General
This product is equipped with the LCD display of text height
20mm, is a 3 3/4 digital multimeter which has the merits of clear
reading, stable performance and high reliability. It can be used to
measure DC voltage, AC voltage, DC current, AC current,
resistance, capacity, frequency/duty cycle, diode and make
on-and-off test. Meanwhile, it is available for unit symbol display,
automatic/manual range switching, automatic power off and
alarm function. The multimeter has high resolution and precision,
due to its complete functions, high measurement accuracy and
convenient operation,
troubleshooting
ideal for general electrical testing and
1
Open-package Inspection
Open the package box and take out the meter, check carefully
if the following accessories are absent or damaged. If there
is any absence or damage, please contact the distributor
immediately.
Digital multimeter 1 PC
User’s manual 1 copy
Test leads 1 pair
Temperature Probe (K-Thermocouple) 1 PC
Test socket 1 pc
Safety Note
The design of this meter is accordance with IEC1010 clause (the
safety standard issued by International Electro technical
Commission). Prior to the operation of the instrument, please
read the safety considerations before use.
1. When DC voltage above 30V, AC voltage above 25V, current
above 10mA, AC power line with inductive load or power line
during electric fluctuation is measured, please beware of
electric shock.
2. Prior to measurement, check if the measurement function
switch is at the correct position. Check if the test lead is
contacted reliably, connected correctly, and grounded well
and etc. in order to avoid electric shock.
3. Only if the meter is used with the matched test lead, will it
can of the test lead is damaged, it is necessary to replace
with another one of the same model or the same electrical
specification.
4. Don’t use other unconfirmed or disapproved fuse to replace
the fuse inside the meter. Only the same model or same
specification fuse can be replaced. Before the replacement,
the test leads must be removed from the measuring point to
ensure there is no longer any signal at the input terminal.
2
5. Don’t use other unconfirmed or disapproved battery to
replace the battery inside the meter. Only the same model or
same electrical specification battery can be replaced. Before
the replacement, the test lead must be removed from the
measuring point to ensure there is no longer any signal at the
input terminal.
6. When the electrical measurement is made, never let your
body get in touch with the ground directly, and don’t touch
uncovered metal terminal, output port, lead clamp and etc.
where earth potential may exist. Dry clothes, rubber shoes,
rubber cushion and other insulating material are usually used
to keep your body insulated against the ground.
7. Don’t store and use it in the high-temperature, high-humidity,
inflammable and strong magnetic field environment.
8. It may damage to the meter and endanger the operator’s
safety if the voltage value beyond the permitted ultimate
voltage value is measured. The ultimate voltage value
permitted for measurement is marked on the instrument
panel, and never measure the value exceeding the standard.
Don’t input the ultimate value out of regulation in order to
avoid electric shock and the damage to the meter.
9. When the test lead is inserted into the current socket, don’t
measure any voltage for fear that the meter should be
damaged and the operator’s safety be endangered.
10. Don’t try calibrating or repairing the meter. When necessary,
only the qualified professional personnel who have had
special training or gained approval can make it.
11. During measurement, the requirement of measurement
function must be in accordance with LCD display. Please be
sure to disconnect the line of the test lead with the measured
object first and ensure there is no any input signal. It is
forbidden to switch the function/range selection switch during
measurement
3
12. When “ ” is shown on LCD display, please replace
battery immediately to ensure the measurement precision.
13. It is not allowed to insert the test lead into the current
terminal to measure voltage!
14. Please don’t change the circuits of the meter freely for fear
that the meter is damaged and the safety be endangered.
15. Description of Safety Symbols
Warming!
Double
Insulation
Fuse
Low
Battery
Instrument Panel & Function Button Description
1. Instrument model number
2. LCD Display: Display the measured data and unit.
3. Function Button
3.1 Hz/% (Freq
Cycle): Press this button to
select the frequency or duty
cycle mode. The measurement
mode of Vo
duty cycle could be selected by
pressing this button in AC
voltage setting
3.2 RANGE: Press this button to
select the voltage/current test
range
uency / Duty
ltage / frequency /
In accordance with the
instruction of European
DCV
ACV
DCA
ACA
Trade Union
4
3.3 HOLD: date hold
3.4 REL △(Relative Value Measurement): The relative value
measurement of Capacity function could be conducted by
pressing this button
3.5 Select(Function Switch): Press this button, the function
could switch between resistance, capacity and
AC/DC current。
3.6 LCD backlight button
4. Knob Switch: It could be used to change the measurement
function and range.
5. Input Terminals
5.1. 10A “+”input terminal.
5.2. mA “+” input, Temperature “+” Input, Capacity and HFE
Frequency, Buzzer, Temperature “-”, HFE, Input terminal.
