Read this manual before working with these products. For personal and system safety, and for the
optimum product performance, make sure you thoroughly understand the contents before using or
servicing this product.
For technical assistance from the factory please contact:
DYNISCO INSTRUMENTS
38 Forge Parkway
Franklin, MA 02038
Telephone: (508) 541-9400 or 1-800-221-2201
Facsimile: (508) 541-9436
Internet Address: www.dynisco.com
E-mail: salesinst@dynisco.com
1.INTRODUCTION
The TX180 is a transmitter that accommodates any one of eleven types of thermocouples, six types
of RTD’s, millivolt or ohm inputs. The unit is precision linearized to the measured temperature over
the entire usable range of the selected sensor. This transmitter is simple to set up and operates much
like high performance analog transmitters.
The TX180 also has numerous advanced features that are achieved through the use of digital signal
processing and micro-controller technologies. Typical of these features are the precision
linearization, the independent zero and full scale settings, digital filtering, etc. Other advanced
features, such as the automatic self diagnostics, and the exceptional stability are transparent to the
user and are continuously active.
The TX180 transmitter can also accept one of two optional plug-in displays. The LI-1 is an
inexpensive, single line display that is intended to give a low-cost, local indication of the measured
temperature. The LI-2 two line display will give a local indication and functions as a very easy-to-use
set-up tool. Both displays facilitate local configuration and ranging of the transmitter.
This manual is divided into several sections. After a brief INTRODUCTION, the section on
UNPACKING AND INSTALLATION contains much useful information for the first time installer. The
section called IN A HURRY? helps get the system operating provided the sensor and transmitter
were purchased at the same time and thus most of the set up was completed at the factory. The next
two sections explain the method of CONFIGURATION using either display. Finally, there is
additional APPLICATION INFORMATION and the TECHNICAL SPECIFICATIONS included in the
Page 6
6
sections under those headings.
The TX180 temperature transmitter does not have any potentiometers or switches to set and there are
no user serviceable components inside the transmitter. Opening the enclosure will void the
manufacturer’s warranty. All reconfiguration, re-ranging and “calibration” can be done in the field
using either one of the displays. Any of the communication methods provides reconfiguration and
re-ranging capabilities without other external calibration tools.
2.UNPACKINGAND INSTALLATION
2.1UNPACKING
Remove the Packing List to check off the actual equipment received. If you have any questions on
your shipment, please call DYNISCO Customer Service at (800) 332-2245. Upon receipt of
shipment, inspect the container for any signs of damage in transit. Especially take note of any
evidence of rough handling. Report any apparent damage immediately to the shipping agent.
NOTE: The carrier will not honor any claims unless all shipping material is saved for their
examination. After examining and removing the contents, save the packing material and
carton in the event reshipment is necessary.
2.2MECHANICAL INSTALLATION
Proper installation of the transmitter will assure highest performance and minimize errors of the
measured variable. The transmitter should be mounted in a location that minimizes temperature
extremes, vibration and shock. It is important to survey the area to ascertain the best location for
installation. Will the location be subjected to flooding? Is the location directly above, below or in
proximity to a known high heat source? Does the location make the transmitter unserviceable?
The installation recommendations outlined in this section are provided to act as a guideline only
and cannot cover all possible variations. The final installation must be made at the discretion and
approval of the user.
2.2.1 WEATHER PROOF/EXPLOSION PROOF HOUSING
Several optional transmitter housings are available. The XP-FN is an Explosion Proof / Weatherproof
housing that accommodates a transmitter when the local display option is not required. The XP-FG,
with its glass window, may be used in hazardous and wet locations when the display option is
desired. The XP-HEAD is a small electrical head that can be used without local indication in
Explosion Proof environments. The WP-HEAD is a small aluminum Weatherproof head designed for
head mounting the transmitter without local indication. These housings have appropriate mounting
means in the bottom to attach the TX180 in any of four orientations 90° apart. Captive 8-32 machine
screws are installed on the transmitter to facilitate installation and removal to either a housing
bottom plate or to a mounting panel.
Page 7
TX180 Temperature Transmitter7
Please note that condensation often occurs inside conduit attached to Weather Proof or Explosion
Proof housings. Care must be taken so that liquid condensation does not accumulate and fill the
transmitter housing with liquid. While the transmitter is sealed, we do not recommend operating it
immersed in liquid. Conductive liquids across the top of the transmitter will short the input and loop
terminals. This installation problem can appear to a control system as a transmitter failure.
2.2.2 MOUNTING
The TX180 transmitter may be mounted on a 2 inch pipe (vertical or horizontal), on a bulkhead, a
panel, a DIN rail or other rigid support members utilizing the various mounting brackets and
associated hardware available from Dynsico. These types of mounts provide greater flexibility in
installation and removal of transmitter for service. In locations where extreme temperature variations
are encountered, it is strongly recommended that enclosures be provided to maintain a somewhat
constant temperature at the transmitter. Heaters or steam tracing should be provided if the ambient
temperature variations are extreme.
Fig. 1Optional Weather-Proof Housing
2.2.2.1 DIN RAIL MOUNTING
A bracket is available if DIN rail mounting of the TX180 is desired. See Figure 2.
Page 8
8
Fig. 2Optional DIN Rail Mounting Bracket
2.2.2.2 HEAD MOUNTING
For head mounting, all the Explosion Proof / Weatherproof housings can be used for head mounting.
All heads have two 1/2" female NPT conduit entries. One of these conduit entries can be used to
mount directly onto a 1/2" male NPT extension of sensor. Alternatively, a 1/2" NPT union coupling
can be placed between the weatherproof housing and the temperature sensor. These heads are
shown in Figure 3. Note that the XP-HEAD and WP-HEAD cannot be used with a pipe mount
bracket. For non-display pipe mount bracket installations, use the XP-FN Housing.
Fig. 3The XP-FG with Bracket and WP-HEAD
2.2.2.3 SURFACE MOUNTING
The transmitter has two mounting holes through the body of the transmitter. These mounting holes
allow the transmitters to be attached to any flat surface by means of two bolts or screws. The
transmitter is provided with 8/32 captive screws already installed.
Page 9
TX180 Temperature Transmitter9
2.2.2.4 PIPE MOUNTING
A stainless steel bracket is available for pipe mounting. Use the PY-2 for mounting the XP-FN or XPFG housing onto any 2" pipe. Note that the XP-HEAD and WP-HEAD are not suitable for Pipe
Mounting. The housing is attached to the bracket as shown in Figure 3.
2.3ELECTRICAL INSTALLATION
The TX180 has two groups of terminals. One terminal group is for the sensor input. The second
terminal group is for transmitter output. The terminals labeled “+” and “-” are the 4 to 20mA output
terminals. These are normally connected to the corresponding polarity terminals of the power supply
of the current loop. Refer to Figure 4 for the arrangement of the output terminal connections.
Terminals labeled “1, 2, 3 and 4” are used in various connections to accommodate the different
sensor inputs. Refer to Figure 5 for the arrangement of the input terminal connections.
2.3.1 OUTPUT TERMINALS
The output terminals, labeled “+” and “-”, are generally connected to a power supply having a
nominal 24 Volt DC voltage and capable of supplying 23mA for the TX180. The “+” and “-”
terminals of the transmitter are connected to the corresponding polarity terminals of the power
supply.
Fig. 4Output Terminal Connections
A load resistor, typically 250 ohms, may be connected in series with either terminal of the
transmitter. For Digital communications, 250 ohms must be connected in the loop. The maximum
series resistance in the circuit (including wiring lead resistance) can be calculated using the formula:
Vs - 12
Rs = ––––––––––
0.023
Page 10
10
The following chart gives maximum series resistance:
Max. Series Resistance, RsSupply Voltage, Vs
1300 ohms42.0 Volts
520 ohms24.0 Volts
417 ohms21.6 Volts
250 ohms18.0 Volts
0 ohms12.0 Volts
2.3.2 INPUT TERMINALS
See Figure 5 for sensor input connections. Be certain to include the proper jumpers for
thermocouple sensors and for two or three wire RTD inputs. Any sensor other than the four-wire
RTD requires at least one external jumper. A jumper is supplied with the unit and is attached to
terminals 3 & 4.
