EDAN INSTRUMENTS AX8EDAN User Manual

1
Preface
This manual contains necessary and sufficient information to use the Acclarix AX8 system safely for the purposes for which it is intended. It applies to all 1.3X releases.
Read and understand all instructions in this manual prior to using the Acclarix AX8 ultrasound system.
Contact Information:
For sales or service information please contact your local distributor or the EDAN service department at: support@edan.com.cn
P/N: 01.54.002481 MPN: 01.54.002481014 Release Date: Dec., 2016 © Copyright EDAN INSTRUMENTS, INC. 2015-2016
I
Contents
1 Introduction ................................................................................................................................ 1
1.1 Intended Use/ Indications for Use .......................................................................................... 1
1.2 Contra-indications .................................................................................................................. 1
1.3 Device Description ................................................................................................................. 1
1.4 Labeling Symbols .................................................................................................................. 2
2 Safety ......................................................................................................................................... 5
2.1 Warnings ................................................................................................................................. 5
2.2 Cautions ................................................................................................................................. 7
3 Getting Started ......................................................................................................................... 10
3.1 System Configuration .......................................................................................................... 10
3.2 System Overview ................................................................................................................. 11
3.3 Battery Use ........................................................................................................................... 13
3.4 Connecting and Disconnecting a Transducer ....................................................................... 15
3.5 Powering on/ off ................................................................................................................... 17
3.6 Screen Layout....................................................................................................................... 18
3.7 Control Panel ........................................................................................................................ 21
3.8 Touch Screen ........................................................................................................................ 24
3.9 Trackpad ............................................................................................................................... 25
4 Exam Operation ....................................................................................................................... 27
4.1 How to Start an Exam .......................................................................................................... 27
4.2 How to End an Exam ........................................................................................................... 27
4.3 How to Restart an Exam ...................................................................................................... 27
4.4 The Patient Information Page ............................................................................................... 28
4.5 Modality Worklist ................................................................................................................ 30
5 Imaging .................................................................................................................................... 32
5.1 B-mode ................................................................................................................................. 32
5.1.1. B-mode Touch Screen Controls ................................................................................. 32
5.1.2. Dual Imaging ............................................................................................................. 35
5.1.3. Quad Imaging ............................................................................................................ 35
5.1.4. Panorama ................................................................................................................... 35
5.1.5. Zoom .......................................................................................................................... 36
5.2 PW-mode .............................................................................................................................. 37
5.2.1. PW Touch Screen Controls ........................................................................................ 37
5.2.2. HPRF ......................................................................................................................... 39
5.2.3. Update ........................................................................................................................ 39
5.3 CW-mode ............................................................................................................................. 39
5.3.1. CW Touch Screen Controls........................................................................................ 39
5.4 Color-mode .......................................................................................................................... 41
5.4.1. Color Touch Screen Controls ..................................................................................... 41
5.4.2. Color Mode Variants .................................................................................................. 43
5.5 M-mode ................................................................................................................................ 43
5.5.1. M-mode Touch Screen Controls ................................................................................ 43
5.6 3D/4D Mode* ...................................................................................................................... 44
II
5.6.1. Pre-3D ........................................................................................................................ 45
5.6.2. 3D Volume Sweeping ................................................................................................ 46
5.6.3. 3D Image Review ...................................................................................................... 46
5.6.4. 4D Volume Acquisition .............................................................................................. 51
5.6.5. 4D Live Volume ......................................................................................................... 52
5.6.6. 4D Cine ...................................................................................................................... 52
6 Transducers .............................................................................................................................. 53
6.1 Transducer Model ................................................................................................................ 53
6.2 Name and Function of Each Part of the Transducer ............................................................ 54
6.3 Transducer Cleaning and Disinfecting ................................................................................. 54
7 Features .................................................................................................................................... 58
7.1 Comments ............................................................................................................................ 58
7.2 Body Mark ........................................................................................................................... 59
7.3 Measurements ...................................................................................................................... 60
7.3.1. Generic Measurements .............................................................................................. 63
7.3.2. Application Measurements and Calculations ............................................................ 69
7.4 Patient Information Management ......................................................................................... 73
7.4.1. Storing Images ........................................................................................................... 73
7.4.2. Reviewing Images ...................................................................................................... 73
