2. Never store the unit with a discharged battery!
The Ecom-B is not suitable for continuous emissions testing. Please allow at least 10 minutes of
fresh air purging for each hour of testing.
The following substances impair the instrument´s operation:
- cleaning agents
- degreasers
- wax polishes
- adhesives
- anything containing formaldehyde
Fresh Air Opening
-allow 1 min for auto-zero in fresh air
-allow at least 2 mins. for stable readings before
should be made only by specialists
who are familiar with these installations.
once per month.)
Ecom-D Page 3
fresh air pump!
Page 4
1. Instrument Design
Air temperature
Connection
Slot for multi-media
card
Connections for
the pressure sensor
Gas temperature
Connection
USB connection
Condensation trap
(with Peltier cooler)
with fine dust filter
(Option)
Sample line
Connection
Draft
Connection
Pressure
Connection
AUX connection
Printer
(Option)
Graphic
display
Charger connection
Charging LED
Antenna for Bluetooth
or WiFi
(Option)
Arrow Keys
Page 4 ECOM-D
Page 5
Keyboard
2 GB SD Card
Part No:3014584G
Thermal IR Portable Printer
Paper (5/pk)
Part No: 6951801H
Water Trap Filter (10/pk)
ESC key
Enter key
Arrow keys
Store Key
Print key
ON / OFF key
Info key
Function keys
During input mode, the keys are used for numerical values
(function shown on display)
(access control menu)
(quit/
escape menu)
(confirm
selection)
(Up/Down/Right/
Left/Scroll)
(access
printing menu)
2. Ecom-D Accessories
Ecom-D Page 5
Page 6
WARM
COLD
Fan
Peltier
Condensate evacuation
3. Gas Cooler (option)
element
How it works:
Exhaust gas with a temperature over the steam dew point (35 - 65 °C)
is drawn into a long gas path. Then, it is pushed through a metal coated surface with strong thermal conductivity. The gas radiates its heat
to this metal surface. A Peltier element (semi-conductor cooling element) operated by a continuous current is connected with the metal
surface and a second metal body with cooling ribs and ventilation
slots.
The flow through the Peltier element creates a heat transfer from
WARM to COLD. This drains the heat from the metal body and from
the gas then transfers it to the outer cooling body. This heat is conveyed thru vertical forced ventilation to the surrounding air.
The condensation created by the heat loss of the gas drops in a receptacle and is pumped out when needed by a hose pump.
The Peltier cooler can be switched off in the following manner:
In the measurement menu, press <Enter>, select <Peltier I/0> and
press <Enter> again (To turn the Peltier cooler switch on, just repeat
this procedure).
Page 6 ECOM-D
Page 7
4. Power Supply
Used batteries can either be returned
to Ecom or taken to local recycling centers!
The Ecom-D comes with an external charger. The analyzer can also
be powered via the internal battery (6 V; 3.8 Ah). The battery should
be recharged as needed (Indicted by a warning sound and display
light). The battery charging status can be by checked looking at the
voltage indication on the display (menu "Control" - Info key). The battery warning is activated when <Accu> is lower than 5.9 V.
At 5.8 V, the battery does not have enough power to operate the analyzer. At this point, the instrument must be charged via the external
charger.
5. Data Memory
The multi-media card enables the storage of measurements. The
measurement results are written in a text file (J2KDV.txt). Data format
information can be found in the appendix. The files can be transferred
to a PC via the memory card.
In order to use the memory card function:
- The card must have at least 32 MB of memory, with a maximum of 2
GB
- The card must be formatted on 16 bit FAT
- It must be either an SD or MM card from rbr
- The PC must have a card reader from ECOM, Belkin, or SanDisk
Ecom-D Page 7
Page 8
Inserting the card
Do not remove the card while opening/saving
Always position the probe in the exhaust pipe
Always use a filtering system as per our rec-
Gas analysis
Soot..Oil trace
Data processing
Adjustments
Control
Insert the multi-media card as shown. Make sure
that the card is inserted completely (so that none of
the card is seen) and is secured in place.
This can cause damage to the card, card
reader, or data loss.
data.
6. Analyzer Start-Up
once the auto-zero phase is over!
ommendation for combustion plants
firing solid combustibles!
Once the instrument is switched on (key <I/0>), the
main menu is displayed.
6 sub-menus with the following functions (nonvisible sub-menus can be viewed with the arrow
keys) are displayed:
- Gas analysis : Perform gas analysis
- Soot...Oil trace : Input of soot measurement results
- Data processing : Save measurements / Data transfer
(Multi-media card required)
- Adjustments : Modify instrument settings
- Control : Check operation state of instrument
Page 8 ECOM-D
Page 9
To perform measurements, select "Gas analysis" with the arrow keys
Do you wish to
use the data
processing?
JA
Auswählen: <>!
