Dynatron AG OxySpy Operating Manual

Page 1
English
Operating Manual
English
R
Page 2
Important remarks concerning the manual
This operating manual makes use of the fol­lowing icons to indicate especially important comments:
Information and tips which are important for opti­mal use of the OxySpy.
Indicates information about details which are important to prevent a risky situation.
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
The following symbols are used in the operating manual:
LCD display samples
Grey or red: part of the display actually being described
Flashing part of the display
Remarks
Attention!
Danger!
English
Page 3
List of chapters
Introduction 4
Warnings and Safety Considerations 5
OxySpy Parts List 6
The OxySpy 6
Set Up Procedure / Maintenance
Before first use 6 Switching the OxySpy on and off 7 Opening the OxySpy 7 Fitting the
O2sensor 8
Fitting the battery 10
Calibration 11
Measurement 14
Service Indications / Trouble Shooting
CAL symbol is displayed 16 O2sensor not connected 16 O2sensor symbol is displayed 17 Battery symbol is displayed 18 O2% indication not stable in the air 18
O2sensor 19
Maintenance and Storage 20 Operational limits 20 Technical Information 20
Limited Warranty 21 Notes 22 FCC / Copyright © / Waste disposal 23 Warranty paper 24
English
Page 4
Introduction
The OxySpy is a portable Oxygen Analyser for the trained and certified mixed gas diver. The OxySpy can be held directly against the slightly opened diving tank and displays within seconds the oxygen percentage of the gas. By pressing the push button the display can be frozen. By pressing the push button once more the maximum operation depth (MOD) is dis­played. The calibration of the OxySpy is as easy: The unit detects itself whether the calibration is made in air or with pure oxygen. A calibration can only be terminated if the
O
2
percentage does not change during the calibration. If the OxySpy is not used, it switches off auto­matically. Thanks to its ergonomical design and its single button operation, it can be operated by one hand. The OxySpy displays the sensor or battery sym­bol, if the
O2sensor or the battery are used up.
Both are easily changed by the user.
English
Page 5
Warnings and Safety Considerations
• If you are unsure about the displayed
O
2
per­centage: Repeat the measurement or check the value with another system. For measure­ments after filling the tank and before diving always use two different measuring units.
• Check the OxySpy in air before each series of measurements. If necessary, calibrate the OxySpy. See page 5.
• Using pure oxygen or oxygen enriched gas needs special safety measures! Please consi­der the general safety considerations of using pure oxygen or oxygen enriched gases!
• If the
O2sensor symbol is shown, the display-
ed values can be wrong.
• If the diver uses the wrong value of the gas mixture he/she or the dive computer could cal­culate insufficient decompression time (EAD etc.) or oxygen toxicity (MOD, CNS% etc.). Both situations could result in death or serious injury.
• The maximum partial pressure of oxygen (pp
O2limit) necessary for the calculation of the maximum operation depth (MOD) should not exceed 1.6 bar (the max. value for sports divers). Higher values may result in death or serious injury.
• Avoid risky situations which are marked in the this manual with and .
• Do not use the OxySpy in pressure chambers.
• Do not submerge the OxySpy in water.
• The battery and the
O
2
sensor contain corrosi-
ve liquid. If the battery or
O
2
sensor leaks: Do not let the liquid come into contact with skin, eyes or mouth. If it does, flush the affected area with water for at least 15 minutes. Consult a doctor. Do not inhale vapor. Immediately clean the leakage from the instrument with a soft towel and dispose of the towel.
Important safety con­siderations
English
Page 6
English
1 OxySpy with case 1 OxySpy Operating Manual 1 Battery CR 2430 1
O2Sensor in a sealed package
1 Philips-type Screwdriver
The supplied battery and the
O2 sensor have to be fitted. Similar to the finite life of a battery, the O2sensor has a certain lifetime. From the first time that a
O
2
sensor contacts oxygen it starts to
be used up. To keep the lifetime of the
O
2
sensor, its airtight package should only be opened before the sensor is used the first time.
OxySpy Parts List / The OxySpy / Set Up
OxySpy Parts List
The OxySpy
Before first use
Display Push button
Handgrip
Gas Input /Exit openings with rubber rings
Battery symbol MOD symbol
(maximum operation depth) Calibration symbol O
2
Sensor symbol
Symbols for ppO
2
,% O2, mtr. and ft.