Fig. 5Input Terminal Connections
2.3.2.1 MILLIVOLTAND THERMOCOUPLE INPUT
Apply signal to terminals 1 and 2. Terminal 1 is the negative and terminal 2 is the positive. Terminals
3 and 4 must be jumpered together for proper operation as well as to prevent any build-up of
electrostatic charge on these terminals which could affect the transmitter readings.
2.3.2.2 TWO-WIRE RTD INPUT
Apply signal to terminals 1 and 3. Jumpers must be installed on terminals 1 and 2 as well as on 3
and 4 for proper operation and to prevent any build-up of electrostatic charge on these terminals
which could affect the transmitter readings.
Page 11
TX180 Temperature Transmitter11
2.3.2.3 THREE-WIRE RTD INPUT
Apply the common legs from the RTD (generally the same color RTD leads) to terminals 1 and 2.
Apply the other signal lead to terminal 3. Terminals 3 and 4 must be jumpered together for proper
operation and to prevent any build-up of electrostatic charge on these terminals which could affect
the transmitter readings.
2.3.2.4 FOUR-WIRE RTD INPUT
Apply one set of the common legs from the RTD (generally the same color RTD leads) to terminals 1
and 2. Apply the other signal lead pair to terminals 3 and 4. No jumpers are necessary for a 4 wire
RTD input.
3.TRANSMITTER OPERATION
3.1INAHURRY?
When in a hurry, this short set of instructions and references will help get the transmitter running.
3.1.1 FACTORY CONFIGURATION
Input=Type J Thermocouple
Output=Analog
4.00mA=40°F
20.00mA=200°F
Sensor Fail-safe=23.00mA (High)
On special request the factory will set the transmitter to any desired configuration. Special
configurations are identified on a tag attached to the unit.
3.1.2 OPERATION WITHOUTA DISPLAY
If the unit was ordered with the standard factory configuration, the sensor required is a Type J
thermocouple. The packing slip and a tag on the unit will indicate if the unit was set up to any other
customer requested special configuration. If there is a need to change the configuration of the
transmitter, or to re-range it, use either the LI-1 or LI-2 Display and refer to the procedures described
in SECTIONS 4 (for LI-2), 5 (for LI-1).
NOTE: Even when “In a Hurry”, the use of an appropriate power supply is important. A 24V DC
supply having a current handling capacity of at least 0.1A is commonly used. Always use a
DC (direct current) supply, or suitable size battery. NEVER CONNECT THE TRANSMITTER
DIRECTLY TO 115VAC.
With the power supply off, connect the + side of the power supply to the + terminal of the
Page 12
12
transmitter. Connect the - side of the power supply to the - terminal of the transmitter.
Connect a Type J thermocouple to the transmitter input.
Thermocouple high (+) (input terminal 2)
Thermocouple low (-) (input terminal 1)
Jumper terminals 3 & 4 together
Unlike conventional electrical wiring, (on a J thermocouple the red lead is negative). This should be
checked and verified with the particular sensor to be used.
To connect other sensors to the input refer to Section 2.3.2 for the proper sensor connections.
The output can be monitored by connecting a milliammeter in series with either of the two output
terminals, or by connecting a high impedance voltmeter across the optional 250 ohm resistor. Now
turn on the power supply. In about 5 seconds the TX180 loop current will settle to its normal value
in the range of 4 to 20mA, unless the input terminals are open, in which case the output current will
be 23.00mA. Note that for a Type J thermocouple, if 4mA = 40°F and 20mA = 200ºF, each
additional 10°F increases the current by 1.0mA.
3.1.3 OPERATION WITHA DISPLAY
If the transmitter was ordered with either display option, it will have a small local LCD display
module (with two integral buttons) plugged in to the top of the unit. Either display option can be
ordered already installed on the TX180 transmitter. Alternately, either display can be ordered and
field installed at any time.
Having the display option as part of the transmitter does not affect its operation in the analog mode
and the description of the previous section applies. However, the display option does provide some
very useful local indication of the measured temperature and other diagnostic functions. Figure 6
below indicates the arrangement of the display screen.
Fig. 6Local Displays, LI-1 and LI-2
Page 13
TX180 Temperature Transmitter13
In operation, the LI-1 and LI-2 displays both give the process temperature.The LI-2 provides some
additional information The LI-1 displays the process temperature and a minus sign if applicable. The
temperature is displayed with a floating decimal point. For measured temperatures over 999.9 no
decimal point will be displayed. Otherwise, the LI-1 will show one tenth degree increments. Unlike
the more capable LI-2 display, the LI-1 does not show the units of measurement “C”, “F”, “R”, or
“K”. If it is necessary to display the temperature units on the LI-1, note by hand or apply a separate
label on the face of the display.
The LI-2 has more display capabilities. With the LI-2, the top display row shows the process
temperature, the units of measurement, “C”, “F”, “R”, or “K” and a minus sign if applicable. The mid
portion is an analog bar graph display showing the % of range based on the ZERO and FULL SCALE
setting of the transmitter. When power is applied the leftmost segment of the bar graph, the 0% and
the 100% become energized momentarily. If the measured temperature is below what the ZERO is
set to (below LRV), then the left arrow is energized. If the measured temperature is above the FULL
SCALE setting (above URV), then the right arrow becomes energized. The bottom portion of the LI-2
display is capable of displaying an alphanumeric message up to 7 characters long. In normal
operation this row shows a label, which is factory set to display “TX180”.
Note that the process temperature displayed on the LI-1 and LI-2 is the actual temperature as
measured by the transmitter, it is not affected by the analog output range settings. This is particularly
useful in startup or operation where the measured temperature is temporarily outside the normal
operating range.
When the unit is first turned on, the display will show the measured temperature. It is frequently the
case that no sensor is connected when the transmitter is first turned on. In this case, the display will
show a sensor failure. In the event of a sensor or transmitter failure, the indication on the L-1 display
changes to read:
FAILSAFE
The words “FAIL” and “SAFE” will alternate in the display window to let you know that a failure
condition has occurred.
In the event of a sensor failure, the indication on the bottom line of the LI-2 display changes to:
<0%100%>
FAIL
The words “FAIL” and “SAFE” will alternate in the display window to let you know that a failure
condition has occurred. The Percent of Output Bar Graph will indicate the output level of the
transmitter. If the transmitter Failsafe Report value is set to “Fail High” (23mA), the display will be as
shown, at over 100% of output. If the Failsafe Report is set to “Fail Low” (3.8mA), the Percent of
Output Bar Graph would indicate the output level at under 0% of output. See sections 4.8, 5.8 or
<0%100%>
SAFE
Page 14
14
6.2 for further information on setting Failsafe Reporting.
Once the proper sensor is connected the fault message on the display should clear and the
transmitter output should go to the proper value.
Both LCD displays take full advantage of the precision of these transmitters. The digital display of
measurement does not include the small D/A error otherwise present in the analog output. It
provides highly accurate local indication of the measurement, local fault diagnostics, and transmitter
identification. The LCD continues to display the measured temperature even if it is beyond the zero
and full scale limits set for the analog output.
If you should desire to change the sensor input or to re-range or reconfigure the transmitter, please
refer to Sections 4, 5 or 6 of this manual, which show you how to set-up the transmitter with the LI-1
or LI-2 displays.
4.CONFIGURATION USINGTHE LI-2, TWO-LINE DISPLAY
To configure a transmitter using the DISPLAY MODE, either the LI-1 or LI-2 local LCD display is
required. These displays are available as an option and can be plugged into the top of the TX180
transmitter. The transmitter can also be purchased with these options already installed. These
inexpensive options make the reconfiguration, or re-ranging of the transmitter very simple and easy
to follow. Without the use of a calibrator, or any other tools, the transmitter can be set up for a
different sensor, or the new range limits can be set much like one would set the time on a digital
watch.
In the event that the LI-1 or LI-2 Display / Keyboard are not purchased at the same time as the
transmitter, the one piece display design allows for easy field installation by simply plugging the LI-1
or LI-2 into the top of the transmitter.
4.1ENTERINGTHE DISPLAY MODE
To start the DISPLAY MODE, first connect the transmitter to an appropriate DC power supply.
Typically a 24VDC supply is connected with the + side of the power supply connected to the
transmitter’s output “+” terminal and - side of the power supply connected to the transmitter’s output
“-” terminal. A sensor may be connected to the transmitter’s input terminals, but this is not required
for setting up the transmitter.