NEIN
and confirm with <Enter>. The instrument starts then a 1- minute auto-zero phase and the fuel type selection list is displayed. The following fuel types are available:
Natural Gas
Propane
Butane
Number 2 Oil
Number 4 Oil
Number 6 Oil
Coal
Wood (0% H20)
Wood (17% H20)
Wood (50% H20)
North Sea Gas
Diesel Oil
Use the arrow keys to select the desired fuel type
and confirm with <Enter>. The instrument will then
ask if you wish to use the data bank. If you want to
assign the sampled data to a specific file, so press
<F1>. Press <F4> to perform the measurment
without selecting a save file.
Ecom-D Page 9
Page 10
7. Input/ Selection of measurement data
Selection upon:
Search word
Memory
Create new
Quit with: ESC
To view data already recorded or to create a new file, the following options are available:
Create new (is automatically selected the first
time an MM card is used): When creating a new
file, a number should be assigned to it.
number
Select <Create new> and confirm with <Enter>. Input a number designation for this file (max. 16 figures) using the keyboard:
- Use the arrow keys <Up/Down/Right/Left> to select
the character (selected character is outlined by a black background)
- Press <Enter> to confirm the character selection (press <F2> to delete the last character if needed)
- Repeat this procedure until the desired number is complete.
Example: "25.09.2009"
Tip: We suggest a date-related file number to easily find the data file
later on via the search function.
After confirming with <F1>, it is possible to enter a text via the key-
board (max. 6 lines with 20 characters each) which is printed out and
can be used for data processing purposes. Proceed as follows:
- Select text line 1 using the arrow keys <Up/Down> and
confirm with <Enter>
- Select the keyboard mode with <F3> (4 keyboard modes
are available)
- Select with the desired character with <Up/Down/Right/Left> (selected character is outlined by a black background)
- Press <Enter> to confirm the character (press <F2> to delete the
last character)
- Repeat this procedure until the desired text is complete.
Page 10 ECOM-D
Page 11
Proceed as follows to change a character:
Search word
Memory
Create new
Selection upon:
Quit with: ESC
Input number
3
Please use the
numerical keys!
- Press <F4> to stop the character selection
- Use the arrow keys <Right/Left> to select the
character that needs to be changed
- Confirm the character selection and correct character with <F4>
- Return to character selection and open the next line in the
process with <F1>
Close the input menu with <ESC>, and the next available data record
is activated. Press <ESC> to return to the gas analysis menu.
Memory number: Selection by file number is the
number
easiest way to view data already stored in the analyzer.
Select <Memory number> and confirm with <En-
ter>. Input the desired data file number:
Example: "3" for data file number 3
- Put in the desired file number and confirm with <Enter>. You can
use the arrow keys <Up/Down> to view a record of all saved files.
- Press <F1> to select the first file number and <F2> for the last.
- Press <F4> to delete the content of the selected file number.
- Press <Enter> to open the selected file number.
- Finally, press <ESC> to start the gas analysis.
Ecom-D Page 11
Page 12
Search word: It is also possible to search for a specific saved record,
Memory number 1
25.09.2009
Memory number 1
25.09.2009
12:15 25.09.09
O2 value in air
O2 value in air 12:15 25.09.09
Memory number 1
O2 19.5
CO
Draught
dT-measurement
dT-measurement 12:15 25.09.09
Memory number 1
T1 70.4
T2
dT
Soot..Oil trace
Soot..Oil trace 12:15 25.09.09
Memory number 1
Boiler temp. : 65°C
1.Soot meas. : 0.5
2.Soot meas. : 0.3
3.Soot meas. : 0.7
Oil trace : NEIN
Gas analysis 12:15 25.09.09
Memory number 1
O2 3.2 %
CO2 13.1 %
CO 0 ppm
Eff.
Losses
Exc. air
T.Gas 184 °C
T.Air
CO measurement 12:15 25.09.09
Memory number 1
O2 17.5
CO 0%
CO
Exc. air
if you know the file number.
Select "Search word" and confirm with <Enter>. Using the software
keyboard, type in at least 3 characters of the file number.
Example:"25.09" for file number 25.09.2009
Next, press <F1> to start the search. All files match-
ing this number will be shown. The results can be
scrolled using the arrow keys (F1 for the first file, F2
for the last). Once found, you can view the desired
data file with <Enter>. The last measurement on
this in this file can be viewed pressing <Enter> /<View memory> /<Enter>. All measured and calculated values can are viewed on 5 display pages,
using the arrow keys to scroll between pages.
20 °C
92.5 %
7.5 %
1.18
7.00
%
3 ppm
0.01 hPa
56.3 °C
14.1 °C
%
738 ppm
123 ppm
°C
Press <ESC> to leave measurement data and the recording of the
current data can begin.
