Dynisco PT467XL-XM-10/18, PT460XL-XM-6, TPT463XL-XM-6/18, PT462XL-XM-6/18 User manual

OPERATING INSTRUCTIONS FOR MELT PRESSURE TRANSDUCERS
INSTRUCCIONES DE OPERACION PARA TRANSDUCTORES DE PRESION DE POLIMERO FUNDIDO
Operating Principle
The Dynisco Melt Pressure Transducer for space restricted areas is used to make pressure measurements of molten polymers up to 750˚F (400˚C). These models incorporate a 350-Ohm, bonded foil strain gage Wheatstone Bridge. This proven technology provides an output of 0 - 3.33 mV/V (nominal), proportional to melt pressure (within the specified error band).
Most models include an internal shunt calibration ("Rcal") function that is used to simulate a signal of 80% of full scale. This eliminates the need for a cumbersome calibrated pressure source when scaling associated instrumentation.
Installation
Do not remove protective cap until ready to install.
Prior to initial installation, verify correct machining of mounting hole per Figure 1. Gauge Plug, P/N 435901, is available for this purpose.
When reinstalling make sure that mounting hole is clear of material. Dynisco Cleaning Tool Kit, P/N 200100, should be used.
To prevent galling, lightly coat transducer threads with a high temperature anti-seize material. An adequate seal, in a properly machined and maintained mounting well, is obtained with 100 in-lbs. mounting torque. Maximum recommended torque is 500 in-lbs.
The electronics housing should be secured, with the enclosed mounting bracket (P/N 200941), in an area where the ambient temperature will not exceed 250˚F (121˚C).
Wiring
Use 6 conductor, shielded cable. Attach cable shield to ground at one end only. NOTE: DYNISCO cable assemblies are constructed with shield wired to transducer mating connector, so do not attach shield to instrument.
Excitation is from a well regulated 10 VDC, recommended, (12 VDC max.) power supply. (Lower voltage can be used.)
Principio Operativo
Los Transductores de Presión de Polímero fundido de Dynisco se utilizan para tomar medidas de Presión sobre termoplásticos fundidos hasta temperaturas de 750˚F (400˚C). Estos modelos incorporan una galga extensiométrica con un “Puente de Wheatstone” de 350-Ohmios. Esta probada tecnología proporciona una señal de salida de 0-3.33 mV/V (nominal)
proporcional a la presión (dentro de la banda de error especificada).
La mayoría de los modelos incorporan una calibración interna (“Rcal”) función que se utiliza para generar una señal del 80% del rango total. De este modo se elimina la necesidad de disponer de una fuente calibrada de presión para el escalado del instrumento.
Instalcion
No extraer el capuchón protector hasta que no esté listo para instalar.
Antes de la instalación inicial, comprobar que la mecanización del agujero sea correcta según Figura 1. Para esta verificación disponemos del utillaje “Gauge Plug, P/N 435901.”
En sucesivas reinstalaciones se deberá asegurar que el agujero para alojamiento esté limpio de Material. Para ello deberán utilizar el equipo “Dynisco Cleaning Tool Kit P/N200100.”
Para evitar “gripaje” de la rosca, recubrir los pasos de la rosca ligeramente con un material anti-gripaje para alta temperatura. En un agujero de alojamiento bien mecanizado, con un buen mantenimiento, un adecuado cierre ó estanqueidad se obtendrá con un par de clerre de 100 pulgada-Libra. El Par máximo de cierre es de 500 pulgada-Libra
El cabezal electrónico deberá asegurarse mediante el soporte de montaje (P/N200941) incluido, en una zona donde la temperatura ambiente no exceda los 121˚C (250˚F).
Cableado
Utilizar cable apantallado de 6 conductores. Conexionar a Tierra en un solo extremo. NOTA: El cable de DYNISCO hacia el conector está construido con cable apantallado, por lo tanto, no sujete este cable al instrumento.
Se recomienda que la Excitación proceda de un suministro eléctrico bien regulado de 10 VDC, (12 VDC máxima). (Voltajes inferiores pueden ser utilizados).
Figure 1 - Recommended Mounting Well (1/2-20 UNF) Figura 1 - Agujero de Montaje Recomendado (1/2-20 UNF)
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Installation Instructions for These Models:
PT435A, PT465XL, PT467E, PT467XL
Start-Up
Bring system to operating temperature and with no pressure, follow recommended procedures with instrumentation for zero and span adjustment.
Make sure that there is sufficient "soak time" to assure that any material at the tip of the transducer is molten before extruder drive is started.
Removal
Transducer should only be removed when polymer is hot and liquid. Wipe tip with a soft cloth immediately. The melt pressure transducer must be removed before using an abrasive material or wire brush to clean the extruder barrel. Clean mounting well before attempting to reinstall transducer with Dynisco Mounting Well Cleaning Tool Kit, P/N
200100.
Transducer Repair
Damaged transducers should be returned to:
Dynisco Instruments
Questions concerning warranty, repair cost, delivery, and requests for a RA#, should be directed to the Dynisco Repair Department,
Please call for a return authorization number (RA#) before returning any product.
Technical Assistance
Warranty
This Dynisco product is warranted under terms and conditions click on “Warranty” at the bottom of any page for complete
details.
Puesta En Marcha
Llevar el sistema a la temperatura de trabajo y sin presión, seguir los procedimientos recomendados para el ajuste del “cero” y el “span” para el instrumento en cuestión.
Asegurarse de que el tiempo de espera desde que la instalación está a temperatura de trabajo hasta que se ponga en marcha la extrusora, sea el suficiente para que cualquier material en contacto con el diafragma del transductor, esté completamente fundido.
Extraccion
Sólo debería extraerse el Transductor mientras el polímero está caliente y líquido. Una vez extraído y mediante un suave textil limpiar el diafragma con mucha delicadeza. El transductor deber· extraerse ó desmontarse antes de que se utilicen materiales abrasivos ó cepillos de púas para la limpieza
de la extrusora/husillos. Antes de reinstalar el Transductor limpiar correctamente el aguajero de alojamiento mediante el equipo “Dynisco Mounting Well Cleaning Tool Kit, P/N200100.”
Servicio
Los transductores dañados deberán ser devueltos a: Dynisco Instruments
Cuestiones relacionadas con la Garantía, Coste de reparación, entrega, y peticiones para un/una (RA#) deberían dirigirse a
Agradeceremos pidan el Número de Autorización de Devolución (RA#) antes de proceder a su devolución.
Asistencia Tecnica
Garantía
Este producto Dynisco está garantizado bajos los términos y condiciones establecidas en las Páginas Web Dynisco. Diríjase a
cualquier página para obtener información completa.
Mating Connector
Bendix PT06A-10-6S(SR) Bendix PT06A-12-8S(SR) Dynisco P/N 711600 Dynisco P/N 710700
8-Pin Connector PT420 Series
A = Sig + Red/Rojo
B = Sig - Black/Negro
C = Exc + White/Blanco
D = Exc - Green/Verde
E = R-Cal Blue/Azul
F = R-Cal Orange/Naranja
A = Exc + White/Blanco
B = Sig + Red/Rojo
C = Exc - Green/Verde
D = Sig - Black/Negro
E = R-Cal Blue/Azul
F = R-Cal Orange/Naranja
G = No Connection
No Conexión
H = No Connection
No Conexión
6-Pin Connector PT460 Series
Conector 8-Pin Serie PT420
Conector 6-Pin Serie PT460
P/N 974114
01/02 REV. A
ECO #26421