7.4.3. Patient Database ......................................................................................................... 75
7.4.4. Archiving Studies....................................................................................................... 77
7.5 Needle Biopsy Guide ........................................................................................................... 78
7.5.1. Installing Needle Guide Bracket ................................................................................ 78
7.5.2. Activating Needle Guide Function ............................................................................ 82
7.5.3. To Adjust the Needle Guide Line .............................................................................. 83
7.6 Needle Visualization ............................................................................................................ 83
8 Presets ...................................................................................................................................... 85
8.1 Preset Organization .............................................................................................................. 85
8.2 Selecting a Preset ................................................................................................................. 86
8.3 Storing a Preset .................................................................................................................... 86
8.3.1. Presets ........................................................................................................................ 87
8.3.2. Comments Presets ...................................................................................................... 90
8.3.3. Bodymarker Presets ................................................................................................... 91
8.3.4. Measure Presets ......................................................................................................... 93
9 Utilities .................................................................................................................................... 97
9.1 Set-up ................................................................................................................................... 97
9.1.1. General Set-up ........................................................................................................... 97
9.1.2. Patient Set-up ............................................................................................................. 99
9.1.3. Store/Print Set-up ..................................................................................................... 100
9.1.4. Miscellaneous Set-up ............................................................................................... 101
9.1.5. User Set-up .............................................................................................................. 102
9.2 Connectivity ....................................................................................................................... 103
9.2.1. TCP/IP ..................................................................................................................... 103
9.2.2. DICOM .................................................................................................................... 104
9.2.3. Network Store .......................................................................................................... 107
9.3 Maintenance ....................................................................................................................... 108
9.3.1. License ..................................................................................................................... 108
9.3.2. Version ..................................................................................................................... 108
III
9.3.3. Demo ........................................................................................................................ 109
9.3.4. Export/Import ........................................................................................................... 109
9.4 Screen Adjust ..................................................................................................................... 111
10 In Between Exams ................................................................................................................. 113
10.1 Unpacking ........................................................................................................................ 113
10.2 Transport .......................................................................................................................... 113
10.3 Storage .............................................................................................................................. 113
11 Order List ............................................................................................................................... 114
12 Troubleshooting and Maintenance ......................................................................................... 116
12.1 Daily Checklist ................................................................................................................. 116
12.2 Troubleshooting ............................................................................................................... 116
12.3 Maintenance ..................................................................................................................... 117
12.4 Cleaning the System ......................................................................................................... 117
12.4.1. Cleaning the System Surface ................................................................................... 118
12.4.2. Needle Guide Bracket Cleaning and Sterilization ................................................... 118
13 Specifications ......................................................................................................................... 119
13.1 Electrical Safety Classifications ....................................................................................... 119
13.2 Power Supply ................................................................................................................... 119
13.3 Machine Specifications .................................................................................................... 120
13.4 Display Specifications ...................................................................................................... 120
13.5 General Technical Specifications ..................................................................................... 120
13.6 Operating, Storage and Transportation Environment ....................................................... 121
13.6.1. Operating Environment ............................................................................................ 121
13.6.2. Storage and Transportation Environment ................................................................. 121
14 EMC Information ................................................................................................................... 122
IV
Acclarix AX8 Diagnostic Ultrasound System User Manual Introduction
1 Introduction
1.1 Intended Use/ Indications for Use
For the United States of America The Edan Acclarix AX8 Ultrasound system is intended for use by a qualified physician or allied
health professional for ultrasound evaluations. Specific clinical applications include:
Abdominal Gynecology (including endovaginal) Obstetric Cardiac Small parts (Breast, Testes, Thyroid, etc.) Urology Musculoskeletal Peripheral vascular Intra-operative
For the EU and other non-USA regions The Edan Acclarix AX8 Ultrasound system is intended for use by a qualified physician or allied
health professional for ultrasound evaluations. Specific clinical applications include:
Abdominal Gynecology (including endovaginal) Obstetric Cardiac Small parts (Breast, Testes, Thyroid, etc.) Urology Musculoskeletal Peripheral vascular Intra-operative Pediatric Neonatal (including abdominal and cephalic)
Adult Cephalic*
1.2 Contra-indications
The Edan Acclarix AX8 Ultrasound system is not intended for ophthalmic use.
1.3 Device Description
The Edan Acclarix AX8 Ultrasound system consists of a main system along with associated transducers. The system circuitry generates an electronic voltage pulse, which is transmitted to the transducer. In the transducer, a piezo electric array converts the electronic pulse into an ultrasonic pressure wave. When coupled to the body, the pressure wave transmits through body tissues. The waves are then reflected within the body and detected by the transducer, which
- 1 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Introduction
No.