O
2
value, ppO2, MOD, messages
Page 7
English
Set Up Procedure / Maintenance
After a short push on the button, the OxySpy will display the measured
O
2
percentage. If need be, the unit indicates that it requires calibration or indicates that no
O2sensor is connected.
The OxySpy switches off automatically, at the latest 30 minutes after switching on. The unit cannot be switched off manually.
• Open the OxySpy only in a clean and dry area
away from water, dust and other contaminants. It prevents soiling and damage to the electro­nics and the mounting ring.
• Open the OxySpy only if the battery or the
O
2
sensor must be replaced.
• Do not touch the electronics.
1. Pull the rear rubber ring off the housing.
2. Remove the 7 screws with the enclosed scr-
ewdriver and separate the two parts of the hou­sing.
Switching on and off
Opening the Oxy­Spy
Battery
O2Sensor
Plug
Page 8
Set Up Procedure / Maintenance
1. Open the rear side of the housing according to the chapter "Opening the OxySpy" (page 7).
Consider the safety considerations on page 5 and the ones on the
O2sensor and its package.
2. Take the
O2sensor (if there is already one) out
of its position and pull off the mounting ring.
If the
O2sensor is disconnected while the unit is
on, the display shows the word "no" and the
O
2
sensor symbol. In this situation the OxySpy switches off after 5 minutes automatically.
3. Carefully pull the plug out of the socket. The tab, which stops the plug from being disconnec­ted, has to be pulled back while pulling out the plug.
Do not pull on the wires!
• Do not open the airtight container before you need to use the
O
2
sensor.
• After opening the sealed package the
O
2
sen­sor needs 30 minutes start-up time. This means that calibration cannot be made until 30 minutes after opening the package. See "Cali­bration" page 11.
• Take care of the environment; properly dispose of used
O
2
sensor! The sensor should be dis-
posed of according to local laws.
Fitting the O2Sensor
English
O2sensor
Mounting ring
Page 9
English
Set Up Procedure / Maintenance
4. Take the new
O2sensor out of its package.
Take the O ring off the
O2sensor (if there is any
O ring).
Attach the mounting ring by twisting it onto the O2sensor, exactly as shown in the picture on the left. Maneuver until it fits snugly.
Write the date of the fitting onto the
O
2
sensor.
5. Insert the plug into the socket of the
O
2
sen­sor. Make sure that the tab, that secures the plug, is back in place. The lug of the mounting ring must fit into the notch of the housing.
6. If a battery is mounted: Switch the unit on and check that the display is on.
The unit will show the CAL symbol, because the OxySpy requires calibration. The displayed
O
2
percentage may be incorrect.
If "no" and the
O2sensor symbol is displayed
check the chapter "Trouble Shooting", page 16.
7. Carefully refit together the 2 parts of the case. Make sure that none of the wires get trapped and damaged by the case. Take care that the two parts fit together well and carefully tighten the 7 screws.
8. Locate the rubber ring into the housing so that its opening does not close the gas flow to the
O
2
sensor.
9. Calibrate the OxySpy or fit a new battery. See "Calibration" or "Fitting the Battery" on page 11 or page 10, respectively.
Opening
Mounting ring
Page 10
Set Up Procedure / Maintenance
10
1. Open the rear side of the housing according to the chapter "Opening the OxySpy", (page 7).
Consider the safety considerations on page 5.
2. Pull the battery (if there is any) out of the bat­tery holder, by hooking a finger under the bat­tery.
Take care of the environment: Dispose of the used battery properly!
3. Wait a minimum of 2 minutes before installing the battery. Check that it is the correct type of battery (see "Technical Information" page 20). Press the battery into the battery holder until it snaps in place ( + on the top). The battery hol­der prevents an incorrect fitting.
4.
Switch the unit on and check if the display is on. If the unit cannot be switched on, repeat step 2 and 3.
Choose the correct depth unit (mtr. or ft.) to dis­play the maximum operation depth (MOD) by quickly pressing the push button. To confirm the depth unit press the push button for at least two seconds.
Check the chosen metric system before refitting the parts of the case!
The unit will show the CAL symbol, because the OxySpy requires calibration. The displayed
O
2
percentage may be incorrect.
5. Carefully refit together the 2 parts of the case. Make sure that none of the wires get trapped and damaged by the case. Take care that the two parts fit together well and carefully tighten the 7 screws.
6. Calibrate the OxySpy or mount a new sensor. See "Calibration" or "Fitting the
O
2
sensor" on
page 11 or page 8, respectively.
