CAUTION: Federal law restricts this device to sale by or on the order of a physician,
chiropractor, physical therapist, or dentist licensed by the law of the state in which said
person practices to use or order the use of the device.
IMPORTANT: Before treating a patient with the Dynatron 850plus or the Dynatron
550plus, see the “Contraindications, Warnings, and Precautions” in this manual.
INDICATIONS FOR USE:
Muscle stimulation therapy delivered by the Dynatron 850plus and 550plus is indicated
for the following uses:
1. relaxation of muscle spasm;
2. prevention or retardation of disuse atrophy;
3. increasing local blood circulation;
4. muscle re-education;
5. immediate postsurgical stimulation of calf muscles to prevent venous thrombosis; and
6. maintaining or increasing range of motion.
Interferential, premodulated, high volt, and microcurrent therapy are indicated for
providing temporary relief of chronic intractable pain.
INDICATIONS FOR ULTRASOUND USE:
Ultrasound therapy is intended to generate deep heat within body tissues for the treatment
of selected medical conditions such as relief of pain, muscle spasms, and joint
contractures, but not for the treatment of malignancies.
(21 CFR 890.5300)
Dynatron® 850plus and Dynatron® 550plus Service Manual
Introduction to the Dynatron 850plus™ and 550plus™ ........................................................................ 2
New Features..........................................................................................................................................2
Before You Treat a Patient.....................................................................................................................4
How to Use This Manual........................................................................................................................ 4
Installation and Features............................................................................................................................ 5
High Volt Probe (optional accessory) ..................................................................................................17
Quick Reference of Special Key Presses .................................................................................................18
Section II
Operation and Treatment Instructions
General Operating Instructions............................................................................................................... 20
Power On/Off ....................................................................................................................................... 20
High Volt Instructions...............................................................................................................................35
Set Up High Volt Treatment with Electrodes .......................................................................................35
Set Up High Volt Treatment with Optional Probe................................................................................36
Basic High Volt Setup ..........................................................................................................................37
Detailed High Volt Setup......................................................................................................................37
High Volt Default Settings....................................................................................................................41
Audio Tone ...........................................................................................................................................46
High Volt Therapy .................................................................................................................................... 76
High Volt Therapy with Electrodes...................................................................................................... 76
High Volt with Optional Probes........................................................................................................... 76
High Volt Waveform............................................................................................................................ 76
High Volt Settings................................................................................................................................ 76
Lead Wires............................................................................................................................................92
Test Leads .............................................................................................................................................93
vi
Dynatron® 850plus & 550plus
Current Limit........................................................................................................................................ 93
Current Limit Warning......................................................................................................................... 94
Problem Solving ...................................................................................................................................... 109
Soundhead Temperature Too Hot ......................................................................................................109
Cooling the Soundhead ...................................................................................................................... 109
Soundhead Temperature Too Cold..................................................................................................... 110
No Soundhead ....................................................................................................................................110
Other Error Messages in the Display.................................................................................................. 110
Section VI
Technical Information
General Specifications ............................................................................................................................ 114
Dynatron 850plus and 550plus Specifications ................................................................................... 114
The Dynatron 850plus and Dynatron 550plus offer the practitioner a wide range of treatment
options in a single device. Both devices provide interferential and premodulated therapy,
high voltage pulsed stimulation, Russian and biphasic stimulation, and microcurrent. The
850plus also includes ultrasound and combination therapy.
The devices include the advantages of earlier Dynatronics models, such as customizable
treatments with convenient setup. In addition, these units are smaller and lighter in weight
than earlier models and now offer the option of battery operation, making the devices truly
portable. One additional channel is now dedicated to high volt therapy, allowing a greater
variety of simultaneous treatments. The manufacturer’s warranty for these devices is now
two years (see full warranty details at the back of this manual).
This manual provides operator information and instructions for both of these models. The
sections that discuss ultrasound and combination treatments apply only to the Dynatron
850plus model. All other sections of this manual apply to both devices. Sections of this
manual that apply only to the Dynatron 850plus are clearly marked.
