DYTRAN 3053B DATA SHEET

S/N
3053B
DYTRAN
.50
CONNECTOR MATES WITH DYTRAN CABLE 6430AXX
2
USA
MOUNTING SURFACE
.50
.35
1
3
A
4
1
3
2
A
VIEW A-A LOOKING
INTO CONNECTOR
PINOUT
PIN #1 SIG/PWR AXIS 1 (X) PIN #2 SIG/PWR AXIS 2 (Y) PIN #3 SIG/PWR AXIS 3 (Z) PIN #4 GND RETURN, COMMON
4 WEIGHT: 6.0 GRAMS. 3 XX=LENGTH IN FEET.
2. + SIGNS INDICATE SENSE AND DIRECTION OF ACCELERATION FOR POSITIVE OUTPUT.
1. HOUSING MATERIAL: TITANIUM ALLOY
ALL DIMENSI0NS IN INCHES TOLERANCE: .XXX = ± .XX= ±
SURFACE FINISH EXCEPT AS NOTED
BREAK EDGES TO DEBURR RADIUS OR CHAMFER
THESE DIAS TO T.I.R.
FILLETS - MAX RAD.
.005 .01
INSTRUMENTS, INC.
SCALE
DATE
DRAWN
APPROVED
TITLE
2X
8/11/00
N.C.
OUTLINE/INSTALLATION DRAWING,
REV
PART NO.
MAT'LCHECKED
R.A.
NEXT ASSEMBLY USED ON
DATE
-
-
MODEL 3053B
MODEL 3053B
ECN
CHATSWORTH, CA.
-
3053B
DWG NO.
127-3053B
SHEET
OF
11
INSTRUMENTS,
INC.
SPECIFICATIONS, MODEL 3053B TRIAXIAL ACCELEROMETER
SPECIFICATIONS VALUE UNITS
PHYSICAL
WEIGHT 6.0 GRAMS SIZE (HEIGHT x WIDTH x DEPTH) 0.35 x .50 x .50 INCHES MOUNTING [1] ADHESIVE MOUNT CONNECTOR 4-PIN [2] MATERIAL, HOUSING/CONNECTOR TITANIUM ALLOY
PERFORMANCE
SENSITIVITY, -10 +15% [3] 10.0 mV/G RANGE, F.S. (each axis) +/- 500 G FREQUENCY RESPONSE, +/- 10% 2 to 5000 Hz ELEMENT NATURAL FREQUENCY, NOM. 30 kHz EQUIVALENT ELECTRICAL NOISE .007 G, RMS LINEARITY [4] 1 %F.S. TRANSVERSE SENSITIVITY, MAX, 5 % SIGNAL POLARITY POSITIVE FOR MOTION IN DIRECTION OF ARROWS ETCHED ON HOUSING
ENVIRONMENTAL
MAXIMUM VIBRATION +/- 600 G MAXIMUM SHOCK 5000 G TEMPERATURE RANGE -60 to +250 ENVIRONMENTAL SEAL HERMETIC COEFFICIENT OF THERMAL SENSITIVITY .03 %/oF
o
F
ENVIRONMENTAL
SUPPLY CURRENT RANGE, (each axis) [5] 2-to 20 mA COMPLIANCE (SUPPLY) VOLTAGE RANGE (each axis) +18 to +30 VDC OUTPUT IMPEDANCE, TYP. 100 OHMS OUTPUT BIAS VOLTAGE, NOM. +10 VDC DISCHARGE TIME CONSTANT, NOM. 0.5 SEC GROUND ISOLATION 10 MEGOHMS (MIN)
[1] Case ground isolation is achieved by internal means. [2] Connector mates with Dytran cable assy. Model 6430AXX. (XX = length in feet) [3] Reference sensitivity measured at 100 Hz, 1 G RMS per ISA RP 37.2 [4] Linearity is % of specified full scale (or any lesser full scale range), zero-based best fit straight line method. [5] Power only with Dytran LIVM power unit or other Dytran-compatible constant current type power unit. If power is applied without current limiting protection, the internal amplifier will be immediately destroyed.
2
Loading...