Fitting the battery
English
Page 11
Calibration
11
The calibration can be made either in air (20.9%
O2) or pure oxygen (100%
O2). The Oxy­Spy will automatically detect which gas is being used for calibration. The
O2sensor measures the partial pressure of
the oxygen and the OxySpy calculates the
O
2
percentage (
O
2
%) from it. Changes of the ambient pressure or temperature will influence the
O2measurement values. This is the reason why the OxySpy has to be calibrated frequently. Once the
O2sensor has been calibrated, the maximum allowed partial pressure of oxygen (pp
O2limit)
can be set. This value is needed by the OxySpy to calculate the maximum operation depth (MOD).
A new calibration is required when:
• The measured
O
2
percentage in the air is
below 20.7%
O
2
.
• The measured
O2percentage in the air
exceeds 21.1%
O2.
• The CAL symbol is displayed.
• If you are unsure about the displayed
O
2
per-
centage.
The
O
2
percentage of the gas mixture (used for the calibration) must not change during the cali­bration.
The duration of an OxySpy calibration is a maxi­mum of 30 seconds.
The calibration can be interrupted at any time by pressing the button.
In hot or humid climates, the OxySpy should be calibrated either in pure oxygen (100%) or in compressed air from a tank.
What you have to know about calibration
English
Page 12
Calibration
12
1. Before the calibration of the OxySpy in air (20.9%
O2) it has to be moved slowly back and forth, through the air, for about 1 minute to flush the
O2sensor. Without this flushing, the calibra­tion may not be successfully completed. Do not blow through the OxySpy as the C
O
2
in your
breath would cause an incorrect calibration.
2. Press the button until "CAL" and "run" appear. "CAL" and "run" will flash during the calibration. Move the OxySpy back and forth, slowly, during the entire 30 second calibration.
3. After the
O2sensor has been calibrated, the
OxySpy will display the pp
O2limit (necessary for calculating the MOD). By quickly pressing the push button the setting may be changed in the range of 1.2 to 1.8 bar, displayed in 0.05 bar increments.
To confirm the indicated pp
O
2
setting press the
push button for at least 2 seconds.
If neither changes nor confirmations are made within 30 seconds, the OxySpy automatically switches back to the default setting of 1.4 bar. If the selected pp
O
2
limit is higher than 1.6 bar,
the pp
O2symbol will flash to warn the diver.
• The maximum partial pressure of oxygen
(pp
O
2
limit) for calculating the maximum opera­tion depth (MOD) should not exceed 1.6 bar (the max. value for sports divers). Higher values may result in death or serious injury.
• For strenuous dives or diving in cold water the chosen pp
O
2
limit should not exceed 1.4 bar.
The OxySpy switches automatically into the nor­mal measurement mode.
Calibrating in air
English
Page 13
Calibration
13
Calibration was not successful
English
Calibrating with pure oxygen
The OxySpy can also be calibrated with pure medical oxygen (100%
O2).
• Please consider the general safety con­siderations of using pure oxygen!
• The calibration gas has to be 100% medical oxygen.
• At the gas input make sure that there is a good seal and that air cannot be sucked in.
The calibration procedure with pure oxygen is similar to the one described in the chapters "Calibrating in air" (page 12) and "Measure­ment" (page 14).
After the pure oxygen calibration: "Flush"(move it back and forth slowly) the OxySpy for 1 minu­te in air and check the indicated value. The value must be between 20.7 and 21.1%
O2. If
not, redo the calibration.
If the calibration has not been successfully com­pleted or the calibration has been interrupted by pressing the button switch, the display will show a "- - -", for a short time. The OxySpy will then change automatically into the measurement mode. No new calibration values have been entered and the indicated
O
2
% values are still based on the old calibration data. The CAL sym­bol is displayed to inform you about the unsuc­cessful calibration.
Repeat the calibration if the CAL symbol is still displayed after the calibration. Several calibrati­ons without success could be an indication that the
O
2
sensor needs to be replaced or a fluctua­tion of the mixture of the calibration gas during the calibration. The fluctuation of the gas mixtu­re might have been caused by air leaking in at the gas input due to a bad seal.
Page 14
Measurement
14
Consider the following points if you measure with the OxySpy:
•
The gas must be able to flow through both elas­tic rings without resistance.
• Check the OxySpy in air before each measu­rement by "flushing" it (moving it back and forth for 1 minute in the air). If the displayed value is not between 20.7 and 21.1%
O2, the OxySpy
has to be calibrated.