New Features
For models having software greater than Rev. 1.01 or later, the following new features have
been added (effective January 1998).
• Conductance Bar Graph. A graphical representation of conductivity during and
• Ultrasound Coupling Bar Graph. Shows soundhead coupling during an ultrasound
• Soundhead Temperature Bar Graph. Indicates when soundhead temperature is too hot
• “Lead” warning. A new warning displayed if the device detects an open circuit,
• New warnings notifying the user when the soundhead is too hot, or when the soundhead
These new features are described in greater detail in the treatment instructions later in this
manual. The determine the software revision level for the Dynatron device, power the device
on and observe the number shown in the Time display at the end of the device’s startup
routine.
Dynatron® 850plus & 550plus
Introduction to the
interferential, premodulated, or microcurrent treatment.
treatment.
during a treatment.
indicating no patient current is being delivered and the treatment setup must be inspected.
is not detected by the device
Introduction
2
Treatment Options
The following modalities are provided by both the Dynatron 850plus and the Dynatron
550plus:
Interferential Therapy: This traditional therapy, delivered with two channels and
four electrodes, also includes Dynatronics’ patented Target feature for easy location
of the treatment site and delivery of the full interferential beat where it is needed.
You can select the high frequency range, the low frequency range, or both, along
with Target. Or select Sweep when you want to deliver the current over a more
general area. You can also customize the frequency settings for individual treatments
or save a unique default setting.
Premodulated Therapy: Ideal for treating areas where four electrode-placement is
not practical, this modality extends the versatility of the device by offering a
premodulated interferential current using one channel and two electrodes.
Biphasic Stimulation: This option provides muscle stimulation using a sinusoidal
biphasic waveform with pulse widths, ranging from 50 to 400 microseconds. It
offers normal, co-contraction and reciprocal contraction modes, selectable on/off
cycle times and ramp times, and customizable pulse rate and pulse width.
Russian Stimulation: With this modality, you can deliver muscle stimulation
treatments selecting from an array of options including normal, co-contraction and
reciprocal contraction, selectable on/off cycle times and ramp times, and
customizable pulse rate and pulse width.
High Voltage Pulsed Stimulation: This traditional electrotherapy modality offers a
pulsed, monopolar current with pulse widths in the microsecond range and pulse rates
ranging from 1 to 200 Hz utilizing a twin-peak monophasic waveform. Polarity,
on/off cycle time, and ramp time are selectable, and pulse rate may be customized for
each treatment. Treatment may be delivered using either electrodes or optional
probes.
Microcurrent Therapy: Microcurrent treatment may be delivered using two
electrodes for unattended therapy, or the optional hand-held probes for manual
treatment. You may select the desired polarity and customize the frequency and
intensity. An audible tone allows you to monitor conductance.
The following additional modalities are provided with the Dynatron 850plus (not included
with the Dynatron 550plus):
Ultrasound Therapy: The Dynatron 850plus includes Dynatronics’ patented multifrequency ultrasound capability. Soundheads are available in 1 cm
and 10 cm
offers 2 and 3 MHz.) Duty cycle options are 10, 20, or 50 percent, or a continuous
duty cycle.
Combination Therapy - Comboplus™: Dynatronics’ new Comboplus feature gives
you almost unlimited options in setting up a combination treatment with the
Dynatron 850plus. Now you can combine an ultrasound treatment with any singlechannel electrotherapy modality provided by this device, including Premodulated,
Biphasic, Russian, High Volt, or Microcurrent. In addition, combination therapy is
Dynatron® 850plus & 550plus
2
2
sizes. Each soundhead offers 1, 2, and 3 MHz. (The 1 cm2 soundhead
, 2 cm2, 5 cm2,
3
no longer restricted to just one channel—any channel may be used. This means you
may set up a combination treatment using any of the three electrotherapy output
jacks. Combination therapy offers simultaneous treatment time and allows you to
select from all the available options for both modalities used in the combined
treatment.