Page 12 ECOM-D
Page 13
8. Flue Gas Analysis
Re-zero the instrument after each
measurement (at least once per hour of meas-
Hotkey
Key <F1>
Hotkey
Taste <F4>
Store and print
values
key <F2>
Switch-off CO sen-
sor
key <F3>
Symbol
cooler
ON/OFF
O2 3.2
CO2
T.Gas
T.Air
BImSchV
8.1. Gas analysis
for the most accurate results!
urement)
After the 1-minute auto-zero phase, the instrument switches over to
measurement mode. The gas measurement values are seen on 4 display pages. Use the arrow keys to scroll the pages.
20 °C
%
13.1 %
184 °C
Pressing <F1> allows you to switch from the values display to a previous sub-menu (see chapter Adjustments). Possible sub-menus are:
Soot...Oil trace, Data processing, View memory, Display values, Fuel
type, Efficiency (K), Internal, Adjustments. You can also switch back
from any menu to the values display with <F1>.
Press <F2> to print out the values as they are recording in the intermediate memory.
Press <F3> to initiate the CO bypass, so that the CO sensor is protected against gas concentrations that are too high.
Ecom-D Page 13
Page 14
The position of the measured and calculated values (gas analysis
Probe tip
Gas stream
Protection bow
Core stream search
O2 3.2 %
CO2 13.1 %
CO 0 ppm
Eff. 92.5 %
Losses 7.5 %
Exc. air
T.Gas 184 °C
T.Air 20 °C
BImSchV
sub-menu) on the display pages may be changed at any time. To
change these positions:
- Press <Enter> / <Display values> / <Enter> to begin
- Select the line with the arrow keys (Up/Down)
- Select desired parameter with the arrow keys (Right/Left)
- Repeat procedure until desired layout is completed
- Press <Enter> when finished
1.18
Connect the sample line to the <Gas Connection> plug, and position
the sampling probe in the exhaust channel, so that the thermocouple
is fully surrounded with the gas (see drawing).
Perform the measurement in the core stream of the exhaust gas
channel with the probe placed in the hottest gas temperature area. As
you move the probe toward the center of the exhaust, the display will
show an up arrow, indicating an increase in the measured temperature. If the probe is pulled away from the exhaust, an arrow pointing
down is displayed, indicating a decrease in temperature. If no temperature change is shown for at least 3 seconds, the arrow display will
disappear, indicating the lack of change.
Page 14 ECOM-D
Page 15
CO2, efficiency, losses, excess air and dew point are calculated val-
Measurement stored in
intermediate memory
O2 3.2 %
CO2 13.1 %
CO 0 ppm
Eff. 92.5 %
Losses 7.5 %
Exc. air 1.18
T.Gas 184 °C
T.Air 20 °C
BImSchV
ues. They can only be calculated if values for the basic parameters
like O2 and the temperatures are available. Make sure that:
O2 < 20.5 % and T.Gas - T.Air > + 5 °C
The current barometric pressure must be typed in (in the <Adjust-ments> menu) for the dew point to be calculated accurately. If the gas
temperature falls below the dew point (between 25 and 65 °C), then
efficiency will be calculated with condensation gain. In this case, a (K)
is displayed after <Efficiency>.
Correct measurement values are displayed after a short delay. This is
necessary so the gas has time to reach the sensors and to build-up a
stable electrochemical reaction for the sensors. This time period lasts
approx. 1 to 1.5 minute. For recording, printout and evaluation wait
until the values do not change anymore. If deviations higher than 2
ppm still occur with the gas values, it may be due to unstable pressure
conditions in theexhaust channel.
If the measurement values are stable, press the key
<Record> (disk symbol) to transfer the values in the
intermediate memory (caution: store gas analysis,
CO, O2 check and draft measurements separately).
They will be kept there for a later printout and, if
need be, for a final data record storage.
If a printout of the values needs to be made as the measurement is
taking place, simply press <F2> (the completecontent of the intermediate memory will beprinted).
Ecom-D Page 15
Page 16
The CO sensor is protected against over-saturation thanks to an in-
Pressure
Draught -0.12 hPa
Recorded value:
ternal program. If the 4000 ppm limit is exceeded, a purging pump is
activated which provides the sensor with fresh air. After sufficient
purging time (X behinds CO disappears), the sensor can be reintegrated into the measurement system with <F3>. You can do this automatically by selecting <Yes> in <Adjustments> / <Internal> / <CO-Automatic>. The sensor can also be manually excluded from the
measurement system by pressing <F3>.
8.2. Draft Measurement
A trend indication for the draft conditions in the exhaust channel can
be viewed during the gas analysis. However, the draft will not be
stored together with the gas values when pressing <Record> because the differential pressure sensor tends to drift.
For an exact measurement, auto-zero this sensor just before sampling
and document the value.
