![](/html/6c/6cf3/6cf3de4af0101a517c365cca8e07490e5dad36fcd6d71ab961ca40fc32cd1cab/bg2.png)
PERFORMANCE SPECIFICATION
MODEL 3092C
HIGH TEMPERATURE, CHARGE MODE ACCELEROMETER
FEATURES:
• HIGH TEMPERATURE 10-32 COAXIAL CONNECTO
• HIGH TEMPERATURE RANG
Units Minimum Typical Maximum
PHYSICA
Weight grams 42
Mounting #10-32
Connector Type 10-32 coaxial
Base Material 17-4 PH
Isolation case ground
PERFORMANC
Sensitivity pC/g 3.5
Frequency Range, ±10% Hz see note 2 5000
Resonance Frequency kHz >40
Transverse Sensitivity % 3
Capacitance pF 80
Non-linearity % 1
Insulation Resistance (Room Temp.) GΩ 300
Insulation Resistance (1000°F) MΩ 1
Insulation Resistance (1200°F) kΩ 75
Actual Size
ENVIRONMENTA
Shock/Vibration g pk 5000/3000
Operating Temperature °F -60 see note 1
Seal hermetic
Coefficient of Thermal Sensitivity %/°F 0.01
Magnetic Sensitivity g/G 0.08
Base Strain Sensitivity g/µε 0.06
ACCESSORIES SUPPLIE
none
NOTES
1.) MAXIMUM TEMPERATURE OF OPERATION IS DETERMINED BY THE ABILITY OF THE CHARGE AMPLIFIER TO ACCEPT
LOW INSULATION RESISTANCE OF THE SENSOR (SEE GRAPH)
2.) THE LOW FREQUENCY RESPONSE IS DETERMINED BY THE CHARGE AMPLIFIER'S TIME CONSTANT USED IN THE
APPLICATION.
3.) EACH INSTRUMENT IS SUPPLIED WITH AN ACCREDITED CALIBRATION CERTIFICATE (ISO 17025 BY A2LA), CALIBRATED
AT THE RANGE OF 20Hz TO 10kHz AT ROOM TEMPERATURE.
75 300000000
300GΩ
700 100000000
TYPICALINSULATIONRESISTANCE
1000 1000000
12000 75000
3GΩ
INSULATION RESISTANCE
50kΩ
75 750 1200
REV A ECN 5482 PS3092C ENGLISH VERSION
REV A ECN 5482 PS3092C ENGLISH VERSION
TEMPERATURE (°F)