Simplified Setup
The unique design of the Dynatron 850plus and 550plus front panels means treatment setup
has never been easier. A few simple key presses are all you need to fully set up a treatment.
The careful grouping of available options for each modality ensures that you can easily see
and select from the appropriate options for that modality.
Each modality offers default settings which are automatically selected when the modality is
selected—saving time in the treatment setup. You can change these defaults to match your
own most common treatment setups to reduce setup time to a matter of seconds.
Before You Treat a Patient
Before administering a treatment to a patient with the Dynatron 850plus or 550plus, you
should familiarize yourself with all the operating instructions for the modality used, as well as
the contraindications, warnings, and precautions for that modality.
You should also read in this manual general information about each of the modalities. In
addition to the information provided in this manual, consult other published sources for
additional application and safety instructions regarding use of each type of therapy.
How to Use This Manual
This manual is organized to help you learn how to use your Dynatron 850plus and 550plus
device, and to provide easy reference for looking up specific instructions or information. The
information in the manual is grouped as listed below. Note that all treatment instructions for all
modalities are in one section, and that all contraindications, warnings, and precautions for all
modalities are provided together in another section.
This grouping of information makes it easy for you to find what you are looking for. To assist
you further, descriptive titles appear at the foot of each page to help to identify immediately
what section you are in. Familiarize yourself with the organization and presentation of this
manual to make best use of the information provided.
Section I: Introduction, Installation and Features
Section II: Operation and Treatment Instructions
Section III: General Modality Information
Section IV: Contraindications, Warnings, and Precautions
Section V: Ultrasound Technical Information
Section VI: General Technical Information
Section VII: Patient Information
Dynatron® 850plus & 550plus
Introduction
4
Installation and Features
Unpacking
When you receive the unit, immediately unpack it and all accessories and check for possible
damage, obvious or concealed. In case of damage, immediately notify the carrier and take
any steps necessary to file a claim for the damage sustained. Do not destroy the shipping
carton. The carton should be reused if the device must be shipped for any reason. The carton
is specially designed to protect the unit from shipping damage. Improper packaging of the
unit during transport can result in damage and invalidate the warranty.
Complete the warranty registration form and return it to Dynatronics within 10 days of
purchase. This is essential to insure you are not billed for services that are covered by
the warranty policy. Warranty registration should include serial numbers for both the
device and soundheads.
Connect the AC power cord, which is equipped with a hospital grade, UL listed plug, to a
properly grounded 110/120V 60 Hz AC outlet (the device will automatically switch to
220/240V 50 Hz when connected to a power source with that voltage). The power cord must
also be firmly plugged into the device itself. When the cord is properly connected, it can not
be easily pulled out. Do not place the cord or the device in a place where the cord could be
tripped over or accidentally pulled out of its socket during a treatment.
Read the operating instructions in this manual before proceeding with a treatment.
Standard Accessories
The following accessories are included with the Dynatron 850plus or 550plus units:
Qty Part No. Description
1 D851/ D551 Dynatron
1 9B0031 Power Cord
1 8A0061 Patient Remote Stop Cable
1 8H0006 Operator's manual
2 7B0230/ 7B0231 72” double leads (1 red, 1 black)