The display shows the current value as well as the
recommendation to reset the zero point of the sensor. Disconnect the draft hose from the instrument
for a short period and press <F4>. The analyzer will
then reset the sensor’s zero point.
--.-- hPa
Page 16 ECOM-D
Page 17
Re-connect the draft tubing. The display shows the
Pressure
Draught -0.12 hPa
Recorded value:
Measurement stored in
intermediate memory
Measurement stored in
intermediate memory
Flow measurement
Gas speed 0.3
M.Flow
Connections for
exact measurement value which can be stored by
pressing <Record> and added to those results previous stored in the intermediate memory. The stored
value is shown on the display. Press <ESC> to exit
the draft measurement menu.
8.3. Flow measurement (option)
This measurement can be done with a pitot tube.
First, the pitot factor of the pitot tube must be entered by going to <Adjustments> / <Internal> /
<Pitot factor>). After connecting the pitot tube to
the analyzer, the zero point of the sensor can be set
with <F4>. Enter the cross section of the flow channel with <F1>. This is needed for the calculation of
the flow rate. After the pitot tube is positioned in the
-0.12 hPa
flow channel, the display shows the speed (m/s) and
the flow rate (Nm3/h). If the value is stable, press
<Record> to store the value in the intermediate
memory. If a printout of the values is needed during
the recording to the intermediate memory, press
<F2>. The complete content of the intermediate
memory will be printed out.
pitot tube
m/sek
44 Nm3/h
Ecom-D Page 17
Page 18
8.4. Soot...Oil trace
Soot..Oil trace
Boiler temp. : 66°C
1.Soot meas. : -.-
2.Soot meas. : -.-
3.Soot meas. : -.-
Oil trace : ----
Soot..Oil trace
Boiler temp. : 66°C
1.Soot meas. : 1.0
2.Soot meas. : 0.5
3.Soot meas. : 1.5
Oil trace : NO
Make sure to let the probe cool down before
The "Soot...Oil trace" menu enables the input of
measured results for boiler temperature, soot dots
and oil trace. Select the desired line on the display
and confirm the input with <Enter>. The input for
boiler temperature and soot measurements 1-3 can
be made one after the other using the instrument
keyboard. Press <Enter> to store the value in the
Select :
data record of the measurement.
The result of the oil trace check will be documented as follows:
- Move cursor to "Oil trace"
- Confirm the result with <Enter>
("No", "Yes" or "- - - ")
Select :
Once all inputs have been entered, press <ESC> to exit the menu.
The measurement is now complete.
securing it for transport!
Page 18 ECOM-D
Page 19
8.5. Measurement record and printout
Start printout
View memory
Memory -> M
Input text
--ECOM-D--
Select
Start printout
View memory
Memory -> M
Input text
--ECOM-D--
Select
Start printout
View memory
Memory -> M
Input text
--ECOM-D--
Select
Start printout
View memory
Memory -> M
Input text
--ECOM-D--
Select
Important: Once the gas analysis is completed, transfer the values recorded in the intermediate memory on the MM card, otherwise they will be deleted when the analyzer is turned off!
Press <Print> (printer symbol) to enter the printing
menu. The sampled data can be checked again
(<View memory>, <Enter> and scroll with the arrow keys).
The software keyboard enables input or correction
of the 6 lines of text in the file identification. (<Input text>, <Enter>, to write text, see chapter 7).
If all the data is correct, press <Memory> -> <M>
and <Enter> to save it in the internal memory or on
the MM card ("Disk symbol" is shown in black in
measurement mode). The text is saved into the data
record only by saving on an MM card.
Select <Start printout> and <Enter> to print out
the data (only instruments with printer).
Press <ESC> to return to gas analysis mode.
<View memory>, <Memory -> M> and <Input
text> can also be selected outside of measurement
:
:
mode with <Enter>.
Ecom-D Page 19
:
:
Page 20
9. Mean Value Measurement (option)
Gas analysis
Mean value
Soot..Oil trace
Data processing
Adjustments
Control
Diagnostics
The mean value measurement function allows you to sample measurements in an adjustable time frame, and to calculate the average of
those results. Should the several measurement values or the average
calculation need to be saved, a save file has to be selected as described in chapter 7. If the “Store” function is used, all measurements
will be placed in numeric order starting with the number of the first
saved file. If the “Store” function is not used, the results can be
stored on MMC with <Print> / <Memory -> M>.