With the standard factory set-up and no sensor connected, the LI-2 display will give the following
indication:
<0%100%>
FAIL
<0%100%>
SAFE
Page 15
TX180 Temperature Transmitter15
The transmitter is indicating FAILSAFE, since no sensor is connected, and the analog output is
indicating greater than 100%, loop current at 23.00mA, which is the standard Failsafe report
condition. Please note that the display / keyboards can be plugged into the transmitter while the
transmitter is powered up. There is no need to disconnect power before plugging the LI-1 or LI-2 into
the TX180.
Press the key marked NEXT. The display starts to alternate asking if the user wishes to return to the
Operate Mode?
RETURNTOOPERATEMODE?
To activate the NEXT and ENTER; keys a slow, deliberate push of the key is required. This prevents
any casual, inadvertent activation of the transmitter into one of the configuration modes.
The answer would be “No”, therefore, press the NEXT key. This will enter you into the DISPLAYMODE configuration menu. If you wish to answer a question “Yes”, press the ENTER key. A flow
chart summarizing the operation of the DISPLAY MODE appears at the end of this manual.
Note that when more than seven characters are required to describe a function, the display keeps
sequencing through two or more screens or may use common abbreviations. In this manual, the
sequencing of the display is indicated by placing the two or more parts of the message adjacently.
With some functions, the LI-2 display indicates a numeric value and unit of measurement on the top
line of the display in addition to the message on the lower display line.
4.2DISPLAY MODE CONFIGURATION
The DISPLAY MODE will allow the user to do the following:
•Select a Sensor Input (Select Input)
•Select a desired temperature unit, such as ºF (Select Units)
•Change the 4mA Lower Range Value (Change Zero)
•Change the 20mA Full Scale Value (Change Full Scale)
•Change the Sensor Fail Safe detection (Select Sensor Fail Safe)
•Change the Fail Safe reporting (Select Fail Safe Report)
•Trim the 4.0mA output current (Trim 4 MA)
•Trim the 20.0mA output current (Trim 20 MA)
•Trim the display value (Trim Display)
•Change the Language of the display
Each of these functions is presented in sequence on the LCD display. If the indicated function need
not be performed, press Next, and the next function is displayed on the screen. To perform any
function press the Enter key. This will cause additional screens to be displayed which enable you to
perform the function. These are described in detail below and summarized on the LI-2 Two-Line
Display / Keyboard Flow Chart found in the rear of this booklet.
Page 16
16
4.3SELECTA SENSOR INPUT
The SELECT SENSOR is the first function in the sequence. Virtually any thermocouple, RTD or
millivolt input can be selected. The display will read as follows to indicate this position on the
menu:
SELECTINPUT
If the sensor is set correctly, press NEXT and skip to Section 4.4 of this manual; otherwise press
Enter. After pressing the ENTER key, the display will change to:
T/C J
Indicating that the transmitter is set to a Type J thermocouple input. If this is the desired sensor, then
press ENTER, otherwise press NEXT repeatedly to sequence through the available sensors. Each
time NEXT is pressed, the next available sensor selection is displayed.
T/C J
Press the NEXT key to go the next sensor.
T/C K
Press the NEXT key to continue through the different sensor selections.
T/C L
T/C N
T/C R
T/C S
T/C T
T/C U
T/C SPEC
NOTE: The T/C SPEC or Special Thermocouple input is reserved for a special thermocouple input,
Page 17
TX180 Temperature Transmitter17
should one be desired. This Special must be ordered from the factory.
2W OHMS
2W DINP
NOTE: This is the 100⍀ Platinum DIN Curve with ␣ = 0.00385.
2W SAMP
NOTE: This is the 100 ⍀ SAMA Platinum Curve, known variously as the SAMA RC21-4 or SAMA
PR-279. Constants are 98.13⍀ @ 0°C, ␣ = 0.003923.
2W SPEC
NOTE: The 2W SPEC or Special 2 wire RTD input is reserved for a special RTD input, should one
be desired. Any special 2-wire RTD curve must be ordered from the factory.
3W OHMS
3W DINP
NOTE: This is the 100⍀ Platinum DIN Curve with ␣ = 0.00385.
3W SAMP
NOTE: This is the 100 ⍀ SAMA Platinum Curve, known variously as the SAMA RC21-4 or SAMA
PR-279. Constants are 98.13⍀ @ 0°C, ␣ = 0.003923.
3W SPEC
NOTE: The 3W SPEC or Special 3 wire RTD input is reserved for a special RTD input, should one
be desired. Any special 3-wire RTD curve must be ordered from the factory.
4W OHMS
4W DINP
Page 18
18
NOTE: This is the 100 ⍀ Platinum DIN with ␣ = 0.00385. This sensor will give superior
measurement results in most real-world situations where the measured temperature is under
1,000°F.
4W SAMP
NOTE: This is the 100 ⍀ SAMA Platinum Curve, known variously as the SAMA RC21-4 or SAMA
PR-279. Constants are 98.13⍀ @ 0°C, with ␣ = 0.00385.
4W SPEC
NOTE: The 4W SPEC or Special 4 wire RTD input is reserved for a special RTD input, should one
be desired. Any special 4-wire RTD must be ordered from the factory.
MV
HHTONLY
NOTE: The HHTONLY is for a Hand-Held set-up. This is used for Factory set-up only.
T/C B
T/C C
T/C E
Pressing NEXT key again returns you to the J thermocouple selection. Repeated pressing of NEXT
key will again cycle you through the input selection submenu. You can stop at any one of the
thermocouple, RTD or mV selections by pressing the ENTER key. This action changes the
transmitter mode to that sensor. If no sensor change is desired, then, without sequencing through the
various sensor options, but jus pressing the ENTER key will allow one to confirm the sensor
selection and leave it unchanged. Assume that the sensor is left as T/C J. After pressing ENTER the
display will return to the main menu entry of SELECT INPUT. Pressing the NEXT key then takes
the transmitter to the next main menu selection.
4.4SELECT UNITS
If the selected sensor is a thermocouple or RTD, the next menu entry is SELECT UNITS.
SELECTUNITS?
Page 19
TX180 Temperature Transmitter19
Pressing the ENTER key displays the current units.
DEG F
By repeatedly pressing the Next key, the display will sequence through the following screens:
DEG R
DEG K
DEG C
These correspond to K=Kelvin, R=Rankine, C=Celsius and F=Fahrenheit. Stopping the selection at
any one of these units and then pressing ENTER will set the transmitter to the corresponding new
units. For the purposes of this example the units of measure can be left at DEG F by pressing
ENTER. Advancing the menu selection with the NEXT key lets you change the zero.
4.5CHANGE ZERO
The display will then alternate between the following screens to indicate that one may now change
the zero, or 4mA output point. The numeric value seen on the upper portion of the screen is the
ZERO value of the transmitter. One can now change this Zero, or LOWER RANGE VALUE, (LRV),
totally independent of the FULL SCALE, or UPPER RANGE VALUE, (URV), without the use of any
calibrators or external sensor inputs.
40.0°F
CHANGE
To change the zero, press ENTER. The display changes to read:
40.0°F
ZERO?
0040.0°F
PLUS?
indicating that the existing zero is set to “plus” 0040.00ºF. The question mark “?” indicates a
question asking if this value is to remain positive (PLUS ?). By repeatedly pressing the NEXT key the
display will alternate
-0040.0°F
MINUS
0040.0°F
PLUS?
Page 20
20
After deciding whether the zero value, LRV, is to remain positive (PLUS), press the ENTER key. In
this example assume it is to remain positive. The display changes to read:
0
040.0°F
THOUSN?
and the left most digit position will start blinking (shown here in italics) asking if the thousands
position needs to be changed. To change the thousands position, start pressing the NEXT key and
the left most digit will increment through 1 2 3 4 5 6 7 8 9 0. Stop pressing the NEXT key at any of
the numerals desired, then press Enter to accept the selection. If the numeral selected before
pressing ENTER was 0, then the display would change to read:
0
0
40.0°F
HUNDRD?
and the second digit from the left will start blinking (shown here in italics) asking if the hundreds
position needs to be changed. As before, to change the number in this digit position repeatedly press
the NEXT key until the desired numeral is reached. Then press ENTER to go to the next lower
significant digit position. Each time the NEXT key cycles through the ten choices for that digit
position and the ENTER key enters the selected number. The digit position being changed is the one
that is blinking. The legend on the display will change successively to read:
00
4
0.0°F
TENS?
