P/N: 974022 |  Rev: 100517 |  ECO: 30965 
providing a window into the process 
PT130, PT140, PT150 & PT160 
Hydraulic Sensor 
Operating Manual 
www.dynisco.com 

P/N: 974022 |  Rev: 100517 |  ECO: 30965 
providing a window into the process 
10 V dc Recommended, 
15 V dc Max. 
12-32 V dc ± 12 to ± 16 V dc 
12-32 V dc
16-32 V dc, ± 14 to ± 16 V dc 
16-32 V dc
“Product Warranty” at the bottom of any page for 
complete details.
OPERATING PRINCIPLE 
The PT100 Series Industrial Transducer is used to 
make pressure measurements for hydraulic 
applications and corrosive media up to 10,000 psi. 
These models incorporate a four-arm, 350-ohm bonded 
foil Wheatstone bridge strain gage. An amplifier can be 
added to this proven technology to provide a high-level 
output signal (e.g. 0 – 5 V dc or 0 – 10Vdc) proportional 
to the pressure (within the specified error band). 
SHUNT CALIBRATION 
Internal – All 100 Series transducers contain an 
internal Shunt Calibration resistor, R-CAL, which 
provides a simulated output of 80% of full scale when 
pins E and F are shorted. 
External – The model PT130 also provides an output of 
80% of full scale when a 34,935-ohm Shunt Calibration 
resistor is placed across pins B and D. External R-CAL 
feature is not available on models PT160 and PT150. 
NOTE: The internal and external R-CAL should not be 
used at the same time because incorrect readings will 
result. 
PRESSURE FITTING 
The pressure fitting on all 100 Series transducers 
contains a 1/8”-27 NPT internal pressure port. The 
fitting is fabricated from high-strength stainless steel. 
ELECTRICAL CONNECTORS  
All models are equipped with a six pin, Bendix PT02A10-6P receptacle. The customer supplied mating plug is 
a socket, Bendix PT06A-10-6S, or equivalent.
TRANSDUCER REPAIR 
Questions concerning warranty, repair cost, delivery, 
and requests for a RA# should be directed to the 
Dynisco Repair Department, 508-541-9400 or email: 
repair @dynisco.com. Please call for a return 
authorization number (RA#) before returning any 
product. Damaged transducers should be returned to: 
Dynisco Instruments 
Attn: RA #_____________ 
38 Forge Parkway 
Franklin, MA 02038 
TECHNICAL ASSISTANCE  
Please call 800-221-2201 or 508-541-9400 or fax 
508-541-9206.
ELECTRICAL CHARACTERISITCS 
WARRANTY
This Dynisco product is warranted under terms and 
conditions set forth in the Dynisco Web Pages. Go to 
www.dynisco.com 
www.dynisco.com 
and click on “Instruments”. Click 

P/N: 974022 |  Rev: 100517 |  ECO: 30965 
providing a window into the process 
POWER SUPPLY/TRANSDUCER WIRING DIAGRAMS 
Dynisco 
38 Forge Parkway, Franklin, MA 02038 
Tel: 508-541-9400 Fax: 508-541-9206 01/06 
www.dynisco.com
www.dynisco.com