ECOM CN User Manual

ECOM CN
Portable Emission Analyzer
Measures O2, CO & NO (NOx)
CO2, efficiency & excess air calculations
Stack temperature and draft measurement
Backlit LCD display
IR printer port
Gas cooler & moisture alarm with shutdown
1.5+ lpm pump
Automatic purge /
Sensor overrange protection
On-board data storage
The ECOM CN is an easy to use, handheld gas analyzer designed for basic combustion monitoring in a variety of applications.
V E T S
The ECOM CN is fitted with 2-3 gas sensors, measuring O2, CO & NO plus Stack and
Ambient Temperatures, and Stack Pressure and Draft. The CN also calculates CO2,
The instrument is housed in a robust composite case suitable for rugged field use. On-board
data storage for up to 500 separate readings. An IR printer port transmits data to a portable
printer for on-the-spot print-outs. A large LCD displays all measured and calculated
parameters at once, or can be “zoomed” to show only those particular measurements the
operator wants to see. The tactile keypad guides the operator through the various functions
of the analyzer.
The ECOM CN sample conditioning system consists of a high flow pump and radiant gas
cooler with moisture trap. An automatic shut-off activates when the gas cooler filter becomes
saturated. Another unique feature of the CN is an Automatic and Manual CO bypass function
which protects the CO sensor from over exposure.
Technical Specifications
MEASUREMENT (KEY) RANGE ACCURACY RESOLUTION SENSOR LIFE SENSOR TYPE
Oxygen (O) 0-21% vol. ± 2% Measured 0.1% vol. 2 Years Electrochemical
Carbon Monoxide (C) 0-4,000 ppm ± 2% Measured 1 ppm 3 Years Electrochemical
Carbon Monoxide (V) 0-40,000 ppm ± 2% Measured 1 ppm 3 Years Electrochemical
Nitric Oxide (N) 0-4,000 ppm ± 2% Measured 1 ppm 3 Years Electrochemical
Gas Temperature 32-1800 F ± 2% Measured 1 deg F 10 Years NiCrNi
Ambient Temperature 0-250 F ± 2% Measured 1 deg F 10 Years Semi-conductor
Draft / Pressure ± 40” H2O ± 2% Measured 0.1% H2O 10 Years DMS
O2 Correction 0-20% Oxygen
NOx Calculated NOx by programming NO2 factor
Carbon Dioxide CO2 0-CO2 max of fuel Calculated
Efficiency 0-99.9% Calculated
Excess Air (Lambda) 1-infinity Calculated
*ACCURACY: When calibrated prior to use per ECOM America, Ltd. specifications.
ECOM CN
Physical
Instrument: 3” x 10” x 4” Probe: 10” Length x 3/8” OD Inconel w/ Pistol Grip Handle* Sample Line: 15’ high-temp flex hose w/thermocouple wire* Weight: 1 lb.
Electrical
Batteries: 6-8 hour battery life, 6V lead-acid Pump: Flow rate of 1.5+ lpm Display: Backlit, adjustable contrast & zoom, displays all
parameters simultaneously.
Operating Temperature
Core temperature of the instrument is monitored continuously. Internal Temperature Compensation software assures accurate sensor response over the range of 20°F to 104°F.
*Optional probe & sample line lengths are available.
Customer Service / Technical Support
ECOM America Ltd. has built an outstanding reputation in the industries we serve by providing a high quality product and responsive technical support to our customers. We maintain a large inventory of analyzers and parts to assure fast delivery of new items and quick turnaround on service. Always Expect Responsive Customer Support & Service from ECOM!
Order Information
ECOM CN
CN is available with 2 sensors. Common configurations include:
2 Gas 3 Gas
OC OLN OV OVN
Accessories & Options:
Infrared Printer, Thermal Printer Paper
Consumables:
Condensate / Particulate Filters
Parts & Accessories:
Refillable Calibration Gas Bottles
ECOM America, Ltd. 1628 Oakbrook Drive Gainesville, GA 30507
Toll Free: 877.326.6411 Tel: 770.532.3280 Fax: 770.532.3620
info@ecomusa.com www.ecomusa.com
©2000 ECOM America, Ltd. All rights reserved. All other trademarks belong to their respective owners.