Once the fresh air calibration is completed, select
<Mean values>. Before starting, <Measurement
time>, <Scanning>, <Printer> and <Record>
should be modified as needed. See their use below:
- Measurement time = the length of time the measurements will occur
- Scanning = Time interval between the
measurements considered in the average value
calculation
- Printer = Documentation of
measurements of the average value calculation
- Store = All measurements for average
value calculation will be stored
<Measurement time> and <Scanning> can be adjusted as follows:
- Select value you want to change and confirm with <Enter>
- Set the desired time using the numerical keys:
0.01 = 1 sec = minimal value
59.59 = 59 min: 59 sec = maximum value
- Confirm with <Enter>
Page 20 ECOM-D
Page 21
The settings for <Printer> can be changed as follows:
O2 3.2 %
CO2 13.1 %
CO 0 ppm
Eff. 92.5 %
Losses 7.5 %
Exc. air 1.18
T.Gas 184 °C
T.Air 20 °C
Mean value
15:35min
- Select value you want to change and confirm with <Enter>
- Make the desired adjustment with the arrow keys
- Confirm with <Enter>
The setting for <Store> can be changed as follows:
- Select value you want to change and confirm with <Enter>
- Activate the memory function with <F1> or deactivate it with <F4>
Press <Start measurement> / <Enter> to start the
evaluation of the measurement values. On the display, the actual average values will be shown and
will be updated as new measurements come in. You
can switch to the actual values with the arrow keys
<up/down>. You can also scroll through the values
with the arrow keys <right/left>. You can interrupt
the measurement with <F2> or stop it completely
with <F4>.
After the measurement time has elapsed, you can print the results
containing all average values (only instruments with printer - key
<Print>).
Ecom-D Page 21
Page 22
10. Adjustments
E
= E
*
21 - O
21 - O
Einheit
Ref. O2
Fuel type
Air pressure
Set clock
Paper feed
Internal
Select :
Adjustments
In addition to the ECOM-D functions described previously, various adjustments can be made to the instrument.
From the main menu, select "Adjustments" and confirm with <En-ter>.
A selection of modifiable parameters, adjustable according to the application, is displayed.
Place the cursor on the desired line and press <En-ter> to view or modify the adjustment. The modifiable parameters are:
Unit (adjustment with arrow keys):
- Calculation of gas concentrations in:
- ppm = volume concentration (parts per million)
- mg/m
3
= mass concentration per volume unit
- mg/kWh (undiluted) = mass concentration per power unit
- mg/MJ (undiluted) = mass concentration per power unit
- ppm (undiluted) = volume concentration (parts per million)
- mg/m3 (undiluted) = mass concentration per volume unit
Undiluted:
Conversion of the gas concentration on selected reference oxygen:
- mg/kWh and mg/MJ are always calculated on 0% O2 basis
- Conversion formula
ref
meas
O2 reference
(for ppm and mg/m3 - Input after pressing <Enter>):
- Input of O2 reference value O
Page 22 ECOM-D
2ref
2meas
2 ref
Page 23
Fuel type (press <Enter> to access selection list):
Print contrast
Displ. contrast
Key beep
Language: English
F1 Hotkey
F4 Hotkey
Eff. (C)
CO-Automatic
USB
Bluetooth
WLAN
Pitot-factor
Printout
Select:
Internal
- Modification of adjusted fuel type
(e.g. by measurements at combi-plants)
Air pressure (press <Enter> to access menu):
- Input of barometric air pressure for dew point calculation
Clock set (press <Enter> to access setting menu):
- Correction of internal clock with the arrow keys
Paper feed (only instruments with printer)
(press <Enter> to activate paper feeding):
- Paper feed line by line
Internal (press <Enter> to open menu):
- Further instrument settings:
Print contrast
(only instruments with printer)
(press <Enter> to access setting menu):
- Print contrast adjustment with
the arrow keys
Display contrast
(press <Enter> to access setting menu):
- Display contrast adjustment with the
arrow keys
Key beep
(<F1> for YES / <F4> for NO):
- Keys beep when pressed
Language: English (change with <Enter>):
- Info about selected language (3 languages selectable)
Ecom-D Page 23
Page 24
F1 Hotkey (selection after pressing <Enter>):
With first using the Bluetooth connection to your
- Change the menu that pressing <F1> will open
F4 Hotkey (selection after pressing <Enter>):
- Change the menu that pressing <F4> will open
Efficiency (C) (<F1> for YES / <F4> for NO):
- Efficiency calculation with or without condensation gain
CO-Automatic (<F1> for YES / <F4> for NO):
- Adjustment for CO purging
- YES = CO sensor switches on after purging automatically
- NO (default) = CO sensor has to be switched on manually with
<F3> after purging.
USB (selection after pressing <Enter>):
-Adjustment of transfer speed (arrow keys <Up/Down>) and protocol
(arrow keys <Right/Left>) for the USB interface (connection USB):
-Protocol DAS = Protocol for the program DASNT
-Protocol Enhanced = Protocol for the program DAS5
Bluetooth (selection after pressing <Enter>):
- Adjust the protocol for the Bluetooth interface with the arrow
keys <Right/Left>:
-Protocol DAS = Protocol for the program DASNT
-Protocol Enhanced = Protocol for the program DAS5
PC, type in the password shown!