004
0
.0°F
ONES?
0040.
0
°F
TENTHS?
After the tenth’s digit position has also been changed to the desired value, the next pressing of the
ENTER key returns the transmitter to the alternating display of CHANGE ZERO. Since changing of
the zero has just been completed, press the NEXT key to proceed to the next menu selection,
CHANGE FULL SCALE.
4.6CHANGE FULL SCALE
200.0°F
CHANGE
200.0°F
FULL
200.0°F
SCALE?
Page 21
TX180 Temperature Transmitter21
To change the full scale value press ENTER. The procedure for selecting Plus or Minus is identical
to that described for changing the zero. Similarly, the procedure for changing each of the digit
positions is identical to that described for changing the zero. Once the steps of changing the Full
Scale have been completed and the ENTER key is pressed at the end of the procedure, the display
returns to CHANGE FULL SCALE. Press NEXT for the next function SELECT SENSOR FAILSAFE DETECTION.
4.7SELECT SENSOR FAIL-SAFE DETECTION
SELECTSENSORFAILSAFE?
If you want to change the SENSOR FAIL SAFE detection press ENTER. The present status of the
Sensor Fail Safe is displayed. It is recommended that one turns off the Sensor Fail Safe System when
using the TX180 with an input simulator. It should then be turned on when reconnecting the
transmitter to the actual sensor.
ON
OFF
When the desired Fail Safe condition is displayed, pressing the ENTER key will change to the new
setting and the screen returns to the SELECT SENSOR FAIL SAFE display. Pressing the NEXT
key will bring up the Fail Safe Reporting selection screen.
4.8SELECT FAIL SAFE REPORTING
SELECTFAILSAFEREPORT?
Fail Safe reporting allows the transmitter to change the 4-20mA loop to indicate a failure condition.
This failure may be a sensor failure or a transmitter failure. In any event, the user may select to drive
the loop to 23.0mA, corresponding to the “HI” selection; or 3.6mA, corresponding to the “LO”
selection or to turn the function “OFF”.
HIGH?
OFF?
LOW?
Page 22
22
4.9TRIM 4.0MA
TRIM4 MA?
This allows trimming of the 4.00mA output current.
NOTE: This function is only for the purpose of adjusting the 4.00mA limit of the transmitter loop
current to be exactly 4.00mA according to the plant’s local standard. This is NOT for the
purpose of ranging the transmitter!
If trimming the 4.00mA limit is still desired then press ENTER. The transmitter will now output a
milliamp current equal to its internally set 4mA. This 4 mA value should be read on an external
meter and compared to a local standard. It is advisable to use a very good voltmeter to make these
comparisons. It is very possible that the transmitter will be more accurate than a great many
voltmeters. In this case, trimming will make the transmitter less accurate rather than more accurate!
Once trimming the 4.00mA value has been selected, the display will alternate as follows:
RAISEMA OUT?
By pressing the NEXT key the display then alternates as
LOWERMA OUT?
When it is decided whether to raise or lower the output current, then press ENTER and the display
changes to one of the following depending on whether the raise or lower function has been
selected.
NEXT = +NEXT = -
Now every time the NEXT key is pressed, the display blinks, and the 4.0mA output limit decreases
(-), or increases (+). The decrease or increase is in approximately 3.5 micro ampere increments.
NOTE: The 4.00mA limit is factory calibrated to a precision standard. Using the Output Trim
function voids the NIST traceability of calibration. Do not arbitrarily trim the output unless a
qualified and accurate local standard is available to measure the adjusted 4.00mA output!
Also note that the 4.0mA limit should not be trimmed by more than about 50A, or
transmitter operation may be impaired.
Once the desired trim is reached, pressing ENTER will return to one of the corresponding TRIM4mA screen. At this point one may still go back and do further trimming of the 4.0mA limit by
Page 23
TX180 Temperature Transmitter23
pressing the ENTER key, or pressing the NEXT key changes to the next function.
4.10TRIM 20.0 MA
TRIM20 MA?
Trimming of the 20.0mA current limit is done in exactly the same manner as was described for
trimming the 4.0mA point. The same precautions apply. After completing the trim 20.0mA pressing
the Next key brings up the display trim.
4.11TRIM DISPLAY
The display trim allows the display to be trimmed by a desired offset amount. The transmitter display
will display its value based upon its internal standards. It is often desirable to alter this display to
make it agree with another external instrument at a critical measurement point. If this is desirable,
the display can be trimmed. The display trim operates as a zero shift. It shifts the display readings by
the same amount at every point.
You can enter a single point offset to the display. Be certain before making a display trim correction
that you have made good electrical connections to the transmitter and the sensor. In the 2 or 3 wire
RTD input, or thermocouple input modes, it is possible to produce an error of a few degrees with a
fraction of an ohm in any one of the connections. Please be careful when tightening down the input
connections. These can be easily broken if a lot of torque is applied. The idea is to make a good
electrical connection without breaking the connections. When you press the ENTER key the display
changes to
0.0°F
TRIM
You can now enter an offset to the display. Suppose that the display reads 530°F, at a time when an
external device that you want to agree with reads 525°F. You would then want to enter a -5°F offset
in the display trim. This is done exactly the same way as setting the zero and full-scale values:
The numeric value seen on the upper portion of the screen is the existing Display Trim Value.
Normally this is set to zero. One can now change this Offset totally independent of the ZERO, or
LOWER RANGE VALUE, (LRV) or the FULL SCALE, or UPPER RANGE VALUE, (URV), without the
use of any calibrators or external sensor inputs. To change the display offset, press ENTER. The
display changes to
0.0°F
DISPLY?
0.0°F
PLUS?
Page 24
24
indicating that the existing offset is set to “plus” 0000.0°F. The question mark “?” indicates a
question asking if this value is to remain positive (PLUS ?). By repeatedly pressing the NEXT key the
display will alternate
-0000.0°F
MINUS?
After deciding whether the offset value is to become negative (MINUS), press the ENTER key. In this
example the offset is assumed to be negative and a minus sign will be carried through this example.
The display then changes to read:
0
000.0°F
0000.0°F
PLUS?
THOUSN?
and the leftmost digit position will start blinking (shown here in italics) asking if the thousand’s
position needs to be changed. To change the thousands position, start pressing the NEXT key and
the leftmost digit will increment through 1 2 3 4 5 6 7 8 9 0. Stop pressing the NEXT key at any of
the numerals desired, then press ENTER to accept the selection. If the numeral selected before
pressing ENTER was 0, then the display would change to
0
0
00.0°F
HUNDRD?
and the second digit from the left will start blinking (shown here in italics) asking if the hundreds
position needs to be changed. As before, to change the number in this digit position repeatedly press
the NEXT key until the desired numeral is reached. Then press ENTER to go to the next lower
significant digit position. Each time the NEXT key cycles through the ten choices for that digit
position and the ENTER key enters the selected number. The digit position being changed is the one
that is blinking. The legend on the display will change successively to
00
0
0.0°F
TENS?
000
0
.0°F
ONES?
0000.
0
°F
TENTHS?
After the tenth’s digit position has also been changed to the desired value, the next pressing of the
ENTER key returns the transmitter to the alternating display of TRIM DISPLAY. Since changing of
Page 25
TX180 Temperature Transmitter25
the offset value has just been completed, press the NEXT key to proceed to the next menu selection.
Note, if trimming the transmitter to external devices is desirable, it may be necessary to trim the 4
and 20mA output after setting the display offset.
4.12SELECT LANGUAGE
The 2-line display provides the option of changing the language from English to French, German or
Spanish.
To change the language of the unit.
SELECTLANGUAGE
ENTER
Sequence through the screen until you reach the language you would like, After selecting the
language, press the ENTER Key to confirm.
English Deutsch French Espanol
RETURNTOOPERATEMODE?
If all of the set-up and re-ranging operations have been satisfactorily completed, then pressing
ENTER will return the transmitter to the normal operate mode. Pressing the NEXT key at this point
will return the display to the first screen in the sequence, Select Input.
Note again, that whenever the transmitter is in the display set-up mode, if no activation of the
pushbuttons occur for approximately 2-1/2 minutes, the transmitter returns to the operate mode.