• The OxySpy must have a steady stream of the gas until the displayed
O2% value has stabili-
zed.
• At the gas input (the rubber ring) make sure that there is a good seal and that air can not be sucked in.
1. Press one of the two rubber rings against the
tank valve.
2. Open the valve until you hear a gentle, uni-
form hiss and wait until the displayed
O2% value
has stabilized.
3. "Freeze" the display by briefly pressing the
push button. For confirmation the "%
O2" starts
flashing.
4. Close the tank valve.
What you have to under­stand
English
Procedure to measu­re on diving tanks
Page 15
Measurement
15
5. If the displayed value does not match with your expectation, repeat the measurement or calibrate the OxySpy. See "Calibration" page 11.
6. Record the results of your measurement.
7. After quickly pressing the push button, the
OxySpy will calculate the MOD based on the measured
O2values and the chosen pp
O2limit. The indicated maximum operation depth will be rounded to 0.1 mtr. or 1 ft.
If the selected pp
O2limit is higher than 1.6 bar,
the pp
O2symbol will flash to warn the diver.
8. By pressing the push button once more the OxySpy returns to measuring mode. Repeat the measurement until you are sure
that
you measured the actual oxygen percentage
of
the gas mixture.
English
Page 16
Service Indications / Trouble Shooting
16
If the CAL symbol appears, the OxySpy has to be calibrated. The displayed
O2% value could be
wrong.
If the OxySpy does not "see" an
O
2
sensor, the
display shows "no" and the
O2sensor symbol.
The CAL symbol is displayed
English
Possible reasons:
New battery has been installed.
New
O
2
sensor has
been installed.
Calibration was not successful.
How to solve:
Calibrate the OxySpy. See page 11
Calibrate the OxySpy. See page 11
Calibrate the OxySpy. See page 11
Possible reasons:
No
O
2
sensor
connected
Water in the housing
Plug or wire defect
Electronic defect.
How to solve:
Install an O2sensor and plug in. See
"
Fitting the O
2
sensor
"
, page 8.
Remove the battery and the
O2sensor. Dry the case with a towel and warm air (max. 50 °C / 122 °F).
Take the OxySpy to the store that it was purchased from.
Take the OxySpy to the store that it was purcha­sed from.
The O
2
sensor not connected
Page 17
Service Indications / Trouble Shooting
17
If the
O2symbol appears, the indicated
O2% value could be incorrect.
The
O2 sensors are available at OxySpy sales points. There is more information, regarding the O2 sensor, on page 20.
The O
2 sensor
symbol is displayed
English
Possible reasons:
The ambient air pressure is excep­tionally low. (for example in high altitude)
The
O2sensor is
used up
Water in the hou­sing
Electronic defect.
How to solve:
At high altitudes the
O
2
sensor symbol could be displayed by mistake. If the
O
2
sensor symbol had not been displayed before ascending (to the higher altitude), than it is almost certain, that the low ambient pressure of the altitude is the reason for displaying the O
2
sensor.
Replace the O2 sensor. See "
Fitting the O
2
sensor
"
,
page 8.
Remove the battery and the
O2 sensor. Dry the case with a towel and warm air (max. 50 °C / 122 °F).
Take the OxySpy to the store that it was purcha­sed from.
Check the OxySpy before ascending to altitude.
Page 18
18
At a low battery voltage, the OxySpy shows the battery symbol.
If the battery symbol appears, the indica­ted
O2% value could be wrong.
The battery is sold in camera and electronic sto­res. For more information about the battery, refer to page 20.
The displayed oxygen percentage of the air goes up and down, even if the actual percenta­ge does not change.
The Battery Symbol is displayed
English
Possible reasons:
Extremely low temperature
The battery is weak.
How to solve:
Warm up the OxySpy and check again.
Replace the battery. See page 10.
Possible reasons:
After a measure­ment of an
O
2
enri­ched gas, the high concentration of oxygen in the Oxy­Spy has first to be reduced.
O2sensor is defec­tive or used up.
How to solve:
"Flush" (move it back and forth) the OxySpy in air for approximately 30 seconds. Try to measure again.
Replace the O2sensor.
O2% indi­cation not stable in the air
Service Indications / Trouble Shooting
Possible reasons:
The OxySpy has been calibrated in extremely humid air.
How to solve:
Calibrate the OxySpy in pure oxygen (100%) or in compressed air from a tank!
