and hold Mode for at least 10 seconds, The “CALIBRATING” and “ZEro” icons will flash simultaneously. The calibration will be done automatically. After about 3 minutes the LCD will display “Pass” or “Fail”. If “Fail”, please try again.
3. Press Enter until th e “CO2”, “CALI” and “SPAn” icons will show on the LCD. Press and hold Mode for at least 10 seconds. The “CALIBRATING” and “SPAn” icon will flash simultaneously and the calibration will be done automatically. After about 3 minutes the LCD will display “Pass” or “Fail”. If your unit fails calibration, please try again before calling support.
Note: Before starting either a “ZEro” or ”SPA n” calibration, wait at least 5 minutes for the CO2 level inside the SEU to stabilize. For calibration, gas cylinders connected to calibration port Z should be used.
The O2 CALI function is as same as CO2, except that the “O2” icon flashes to signify calibration of the oxygen sensor.
Note: Before doing either a “ZEr” or ” SPA” calibration, wait at least 5 minutes for the O2 level inside the SEU to stabilize. For Zero calibrations, a calibration bag connected to a nitrogen gas cylinder is typically used
Using the ReFactSet function:
PRESS MODE
PRESS
ENTER
PRESS
MODE
1. Press Mode until the “ReFactSet” icon flashes.
2. Press Enter, and then press Mode to choose either “Yes” or “No”.
3. Press Enter again to save the setting. Note: If the user sets the data or calibrates the sensor incorrectly, use the ReFactSet (recover the factory Setting) to come back the default factory setting.
10 .Specifications
CO2 & O2 & Temperature specifications:
CO2 & O2 Specification
Measurement Range
Display Resolution
Accuracy
Repeatability Temperature
Dependence Pressure Dependence
Response Time AL1 (First Alarm
Level) AL2 (Second Alarm
Level) Sound Alarm 80db@10cm Warm-Up Time <60 seconds at 22°C
Temperature Specifications:
Temperature Range 0°C to 50°C (32°F to 122°F) Display Resolution 0.1°C (0.1°F) Display Options °C/°F
Accuracy Response Time 20-30 minutes (case must equalize with environment)
CO2: 0 - 50,000ppm (5%) display O2: 0- 30% display CO2: 10ppm at 0~10,000ppm;
100ppm at 10,001~50,000ppm O2: 0.1% CO2: ±100ppm or ±5% of reading, whichever is
greater O2: Better than ± 3% of FS over 0.1 to 30% CO2:±20ppm @400ppm
O2: Less than ±1.0% ±0.2% of reading per °C or ±2ppm per °C, whichever
is greater, referenced to 25°C
0.13% of reading per mm Hg CO2: <60 seconds for 90% response to step change
O2: <30 seconds for 90% response to step change CO2: 5000ppm, 1 / 1.5 / 2 % , Default AL1= 1.5% O2:18%,18.5%,19%,19.5%,20%, Default AL1=19% CO2: 1.5 / 2 / 2.5/ 3 / 3.5 / 4 % , Default AL2= 3% O2: 16%,16.5%,17%,17.5%,18%, Default AL2=17%
±1°C(±2°F) when CO2 concentration is below first alarm level
Operating Conditions: Operating
Temperature
0°C to 40°C ( 32°F to 104°F)
Humidity Range 0 ~ 95% RH non-condensing
Storage Conditions: Storage T emperature -20°C to 60 °C (-4°F to 140°F)
Power Supply &
Relay Output:
Power Supply AC adapter 110/220 VAC
Voltage 100 ~ 240 VAC
AC Input
AC/DC Output
Frequency 50 / 60 Hz Power
Requirement
1 W maximum @ 115 VAC 60 Hz
2 W maximum @ 230 VAC 50 Hz Voltage 6VDC Power 3.0 W
Peak Input Current 0.5 A from 6 VDC
Two Relay output: Relay 1 operates at alarm 1 for
Relay Output
CO2 or O2, Relay 2 operates at alarm 2 for CO2 or
O2. Peak Current< 2A@30 VDC or 250 VAC,SPDT.
11.Relay outputs
There are two relay output for this item. Relay 1 operates at alarm 1 for CO2 or O2, Relay 2 operates at alarm 2 for CO2 or O2. The relay 1 will be triggered when CO2 concentration exceeds the first alarm level or O2 concentration drops the first alarm level. The relay 2 will be triggered when CO2 concentration exceeds the second alarm level or O2 concentration drops below the second alarm level.
12. Weight & Dimensions
Weight: SEU (Sensor Unit):1200 g (include cables)
RDU
(Remote Display Unit): 120 g
Dimensions:
SEU (Sensor Unit)
13. Fault Codes &Troubleshooting Guide
This section includes a list of Frequently Asked Questions for problems yo u may encounter with the RAD-0200 CO2 &O2 Monitor.
LCD
No
Fault
Icon
1 Er3
2 Er4
Er5
3
Er6
4 Er7
5 Er8
Description
(of the fault)
The ambient temperature has exceeded the temperature range 0°C to 50°C ( 32°F to 122°F)
some wrong measurement or the sensor has exceeded its expected life
EEPROM System Problem
Internal Data Transmission Error
The accuracy of CO2& O2 sensor may deviate from the actual concentration.