Symbol
Definition
1 Serial Number
2
P/N
Part Number
3 Date of Manufacture
4 Manufacturer
5 Operating instructions
6
Warning (Background: Yellow; Symbol & outline: Black )
7
Refer to User Manual (Background: Blue; Symbol: White)
8
Caution
9 Biological Risks
10
The symbol indicates that the device complies with the European Council Directive 93/42/EEC concerning medical devices.
then converts back to an electrical signal. The Acclarix AX8 system then analyzes the returned signal to generate an image or conduct Doppler processing. The Acclarix AX8 system gives the operator the ability to measure anatomical structures, and offers analysis packages that provide information used by competent health care professionals to make a diagnosis. The system provides both touch screen and hard buttons for the User Interface. *Descriptions of transducer C5-2MQ, adult cephalic application, 3D/4D function, PW one-key
optimization, acoustic power control, panorama measurement and wireless function are not
approved for use until finishing the related registration.
1.4 Labeling Symbols
The following labels are used on the Acclarix AX8 system:
- 2 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Introduction
11
Authorized Representative in the European Community
12
Disposal method. Indicates that the equipment should be sent to special agencies according to local regulations for separate collection after its useful life.
13
General Symbol for Recovery / Recyclable
14
Caution: Federal (U.S.) law restricts this device to sale by or on the order of a physician.
15
Conforms to AAMI Std. ES 60601-1, IEC Std. 60601-2-37
Certified to CSA Std. No. 60601-1, No 60601-2-37
16
IPX7
No harm for short time immersion
17
Type BF Applied Part
18
Transducer connector
19
Pencil Transducer connector (reserved)
20
ECG signal input connector (reserved)
21
Transducer lock
22
Transducer unlock
23
Network port
24
Trademark
25
Direct current
26
Video Output port
- 3 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Introduction
27
USB 2.0 port
28
USB 3.0 port
29
Display port
30
Microphone input
31
Headphones
32
FCC ID:
SMQAX8EDAN
Federal Communication Commission: FCC ID: SMQAX8EDAN
NOTE:
The user manual is printed in black and white.
- 4 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Safety
2 Safety
Throughout this document the following terms are used:
Warning: Advises against certain actions or situations that could result in personal injury
or death.
Caution: Advises against actions or situations that could damage equipment, produce
inaccurate data, or invalidate a procedure.
Note: Provides useful information regarding a function or a procedure.
Please read all warnings and cautions prior to using the Acclarix AX8 ultrasound system. For your convenience all warnings and cautions are provided in this section. These may be duplicated elsewhere in this document in the context of the instructions for use.
2.1 Warnings
Only use an Edan supplied power adapter and power cord. Only use an Edan supplied battery. Read and understand the battery installation
instructions prior to changing the battery.
Only use Edan supplied transducer. Use of other transducers may result in electric
shock or system malfunction.
Only use a hospital grade, grounded, power outlet and plug. Do not use with an
ungrounded outlet.
The system is ordinary equipment (Sealed equipment without liquid proof). But the
console panel including touch screen, track pad and hard keys is IPX1. Transducer (not including the transducer connector) is IPX7. Footswitch is IP68. Do not immerse or expose any of the parts to extended moisture. Splash resistance does not extend to transducer connectors. Keep connectors dry.
Do not use in a wet environment or when the relative humidity exceeds 95%. Do not reverse the positive and negative poles when installing a battery. Do not use the battery near heat sources or when the ambient temperature is over
40oC. Do not heat or dispose of in fire.
Do not destroy the battery; do not pierce or cause a strong impact to the battery. Do not touch the connector pins on the transducer port. Parts and accessories used must meet the requirements of the applicable
IEC/EN60601 series safety standards, and/or the system configuration must meet the requirements of the IEC/EN60601-1.
Use protective barriers (gloves and transducer sheaths) whenever possible.
Follow sterile procedures when appropriate. Thoroughly clean Transducers and reusable accessories after each patient examination and disinfect or sterilize as needed. Refer to transducer use and care instructions. Follow all infection control policies established by your office, department or institution as they apply to personnel and equipment.
- 5 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Safety
Not intended for Ophthalmic use. If a sterile transducer cover becomes compromised during an intra-operative
application involving a patient with transmissible spongiform encephalopathy, such as Creutzfeldt-Jakob disease, follow the guidelines of the U.S. Disease Control Center and this document from the World Health Organization: WHO/CDS/APH/2000/3, WHO Infection Control Guidelines for Transmissible Spongiform Encephalopathies. The transducers for your system cannot be decontaminated using a heat process.
