D Abbreviations..............................................................................................................................200
XIV
Patient Monitor User Manual Intended Use and Safety Guidance
Chapter 1 Intended Use and Safety Guidance
1.1 Intended Use
The iM50 patient monitor is intended to be used for monitoring, storing, reviewing, recording,
and generating alarms for multiple physiological parameters of adults, pediatrics and neonates in
hospital environments. The monitored physiological parameters include: ECG, respiration
(RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO
pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2) and quick temperature
(Quick TEMP). This monitor is suitable for use in intra-hospital transport and hospital
environments including clinic, emergency department, wards, PACU and NICU.
The iM60 patient monitor is intended to be used for monitoring, storing, reviewing, recording,
and generating alarms for multiple physiological parameters of adults, pediatrics and neonates in
hospital environments. The monitored physiological parameters include: ECG, respiration
(RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO
pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2) and cardiac output (C.O.).
This monitor is suitable for use in hospital environments including OR, ICU, NICU, PACU and
wards.
), non-invasive blood
2
), non-invasive blood
2
The iM70 patient monitor is intended to be used for monitoring, storing, reviewing, recording,
and generating alarms for multiple physiological parameters of adults, pediatrics and neonates in
hospital environments. The monitored physiological parameters include: ECG, respiration
(RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO
), non-invasive blood
2
pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), cardiac output (C.O.) and
anesthetic gas (AG). This monitor is suitable for use in hospital environments including OR, ICU,
NICU, PACU and wards.
The iM80 patient monitor is intended to be used for monitoring, storing, reviewing, recording,
and generating alarms for multiple physiological parameters of adults, pediatrics and neonates in
hospital environments. The monitored physiological parameters include: ECG, respiration
(RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO
pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO
2
), non-invasive blood
2
), cardiac output (C.O.) and
anesthetic gas (AG). This monitor is suitable for use in hospital environments including OR, ICU,
NICU, PACU and wards.
The arrhythmia detection and ST Segment analysis are not intended for neonatal patients.
- 1 -
Patient Monitor User Manual Intended Use and Safety Guidance
1.2 Safety Guidance
WARNING
1Before using the device, the equipment, patient cable and electrodes etc. should be
checked. Replacement should be taken if there is any evident defect or signs of aging
which may impair the safety or performance.
2Medical technical equipment such as these monitor/monitoring system must only be
used by persons who have received adequate training in the use of such equipment
and who are capable of applying it properly.
3EXPLOSION HAZARD-Do not use the device in a flammable atmosphere where
concentrations of flammable anesthetics or other materials may occur.
4SHOCK HAZARD-The power receptacle must be a three-wire grounded outlet. A
hospital grade outlet is required. Never adapt the three-prong plug from the monitor to
fit a two-slot outlet.
5Extreme care must be exercised when applying medical electrical equipment. Many
parts of the human/machine circuit are conductive, such as the patient, connectors,
electrodes, transducers. It is very important that these conductive parts do not come
into contact with other grounded, conductive parts when connected to the isolated
patient input of the device. Such contact would bridge the patient's isolation and
cancel the protection provided by the isolated input. In particular, there must be no
contact of the neutral electrode and ground.
6Magnetic and electrical fields are capable of interfering with the proper performance
of the device. For this reason make sure that all external devices operated in the
vicinity of the monitor comply with the relevant EMC requirements. X-ray equipment
or MRI devices are a possible source of interference as they may emit higher levels of
electromagnetic radiation.
7Route all cables away from patient’s throat to avoid possible strangulation.
8Devices connecting with monitor should be equipotential.
9Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing
equipment and IEC/EN 60601-1 for medical equipment). Furthermore all
configurations shall comply with the valid version of the standard IEC/EN 60601-1-1.
Therefore anybody, who connects additional equipment to the signal input or output
connector to configure a medical system, must make sure that it complies with the
requirements of the valid version of the system standard IEC/EN60601-1-1. If in doubt,
consult our technical service department or your local distributor.
10 Only patient cable and other accessories supplied by EDAN can be used. Or else, the
performance and electric shock protection cannot be guaranteed, and the patient may
be injured.
- 2 -
Patient Monitor User Manual Intended Use and Safety Guidance
WARNING
11 Do not rely exclusively on the audible alarm system for patient monitoring. Adjustment
of alarm volume to a low level or off during patient monitoring may result in a hazard
to the patient. Remember that the most reliable method of patient monitoring
combines close personal surveillance with correct operation of monitoring equipment.
