Eclipse Eclipse MT-5110 Manual

Test Equipment Depot - 800.517.8431
99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
MT-5110 3 1/2 Capacitance Meter
User
’s Manual
st
Edition, 2012
1 ©2012 Copyright by Prokit’s Industries Co., Ltd.
2
1. FEATURES
Easy High measuring accuracy. Measurements are possible even under a strong magnetic field. LSI-circuit provides high reliability and durability. Input overload protection is provided. LCD display for clear readout even in bright ambient light
Light-weight and compact construction for easy operation. Low battery condition is indicated on the LCD display. Low power consumption.
2. SPECIFICA
2-1.GENERAL SPECIFICATIONS
Display: LCD (Liquid Crystal Display) Max. Indication 1999. Measurement: C (Capacitance) Range: single 9 position, whole range value (from 0.1pF to
Zero Adjustment :Manual (range:±20pF) Calibrate Adjustment: Have two internal adjustments. Over-input: Display shows “1”. Backlight Function: it went out by itself within 8 seconds. Sampling Time: 0~5second Operating Temp: 0 to 40℃. Operating Humidity: 80% MAX.R.H. Power Supply: Single, standard 9 volt battery. NEDA
Typical consumption current: 3~4mA Standard Accessories: Test alligator clips (red & black)…1 pair.
and correct readout.
conditions.
TIONS
20000uF)
1604IEC6F22
Test leads (red & black)………….1 pair.
Test socket …………………… … 1 pc
Instruction manual………………………….………… ……..1 pc.
3
2-2. ELECTRICAL SPECIFICATION
Accuracy is ±(percentage of reading + number of digit) at 23±5 ,<80%RH
.
Range Accuracy Resolution Test Frequency
200pF ±(0.5%+7) 0.1pF 800Hz 199.9pF
2nF 1pF 800Hz 1.999nF
20nF 10pF 800Hz 19.99nF
200nF 100pF 800Hz 199.9nF
2uF 1000pF 800Hz 1.999uF
20uF 0.01uF 80Hz 19.99uF
200uF
2000uF ±(2%+5) 1uF 8Hz 1999uF
20000uF ±(3%+10) 10uF 8Hz 1999(×10)uF
pF= Pico Farad(10
±(0.5%+5)
0.1uF 8Hz 199.9uF
-12
F )nF= nan Farad(10-9F). uF= micro Farad(10-6F)
Max indication
value
Excitative voltage: Max.2.8Vrms
Overload Rating: Protection by a 0.1A/250V fuse.
3. O
PERATION PA
1. Model
2.
LCD display: display
3.
Backlight key: pr
NEL
Number
the test value and unit.
ess the button lightly; it was turning off by itself about 8 seconds. Capacitance “+”i
4.
5. Cap
acitance””i
nput terminal.
nput terminal.
6. Function Selector: It is used for power on and changes the
7.
Zero knob: Knob to zero when test low
range of fu
nction.
capacitance.
4
4. CONSIDERATION OF MEASUREMENT
(1) This C METER is intended for measuring the capacitance value of
a capacitor. It is not intended for determining the “Q” factor for above reactive components. Misleading readings may be obtained if the measurement of capacitance of a resistor is attempted.
(2) When measuring components within circuit that circuit must be
switched off and de-energized before connecting the test leads.
(3) Do not damage (black & red) test leads, if any worn please change (4) Instruments used in dusty environments should be stripped and
cleaned periodically.
(5) Do not leave the instrument exposed to direct heat from the sun for
long periods.
(6) Before removing the battery and fuse compartment cover, ensure
that the instrument is disconnected with any circuit and the power switch is in the off position.
(7) For all measurements, should connect BLACK test lead into “-”
terminal and RED test lead into “+” terminal.
5. CAPACITANCE(C) MEASURING PROCEDURE
(1) Select the range selector for the maximum expected capacitance. (2) The power will on when the selector switch away from OFF
position.
(3) Check "0" indication: If test range is 200pF, 2nF, 20nF, should
check "0" indication before test. (4) Observe polarity when connecting polarized capacitors. (5) Full discharge any capacitors. (6) Connect the alligator clips to the capacitors leads. (7) Read the display. The value is direct reading in the electrical unit
(pF, nF, uF) indicated at the selected range switch. If display show
“1”, It indicate on Out-of-Range measurement. If the display
indicates one or more leading zeros, shift to the next lower range
scale to improve the resolution of the measurement.