Contact with natural rubber latex may lead to a severe anaphylactic reaction in
persons sensitive to the natural latex protein, Sensitive users and patients must avoid contact with these items. EDAN strongly recommends that health-care professionals identify their latex-sensitive patients, and refer to the March 29, 1991 Medical Alert on Latex products. Be prepared to treat allergic reactions immediately.
Improper operation may cause the internal lithium battery (hereinafter called
battery) to be hot, ignited or exploded, and it may lead to the decrease of the battery capacity. It is necessary to read the user manual carefully and pay more attention to warning messages.
Do not touch accessible contacts of electrical equipment and the patient
simultaneously.
This device is not suitable for intra-cardiac use or direct cardiac contact. The system shall not be serviced or maintained while in use with a patient. Install the system according the EMC guidance provided in Chapter 14 Do not stack the system on other electronic equipment. The use of transducer and connecting cable not supplied by the manufacturer may
result in increased emissions or decreased immunity of the equipment.
Refer to Chapter 14 for recommended separation distances from other equipment,
including portable and RF communication devices.
The power adapter is used to isolate the system from main power. Position the
system so that it is easy to disconnect the device.
No modification of this equipment is allowed. The system should be maintained regularly, at least annually, by a qualified
technician who has adequate training, knowledge and experience. That person should be familiar with the AX8 Service Manual (P/N: 01.54.002483), available from your Edan representative.
Keep non-medical equipment (such as the external printer) out of the vicinity of the
patient. (1.5m/6ft.)
Use of an extension cord or multi-socket outlet setup to provide power to the
ultrasound system or to the system‟s peripheral devices, may compromise the
- 6 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Safety
system grounding and cause the system to exceed leakage current limits.
SHOCK HAZARD – Do not connect non-isolated electrical equipment to the same
circuit being used to power the system.
Edan recommends the use of isolated connectors on any electrical equipment
attached to the system, and/or using isolation transformers that complies with IEC60601-1 to power that electrical equipment.
Always use sterile technique during a biopsy procedure. Sterilize the needle
guide assembly between uses.
Use a sterile needle with each use. Always confirm the clinical accuracy of the IMT result before entering it into the
report.
Transducer Warnings
To avoid infection, always use protective gloves when cleaning or disinfecting Read and follow all manufacturer instructions for disinfection agents. To avoid infection, ensure that expiration date of the disinfecting solution has not
passed.
Disinfect the transducer after each intra-cavity or intra-operative procedure. Use a
new sterile sheath for each such procedure.
Use a pyrogen-free transducer sheath for neurological intra-operative procedures. Unplug the transducer from the system prior to cleaning or disinfecting. Do not immerse the transducer beyond the point indicated in Figure 6-2. Do not allow the transducer connector to get wet.
2.2 Cautions
The system contains no user serviceable components other than the battery. Do
not remove any covers.
Excessive dust and dirt could clog internal airflow and cause overheating. Do not
use in a dusty environment.
Do not use a battery that leaks, emits an odor, appears deformed, or discolored.
Immediately replace it with a new Edan-supplied battery and dispose of the old battery according to local regulations. Replace a battery that has reached the end of its service life.
Use care when storing or disposing of batteries. Do not allow the leakage to come
in contact with each other. Do not dispose of them together with household garbage. At the end of their life hand the batteries over to the applicable collection points for the recycling of waste batteries.
Inspect the system regularly, at least weekly. Before use ensure there is no visible
- 7 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Safety
evidence of damage to the equipment, cables, and transducers. If a component is damaged replace it before use.
Do not use in locations subject to vibration. Read and understand the section A1.2 Ultrasound Safety and the ALARA Principle
of Acclarix AX8 Advanced User Manual before using the system. Do not expose a patient to ultrasound energy longer than clinically reasonable.
Practice ALARA principle when operating ultrasound system. Minimize the
acoustic power without compromising the image quality.
Do not use in the presence of a flammable anesthetic. The system generates radio frequency energy, which may cause interference with
other devices in the vicinity. If interference is suspected, try re-orienting or relocating the equipment.
The use of electrosurgical units or other devices that generate radio frequency
interference may cause image distortion or other malfunction.
During long term storage the battery should be charged at least once every 3
months to ensure battery capacity.
The system should only be used by a qualified physician or allied health
professional for ultrasound evaluations.