1 7B0234 Special lead wire for combination treatments (with Dynatron
1 7B0204 POLYS™ self-adhesive electrodes 1.75” x 3.75” w/pin
4 7B0063/ 7B0065 3” round carbon electrodes (2 red, 2 gray)
4 7B0191 Sponge fabric for use with carbon electrodes
2 DW248 2” x 48” straps
1 7B0191 5” x 8” dispersive electrode for high volt w/ sponge fabric
1 7B0202 Applications manual
1 7B0217 Ultrasound gel sample (with Dynatron 850plus only)
Dynatron® 850plus & 550plus
® 850plus or Dynatron® 550plus unit
850plus only)
connector (pkg of 4)
Installation & Features
5
Soundheads and Probes
The Dynatron 850plus may be purchased with one or more applicator soundheads in the
following sizes:
Part No. Size Frequencies
SH01CM 1 cm
CM02SH 2 cm
CM05SH 5 cm
CM10SH 10 cm
An optional high volt probe or microcurrent probe may also be purchased for use with this
device. Contact your Dynatronics dealer to order optional accessories
Accessories
The following optional and replacement accessories may be purchased from Dynatronics or
from your Dynatronics dealer:
Part No. Description
8E0026 High volt probe (requires one or more applicators)
8E0017 High volt applicator 2” round
8E0018 High volt applicator 5/8” round
8E0019 High volt applicator 2”x1-1/2”
8D0007 Microcurrent active probe
8D0027 Microcurrent ground probe
7B0208/ 7B0209 2” diameter carbon electrodes (red and gray)
7B0063/ 7B0065 3” diameter carbon electrodes (red and gray)
7B0059/ 7B0061 3” x 5” carbon electrodes (red and gray)
7B0067/ 7B0069 1.5” x 2.0” carbon electrodes (red and gray)
7B0003/ 7B0203 1.75” x 3.75” adhesive electrodes (with snap or pin connector)
7B0005/ 7B0204 1.75” square adhesive electrodes (with snap or pin connector)
7B0205 1.25 round adhesive electrodes (with pin connector)
7B0206 2” round adhesive electrodes (with pin connector)
7B0207 3” round adhesive electrodes (with pin connector)
7B0139 120” lead (black)
7B0141 120” lead (red)
7B0077 Bifurcated extension lead wire
7B0081 Banana-to-pin adapter
7B0001 Snap adapter
7B0161/ 5LTRGEL Ultrasound coupling gel (8 oz.bottle or 5 liter container)
Dynatron® 850plus & 550plus
2
Operates at 2 and 3 MHz
2
Operates at 1, 2, and 3 MHz
2
Operates at 1, 2, and 3 MHz
2
Operates at 1, 2, and 3 MHz
Installation & Features
6
64
Dynatron® 850plus & 550plus
7589123 1011
14
13
12
16
17
18
HV
1
MICRO
2
CH
PREMOD
SOUND
COMBO
SOUND
IFC
CH 1CH 2
TIME
HIGH
LOW
1 MHz
2 MHz
3 MHz
TIME
FREQ
RATE
WIDTH
TARGET
SWEEP
10%
20%
50%
CONT
START
STOP
PAUSE
FUNCTION
TARGET
POWER-INTENSITY
NORMAL
CO-CONT
RECIP
10/10
10/30
10/50
CONT
.5 RAMP
1.0 RAMP
1.5 RAMP
2.0 RAMP
HV
BIPHASIC
RUSSIAN
HI VOLT
MICRO
261525
Dynatron 850plus Control Panel
Dynatron 850plus and Dynatron 550plus Physical Features
Before operating the Dynatron 850plus or Dynatron 550plus, acquaint yourself with the
control panel by reviewing the illustrations and descriptions on this and the following pages.
The numbered features in the diagrams correspond to the numbered descriptions. Before
administering treatment to a patient, read the sections later in this manual that provide
specific instructions for performing treatments, discussions of each modality, definitions of
the available options, along with contraindications, warnings, and precautions for all
modalities.
Note that some options use “toggle” keys for making selections. More specific instructions
for using toggle keys are provided later in this section.
General Selections:
1. START: Pressing this key starts the treatment timer, and treatment proceeds as set up.
2. STOP: Pressing this key during a treatment IMMEDIATELY stops the output and sets
the treatment time to zero for all modalities. Treatments may also be stopped by pressing
the button on the patient REMOTE STOP cable. You can also stop a treatment for one
treatment only by pressing and holding the Function
21
19
+
-
+
-
22
20
23
24
Installation & Features
7
64
Dynatron® 850plus & 550plus
7589123 1011
14
13
12
CH
PREMOD
IFC
HV
1
MICRO
2
Dynatronics
CH 1CH 2
Dynatron 550plus Control Panel
key while you press the Stop key, or by reducing that channel’s treatment time to zero.