WLAN (selection after pressing <Enter>)
Instrument as Access Point (for connection with mobile terminals):
-(Start/Stop WLAN: manual switching of WLAN connection
– available only with if Auto Connect is inactive)
-Access Point: (<F1> for YES / <F4> for NO)
Page 24 ECOM-D
Page 25
-(W.O.) Auto Connect: Automatic connection
(<F1> for YES / <F4> for NO)
-Channel: Input channel (1 – 13)
(selection after pressing <Enter>)
Connection with existing network:
-(Start/Stop WLAN: manual switching of WLAN connection
– available only with deactivated Auto Connect)
-Existing Network: available only with deactivated Access Point
-(W.O.) Auto Connect: Automatic connection
(<F1> for YES / <F4> for NO)
-Network scan: Search for available networks (selection with <Enter>)
-WPA password: Input of password for the selected network
Pitot factor (selection after pressing <Enter>):
- Input of Pitot factor for flow rate calculation (standard = 0.93)
Printout (only instruments with printer)
(selection after pressing <Enter>):
- Text input for printout on measurement protocol (8 x 24 characters)
- Input the text of line 1 as follows:
1. Activate character selection list with <F4>.
2. Select keyboard type with <F3> (4 different keyboards available).
3. Use the arrow keys to select the desired character
(selected character is outlined by black background).
4. Confirm the selection by pressing <Enter>.
5. Repeat procedure until desired text is complete.
6. Once input for line 1 is completed, deactivate the character
selection mode with <F4> and move to the second line with the
arrow key <Down>.
7. Once all lines are complete, exit the menu
with <ESC>.
Ecom-D Page 25
Page 26
11. Control
O2 10744 mV
CO 7 mV
Accu 6.09 V
Further pages:
Operation hours : 8.45
hrs
Total : 18.75 hrs
Next unit check :
01.07.10
Last service (history)
25.08.09 88
14.08.10 145
23.08.11 205
19.08.12 322
--.--.--
--.--.--
Further pages:
Program version :V1.2
19.11.09
Serial no. :D-0001
CO purges : 15
Error counter : 21
Further pages:
Operation hours :
8.45 hrs
Total : 18.75
hrs
Next unit check :
The electrochemical sensors for gas analysis are subject to normal
wear and tear as they age. Their accuracy may decrease as time
goes on, depending on the gas concentration, the exposure time and
the soiling grade of the sampled gas. The analyzer monitors the sensors and corrects drift.
If the drift and the correlated measurement errors increase, an error
message is displayed. In this case, the corresponding sensor must be
changed by an authorized service center. The control menu has information about the current status of the sensors.
Further information is can be seen on 3 display pages (use the arrow
keys to scroll):
- Battery voltage (charging status);
is displayed as a symbol in all menus:
Full charge Half charge Empty
Enables data importation from the rbr software (available on our website <www.rbr.de>). See the chapter on <Technical Data> for data
format information (please observe the transfer options of your software!).
Proceed as follows:
-Connect ECOM-D to your PC via USB cable.
-Select <Load data> and confirm with <Enter>.
-Answer the displayed question with YES (<F1>).
-Decide if the data recorded can be cancelled
(<F1> for YES / <F4> for NO).
-Start the data transfer to your PC.
Send data:
With this function, the data records completed with measurement values can be transferred to the PC.
(the procedure is similar to <Load data>).
Data logger (Option):
Here a Data logger record (Select <Data logger> and press <OK> /
On the top right of the display, the disk symbol appears.) can be started or finished (only available when using the multi-media card). With
<Record>, you can interrupt or continue the record (only in gas analysis). For each recording, one file will be written on the card. The files
will be numbered consecutively (J2KDL-00.csv, J2KDL-01.csv and so
on) and can be transferred to your PC with a card reader. The maximum size of a dataset is 500 bytes which means that on a 32 MB card
64000 measurements can be recorded.
In addition to data logger recordings, the data can also be transferred
with a USB cable (USB Driver / 1200 Baud / Protocol DAS) to the
software “DASNT”. The software “DASNT” and the USB Driver are
available free of charge from the rbr website.
Save to MMC (min. 1 second / max. 255 seconds):
Adjustment of the interval time for data logger recordings.
Page 28 ECOM-D
Page 29
Service by unauthorized service centers will
result in a complete and immediate loss of any
Fine dust filter
13. Maintenance Tips
To secure the accuracy of your analyzer, we recommend an annual
check by an authorized ECOM service center. In the case of heavy
usage (e.g. several hours of measurement per day, rough conditions
etc.) service should be done more frequently. All ECOM service centers can be found at www.rbr.de.
warranty!
The following advice will be of help with maintenance of your ECOMD:
Change Fine dust filter (Instruments with gas cooler)
Screw off the cover of the condensate trap/gas cooler and check the
particle filter. Change it once the filter has a grey color (= number 2 - 3
of the soot comparison scale).