12 When interfacing with other equipment, a test for leakage current must be performed
by qualified biomedical engineering personnel before using with patients.
13 During monitoring, if the power supply is off and there is no battery for standby, the
monitor will be off, and only the patient information and alarm settings can be saved.
After reconnecting the power supply, the user should turn on the monitor for
monitoring.
14 Keep away from fire immediately when leakage or foul odor is detected.
15 The device and accessories are to be disposed of according to local regulations after
their useful lives. Alternatively, they can be returned to the dealer or the manufacturer
for recycling or proper disposal. Batteries are hazardous waste. Do NOT dispose
them together with house-hold garbage. At the end of their life hand the batteries over
to the applicable collection points for the recycling of waste batteries. For more
detailed information about recycling of this product or battery, please contact your
local Civic Office, or the shop where you purchased the product.
16 Dispose of the package material, observing the waste control regulations and keeping
it out of children’s reach.
17 After defibrillation, the screen display recovers within 10 seconds if the correct
electrodes are used and applied based on the manufacturers’ instructions.
18 This equipment is not intended for family usage.
CAUTION
1Electromagnetic Interference - Ensure that the environment in which the patient
monitor is installed is not subject to any sources of strong electromagnetic
interference, such as radio transmitters, mobile telephones, etc.
2Keep the environment clean. Avoid vibration. Keep it far away from corrosive
medicine, dust area, high temperature and humid environment.
3Do not immerse transducers in liquid. When using solutions, use sterile wipes to avoid
pouring fluids directly on the transducer.
4Do not use autoclave or gas to sterilize the monitor, recorder or any accessories.
5The device and reusable accessories could be sent back to the manufacturer for
recycling or proper disposal after their useful lives.
6Disposable devices are intended for single use only. They should not be reused as
performance could degrade or contamination could occur.
- 3 -
Patient Monitor User Manual Intended Use and Safety Guidance
CAUTION
7Remove a battery whose life cycle has expired from the monitor immediately.
8Avoid liquid splash on the device. The temperature must be kept between 5qC and
40qC while working. And it should be kept between -20qC and 55qC during
transportation and storage.
9To ensure patient safety, use only parts and accessories manufactured or
recommended by EDAN.
10 Federal (U.S.) law restricts this device to sale by or on the order of a physician.
NOTE:
1Position the device in a location where the operator can easily see the screen and
access the operating controls.
2The monitor can only be used on one patient at a time.
3If the monitor gets damp or liquid pours on the monitor, please contact the service
personnel of EDAN.
4This monitor is not a device for treatment purposes.
5The pictures and interfaces in this manual are for reference only.
6Regular preventive maintenance should be carried out every two years. You are
responsible for any requirements specific to your country.
7The monitor may not be compatible with all models of USB flash drives. Use the USB
flash drives that are recommended by EDAN.
8It is recommended to format the USB flash drive to the FAT file type via PC prior to
use.
1.3 Explanation of Symbols on the Monitor
This symbol indicates that the equipment is IEC/EN60601-1 Type CF
equipment. The unit displaying this symbol contains an F-Type
isolated (floating) patient applied part providing a high degree of
protection against shock, and is suitable for use during defibrillation.
This symbol indicates that the instrument is IEC/EN 60601-1 Type
BF equipment. The unit displaying this symbol contains an F-Type
isolated (floating) patient applied part providing a high degree of
protection against shock, and is suitable for use during defibrillation.
Symbol for “Caution”
- 4 -
Patient Monitor User Manual Intended Use and Safety Guidance
Equipotential grounding terminal
Alternating Current
Power Supply switch
Serial number
Network port
USB (Universal Serial Bus) Connection
Audio alarm is off
NIBP measurement
Trend graph
Freeze
Record
Menu
VGA output, External Monitor
- 5 -
Patient Monitor User Manual Intended Use and Safety Guidance
RS-232 port
Nurse call port
SD Card port
Signal output port
Signal output
The symbol indicates that the device complies with the European
Council Directive 93/42/EEC concerning medical devices.
Authorized representative in the European community
Date of manufacture
Manufacturer
P/N Part Number
Recycle
The symbol indicates that the device should be sent to the special
agencies according to local regulations for separate collection after
its useful life.