3. PAUSE/FUNCTION: This key is used in combination with other key presses for
accessing unique features including: select polarity (high volt and microcurrent), audio
volume control (microcurrent), and to stop one treatment only. Specific instructions for
using this key are provided later in this manual.
For Dynatron 850plus only: This key is also used to pause an ultrasound treatment. First
press the channel toggle (CH) to select SOUND, then press PAUSE/FUNCTION; the
ultrasound output is stopped, the treatment time is paused, and the light on the Pause key
is on. When this key is pressed again, the ultrasound treatment resumes and the light on
the Pause key is off. The PAUSE option is not available with the Dynatron 550plus since
that model does not offer ultrasound. NOTE: During a Combo treatment, only the
ultrasound output and the treatment timer are stopped when you press Pause; the stim
output continues.
4. TIME ARROW KEYS: These UP/DOWN arrow keys are used to increase/decrease the
treatment time or other parameter that is displayed in the Time display.
5. TIME DISPLAY: This display is used to show the treatment time for one treatment at a
time; the display shows treatment time for the selected channel (the selected channel is
indicated by the green LED—all other channels in use at the
TIME
HIGH
LOW
TIME
FREQ
RATE
WIDTH
TARGET
SWEEP
START
STOP
FUNCTION
TARGET
POWER-INTENSITY
NORMAL
CO-CONT
RECIP
10/10
10/30
10/50
CONT
.5 RAMP
1.0 RAMP
1.5 RAMP
2.0 RAMP
HV
261525
BIPHASIC
RUSSIAN
HI VOLT
MICRO
21
19
+
-
+
-
22
20
23
24
Installation & Features
8
Dynatron® 850plus & 550plus
time will have yellow LEDs). The Time display can also display the pulse rate and width
for Russian and biphasic treatments, as well as the frequencies for interferential,
premodulated, and microcurrent treatments, and pulse rate for high volt. The treatment
parameters for any treatment in progress may be displayed at any time by first using the
channel toggle key to choose the desired channel then using the Time toggle key to select
the desired parameter (Time, Freq, Rate, Width).
6. CHANNEL TOGGLE KEY(CH): When a treatment is in progress, you can press this
key to choose an output channel and display the parameters for the treatment being
delivered by that channel. When a channel light is GREEN, the displays show the
settings for that channel. The available options depend on the modality selected. When
two or more treatments are in progress simultaneously, the toggle key is used to select the
channel you wish to view.
7. CHANNEL SELECTIONS: These lights indicate which output channels are currently in
use. A solid GREEN light indicates current is being delivered to that channel; the time,
intensity and other treatment parameters for that channel are also displayed. A solid
YELLOW light indicates a channel is in use and delivering current, but the time, intensity,
and treatment parameters are not displayed at this time (only one channel’s time and
intensity may be displayed at a time). Flashing green or flashing yellow indicate the “off”
segment of a biphasic, Russian or high volt treatment cycle. The channel’s intensity and
other treatment parameters may only be modified when it has a green indicator light. Press
the Channel Toggle key (CH) to select a channel to be viewed.
8. TIME TOGGLE KEY: Press this key to display various treatment parameters in the
Time display including Time (treatment time), Freq (frequency), Rate (pulse rate), and
Width (pulse width). Available options during a given treatment or treatment setup
depend on the modality selected.
9. TIME GROUP SELECTIONS: These lights indicate the parameters that are displayed
(one at a time) in the Time display. The default selection is the treatment time. Press the
Time toggle key to select the desired option (available options depend on the modality
selected). When a parameter is selected, its indicator light is green, its value is displayed
in the Time display above, and the time arrow keys may be used to change the value.
The device returns to the Time display after 10 seconds with no key presses.
10. POWER-INTENSITY DISPLAY: This display is used to display the treatment power (for
ultrasound) or intensity (for all other modalities) for the currently selected channel (the
selected channel is indicated by the green LED—all other channels in use at the time will
have yellow LEDs). Press the channel toggle key to select the desired channel to be
viewed.