Ecom-D Page 29
Page 30
Condensate pad (Instruments without gas cooler)
Regularly check the condition of the condensate pad. It should be
changed:
- If it is darker than no. 3 on the soot scale
- If it has expanded to the "MAX" mark
To replace the condensate pad:
1. Remove hose from the cap and unscrew the cap
2. Unscrew the retaining bolt with spring
3. Replace old condensate pad with the new
4. Screw on retain bolt with spring again
5. Screw on cap and replace the hose to the cap
Sensors
The sensors are auto-zeroed in fresh air each time the analyzer is
turned on. Their state is constantly monitored by the instrument. New
sensors age as you use the analyzer due to normal wear and tear
(oxygen sensor) and due to measurements exceeding concentrations
beyond the nominal measurement range (toxic sensors). The output
values of the sensors are (menu "Control"):
O2 approx. 20000 mV
Others 0 mV (+/- 150)
If an error message is displayed during the auto-zero phase and does
not disappear after several retries, the instrument must be checked by
an authorized service center. The oxygen sensor must show a value
of >7000 mV, otherwise it must be changed by an authorized service
center.
The CO sensor is protected against over-saturation by an internal
program. If the limit value of 4000 ppm is exceeded, a second pump
switches on and floods the sensor with fresh air.
Page 30 ECOM-D
Page 31
After sufficient purging time (X behinds CO disappears), the sensor
Unlocking
printer cover
can be reintegrated into the measurement system with <F3> (if you
choose <Yes> in <Adjustments> / <Internal> / <CO-Automatic>,
the CO sensor will begin measuring again automatically). The sensor
can also be manually excluded from the measurement system with
<F3>.
Probe and tubing
Depending on the frequency of use, the probe and sample line should
be regularly cleaned in order to prevent particle deposit build-up and
to prevent early wearing due to corrosion.
-Release the connections at the instrument and at the probe grip
to disconnect the tubing.
-Clean it with warm water in then blow dry the water out
-Slightly grease the O-rings of the tubing connections from time to
time with acid-free grease.
Change printer paper roll
-Release the printer cover (press lock downwards).
-If necessary, extract the remaining paper from the printer. To do this,
select
<Adjustments>/<Paper feed>/<Enter>).
-Remove the plastic tube of the previous roll.
-Insert the paper end in the slot under the transport roll (paper
roll on the inner side facing you when inserting the paper).
-Feed approx. 3 cm paper through the printer (<Adjust-ments>/<PaperFeed>/<Enter>).
-Lay the paper roll in the corresponding space.
-Insert the paper through the slot of the printer compartment then cover and close the lid.
Ecom-D Page 31
Page 32
14. Technical Data
Parameter Range PrincipleO2 0 ... 21 vol-% Electrochemical
CO 0 ... 4000 ppm Electrochemical
NO (option) 0 ... 5000 ppm Electrochemical
NO2 (option) 0 ... 500 ppm Electrochemical
Power supply Charger 230 V / 50 Hz~; Accu 6 V / 3,8 Ah
Protocol printer integral; 58 mm paper width
Printout individually programmable
Indication graphic display; backlit
Dim. (W x H x D) 400 mm x 260 mm x 175 mm
Weight approx. 6 kg complete with sampling system
Calculation
2max
Subject to technical changes
Page 32 ECOM-D
Page 33
15. Calibration Procedure
For best results, the calibration gas concentration should be as
close to the expected levels of emissions as possible. Because each
sensor is linear through a nominal range, one calibration gas concentration can be used for a reasonably wide range of emission levels.
Fully charge analyzer:
Place unit on charge the night before you plan to use analyzer to
ensure full charge of battery.
Turn on analyzer and let warm up:
Switch on analyzer. Select Gas Analysis on main menu. Select
Fuel Type and press OK. Select No for Data Processing (if prompted).
Instrument will complete auto-zero sequence. Let the analyzer run for
15 minutes to acclimate to ambient temperature. This is very important, as the accuracy of electrochemical sensors is dependent on a
proper temperature setting. Switch off the analyzer and switch back
on. Go to Gas Analysis mode. The instrument will complete autozero based on the proper temperature.
Calibrate your analyzer:
1.Select Control on the main menu. Swipe calibration magnet over
the Cal Magnet sticker, located above the ESC button. The analyzer is now in calibration mode. You will see real-time readings on
the screen.
2.Connect calibration hose from the regulator to the gas inlet
(O2/CO) on the analyzer.
a. If you are using an on demand regulator, connect it to the gas
bottle and open gas bottle valve completely. The regulator will
hold back the flow of gas until connected to the analyzer.
b. If you are NOT using an on demand regulator, adjust the flow
to match the analyzer pump draw.