Consult Instructions For Use
- 6 -
Patient Monitor User Manual Intended Use and Safety Guidance
Locked position
Gas inlet
Gas outlet (evac)
ISA equipped to measure CO
only.
2
ISA equipped to measure multiple gases.
Federal (U.S.) Law restricts this device to sale by or on the order of a
physician.
(Only applicable to iM50 and iM80) With respect to electrical shock,
fire and mechanical hazards only in accordance with UL 60601-1and
CAN/CSA C22.2 No. 601.1, IEC 60601-2-25*, IEC 60601-2-27,
IEC 60601-2-30,IEC 60601-2-34, IEC 60601-2-49, IEC
60601-2-51* (Symbol * means this standard only applicable to
iM80)
- 7 -
Patient Monitor User Manual Installation
Chapter 2 Installation
NOTE:
1The monitor settings must be specified by the authorized hospital personnel.
2To ensure that the monitor works properly, please read the user manual and follow
the steps before using the monitor.
2.1 Initial Inspection
Before unpacking, check the packaging and ensure that there are no signs of mishandling or
damage. If the shipping cartons are damaged, contact the carrier for compensation and package
them again.
Open the package carefully and remove the monitor and accessories. Check that the contents are
complete and that the correct options and accessories have been delivered.
If you have any question, please contact your local supplier.
2.2 Mounting the Monitor
If all situations are normal, please place the monitor on a flat, level surface, hung on the bed rail,
or mounted on a wall. About how to install the wall mount for the monitor, please refer to the
following content.
2.2.1 Installing Wall Mount for the Monitor
For how to install wall mount for the monitor, please refer to Wall Mounting Bracket Assembly
Instruction.
2.3 Connecting the Power Cable
Connection procedure of the AC power line is listed below:
1Make sure the AC power supply complies with the following specifications: 100V-240V~,
50Hz/60Hz.
2Apply the power line provided with the monitor. Plug the power line to inlet interface of the
monitor. Connect the other end of the power line to a grounded 3-phase power output.
NOTE:
Connect the power line to the jack special for hospital usage.
2.4 Checking Out the Monitor
Make sure there is no damage on the measurement accessories and cables. Then turn on the
monitor, check whether the monitor can start normally. Make sure all alarm lamps light up and
the alarm sound is heard when turning on the monitor.
- 8 -
Patient Monitor User Manual Installation
WARNING
If any sign of damage is detected, or the monitor displays some error messages, do not
use it on any patient. Contact Customer Service Center immediately.
NOTE:
1Check all the functions of the monitor and make sure that the monitor is in good
status.
2If rechargeable batteries are provided, charge them after using the device every time,
to ensure the electric power is enough.
3The interval between double pressing of POWER switch should be longer than 1
minute.
4After continuous 360-hour runtime, please restart the monitor to ensure the monitor’s
steady performance and long lifespan.
2.5 Checking the Recorder
If your monitor is equipped with a recorder, open the recorder’s door to check if paper is properly
installed in the slot. If no paper exists, refer to Chapter Recording for details.
2.6 Setting the Date and Time
To set the date and time:
1.Select Menu > Maintenance > User Maintain > Date/Time Setup.
2.Adjust the date display format based on the user’s habit.
3.Set the correct time of year, month, day, hour, min and sec.
2.7 Handing Over the Monitor
If you are handing over the monitor to the end-users directly after configuration, make sure that it
is in the monitoring mode.
The users must be adequately trained to use the monitor before monitoring a patient. To achieve
this, they should have access to, and read, the following documentation delivered with the
monitor:
zUser Manual (this book) - for full operating instructions.
zQuick Reference Card - for quick reminders during use.
2.8 FCC Statement*
*The statement is not applicable to iM50 or iM80.
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to part 15 of FCC Rules. These limits are designed to provide reasonable protection
against harmful interference in a residential installation. This equipment generates and can radiate
radio frequency energy and, if not installed and used in accordance with the instructions, may
- 9 -
Patient Monitor User Manual Installation
cause harmful interference to radio communications. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by turning the equipment
off and on, the user is encouraged to try to correct the interference by one or more of the
following measures:
1.Reorient or relocate the receiving antenna.
2.Increase the separation between the equipment and receiver.
3.Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected.
4.Consult the dealer or an experienced radio/TV technician for help.
This device complies with Part 15 of FCC Rules.