One can also return to the operate mode at any point while in the DISPLAY MODE by removing
power from the transmitter for about 10 seconds, then reapplying power.
NOTE: Pressing ENTER stores new values in transmitter. To expace at any point, or to retain
original values (before pressing ENTER to accept new value): disconnect power and wait
30 seconds. Reconnect power & transmitter will start up in Operate Mode.
To configure a transmitter using the DISPLAY MODE, either the LI-1 or LI-2 local LCD display is
required. These displays are available as an option and can be plugged into the top of the TX180
transmitter. The transmitter can also be purchased with these options already installed. These
inexpensive options make the reconfiguration, or re-ranging of the transmitter very simple and easy
to follow. Without the use of a calibrator, or any other tools, the transmitter can be set up for a
different sensor, or the new range limits can be set much like one would set the time on a digital
watch.
In the event that the LI-1 or LI-2 Display / Keyboard are not purchased at the same time as the
transmitter, the one piece display design allows for easy field installation by simply plugging the LI1or LI-2 into the top of the transmitter.
5.1ENTERINGTHE DISPLAY MODE
To start the DISPLAY MODE, first connect the transmitter to an appropriate DC power supply.
Typically a 24VDC supply is connected with the + side of the power supply connected to the
transmitter’s output “+” terminal and - side of the power supply connected to the transmitter’s output
“-” terminal. A series resistor in the loop is optional. A sensor may be connected to the transmitter’s
input terminals, but this is not required for setting up the transmitter.
With the standard factory set-up and no sensor connected, the LI-1 display will give the following
indication:
FAILSAFE
The transmitter is indicating a fault. This would be the proper indication, since there is no sensor
connected. The analog output would indicate greater than 100% (loop current at 23.00mA), which
is the standard over range condition. If the proper sensor were connected to the transmitter, the
display would indicate the sensor’s temperature. Please note that the display / keyboards can be
plugged into the transmitter while the transmitter is powered up. There is no need to disconnect
power before plugging the LI-1 or LI-2 into the TX180.
Press the key marked NEXT to begin scrolling through the Display Mode menus.
9900
The 9900 code corresponds to the RETURN TO OPERATE MODE function. At this point,
assuming one does not want to return to the operate mode, the answer should be no, therefore,
press the key marked NEXT. Pressing the key marked Enter at this point will return the transmitter to
the operate mode.
Page 28
28
5.2DISPLAY MODE OPERATION
The one-line, LI-1 display will allow the user to do the following in a manner similar to the two-line
display.
•Select a Sensor Input (Select Input)
•Select a desired temperature unit, such as F or C (Select Units)
•Change the 4mA Lower Range Value (Change Zero)
•Change the 20mA Full Scale Value (Change Full Scale)
•Change the Sensor Fail Safe detection (Select Sensor Fail Safe)
•Change the Fail Safe reporting (Select Fail Safe Report)
•Trim the 4.0mA output current (Trim 4 mA)
•Trim the 20.0mA output current (Trim 20 mA)
•Trim the display value (Trim Display)
Each of these functions is presented with a code in a prescribed sequence on the LCD display. If the
indicated function need not be performed, press NEXT, and the next function will be displayed on
the screen. To perform any function press the ENTER key. This will cause additional screens to be
displayed which enable you to perform the function. These are described in detail below and
summarized in the LI-1 One-Line Display / Keyboard Flow Chart found at the rear of this manual.
5.3SELECTA SENSOR INPUT
The SELECT INPUT is the first function in the sequence. Virtually any thermocouple, RTD or
millivolt input can be selected. The display will read as follows to indicate this position on the
menu:
9000
The factory default sensor input is a J thermocouple. If the sensor does not require changing, then
press NEXT, and skip to Section 5.4 of this manual; otherwise press ENTER. After pressing the
ENTER key, the display will change to:
9004
Indicating that the transmitter is set to a Type J thermocouple input. If this is the desired sensor, then
press ENTER, otherwise press NEXT repeatedly to sequence through the available sensors. Each
time NEXT is pressed, the next available sensor selection is displayed.
9004
The 9004 Code corresponds to a J thermocouple
Page 29
TX180 Temperature Transmitter29
9005
The 9005 Code corresponds to a K thermocouple
9006
The 9006 Code corresponds to an L thermocouple
9007
The 9007 Code corresponds to an N thermocouple
9008
The 9008 code corresponds to an R thermocouple
9009
The 9009 code corresponds to an S thermocouple
9010
The 9010 code corresponds to a T thermocouple
9011
The 9011 code corresponds to a U thermocouple
9012
NOTE: The 9012, T/C SPEC or Special Thermocouple input is reserved for a special thermocouple
input, should one be desired. This special curve must be ordered from the factory.
9013
The 9013 code corresponds to a 2-wire ohm input.
Page 30
30
9014
The 9014 code corresponds to a 2-wire 100 ⍀ DIN curve platinum RTD with an = 0.00385
9015
The 9015 code is the 2-wire 100 ⍀ SAMA Platinum Curve, known variously as the SAMA RC21-4 or
SAMA PR-279.
9016
The 9016 code is reserved for a Special 2 wire RTD, should one be desired. Any special 2-wire RTD
curve must be ordered from the factory.
9017
The 9017 code is for 3-wire Ohms.
9018
The 9018 code is for a 3-wire 100 ⍀ DIN curve RTD with ␣ = 0.00385. This is the most commonly
used RTD in industrial applications.
9019
The 9019 code is the 3-wire 100 ⍀ SAMA Platinum Curve, known variously as the SAMA RC21-4 or
SAMA PR-279.
9020
The 9020 code for the Special 3 wire RTD input is reserved for a special RTD input, should one be
desired. Any special 3-wire RTD curve must be ordered from the factory.
9021
The 9021 code is for a 4 wire Ohm input.
Page 31
TX180 Temperature Transmitter31
9022
The 9022 code is for a 4-wire 100 ⍀ DIN curve Platinum RTD with ␣ = 0.00385. This sensor will
give superior measurement results in most real-world situations where the measured temperature is
under 1,000°F.
9023
The 9023 code is the 4-wire 100 ⍀ SAMA Platinum Curve, known variously as the SAMA RC21-4 or
SAMA PR-279.
9024
NOTE: The 9024 code is for Special 4 wire RTD input is reserved for a special RTD input, should
one be desired. Any special 4-wire RTD curve must be ordered from the factory.
9025
The 9025 code corresponds to a millivolt input.
9026
The 9026 code corresponds to an input known as the “HHTONLY”. This is reserved for a Hand-Held
set-up at the factory only.
At this point, the menus recycle to the top and begin with the first sensor input.
9001
The 9001 code corresponds to a B type thermocouple.
9002
The 9002 code corresponds to a C type thermocouple.
9003
The 9003 code corresponds to an E type thermocouple.
Page 32
32
You can stop at any one of the thermocouple or RTD or mV selections by pressing the ENTER key.
This action changes the transmitter mode to that sensor. If no sensor change is desired, then, without
sequencing through the various sensor options, but just pressing the ENTER key will allow one to
confirm the sensor selection and leave it unchanged. Assume that the sensor is left as T/C J. After
pressing ENTER the display will return to the main menu entry of SELECT INPUT. Pressing the
NEXT key then takes the transmitter to the next main menu selection.
5.4SELECT UNITS
If the selected sensor is a thermocouple or RTD, the next menu entry is SELECT UNITS. You will
not see this selection if an ohms or mV input selection is made.
9100
The code 9100 corresponds to the Select Units entry in the main menu. Pressing the ENTER key
takes you to this section of the menu. This screen indicates that the transmitter is currently set to
degrees F. Pressing the NEXT key, the display will sequence through the following screens:
9133
The 9133 code corresponds to units of Degrees Fahrenheit.
9134
The 9134 code corresponds to units of Degrees Rankine.
9135
The 9135 code corresponds to units of Degrees Kelvin.
9132
The 9132 code corresponds to units of Degrees Centigrade.
Stopping the selection process on the LI-1 display at any one of these units and then pressing Enter
will set the transmitter to the corresponding new units. For the purposes of this example the units of
measure can be left at DEG F by pressing ENTER. Pressing NEXT key will bring you to the next
section of the menu, changing the Zero.