Other Functions
Auto Power off
During measurement, the meter will automatically shut down
(enter sleeping mode) to save power if function buttons and knob
switch are not operated in 15 minutes. In auto power off mode,
press any function buttons or rotate the knob switch, the
instrument will get into the auto power on mode (working mode)
Input Impedance: 10MΩ .
Overload Protection: 1000V DC or 750 AC peak value.
Frequency Response: (50~200) Hz
Display: Average value response (RMS of sine wave).
2-2-3. DCA
A) Insert the black test lead into the “COM” input terminal and
red test lead into the “mA “ input terminal. (Max 400mA), or
“10A “input terminal (Max 10A).
B) Rotate function switch to Current setting. The initial state of
the meter is in automatic range status, which shows “DC”
symbol. Then connect the test lead to the tested circuit in
serial, the tested current value and the current polarity of the
point where the red pen is contacted will be displayed on the
screen simultaneously.
Caution:
1. If “OL” is displayed on LCD, it indicates the tested current
value has exceeded the present range limit, please select
higher range to complete the measurement.
2. The Max input value is 400mA or 10A. (Depending on the
terminal where the red test lead is contacted)
Range Accuracy Resolution
400uA 0.1uA
4000uA 1uA
40 mA 10uA
400mA
4A ±(1.2%+10d) 10mA
10A ±(1.2%+10d) 10mA
Max measurement voltage drop: Full Range mA is 0.4V, A is
100mV;
Max input current: 10A (less than 15 seconds);
Overload Protection: 0.4A/250V resettable fuse, 10A/250V
fuse.
±(1.0%+10d)
100uA
8
2-2-4. ACA
A) Insert the black test lead into the “COM” input terminal and
red test lead into the “mA “input terminal. (Max 400mA), or
“10A” input terminal (Max 10A).
B) Rotate function switch to Current setting. Press “SELECT”
button to select the AC measurement mode. Then connect
the test lead to the tested circuit in serial, the tested current
value and the current polarity of the point where the red test
lead is contacted will be displayed on the screen
simultaneously.
Caution:
1. If “OL” is displayed on LCD, it indicates the tested current value
has exceeded the present range limit, please select higher
range to complete the measurement.
2. The Max input value is 400mA or 10A. (Depending on the
terminal where the red test lead is contacted) The overrated
current will lead to fuse melt or even damage the meter.
A) Insert the black test lead into “COM” terminal and red test
lead into “
B) Rotate the Range to “Ω” setting. the initial state of the meter
is in automatic range status, which shows “AUTO” symbol
C) Connect the two test leads to the tested resistor.
D) When measuring the low resistance, please short-circuit the
test leads at first to test the wire resistance, and then deduct
it from the actual resistance.
Caution:
1. If “OL” is displayed on LCD, it indicates the tested resistance
value has exceeded the present range limit, please select
higher range to complete the measurement. When measuring
the Resistor higher than 1MΩ, the instrument will take several
seconds to make the reading stable. It is normal when
measuring the high resistor.
2. When the input terminal is open circuit, it will display “OL”.
3. When measuring in-line resistor, be sure that the power is off
• Overload Protection: 250V DC or AC Peak Value;
Note: When measuring at Range 400Ω, please short-circuit the
test leads at first to test the wire resistance, and then deduct
it from the actual resistance.
“terminal.
±(0.8%+4d)
1kΩ
10
WARNING: DO NOT input any voltage at resistance range
for safety!
2-2-6.Diode and Continuity Test
A) Insert the black test lead to “COM” terminal and the red test
lead to “
B) Rotate the Range to “
to select the Diode measurement mode.
C) Forward Measurement: Connect the red test lead to the
diode positive polarity and the black test lead to the diode
negative polarity. The approximate value of diode forward
voltage drop will show on the display.
4. Backward Measurement: Connect the red test lead to the
diode negative polarity and the black test lead to the diode
positive polarity. “OL” symbol will be displayed on the
screen.
5. The complete diode testing includes forward and backward
measurement, if the result does not meet the above; it
means the diode is bad.
6. Press “SELECT” button to select the Continuity
measurement mode.
7. Connect the test leads to two points of the tested circuit. If
the built-in buzzer sounds, the resistance between the two
points is less than 50Ω.
Range
” terminal. (The polarity of red test lead is“+”).
“setting. Press “SELECT” button
Display Test Condition
Forward Voltage
Drop of Diode
Buzzer makes
a long sound if
resistance is less
than 50Ω
Forward DC Current is
Approx. 0.5mA, Backward
Voltage is Approx. 1.5V
Open circuit voltage is
Approx. 0.5V
11
• Overload Protection: 250V DC or AC Peak Value.
WARNING: DO NOT input any voltage at this range for
safety!
2-2-7. Capacity (F)
A) Rotate function switch to “
B) Or rotate the Range to “Ω“setting. Press “SELECT” button to
select the capacity measurement mode.