Operation is subject to the following two conditions:
1.This device may not cause harmful interference, and
2.This device must accept any interference received, including interference that may cause
undesired operation.
NOTE:
The manufacturer is not responsible for any radio or TV interference caused by
unauthorized modifications to this equipment. Such modifications could void the user’s
authority to operate this equipment.
2.9 FCC RF Radiation Exposure Statement*
*The statement is not applicable to iM50 or iM80.
This equipment complies with FCC RF radiation exposure limits set forth for an uncontrolled
environment. This equipment should be installed and operated with a minimum distance of 20
centimeters between the radiator and your body.
- 10 -
Patient Monitor User Manual Basic Operation
Chapter 3 Basic Operation
This manual is for clinical professionals using the iM50/iM60/iM70/iM80 patient monitors.
Unless otherwise specified, the information here is valid for all the above products.
This user manual describes all features and options. Your monitor may not have all of them; they
are not all available in all geographies. Your monitor is highly configurable. What you see on the
screen, how the menus appear and so forth, depend on the way it has been tailored for your
hospital and may not be exactly as shown here.
3.1 Overview
3.1.1 Front View
1
2
10
3 4 56
7
iM50
89
- 11 -
Patient Monitor User Manual Basic Operation
1
2
10
3 4 5 6 7 8 9
iM60
1
2
10
3 4 5 6 7 8 9
iM70
- 12 -
Patient Monitor User Manual Basic Operation
1
2
3
4
56789
iM80
10
1 Alarm indicator — when an alarm occurs, the alarm indicator will light
or flash. The color of light represents the alarm level.
2 Power supply switch — when the monitor is connected to the AC power
supply, press the key to turn the monitor on. When the monitor is turned
on, press the key to turn the monitor off.
3 Battery indicator, refer to Section Battery Indicator for details.
4 Mute — Press this button to pause the alarm. All the audio alarm will be
closed. At the same time, the message of Temporary Alarm Mute **s
and the symbol
will be displayed in the information area. When you
repress it or the pause time is over, the system will resume the normal
monitoring status and the message of Temporary Alarm Mute **s and
icon will vanish. Symbol
is shown in the information area. Pressing
or holding the button again can resume the alarm.
Further Alarm Mute information can be found in the chapter Alarm
Mute.
5 Start / Stop NIBP measurement — Press this button to inflate the cuff
and start blood pressure measurement. During the measurement, press the
button to stop the measurement.
6 Trend Key — Press this button to enter trend table review interface.
7 Freeze /Unfreeze — In normal mode, press this button to freeze all the
waveforms on the screen. In Freeze mode, press this button to restore the
waveform refreshing.
- 13 -
Patient Monitor User Manual Basic Operation
8 Start / Stop Recording — Press this button to start a real-time recording.
During the recording, press this button again to stop recording.
9 Menu — Press this button to return to the main interface when there is no
menu open.
10 Rotary Knob (hereinafter called knob) — The user can rotate the knob
clockwise or anticlockwise. This operation can make the highlighted item
shift up, down, left or right to choose the desired item. Remember, when
using the knob, rotate this button to highlight, and press it to select the
item.
3.1.2 Rear View
1
2
3
4
10
11
5 6 7 8 9
iM50
1 Equipotential grounding terminal, if the monitor or other processing
unit are used in internal examinations on the heart, ensure that the room
incorporates an equipotential grounding system to which the monitor
and other processing unit have separate connection.
2 Fan
3 Anti-theft lock interface
4 Security lock, used to prevent the power supply cord from falling.
5 Power Supply Inlet
6 USB interface, this port is used to connect the USB device.
- 14 -
Patient Monitor User Manual Basic Operation
7 VGA Interface
8 Network interface, this port is used to connect to the central monitoring
system through the standard network wire.
9 Defibrillator synchronization / analog output. When the user selects
Analog Output, the monitor outputs the waveform through the
auxiliary output port. When the user selects Defibrillation the monitor
outputs the defibrillator synchronization signal through the auxiliary
output port.
10 Speaker
11 SD Card
1
2
3
4
5
6 7
iM60/iM70
8
9
10
11
1 SD Card
2 USB interface, this port is used to connect the USB device.
3 Network interface, this port is used to connect to the central
monitoring system through the standard network wire.
4 VGA output
- 15 -
Loading...
+ 188 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.