Use only Edan supplied or recommended parts and accessories. Verify measurement results prior to entering them into a report. Contact your local distributor or Edan service if there is excessive noise from the
system speaker or fans.
Please read and understand cleaning instructions prior to use. Please read and understand maintenance instructions prior to use. Please read and understand instructions for system operation prior to use. Studies stored on the system hard drive should be archived regularly. The system
is not intended for long term storage of patient information. Confirm successful archiving before deleting a study from the hard drive.
Ensure that the system vents are clear and unobstructed. Confirm patient identification information prior to storing or printing any exam
information.
If you have any questions about maintenance, technical specifications, or system
functionality, please contact your local distributor or Edan service at: support@edan.com.cn
Ultrasound images occasionally have artifacts, and should only be used as one
part of an overall clinical assessment.
To avoid electrical shock, turn off and disconnect the device from the AC power
- 8 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Safety
source before cleaning and disinfecting.
The device and accessories are to be disposed of according to local regulations
after their useful lives. Alternatively, they can be returned to the dealer or the manufacturer for recycling or proper disposal.
Transducer Cautions
Do not use disinfection agents beyond their expiration date. Do not use sterile sheaths beyond their expiration date. Inspect the transducer connector, cable, and head periodically. Do not use if there
is evidence of excessive wear or damage.
Do not operate the transducer to temperatures in excess of 40°C or storage the
transducer to temperatures in excess of 55°C .
Do not kink or pull on the transducer cable. Broken or bent connector pins can cause image artifacts. Do not use a transducer
with broken or bent pins.
- 9 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
Printer Model
Printer Type
Interface
SONY UP-25MD
Color Image
S-Video
SONY UP-D25MD
Color Image
USB
SONY UP-X898MD
B/W Image
USB
HP Officejet Pro 251dw
Color Report
USB
HP Laserjet Pro 200 M251n
Color Report
USB
HP Laserjet CP1525n Color
Color Report
USB
HP Deskjet Ink Advantage 2010
Color Report
USB
HP Deskjet 1010 Color
Color Report
USB
HP Deskjet 1510 Color
Color Report
USB
HP LaserJet 400 M401d
B/W Report
USB
3 Getting Started
3.1 System Configuration
The Acclarix AX8 system is shipped with the following components:
Acclarix AX8 main unit 1 rechargeable lithium battery 1 UPS battery 1 AC adapter 1 bottle of coupling gel 1 basic user manual and 1 advanced user manual
The following options are also available (please see the Order List Appendix for details)
Transducers: C5-2XQ, C5-2Q, L10-4Q, L12-5Q, L17-7HQ, E8-4Q, MC8-4Q, P5-1XQ,
L17-7SQ, C5-2MQ*
1 Trolley MT-807 Footswitch Suitcase The Acclarix AX8 supports the following external DVD drives: SAMSUNG SE-218GN,
SAMSUNG SE-208GN, LENOVO DB75, ASUS SDRW-08D2S-U and PIONEER DVR-XU01C.
The Acclarix AX8 supports the following DVD disks: Philips CD-R 700MB 80min 52x, Sony CD-RW 700MB 1x-4x,Sony DVD-RW 4.7GB 120min 2x/1x, Verbatim DVD+RW 4.7GB 120min 4x.
Note: Only support DVD disk single burning.
The Acclarix AX8 supports the following printers. Please contact the manufacturer for
details.
Table 3-1 Printer List
- 10 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
Model
Name
Description
BGK-C5-2
BGK-C5-2 Needle Guide Bracket Kit
For use with the C5-2XQ/C5-2Q, Supports: 16G, 18G, 20G, 22G
BGK-L40UB
L40UB Needle Guide Bracket Kit
For use with the L10-4Q, Supports: 16G, 18G, 20G, 22G
BGK-CR10UA
CR10UA Needle Guide Bracket Kit
For use with the E8-4Q, Supports: 16G
BGK-R15UB
R15UB Needle Guide Bracket Kit
For use with the MC8-4Q, Supports: 16G, 18G, 20G, 22G
BGK-P5-1X
BGK-P5-1X Needle Guide Bracket Kit
For use with the P5-1XQ Supports: 14G-22G
1
2
3
4
5
6 7
8 9
Note: For more printer models, please see: http://hplipopensource.com/hplip-web/supported_devices/combined.html . Generally, printer models with driver version 3.14.3 or lower can be supported by Acclarix AX8, but some models cannot be perfectly supported.
Needle Guide Bracket Kit
Table 3-2 Needle Guide Bracket Kits
3.2 System Overview
- 11 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
No.