5
NOTE:
(a) If the capacitance value is unmarked, start with the 200pF range a
keep increasing until the over-ra reading is obtained.
(b)
A shorted capacitor will read over-range on all ranges. A capacitanc w
ith low voltage leakage will read over range, or a
value than normal. An open capacitor will read zero on all ranges (possibly i
(c) Measure of very low capacitance should be performed using
extrem inductance.
(d) When using the optioned test leads, remember that the leads
introduce a meas appro opening the leads at the trips, recording the open circuit value and subtracting that value.
(e)
Capacitors, especially electrolytic, often have notoriously wi tolerances. Do not be surprised if the measured value is greater than the value marke However, values are seldom drastically below the ra
(f) If changing range, measured value will be changed; leakage-voltage
cap decreased in lower range.
a few pF on 200pF range, due to stray capacit
rument).
nst
ely short leads in order to avoid introducing any stra
urable capacitance to the measurement.
ximation, the test lead capacitance may be measured by
d on the capacitor, unless it is a close tolerance type.
acitors will be checked also. Leakage-resistance w
nge indication goes of
f and a
much higher
ance of the
y
As a first
de
ted value.
ill be
nd
e
6
6. MAINTENANCE
1) 9-Volt battery replacement a. Ensure the instrument is not connected to any external circuit. Set
the selector switch to OFF position and remove the test leads from
terminals. b. Remove the screw on the bottom case and lift the bottom case. c. Remove the spent battery and replace it with a battery of the same
type.
Fuse replacement
2) a. Ensure the instru
the selector switch to OFF position and remove the test leads from
terminals. b. Remove the screw on the bottom case and lift the bottom case. c. Replace the fuse with the same type and rating: 5×20mm,
100mA/250V, fast-blow fuse or as the replacements.
ment is not connected to any external circuit. Set
The specifications The content of this manual is regarded as correct, error or omits Pls. contact with factory. We hereby will not be responsible for the accident and damage caused by improper operation. The function stated for this User Manual cannot be the reason of special usage.
are subject to change without notice.
7
MT-5110 3 1/2 數位電容錶
1. 特色
◇測量準確, 讀數方便測量精度高在幹擾的磁場下依然可以測量LSI 電路提供高可靠性和耐久性提供輸入過載保護即使在明亮室內光線狀況下, LCD 清楚顯示測量結果結構設計容易操作使用若工作電池電壓低, LCD 上會顯示低電壓符號提醒整機低功耗
2. 說明
2-1.
一般說明
顯示︰LCD(液晶顯示螢幕) 最大讀值 199
測量︰
範圍︰9 個檔位, 全範圍值從 0.1pF to 20000uF
歸零調整
校驗調整
輸入超過︰螢幕顯示
背光功能
採樣時間︰0 ~5second
操作溫度
電源︰
標準消耗
標準配件︰ 測試鱷魚夾子(紅和黑色) 1 對
C(僅電容量
︰手動(範圍︰大約 ±20p ︰內含 2 個調整裝置
8秒內
0 ~40 ℃.最大濕度:
1 個標準
電流:34 毫安培
)
“1”
特電池。 NEDA 1604IEC 6F
9
測試錶棒(紅和黑色) 1 對 測試座 1 只 操作說明書 1 本
F)
80%.R
9
.H.
22
8
2-2. 電氣規格 精度為 ±(讀數的百分比+個位數)在 23 ± 5℃,<80%相對濕度。
範圍 精準度 解析度 測試頻率 最大顯示值
200pF
±(0.5%+7)
0.1pF 800Hz 199.9pF
2nF 1pF 800Hz 1.999nF
20nF 10pF 800Hz 19.99nF
200nF 100pF 800Hz 199.9nF
2uF 1000pF 800Hz 1.999uF
±(0.5%+5)
20uF 0.01uF 80Hz 19.99uF
200uF
0.1uF 8Hz 199.9uF
2000uF ±(2%+5) 1uF 8Hz 1999uF
20000uF ±(3%+10) 10uF 8Hz 1999(×10)uF
pF= Pico Farad(10
-12
F )nF= nan Farad(10-9F). uF= micro Farad(10-6F)
Excitative voltage: Max.2.8Vrms
功率超載:由 0.1A/250V 保險絲保護。
面板
3. 操作
1. 產品型號.