The indi­cated O2-value is inaccurate
The measured O2-value deviates from the actual O
2
-content by more than 1 or 2 percent.
Page 19
O2Sensors
19
The
O
2
sensors have, like batteries, a certain lifetime. From the first time that a sensor contacts oxygen it starts to be used up. They can be replaced by the user (see "Fitting the
O2sensor",
page 8).
Disconnecting the
O
2
Sensor from the socket does not extend it's life. On page 20 you will find more information about storing the OxySpy.
After opening the airtight package the
O2sensor requires 30 minutes to stabilize in air. Calibration should be made only after this "warm-up" time.
The battery and the O
2
sensor contain corrosive liquid. If the battery or O2sensor leaks: Do not let the liquid come into contact with skin, eyes or mouth. If it does, flush the affected area with water for at least 15 minutes. Consult a doctor. Do not inhale vapor. Immediately clean the leakage from the instrument with a soft towel and dispose of the towel.
English
Page 20
Maintenance and Storage
20
• Protect the unit from impact and intensive sun and heat.
• Store the unit in a cool, not too dry place. This helps to preserve the O
2
sensor life.
• Ideal storage temperature: 5 °C up to +15 °C (41 °F up to 59 °F)
• Do not leave the OxySpy in an oxygen enriched atmosphere except when it is necessary.
• Disconnecting the
O2Sensor from the socket
does not extent its life.
• Replace the battery and
O
2
sensor early.
• Do not clean the unit with any solvent (except water) or detergent.
• Clean the OxySpy with a damp, soft towel.
• Dry the instrument with a soft towel.
• Do not scratch the display window.
• Maximum altitude: 4000 m (13,000 feet)
above sea level
• Temperature range: 0 °C up to +50 °C
(32 °F to 122 °F)
• Relative air humidity: 10-90%,
non-condensing
•O
2
Display: • Resolution: 0.1% O
2
• Display range: 0.0 up to 199.9%O
2
•O2Sensor: • Type Dynatron P/N DiveceL3.D1
or other O2sensors if licensed by the manufacturer.
• Lifetime: minimum 1 year*
• Battery: • Type: CR2430
• Lifetime: min. 3 years at 500 on/ off cycles per year
• Linearity: <3% (at constant pressure and temperature)
Maintenance and Storage
English
Operational limits
Technical Information
*
Lifetime depending on environment conditions and the used product.
Page 21
Limited Warranty
21
Please pay attention to the following remarks on warranty claims:
The limited warranty only covers units which were purchased from an authorized retailer.
The manufacturer will repair all defects which are due to faulty materials or workmanship. The war­ranty covers the repair of the unit, free of charge. This includes the repair or replacement of the faul­ty part or the replacement of the entire OxySpy. The
O
2
sensor, the battery and the rubber rings are excluded from the warranty. The manufacturer reserves the right to determine the merits of the warranty claim and to determine whether the unit will be repaired or replaced.
The warranty is given for a period of two years after original date of purchase. Repairs or replace­ments during the warranty period do not extend the warranty period. In order to submit a warranty claim, deliver the unit, along with dated proof of purchase, to your authorized retailer or an authori­zed service location.
Please confirm that you have read the complete manual.
Recognition of warranty
Scope of warranty
English
• improper operation or stress
• exterior influences, e.g. shipping damage, impact, damage due to bumping and hitting
• water damage
• influences of weather or other phenomena
• repairs preformed or attempted by unautho­rized service agencies
• pressure chamber tests
Excluded are faults or defects due to:
Warranty period and claim:
Place: Date: Signature: ____________ ____________
_______________________
Page 22
Notes
22
English
Page 23
FCC / Copyright ©
23
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harm­ful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Copyright © 2003 by
Dynatron AG All rights reserved. No portion of this manual may be reproduced, processed or distributed in any form (print, photocopy, electronic or any other process) without prior, written permission from Dynatron AG.
Protect the environment! When disposing of this OxySpy, do so in an environmentally friendly way. Return old products to an appropriate collection point according (EU Directive 2002/96/EC). The OxySpy must not be placed in the normal household waste.
FCC
English
Copyright ©
Waste disposal
Page 24
English
24
Stamp of the dealer:
Date of Sale:
Serial No. of your OxySpy:
Warranty paper
Operating Manual OxySpy / OS-M-GB-B Rev.A
Printed in Switzerland 09-2006
www.oxyspy.biz
Loading...