Name
Description
1
Monitor
Display the images and parameters during scanning.
2
Touch Screen
Use to control operation and activate functions
3
Control panel
Use to control the operation
4
Handle
Used for carrying the system
5
Hinge
Used to rotate the monitor
6
Transducer locking lever
Use to lock or unlock the transducer
7
Transducer Port
Used for connecting a transducer to the Acclarix AX8 main unit
8
TrackPad
Use to change the cursor position
9
Battery Indicator
LEDs indicate battery charge. Each LED corresponds to 20% of battery power. The LEDs will flicker while the battery is charging. Touch the battery indicator to view the battery power.
10
I/O Ports
Use to connect the I/O extend modules
10
Figure 3-1 Appearance of the System
Table 3-3 Appearance of the System
- 12 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
1. Power supply
5. USB 2.0 port
2. S-Video output port
6. Network port
3. Display port
7. Microphone input port
4. USB 3.0 port
8. Audio output port
1 3 4
5 6 7 2 8
I/O Ports on the left panel:
Figure 3-2 I/O ports of the system
Table 3-4 I/O Ports of the System
NOTE:If an external monitor is connected to the system via display port, configure the resolution to 1920x1080, other resolution may cause abnormal ultrasound machine display.
3.3 Battery Use
The Acclarix AX8 system comes with one lithium-ion battery. When it is fully charged it can run the system for approximately 60 minutes, depending on use. It is automatically charged when the system is plugged in.
CAUTION
1. If the system will remain unused for more than one week, charge the battery to at
least 75%capacity, take the battery out and store the system and battery separately.
2. During long term storage the battery should be charged at least once every 3 months
to ensure battery capacity is more than 75%.
- 13 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
To install the battery:
1. Turn off and unplug the Acclarix AX8 system.
2. Close the monitor, turn the system upside down and rest it on a flat stable surface.
3. Put the battery gently to battery compartment, and the button will lock the battery
automatically.
Figure 3-3 Installation of the Battery
To remove a battery:
1. Turn off and unplug the Acclarix AX8 system.
2. Close the monitor, turn the system upside down and rest it on a flat stable surface.
3. Slide the battery button to unlock the battery.
4. Take the battery out.
Figure 3-4 Battery Removal
To install the UPS battery:
1. Close the monitor, turn the system upside down and rest it on a flat surface.
- 14 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
2. Take out the UPS battery from the package.
3. Put the UPS battery gently to the UPS battery compartment.
Figure 3-5 Installation of the UPS Battery
3.4 Connecting and Disconnecting a Transducer
To connect a transducer:
Figure 3-6 Tansducer Locking Handle
1. Align the connector with the transducer port and carefully push into place.
2. Toggle the locking handle to the bottom position.
3. Do not allow the transducer head to hang free. Impact to the transducer head could result in
irreparable damage.
Figure 3-7 Lock the tansducer locking handle
- 15 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
To disconnect a transducer:
1. Toggle the locking handle to the top position to unlock the connection of the Transducer.
2. Firmly grasp the transducer connector and carefully remove it from the system port.
3. Store transducer in its protective carrying case prior to transport.
Figure 3-8 Unlock the Tansducer Locking Handle
CAUTION
1. Do not touch the pins of the transducer connector.
2. Broken or bent connector pins can cause image artifacts. Do not use a transducer
with broken or bent pins.
Multi-Transducer Connector (MTC)
The Multi-Transducer Connector (MTC) is an optional addition to the AX8 that allows up to three transducers to be connected to the system. The MTC can be mounted to the optional AX8 cart, or lie flat on an adjacent table.
Figure 3-9 Connect to the MTC
Note: Pluging and unpluging the MTC should be done while the system is frozen or turned
off.
- 16 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
3.5 Powering on/ off
Please review and follow the steps described in the Section 12.1.Daily Checklist prior to powering on the device.
To power on:
1. Connect the system to a hospital grade power supply, or use the battery as the power supply.
2. Press the power on/off key on the top left of control panel. To Login
If Password Protection is enabled (see section 9.1.5), the system displays a Login dialog when booting up the system, you can type in or select user name in the drop-down list of User name, then enter password and click Login.
For emergency use, you can click on Emergency to log on directly without entering user name and password.
Note:
When both Admin1 and Admin2 password are forgotten, please contact the serviceman for the system password reset.
To power off:
1. Press the power on/off key on the top left of control panel and the system displays a
confirmation dialog box.