Page 33
TX180 Temperature Transmitter33
5.5CHANGE ZERO (LOWER RANGE VALUE)
The display will then indicate as follows to indicate that one may now change the zero, or 4mA
output point.
9200
The code 9200 indicates that one can now CHANGE ZERO, or Lower Range Value (LRV), totally
independent of the Full Scale, or Upper Range Value, (URV), without the use of any calibrators or
external sensor inputs. To change the zero, press ENTER. The display changes to:
9201
The 9201 code indicates that a positive, or “plus”, number is selected for the 4 mA (LRV) output
point. By repeatedly pressing the NEXT key the display will alternate
92029201
the 9202 code corresponds to a negative number to be selected for the 4 mA output point. After
deciding whether the Zero value, or LRV, is to remain positive (PLUS), press the ENTER key. In this
example assume it is to remain positive. The display changes to read:
0
040
and the leftmost digit position will start blinking (shown here in italics) asking if the thousands
position needs to be changed. To change the thousands position, start pressing the NEXT key and
the leftmost digit will increment through 1 2 3 4 5 6 7 8 9 0. Stop pressing the NEXT key at any of
the numerals desired, then press Enter to accept the selection. If the numeral selected before
pressing ENTER was 0, then the display would change to read:
0040
and the second digit from the left will start blinking (shown here in italics) asking if the hundreds
position needs to be changed. As before, to change the number in this digit position repeatedly press
the NEXT key until the desired numeral is reached. Then press ENTER to go to the next lower
significant digit position. Each time the NEXT key cycles through the ten choices for that digit
position and the ENTER key enters the selected number. The digit position being changed is the one
that is blinking. The legend on the display will change successively to read:
Page 34
34
0040
004
0
After the one’s digit position has also been changed to the desired value, the next pressing of the
Enter key returns the transmitter to the alternating display of Change Zero. Since changing of the
zero has just been completed, press the Next key to proceed to the next menu selection, Change
Full Scale.
5.6CHANGE FULL SCALE (UPPER RANGE VALUE)
9300
The code 9300 corresponds to selection CHANGE FULL SCALE, or Upper Range Value (URV). To
change the full scale value press ENTER. The procedure for selecting Plus or Minus is identical to
that described for changing the zero, with the code 9301 corresponding to a plus (+) number and
the code 9302 corresponding to a minus (-) number. The procedure for changing each of the digit
positions is identical to that described for changing the zero. Once the steps of changing the FULL
SCALE have been completed and the ENTER key is pressed at the end of the procedure, the display
returns to CHANGE FULL SCALE. Press NEXT for the next function SELECT SENSOR FAILSAFE DETECTION.
5.7SELECT SENSOR FAIL SAFE DETECTION
9400
The code 9400 corresponds to selecting the Sensor Fail Safe detection. If one desires to change the
Sensor Fail Safe detection then press ENTER. The present status of the Sensor Fail Safe is displayed.
It is recommended that one turns off the Sensor Fail Safe when using the TX180 with an input
simulator. It should then be turned on when reconnecting the transmitter to the actual sensor.
9401
The code 9401 indicates that the Sensor Fail Safe detection is turned on.
9402
The code 9402 indicates that the Sensor Fail Safe detection is turned off. When the desired Fail Safe
condition is displayed, pressing the ENTER key will change to the new setting and the screen
returns to the SELECT SENSOR FAIL SAFE display, code 9400. Pressing the NEXT key will then
Page 35
TX180 Temperature Transmitter35
bring up the FAIL SAFE REPORTING selection screen.
5.8SELECT FAIL SAFE REPORTING
The code 9500 indicates the main menu entry for setting the transmitter Fail Safe Reporting. Pressing
the ENTER key will bring up the following code.
9501
The code 9501 corresponds to instructing the transmitter to output 3.6mA under a Fail Safe
condition. Pressing the ENTER key at this point sets the Fail Safe LOW. Pressing the NEXT key
brings up the following screen:
9502
The code 9502 corresponds to instructing the transmitter to output 23.0mA under a Fail Safe
condition. Pressing the ENTER key at this point sets the Fail Safe HIGH. Pressing the NEXT key
brings up the following screen:
9503
The code 9503 corresponds to instructing the transmitter to not report a Fail Safe condition. Pressing
the ENTER key at this point turns off this reporting. Pressing the NEXT key brings up the following
screen:
5.9TRIM 4.0MA
9600
This allows trimming of the 4.00mA output current.
NOTE: The 4.00mA limit is factory calibrated to a precision standard. Using the Output Trim
function voids the NIST traceability of calibration. Do not arbitrarily trim the output unless a
qualified and accurate local standard is available to measure the adjusted 4.00mA output!
Also note that the 4.0mA limit should not be trimmed by more than about 50A, or
transmitter operation may be impaired.
If trimming the 4.00mA limit is still desired then press Enter. The transmitter will now output a
milliamp current equal to its internally set 4mA. This 4mA value should be read on an external
meter and compared to the plant standard. It is advisable to use a very good voltmeter to make these
comparisons. It is very possible that the transmitter will be more accurate than a great many
Page 36
36
voltmeters. In this case, trimming will make the transmitter less accurate rather than more accurate!
Once trimming the 4.00mA value has been selected, the display will show:
9601
The code 9601 corresponds to selecting the function to raise mA output.
Pressing the NEXT key the display then shows:
9602
The code 9602 corresponds to selecting the function to lower the mA output. Comparing the
transmitter output to the external device will allow you to decide whether to raise or lower the
milliamp value. When it is decided whether to raise or lower the output current, then press ENTER
and the display changes to one of the following depending on whether the raise or lower function
has been selected.
96109620
(raises output)(lowers output)
The code 9610 confirms that you are in the Raise 4mA output trim function. Each time the NEXT
key is pressed, the display blinks, and the 4.0mA output limit increases (+). The increase is in
approximately 3.5 micro-ampere increments. The code 9620 confirms that you are in the Lower
4mA output trim. Each time the NEXT key is pressed, the display blinks, and the 4.0mA output limit
decreases (-). The decrease is in approximately 3.5 micro-ampere increments.
NOTE: The 4.00mA limit is factory calibrated to a precision standard. Using the Output Trim
function voids the NIST traceability of calibration. Do not arbitrarily trim the output unless a
qualified and accurate local standard is available to measure the adjusted 4.00mA output!
Also note that the 4.0mA limit should not be trimmed by more than about 50A, or
transmitter operation may be impaired.
Once the desired trim is reached, pressing ENTER will return to one of the corresponding Trim 4mA
screen. At this point one may still go back and do further trimming of the 4.0mA limit by pressing
the ENTER key, or pressing the NEXT key changes to the next function.
5.10TRIM 20MA
9700
Page 37
TX180 Temperature Transmitter37
Trimming of the 20.0mA current limit is done in exactly the same manner as was described for
trimming the 4.0mA point. Similarly the same precautions apply. The code 9701 corresponds to
selecting the function to raising the mA output. The code 9702 corresponds to selecting the function
to lower the mA output. The code 9710 confirms raising the 20mA output by approximately 3.5
micro-ampere increments with each push of the NEXT key. The code 9720 confirms lowering the
20mA output by approximately 3.5 micro-ampere increments with each push of the NEXT key.
After completing the trim 20.0mA pressing the NEXT key brings up the TRIM DISPLAY menu.
5.11TRIM DISPLAY
The display trim allows the display to be trimmed to a desired point. The transmitter’s LI-1 display
will show its value based upon the transmitter’s current settings. It is often desirable to alter the
display to make the display agree with another instrument at a critical measurement point. If this is
desirable, the display can be trimmed. The display trim operates as a zero shift and shifts the display
readings by the same amount at every point.
You can enter a single point offset to the display. Be certain before making a display trim correction
that you have made good electrical connections to the transmitter and the sensor. In the 2 or 3 wire
RTD input, or thermocouple input modes, it is possible to produce an error of a few degrees with a
fraction of an ohm in any one of the connections. Please be careful when tightening down the input
connections. These can be easily broken if a lot of torque is applied. The idea is to make a good
electrical connection without breaking the connections.
The display trim allows you to enter an offset correction. For example, suppose that the display reads
530°F, at a time when an external device that you want to agree with reads 525°F. You would then
want to enter a -5°F offset in the display trim. This is done exactly the same way as setting the zero
and full-scale values. Pressing the NEXT key at this point advances the menus and the display will
now read:
9800
The 9800 code corresponds to the display trim.