2.
LCD 螢幕︰顯示測試值和單位
3. 背光按鍵︰輕輕按下按鈕,約 8 秒後自動關閉
4. 電容輸入端 “+”
5. 電容輸入端 “
6. 功能旋鈕︰用於啟動及更改功能範圍
7. 調零旋鈕:低電容測試時調節為零
注意
4. 測量
事項
(1) 這台電容測試器用於測量一般電容器的電容量值。不能確定器件的
“Q”因素。也不能用於測量電壓,電阻等參數,否則獲得誤導性的讀
值。
(2) 當測量零件時,必須確認這個零件在沒有通電的狀態下進行,所連接
的電路必須被關掉並且切斷。
9
(3) 注意檢查(黑色和紅色)測試線, 若損壞請更換. (4) 在佈滿灰 (5) 不要長期把儀器暴露在太陽下。 (6) 取出電池和打開保險絲蓋子之前,確認儀器任何連接或電源開關處於
關閉位置。
(7) 所有的測量,應該連接到黑色測試線 “ - ”和紅色測試線 “+”
5. 電容(C)測量步驟
(1) 將功能旋鈕轉到待測試電容的預期最大值選擇的範圍 (2) 電源會同步開啟。 (3) 檢查“0”位:如果測試的範圍是 200pF,2nF20nF,應測試前檢查
(4) 當測試有
(5) 任何電容 (6) 把鱷魚夾子和電容器各引腳連結起來。 (7) 讀取螢幕顯示的數值,電容值單(pF, nF, uF)裡的直接讀取,如果螢
注意︰
(a) 如果不知道待測零件電容量值,從 200pF 範圍開始並且持續增加,
(b) 一個短路的電容測量時會顯示超過量測範圍的讀值,並顯示“1”。與
(c) 為了避免引起任何雜訊干擾,測量非常低的電容量應該使用極其短
(d) 當使用任選的測試錶棒時, 記住錶棒可能引入一個的電容. 首先
塵的環境裡使用儀器應該定期進行清潔。
儀表顯示為“0”, 若有偏離請調整”ADJ〞調零旋鈕.
極性電容器時,檢查其極性, 以對應測試插座 “+”端,
端連接.
器測量前,應對電容器進行完全放電。
幕顯示“1”表示測量結果超過選定測量範圍。如果螢幕顯示一個或多 個零在數字前面,切換到下一個較低的範圍檔位提高解析度。
直到超量指示熄滅及讀取到讀值。
正常的電容相比較,有低電壓漏電的電容器其測量讀值將會較高或 超過範圍。
的測試線執行。
的電
測試錶棒 值並從測試結果中減去此值。
應在打開筆尖的情況下, 測出該錶棒
容, 記錄其開路
,
10
(e) 電容器,特別是電解電容,往往有廣泛的公差。不要感到驚訝,如
果測量值大於標在電容器上的值,除非它是一個緊公差,不管如何 測量,測量值很少大幅低於額定值。
(f) 如果更改測量檔位,測量顯示值將不同。 電容器漏電壓也將隨檔位
不同而變化, 在低檔測量位置.漏電電阻會變小。
6. 維修
1) 9 伏特
2) 保險絲
的電池更換
a. 確認儀器不連接任何外部電路。確定儀器功能旋鈕轉向為 OFF 位置並
且除去測試線。
b. 取下底蓋的螺絲與蓋子。 c. 取出電池並更換一個同類型的電池。
更換
a. 確認儀器不連接任何外部電路。確定儀器功能旋鈕轉向為 OFF 位置並
且除去測試線。
b. 取下底蓋的螺絲與蓋子。 c. 用相同的類型和等級的保險絲替換︰5×20mm, 10
0mA/250V
本說明書
本說明書的內容被認為是正確的,若用戶發現有錯誤、遺漏等,請與生產廠
家聯繫。
本公司不承擔由於用戶錯誤操作所引起的事故和危害。
本說明書所講述的功能,不作為將產品用做特殊用途的理由。
如有改變,恕不另行通知
11
Test Equipment Depot - 800.517.8431
99 Washington Street Melrose, MA 02176
©2012 Prokit’s Industries Co., LTD. All rights reserved 2012001(T)
TestEquipmentDepot.com
12