2. Select “Shut Down” from the confirmation dialog box. If the system is unresponsive, a long press of the power on/off key will shut down the system
directly.
NOTE:
Please unplug the AC adapter from the power socket and disconnect the battery prior to storage.
Sleep mode The system will enter a sleep mode that maintains exam information while using minimal battery.
There are three events that can invoke sleep mode:
- 17 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
Close the cover of the system without powering-off the system. No user input for a configurable amount of time. Please see System Set-up to configure
this time.
Pressing the Sleep button on the confirmation dialog box when powering off.
Figure 3-10 Confirmation Dialog Box When Power off
3.6 Screen Layout
Figure 3-11 Main Screen Display
Information Field
The top line of this field contains your hospital/institution name. Please see Section 9.1. General Set-up for information on customizing this area.
The second line of this field contains the patient name, gender, age and ID, as entered through the Patient Information screen.
This field also contains data fields for:
The currently active transducer
- 18 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
The currently active preset MI, TI, current date and time.
Image Field:
The ultrasound image appears in the Image Field, under the Information field. The image field also contains information typically associated with the image, such as scales, TGC, and maps. The system supports two different image sizes within the image field. See 9.1.1 for details
Mini Report:
The left side of the screen displays a “mini-report” which provides a synopsis of applications measurements performed in the current exam.
Thumbnail Field:
The right side of the screen displays thumbnail images of all statics and clips captured in the current exam.
Status Bar:
The bottom of the screen is used to display icons that provide system status. These include:
Image Store icon
: Number of static images and clips in the current exam.
Battery icon
: Battery fully charged, symbol in green.
: Battery higher than 80%.
: Battery 60%-80%.
: Battery 40%-60%.
: Battery low, symbol in red.
: Battery removed, outline in red.
: Battery charging.
Network status icon
The network icons show the connection status of the DICOM or FTP Server. If no
Network Server is defined then the icon does not show at all.
- 19 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
: Outline in grey color: Successful connection with a Server. Outline in green color: Data exchange with a Server. Outline in red color: Failure to connect with a configured Server. Hard drive icon.
: Hard drive available
: Hard drive data exchange, symbol in green.
: Hard drive 95% full, symbol in red.
: Hard drive 95% full with data exchange, symbol in red.
USB icon
:
Symbol in grey color: USB available
DVD icon
: Symbol in grey: DVD device is connected
: Symbol in green: DVD device is connected, disc is inserted and data
transmission is available. Printer icon
: Printer available.
Wifi icon *
Wi-Fi available, Subject to regulatory approval. See Connectivity (Section
9.2.1) for enabling wireless if it is licensed. Clicking on this icon shows a list of all Wi-Fi networks that are currently available.
WARNING
The system complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
1) this device may not cause harmful interference, and
2) this device must accept any interference received, including interference that may
- 20 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
No.
Key
Name
Description
1.
Patient
Invokes the Patient Information Screen. Typically used to start/end exams or to modify patient information during an exam. See section 4.1
cause undesired operation.
NOTE:
1. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
2. Any changes or modifications to this unit not expressly approved by the party
responsible for compliance could void the user's authority to operate the equipment.
3.7 Control Panel
Figure 3-12 Control Panel of the System
- 21 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
2.
Preset
Allows you to change the Preset being used. See section 8 for details.
3.
End Exam
Completes an exam without invoking the Patient Information Screen for the next exam. See section 4.2
4.
Utilities
Press this button to get the Set Up, Maintenance, Connectivity and Screen Adjust keys on the touch screen.
5.
Alt Store
One of three keys that can be configured to store images or clips, or to set video printers. See section 7.4 for details.
6.
Report
Press to display the report page
7.
Review
Press to enter exam database or image review mode. See section 7.4 for details.
8.
Zoom
Press to zoom the image in or out. See section 5.1.5 for details.
9.
Depth
Press to adjust the depth of the image displayed.
10.
Body Mark
Enters or exits the Body Mark function. See section 7.2 for details.
11.
Clear
Press to clear all the measurements, calculations, comments, and body marks displayed in the current image.
12.
Comment
Enters or exits the Comment function. See section 7.1 for details.
- 22 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
13.
Update
In Strip Doppler this switches between acquiring the Doppler strip and the reference image. See section 5.2.3 for details.
14.
Cursor
Press to hide or display the cursor
15.
Measure
Invokes the Measure function for Generic and Application Measurements. See section 7.3 for details.
16.