One can set the display trim offset by pressing the ENTER key. The display changes to
9801
The 9801 code indicates that a “plus” number is selected for the display offset. By repeatedly
pressing the NEXT key the display will alternate
98019802
Page 38
38
the 9802 code corresponds to a negative number to be selected for the display trim point. After
deciding whether the display trim value is to remain positive (PLUS), or negative (MINUS) press the
ENTER key. In this example assume it is to be a negative offset. The display changes to
-000
and the leftmost digit position will start blinking (shown here in italics) asking if the hundreds
position needs to be changed. To change the hundreds position, start pressing the NEXT key and the
leftmost digit will increment through 1 2 3 4 5 6 7 8 9 0. Stop pressing the NEXT key at any of the
numerals desired, then press ENTER to accept the selection. If the numeral selected before pressing
ENTER was 0, then the display would change to
-000
and the second digit from the left will start blinking (shown here in italics) asking if the tens position
needs to be changed. Pressing the ENTER key will fix the tens digit and display the ones digit:
-00
0
In this example, we want to enter a -5 degree offset, so we want to cycle the “ones” digit. As before,
to change the number in this digit position repeatedly press the NEXT key until the desired numeral
is reached. Then press ENTER to go to the next lower significant digit position. Each time the NEXT
key cycles through the ten choices for that digit position and the ENTER key enters the selected
number. The digit position being changed is the one that is blinking. The legend on the display will
change successively to:
-005
After the ones digit position has been changed to the desired value, the next pressing of the ENTER
key returns the transmitter to the 9800 code. Note that since the LI-1 will only display in whole
degrees, the display trim is limited to whole degrees. If greater display precision is required, the twoline LI-2 display will give you precision to the tenths of degrees. Since changing of the zero has just
been completed, press the NEXT key to proceed to the next menu selection, RETURN TOOPERATE MODE.
9900
If all of the set-up and re-ranging operations have been satisfactorily completed, then pressing
ENTER will return the transmitter to the normal operate mode. Pressing the NEXT key at this point
will return the display to the first screen in the sequence, SELECT INPUT which corresponds to the
Page 39
TX180 Temperature Transmitter39
code 9000.
Note again, that whenever the transmitter is in the display set-up mode, if no activation of the
keyboard occurs for approximately 2-1/2 minutes, the transmitter returns to the operate mode. One
can also return to the operate mode at any point in the DISPLAY MODE by removing power from
the transmitter for about 10 seconds and then reapplying power:
NOTE: Pressing ENTER stores new values in transmitter. To expace at any point, or to retain
original values (before pressing ENTER to accept new value): disconnect power and wait
30 seconds. Reconnect power & transmitter will start up in Operate Mode.
The TX180 detects a sensor failure condition by making various measurements across its sensor
input terminals. As a result of these measurements, the unit can detect an open thermocouple or
open RTD condition. In addition, the TX180 can detect if an RTD is short circuited, or if any of its
terminal wires (2, 3, or 4-wire RTD’s) are open. Any one of these conditions will cause a “FAIL
SAFE” report indication.
In the process of performing these sensor failure checks, the unit periodically passes small pulses of
current through the sensor and its connecting wires. The transmitter measures the resulting voltage
drop. One of the conditions resulting in a FAIL SAFE reporting condition is if this voltage drop
exceeds 180mV.
In the case of an RTD, the Fail Safe detection is part of the normal excitation for the RTD and
therefore both the temperature measurement and some of the sensor Fail Safe detection routines are
done simultaneously. In the case of a thermocouple, during the temperature measurement cycle,
there is no open sensor test current in the thermocouple. Thermocouple open circuit is detected by
making a second measurement with the test current through the thermocouple.
This method of testing for sensor failure has the following advantages:
1) In the case of thermocouples, there is no steady current through the sensor during measurement
and therefore accuracy is not degraded.
2) During open sensor detection, the test current is sufficiently high that even if there is some
leakage resistance between the sensor leads, an open sensor will be positively detected.
There are certain precautions to be observed when using this method of sensor failure detection. If
the lead wire resistance is too great, then a false FAIL SAFE report could be generated. The
maximum lead wire resistance is dependent on the type of sensor being used and the maximum
temperature expected to be measured. The maximum lead resistance for an RTD is 50⍀ in any one
lead. For a thermocouple, the maximum allowable resistance is 1,000⍀ for a non-grounded
junction Thermocouple and 10⍀ for a grounded junction T/C.
6.2CONFIGURATIONWITHAN EXTERNAL SOURCE USING DISPLAY MODE
With an external source, the basic procedure is to set the external source to the value you require for
4mA or 20mA. Next, read the transmitter value on the LI-1 or LI-2 display. Record these values. Then
follow the display set-up procedure to set the 4mA and 20mA values to the values that you recorded
with the external source.
When attempting to calibrate or check the calibration of the TX180 transmitter with an external
Page 42
42
thermocouple or RTD calibrator, it is generally advisable to disable the “SENSOR FAIL SAFE”
feature. The open sensor test periodically injects about 5A of current into the input terminals, the
millivolts generated by the calibration source is periodically disturbed and depending on the
characteristics of the external calibration source used, erroneous voltages may be applied to the
transmitter. The “SENSOR FAIL SAFE” can be disabled by turning it off in the configuration menus.
After the calibration has been completed, this function can be re-enabled.
6.2.1 THERMOCOUPLE INPUT
Setting the ZERO and FULL SCALE with a thermocouple sensor requires some added steps because
of the automatic cold-junction compensation. Thermocouple tables are normally available for a
reference junction at the ice point of water. These table entries must be adjusted for the actual coldjunction temperature. In the case of the TX180 transmitter, the cold-junctions is measured with an
internal calibrated thermometer.
It is generally good practice to operate the transmitter for 30 minutes or more prior to calibration to
allow it to reach thermal equilibrium.
CALIBRATION USING A MILLIVOLT SOURCE
The procedure starts with the selection of the thermocouple type. Then determine the temperature of
the thermocouple terminals on top of the transmitter. This can be done by measuring with a
thermometer the temperature of the thermocouple terminals on the transmitter. Or one can assume
that the terminals are approximately at room temperature and then determine the room temperature.
Next, locate the appropriate table of temperature versus mV for the selected thermocouple.
Find the table entry corresponding to the terminal block temperature, (mV @ TB°C)
Applying the millivolts (mV applied) to the transmitter and record the temperature displayed on the
LI-1 or LI-2 display for the ZERO (LRV) or 4 mA value. Then record the FULL SCALE (URV) using a
similar procedure. These recorded values will then be set into the transmitter as the zero and fullscale values using the display set-up procedure.
CALIBRATION USING A THERMOCOUPLE CALIBRATOR
Some of the thermocouple calibrators available on the market provide a means of measuring the
temperature of the terminal block and automatically apply the corrected mV to the transmitter. This
procedure is rather simple. However, there can be an appreciable difference between the
temperature of the simulator and the transmitter terminals. With some thermocouple types, this error
could be amplified 5 of 10 fold, resulting in large measurement errors. Use caution so as to not
introduce these possible errors.
Page 43
TX180 Temperature Transmitter43
6.3FOR BEST MEASUREMENT ACCURACY
The TX180 transmitter is a stable instrument, precision calibrated at the factory for any measurement
range the user may select. However, the automatic cold-junction compensation requires certain
precautions to obtain best accuracy when used with a thermocouple sensor.
The cold-junction compensation operates by attempting to measure accurately the temperature of
the thermocouple terminals on top of the instrument. If these terminals are exposed to thermal
radiation or convection, the cold-junction compensation will introduce an error. With certain types
of thermocouples and temperature measurement ranges, the sensitivity of the cold-junction is
greater than the sensitivity of the measurement couple. Under those conditions, a one degree error
in the cold-junction temperature that the transmitter senses can result in a greater than one degree
temperature measurement error.
For best measurement accuracy with thermocouple sensors, it is advisable to shield the top terminals
by placing the transmitter into a housing or enclosure, such as the model XP-FN, XP-FG, XP-HEAD,
or WP-HEAD. In addition, sufficient time should be allowed for the housing and the transmitter to
reach equilibrium temperature in a given operating environment before best accuracy is reached.