C) Insert the black test lead to “COM” terminal and red test lead
to “mA“ terminal;
D) Connect the tested capacity by the test leads to “COM”, “mA
“ input terminal, the screen will show capacitance parameter.
(The relative value measurement could be conducted by
pressing “REL” button.)
Caution:
1. Fully discharge the tested capacitor in case it damages the
meter.
2. When measuring in-line capacitor, the power should be turned
off and all capacitors should be discharged completely.
3. It takes about 30 seconds to input stable reading at 200uF
WARNING: DO NOT input any voltage at this range for
“setting.
safety!
12
2-2-8. Frequency (Hz)
A) Test volt frequency, connect test leads and shielded cable
respectively to “COM” terminal and “
B) Rotate function switch to “
and the cable to the signal source or the tested load. The
tested signal will show on the screen.
Caution:
1. When inputting AC RMS over 10V, it could show reading, but
excess vibration may appear.
2. It is recommended to test weak signals by shielded cable
under noisy circumstances.
3. Select ACV gear when testing the frequency of high voltage.
Then press “Hz/%” button to enter frequency measurement
status.
4. Don’t input voltage of over 250V DC or AC peak value in case
it damages the meter.
Range Accuracy Resolution
400Hz 0.1Hz
4kHz 1Hz
20kHz
0.1-99.9% For your reference0.1V
• Input Sensitivity: >0.7V RMS
• Overload Protection: 250V DC or AC Peak Value.
±(0.5%+4d)
“ gear. Connect test leads
” terminal.
10Hz
2-2-9. Temperature (˚C)
A) Rotate function switch to (˚C) gear.
B) Insert the cathode (black pin) of cold end (free end) of
thermocouple into “COM” jack and anode into “mA” terminal.
Then put the working end (temperature measurement end)
of thermocouple on the surface or inside the object to be
tested. Then you can read temperature from the screen, and
the data is in Centigrade.
13
Caution:
1. When the input terminal is open-circuit, it will display the “OL”
symbol on the screen.
2. Don’t change the temperature probe at random, or the value
accuracy could not be guaranteed.
3. Don’t measure voltage at temperature range.
Range Accuracy Resolution
(-20~1000)˚C<400˚C±(1.0%+5d)
1˚C
≥400˚C±(1.5%+15d)
• Sensor: K Type Thermocouple (Nickel-chromium--nickel
silicon) (banana plug).
WARNING: DO NOT input any voltage at this range for
safety!
2-2-10. Triode hFE test
A) Turn the knob switch to “HFE“Range
B) Insert the test accessory into “COM” and “mA” terminal.
C)Verify the type of the transistor is NPN or PNP, insert the
emitter, basic and collector to the proper jack on test accessory.
Range Displaying value Test condition
hFE NPN or PNP 0~1000 Basic current is
approx.10uA,Vce
is approx.3V
WARNING: DO NOT input any voltage at this range for
safety!
14
Instrument Maintenance
This is a precision instrument and the user shall not modify the
electric circuit as well.
1. Keep the instrument away from water, dust and shock.
2. Do not store and operate the meter under the condition of high
temperature, high humidity, combustible, explosive and strong
magnetic place.
3. Wipe the case with a damp cloth and detergent; do not use
abrasives and alcohol.
4. If the instrument is not operated for a long time, please take
out the battery to avoid leakage.
5. Pay attention to the status of the 9v battery. When the LCD
displays a flashing “
Battery changes steps as below:
5-1. Loosen the screw on the back cover that secures the battery
case and remove the battery case.
5-2. Remove the 9V battery and replace them with new one.
Although a 9V battery of any standard can be used, but in
order to lengthen the operation life, alkaline battery should
be used.
5-3. Mount the battery case back and tighten the screw.
Precaution:
1. Don’t input voltage higher than DC 1000V or AC Peak Value.
2. Don’t measure voltage at current, resistance, diode and buzzer
range.
3. Don’t use the instrument when the battery has not been
mounted properly or the back cover ahs not been tightened.
4. Prior to the replacement of battery or fuse, please remove the
test leads from the measuring point and switch off the meter.
”symbol, the battery shall be replaced.
15
Fault Elimination
If the instrument could not work properly, please try the following
tips to solve some general problems. If the problems still exist,
please contact the maintenance center or the distributor.
Fault
No Display
symbol appearance Replace battery.
No current input Replace fuse
Error value Replace battery
This Instruction is subjected to change without any further
notice.
The content of this Instruction is considered correct, and in
case readers find any errors and missing parts, please contact
the manufacturer.
The Company shall not be held liable for any accidents and
hazards resulted from the mal-operations by the user.
The function elaborated by this Instruction shall not be taken
as the reasons for using the product for special purposes.