Freeze/Cine
Press this key to switch between the frozen and real-time states.
17.
Store 1
One of three keys that can be configured to store or print images or clips, or to set video printers. See section 7.4 for details.
18.
Store 2
One of three keys that can be configured to store or print images or clips, or to set video printers. See section 7.4 for details.
19.
Gain
Controls the overall gain for the imaging mode that currently has the UI focus. In normal imaging the gain can also be adjusted by sliding your finger along the top of the trackpad area
20.
F1
User-defined button
21.
F2
User-defined button
22.
Color
Press to enter or exit Color Mode. See section 5.4 for details
23.
M
Press to enter or exit M Mode. See section 5.5 for details. Use the trackpad to adjust the M sample line.
24.
D
Press to get the sample line, use the trackpad to adjust the position and press the hardkey again to confirm.
For other information, see section 5.2 and 5.3 for details.
- 23 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
25.
B
Press to return to B-mode imaging from any display or imaging mode. See section 5.1 for details.
Table 3-4 Buttons on Control Panel
3.8 Touch Screen
The Touch Screen contains controls that vary depending on the active function. There are several types of controls used by the touch screen, as illustrated below:
Figure 3-13 Touch screen of the System
A. Tabs: Each function that is active has a tab on the touch screen. Usually the function that
was most recently activated is the top tab. Pressing on any other tab will bring it to the top and provide access to that function‟s controls.
B. Slider: Pressing on the top or bottom of a slider changes the control setting by one value.
Pressing anywhere on the control and dragging will continuously change the value.
C. Push button: This can either be an on/off control (like “Colorize”) or a one-shot control
that immediately performs an action (like “Auto”)
D. Radio Buttons: A collection of buttons where only one is active at any time. Activating
one will de-activate all others.
E. Folder: Controls can be grouped together into a folder. Pressing on the folder will open it
to access any of the controls in it.
F. TGC: The B-mode tab has a specialized control for TGC. Each slider can be dragged
horizontally and individually. Dragging vertically down across the sliders will set all sliders.
G. Pages: When a tab has multiple pages of controls each page is represented by a dot at the
top of the page. The current page is indicated by a filled-in dot. You can move between pages by dragging your finger horizontally across the dots. These dots do not appear when there is only one page in the current tab.
- 24 -
Acclarix AX8 Diagnostic Ultrasound System User Manual Getting Started
Customizing the touch screen
The touch screen can be customized to meet your needs. Press and hold any control for about a second to put the touch screen in customization mode. Continue pressing and drag the control to a new location.
Creating Folders: Dragging one control on top of another control will create a folder that
contains both controls. Dragging controls out of a folder until only one exists will automatically delete that folder. Folders cannot contain other folders.
Multiple pages: Dragging a control to the side of the screen will move that control to the
next page.
Radio Button Cluster: There is no restriction on moving a single radio button. However,
we suggest that they are grouped adjacent to each other. When they are grouped in this way the system will automatically draw a border around them to indicate they are a related group of radio buttons.
TGC: The TGC is a special control that cannot be moved. It is always on the right side of
the B-mode tab.
For most tabs there is one arrangement of controls. Whatever way they are customized will be remembered across exams and across power-ups. However, the following tabs can have multiple different arrangements:
Comments Bodymarkers Measurements
These functions are often configured differently for different exams. For example, the measurements for an OB exam are different from those of an abdomen exam. Each configuration of these tabs is stored in an application preset (see Chapter 8 Presets, for details)
3.9 Trackpad
In general the Acclarix AX8 trackpad is used like a trackball or similar pointing device used on most ultrasound systems. Using a trackpad instead of a trackball helps make the console sealed, and makes the system easier to clean. In addition, there are several capabilities that the trackpad supports that would not be possible on a traditional trackball:
The trackpad supports multi-finger gestures. For example:
o While in B-mode a pinch-out gesture can be used to zoom the image. o While in Color a pinch gesture can re-size the ROI.
Dragging along the Swipe button on top edge of the trackpad as a primary sliding control.
For example:
o In most live imaging modes this is Gain.
The top right of the trackpad is reserved as a „Set‟ button. Pressing this corner is generally
equivalent to a single click on a mouse or trackball. For example:
o In CD it is used to toggle between controlling the size and the position of the pan
box.
o In Measure it is used to toggle between controlling one side or the other side of a
caliper pair.
The top left of the trackpad is reserved as an „Enter‟ button. Pressing this corner is
- 25 -
Loading...
+ 101 hidden pages