For best accuracy with any sensor, or in the millivolt mode, it is advisable to allow the transmitter to
operate with the desired fixed input signal for a period of 30 seconds before the reading is taken.
The transmitter periodically measures certain internal references. These internal measurements and
the external signal undergo digital averaging and the full accuracy of the instrument is only achieved
after several readings have been averaged.
When using an RTD sensor, a four-wire connection is generally recommended. With a three-wire
RTD the TX180 makes two separate measurements before calculating the temperature, whereas,
only a single measurement is required when using a four-wire RTD. Conceptually, a better accuracy
is possible using a single measurement as compared with calculating the difference of two separate
measurements.
7.ACCESSORIESAND INFORMATION
Other accessories available from Dynsico are:
LI-1One-line Local Display
LI-2Two-line Local Display
XP-HEADExplosion Proof Housing (No Display Option*)
WP-HEADWeather Proof Housing (No Display Option*)
XP-FNExplosion / Weather proof Housing with mounting lugs
XP-FGExplosion / Weather proof Housing with mounting lugs and Glass Window
PY-2Pipe mount bracket to mount XP-FG or XP-FN on 2" pipe
PS-2412Power Supply, 24/28 VDC, 1.2/0.8A
PS-2405Power supply, 24 VDC, 0.5A
Page 44
44
A variety of signal conditioners, thermocouples are also available.
* The LI-1 or LI-2 can be used for setup and configuration for transmitters installed in XP-HEAD or
WP-HEAD housings, but cannot be installed permanently. The XP-HEAD and WP-HEAD will not
close properly with either of these displays installed.
8.SPECIFICATIONS
THERMOCOUPLE SENSORS:
NIST Types B, C, E, J, K, N, R, S, & T; DIN Types L & U
RTD Resistance Sensors:
100 ⍀ Pt DIN curve (␣ = 0.00385)2, 3 or 4 Wire
100 ⍀ Pt SAMA RC21-4 curve (␣ =0.003923)2, 3 or 4 Wire
100 ⍀ Ni DIN curve2, 3 or 4 Wire
120 ⍀ Ni MINCO curve2, 3 or 4 Wire
10 ⍀ Cu curve2, 3 or 4 Wire
Ni SAMA2, 3 or 4 Wire
MILLIVOLT INPUT RANGE:
-15 to 115mVDC
THERMOCOUPLE AND RTD LINEARIZATION:
Linearization with temperature conforms to NIST & DIN curves within ±0.05°C.
OUTPUT:
Analog, Two wire 4 to 20mA; Digital via HART
OUTPUT RANGING ADJUSTMENTS:
Analog Zero{100% of sensor range, non-interacting}
Full Scale{Normal or reverse acting}
MINIMUM OUTPUT RANGE:
None
OUTPUT RESOLUTION:
Analog, 3.6A
TRANSMITTER ACCURACY:
± 0.05% of the millivolt or ohm equivalent reading, or the accuracy from the table below, whichever
is greater; plus the effect of cold junction measurement error or ± 0.5°C (±0.9°F), if using a
thermocouple sensor; plus ± 0.05% of span.
Accuracy includes repeatability, hysteresis, load and ambient temperature.
TRANSMITTER REPEATABILITY:
One half of accuracy.
REFERENCE CONDITION ACCURACY:
Equal to transmitter repeatability, when set-up under reference conditions to an external source. The
transmitter is then referenced to the prevailing conditions and transmitter accuracy at this reference
condition will include repeatability, linearity, and hysteresis effects. If using a thermocouple add
0.05°F for reference condition accuracy cold junction effect. Reference condition accuracy is
comparable in scope to the accuracy generally specified for analog based transmitters and is
consistent with the ANSI/ISAS51.1-1979 definition of “Accuracy”.
DYNAMIC RESPONSE:
Turn On Time: Less than 5 seconds after power up
Ambient Temperature Gradient: Automatic compensation to 20°C/Hour Change
Update Time: 0.15 Seconds
Response to Step Input: 0.25 Seconds, Typical
COLD JUNCTION COMPENSATION:
Self-correction to ±0.5°C
OPERATING TEMPERATURE RANGE:
-40°F to +185°F (-40°C to +85°C)Electronics
-4°F to +158°F (-20°C to +70°C) Display
STORAGE TEMPERATURE RANGE:
-58°F to +185°F (-50°C to +85°C)
AMBIENT TEMPERATURE STABILITY:
Self-correcting over the operating temperature range.
LONG TERM STABILITY:
Less than 0.05% of reading plus ±3.6A per year.
Page 46
46
AUTOMATIC DIAGNOSTICS:
Every 3 seconds the TX180 transmitter performs self-checks for zero, span, cold-junction
temperature, open T/C, open RTD element, shorted RTD element, each open RTD lead and
transmitter malfunction.
FAILSAFE:
User settable from 3.6mA to 23.0mA, or OFF
INTERCHANGEABILITY:
All units interchangeable without field calibration.
EMI/RFI IMMUNITY:
Less than 0.5% of reading (SAMA PMC 33.1c test method) 20KHz to 1000MHz, 10 V/meter.
ISOLATION:
250 VAC rms or 800VDC
COMMON MODE REJECTION:
120dB
REVERSE POLARITY PROTECTION:
42 VDC applied with either polarity
POWER AND LOAD:
Supply voltage (no load resistance); 12 to 42 VDC
(30 VDC for I/S version);
Supply voltage (with load resistance);
Vsupply=(12)+(Rload in Kohm) x (23mA)
Supply Voltage Effect: < +/-0.005% of Span per Volt
WEIGHT:
4 oz. Electronics only
STANDARD CONFIGURATION:
Factory configured for Type J thermocouple, 40°F=4.0mA, 200°F=20mA, with HI failsafes. Special
configurations are available to suit your requirements. See Price List.
Questions concerning warranty, repair cost, delivery, and requests for a RA# should be directed to
the Dynisco Repair Department, 508-541-9400 or email: repair@dynisco.com. Please call for a
return authorization number (RA#) before returning any product. Damaged products should be
returned to:
DYNISCO INSTRUMENTS
Attn: RA # _______________
38 Forge Parkway
Franklin, MA 02038
For technical assistance please call 800-221-2201 or 508-541-9400 or fax 508-541-9436.
10.WARRANTY
This Dynisco product is warranted under terms and conditions set forth in the Dynisco Web Pages.
Go to www.dynisco.com and click on “Warranty” at the bottom of any page for complete details.
Page 48
48
Fig. 9
NOTES:
TX 180 1/S, ITX 190 I/S, XP-FNI, XP-FGI Factory Mutual and Canadian
Standards Assoc. Approved Intrinsically Safe Installation Control Drawing
1) The electrical circuit in the hazardous area must be capable of withstanding an A.C. test voltage
of 500 volts R.M.S. to earth or frame of the apparatus for one minute.
2) Cable capacitance and inductance plus the I.S. apparatus unprotected capacitance (Ci) and
inductance (Li) must not exceed the allowed capacitance (Ca) and inductance (La) indicated on
the associated apparatus.
3) Any positive polarity shunt zener diode safety barrier approved by FMRC or CSA for groups A,
B, C, D, E, F and G whose output parameters are:
Voc or Vt 울 30 Vdc
Isc or It 울 125 mA
Ca 욷 Ci + C cable
La 욷 Li + L cable
4) The installation including the barrier earthing arrangements must comply with the installation
requirements of the country of use, i.e. ANSI/ISA RP12.6 (installation of intrinsically safe systems
for hazardous (classified) locations) and the national electrical code, ANSI/NFPA 70 or
Canadian electrical code part 1. Installation must be in accordance with the manufacturer’s
guidelines.
Page 49
TX180 Temperature Transmitter49
5) Temp. CodeT4 at maximum ambient temperature of 85°C
T6 at maximum ambient temperature of 40° C
6) Dust-tight conduit seal must be used when installed in Class II and III Environments.
NO REVISIONS WITHOUT PRIOR FACTORY MUTUAL OR CANADIAN STANDARDS
ASSOCIATION APPROVAL.
Page 50
NOTES:
Page 51
WARRANTY REGISTRATION CARD
MODEL NUMBER ______________________________________________________________
SERIAL NUMBER _______________________________________________________________
DATE PURCHASED _____________________________________________________________
PURCHASED FROM ____________________________________________________________
NAME _________________________________________________________________________
COMPANY ____________________________________________________________________