DYRAN 3263AT DATA SHEET

SPECIFICATIONS, MODEL 3263AT TRIAXIAL ACCELEROMETER WITH TEDS
SPECIFICATIONS UNITS PHYSICAL
SPECIFICATIONS 3263A1T 3263A2T 3263A3T UNITS PERFORMANCE
SENSITIVITY, ±10% [2] 10 100 50 mV/g RANGE, F.S. (each axis) ±500 ±50 100 g FREQUENCY RESPONSE, -5 / +15% 1.5 to 10000 1.5 to 10000 1.5 to 10000 Hz ELEMENT NATURAL FREQUENCY, NOM. 40 40 40 kHz ELECTRICAL NOISE FLOOR (25Hz-25kHz) .0008 .0006 .0008 grms
LINEARITY, MAX [3] 1 1 1 % FS TRANSVERSE SENSITIVITY, MAX 5 5 5 % SIGNAL POLARITY POSITIVE FOR MOTION IN DIRECTION OF ARROWS ON HOUSING
(1Hz-10kHz) .001 .0004 .001 grms
ENVIRONMENTAL
MAXIMUM VIBRATION ± 600 gpk MAXIMUM SHOCK 5000 gpk TEMPERATURE RANGE -60 to +180 °F ENVIRONMENTAL SEAL HERMETIC COEFFICIENT OF THERMAL SENSITIVITY .03 %/°F
ELECTRICAL
SUPPLY CURRENT RANGE, (each axis) [4] 2 to 20 mA COMPLIANCE (SUPPLY) VOLTAGE RANGE (each axis) +18 to +30 VDC OUTPUT IMPEDANCE, TYP. 100 150 150 Ohms OUTPUT BIAS VOLTAGE, NOM. +11 +11 +11 VDC DISCHARGE TIME CONSTANT, NOM. .7 0.7 0.7 sec GROUND ISOLATION CASE GROUNDED IEPE SENSOR WITH TEDS FEATURE PER IEEE 1451.4
[1] Connector mates with Dytran cable assy. Model 6811Axx. (xx = length in feet) [2] Reference sensitivity measured at 100 Hz, 1 grms per ISA RP 37.2 [3] Linearity is % of specified full scale (or any lesser full scale range), zero-based best fit straight line method. [4] Power only with IEPE power unit or other Dytran-compatible constant current type power unit. If power is applied without current limiting protection, the internal amplifier will be immediately destroyed.