Ecom-D Page 33
Page 34
3.You must check for air leaks before calibrating. Apply a gas with
0% O2 to check for air leaks (ECOM service dept. uses NO gas
balanced in N2). Flow calibration gas for 3 minutes. Check that O2
reading is 0.0%. If it is reading other than 0.0 %, turn off gas and
troubleshoot the source of air leak. Do not proceed to calibrate
analyzer until air leak is corrected.
4.If there is no air leak, allow gas to flow for another 2 minutes (total
of 5 minutes). After the reading has stabilized, press up/down to
place the arrow beside the target sensor and press OK to go to
input mode. Input concentration of calibration gas using numbers
on keypad. Press OK. The sensor has now been calibrated.
5.Allow the gas to flow for another 45-60 seconds. If it does not drift
more than 2% of the span gas, the calibration is steady and you
may disconnect the gas. (If it does drift more than 2%, allow the
reading to stabilize and reset calibration. If this still doesn’t work,
call ECOM.)
6.If calibrating more than one sensor, disconnect the regulator and
reconnect to the next gas cylinder. Then flow calibration gas for 5
minutes, input calibration setting, & repeat step 5.
7.After all sensors have been calibrated, disconnect gas and allow
the analyzer to flush with fresh air for 10 minutes or until readings
are below 10 ppm.
Questions? Call ECOM at 1-877-326-6411
Page 34 ECOM-D
Page 35
Ecom-D Page 35
16. Memory Card Data
Description of data record ecom-D with Multi Media Card
Format data logger records: J2KDL-xx.csv (separation mark between values = comma)
Format punctual measurements: J2KDV.txt (separation mark between values = comma)
* converted to unit (column AC) and converted on O2 ref. (Column AB) when column AD = N
Page 36 ECOM-D
Page 37
17. Calibration log
Date:
Cal gas concentration
Sensor response after 5 min
CO
NO
NO2
Date:
Cal gas concentration
Sensor response after 5 min
CO
NO
NO2
Date:
Cal gas concentration
Sensor response after 5 min
CO
NO
NO2
Date:
Cal gas concentration
Sensor response after 5 min
CO
NO
NO2
Date:
Cal gas concentration
Sensor response after 5 min
CO
NO
NO2
Date:
Cal gas concentration
Sensor response after 5 min
Ecom-D Page 37
Page 38
Date:
Cal gas concentration
Sensor response after 5 min
CO
NO
NO2
Date:
Cal gas concentration
Sensor response after 5 min
CO
NO
NO2
Date:
Cal gas concentration
Sensor response after 5 min
CO
NO
NO2
Date:
Cal gas concentration
Sensor response after 5 min
CO
NO
NO2
Date:
Cal gas concentration
Sensor response after 5 min
CO
NO
NO2
Page 38 ECOM-D
Page 39
18. FAQ
Where do I find important instrument information?
How long is the life span of the
sensors?
Which sensors can I exchange?
The instrument shows the error
message “O2 sensor 0 mV“!
The instrument shows the message“Check required“!
The instrument shows the error
message “T-Gas“ or “T-Air“!
In the menu “Control“, all important instrument information is
shown (e.g. accu voltage, sensor values, unit number, next
service date, operation hours etc.). With the arrow keys
stands you can switch to the second page.
The life span depends on the operating hours and the instrument equipment. The life span of the toxic sensors (CO,
NO, SO2, NO2) is affected by high gas concentrations and
a not sufficient purging. The life span for these sensors
amounts to on the average between 4 and 6 years. The life
span of the O2 sensor is independent of the operating hours
and amounts to approx. 2 years.
The following sensors are exchangeable:
- O2 sensor
- CO sensor (pre-calibrated)
- NO sensor (pre-calibrated)
- SO2 sensor (pre-calibrated / only together with
CO sensor
The sensor must be renewed.
This message appears automatically every 12 months or
after 250 operating hours. Note: This is a recommendation
to check the instrument. The instrument is
however still ready for use.
Possible reasons could be:
- Cable is broken (at the plug)
- T-Air sensor is broken
- Thermocouple is broken
- Cable is defective
Note: The error messages can be ignored by pressing “En-
ter“. Calculations that depend on these temperatures are
not implemented.
Possible reasons could be:
- O2 is defective (CO2 values are calculated from
the O2 values)
- Please load the accumulator min. 8 hours
(Accumulator could be over-discharged)
Please check whether the printer paper is correctly inserted. The thermal printer writes only on the thermally
sensitive side. Please always use the correct paper for
the printer to prevent defects with the printer. Please
make sure that the printer is clean (no chads in the
drive).
You can change the printout (Menu: Adjustments).
Hint: If you have several instruments of the same type, you can locate an error
by exchanging the accessories (probe, hose, temperature sensor etc.).
If further questions or problems should arise, please contact the next authorized
service center.
Page 40 ECOM-D
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.