SEU
Indication
“Er3” flash Fault
LED blink,
Buzzer beep
“Er4” flash
Fault LED blink,
Buzzer beep
“Er5” & “Er6” flash, Fault LED blink, Buzzer beep
“Er7” flash, Fault LED blink, Buzzer beep
“Er8” flash Fault LED blink, Buzzer beep
RDU
Indication
“Er3” flash Fault LED blink, Buzzer beep
“Er7” flash Fault LED blink, Buzzer beep
“Er7” flash Fault LED blink, Buzzer beep
“Er7” flash, Fault LED Blink, Buzzer beep
“Er8” flash Fault LED blink, Buzzer beep
Suggested Actions
This error will disappear when the temperature returns to the range between 0°C and 50°C ( 32°F to 122°F)
Please unplug the AC adapter and reconnect it. If the “Er4” always appears, please contact with the local dealer.
Please unplug the AC adapter and reconnect it. If the “Er5, Er6” always appear, please contact with the local dealer.
Please unplug the AC
adapter and reconnected it.
Check the RJ45 plug is
connected into the INPUT port of RDU,if the “Er7” displays on the RDU only.
Please unplug the AC
adapter and reconnect it If the “Er8” still appears, please contact with the local dealer.
Please calibrate the unit,
after the calibration, If the “Er8” still appears, please contact with the local
.
dealer
RDU (Remote Display Unit)
mm[inch]
mm[inch]
14.Support & Warranty
Support
The quickest way to obtain technical support is via email. Please send all support enquires to info@edaphic.com.au. In your email, please include a clear, concise definition of the problem and any relevant troubleshooting information or steps taken so far, so we can duplicate the problem and quickly respond to your inquiry.
Warranty
This unit comes with a 1YEAR (warranty period) limited manufacturer’s warranty, starting from the date the unit was shipped to the buyer. During this period of time, Edaphic Scientific warrants our products to be free from defects in materials and workmanship when used for their intended purpose and agrees to fix or replace (at our discretion) any part or product that fails under normal use.
To take advantage of this warranty, the product must be returned to Edaphic Scientific at your expense. If, after examination, we determine the product is defective, we will repair or replace it at no additional cost to you.
This warranty does not cover any products that have been subjected to misuse, neglect, accident, modifications or repairs by you or by a third party. No employee or reseller of Edaphic Scientific’s products may alter this warranty verbally or in writing.
Returns
If the product fails under normal use during the warranty period, an RMA (Return Material Authorization) number must be obtained from Edaphic Scientific. After the item is received, Edaphic Scientific will repair or replace the item at our discretion.
To obtain an RMA number, please call Edaphic Scientific at (61-2) 6584
5768. When requesting an RMA number, please provide the reason for return and original order number.
If we determine that the product failed due to improper use (water damage, dropping, tampering, electrical damage etc.) or abuse, or if it is beyond the warranty period, we will inform you of the cost to fix or replace your device.
If you are returning your device due to a warranty claim (with an RMA number) and you still have the unit original package, please use it to ship your unit to us. Please make sure to include th e provided RMA number on the outside of the box, preferably on the shipping label. Make sure you secure the unit inside the package properly to prevent any damage during transit that could void your device’s warranty. Finally, please ship your device to the address shown under the “Contact Us” section below. Edaphic Scientific will not, under any circumstances, be responsible for your shipment expenses and no refund will be issued for shipping charges necessary for you to ship the unit to us.
Liability
All liabilities under this agreement shall be limited to the actual cost of the product paid to Edaphic Scientific. In no event shall Edaphic Scientific be liable for any incidental or consequential damages, lost profits, loss of time, lost sales or loss or damage to data, injury to person or personal property or any other indirect damages as the result of use of our products.
Contact us: We’re here to help!
If the troubleshooting guide above doesn’t help you solving your problem or for more information, please contact us using th e information below.
info@edaphic.com.au
www.edaphic.com.au
Ref. No.:012013
Appendix A
CO2 level maximum standards in the US and worldwide. For workplaces that fall under OSHA guidelines, you should monitor the display to assure that the CO2 level over an 8-hour day does is not over the 5,000ppm TWA
exposure limit.
nd
Note the 30,000ppm STEL exposure limit is the 2 safety should never be raised.
alarm level by default (3% CO2). This level can be lowered in the settings, but for worker and first-responder
Standards for Oxygen in the Workplace
OSHA does not have a permissible exposure limit (PEL) for oxygen.
Oxygen less than 19.5%
Oxygen less than 18%
Oxygen less than 15%
Oxygen less than 12%
Appendix B
T
o conform with the National Boiler Inspection Code (NBIC) CO2 safety standards, 2 warning signs must be posted. You can download
them here for printing: http://co2meters.com/Documentation/Other/NBIC-CO2-Warning-Sign.pdf
OSHA regulations define an oxygen-deficient atmosphere as one with an oxygen concentration less than 19.5%. Progressive onset of fatigue and other symptoms, Effects accelerate with exertion. O2 levels insufficient to sustain life indefinitely. Loss of reason, judgment, coordination. Dimming vision, nausea, vomiting, gasping for breath. Immediate unconsciousness. Asphyxiation within minutes. Progressive brain damage leading to death unless rescued.
Appendix C
Wiring Diagram for Alarm 1 to control 120VAC, 10A exhaust fan.