11. POWER - INTENSITY ARROW KEYS: The arrow keys are used to increase/decrease
the intensity or power of one treatment. Changes made to power and intensity affect only
the currently selected channel (the selected channel is indicated by the green LED—all
other channels in use at the time will have yellow LEDs). Press the channel toggle key to
select the desired output channel. The arrow keys may then be used to change the
intensity or power for that channel.
Installation & Features
9
Dynatron® 850plus & 550plus
IFC / Premod Selections:
12. IFC/PREMOD: Press this key once to begin setup of an interferential treatment (the IFC
LED is lighted); press this key twice to begin setup of a premodulated treatment (the
Premod LED is lighted). When you select IFC, both channels 1 and 2 are automatically
selected and the green LED lights for those two channels will be on. Connect two leads
to the output jacks for the channels that are selected. When you select PREMOD, a
single channel (CH1 or 2) is automatically selected and that channel’s green LED will be
on. Connect one lead to the output jack that corresponds to the channel indicated by the
green LED.
13. HIGH/LOW TOGGLE (used with interferential, premodulated and high volt): Press this
key one or more times to select the desired frequency range for interferential and
premodulated treatments or the pulse rate range for high volt treatments. The green LED
indicates the option selected. For example, press the High/Low toggle key once to select
High, press again to select Low, press again to select High/Low Alternating, and press
again to select High/Low Consecutive. For high volt treatments, you can select High or
Low only, but not both. During a treatment, the current sweeps through the range(s)
selected.
For interferential and premodulated, the HIGH frequency range is initially set at 80 to
150 Hz; and the LOW frequency range is 0 to 10 Hz. For high volt, the HIGH pulse rate
range is initially set at 80 to 120 Hz; and the LOW pulse rate range is 1 to 10 Hz. These
frequency ranges may be modified for every treatment, if desired and new default settings
for the device may also be saved. See treatment setup instructions later in this manual for
a complete description of the options that may be selected.
14. TARGET/SWEEP TOGGLE: This key is pressed to select either Target, Target Sweep,
or static treatment for an interferential treatment. The Target pad is used to locate the
exact treatment site (if the Target option is selected during an interferential treatment).
15. TARGET PAD: For use during interferential treatments when “Target” is selected.
Touch the TARGET pad at different points on the pad to reach the precise treatment site.
When you lift your finger from the Target pad, the selected point is locked until you
change it again. This patented feature is used to locate the specific treatment site in an
interferential treatment and to place the point of interference at that point.
Ultrasound Selections (DYNATRON 850plus ONLY—the ultrasound feature is
not available on the Dynatron 550plus):
16. SOUND/COMBO: Press this key once to begin setup of an ultrasound treatment (the
Sound LED on this key is lighted as well as the Sound LED in the channel indicator
area); press this key twice to begin setup of a combination treatment (the Combo LED is
lighted as well as the Sound LED and a single Channel LED in the channel indicator
area). When either of these options is chosen, the soundhead should first be plugged into
the ultrasound output jack on the side panel. For combination treatments, the special
combo lead wire should be attached to the output jack selected (see full instructions for
setting up a combination treatment later in this manual). In the COMBO mode, the
electrotherapy treatment is delivered through the soundhead and through a single
electrode which is placed on the patient. Only single-channel electrotherapy options are
available in the COMBO mode.
Installation & Features
10
Dynatron® 850plus & 550plus
17. SOUND FREQUENCY TOGGLE: This key is pressed one or more times to select the
desired ultrasound frequency; 1 MHz, 2 MHz, or 3 MHz.
18. DUTY CYCLE TOGGLE: This key is pressed one or more times to select the desired
duty cycle for ultrasound treatment. Options are 10, 20, or 50 percent, or Continuous.
Russian / Biphasic / High Volt Selections:
19. BIPHASIC/RUSSIAN: Press this key once to begin setup of a biphasic treatment (the
Biphasic LED is lighted); press this key twice to begin setup of a Russian treatment (the
Russian LED is lighted). Biphasic and Russian treatments use a single channel (1 or 2)
when the Normal mode is selected; and a channel pair (1 and 2) when the Reciprocal or
Co-contraction mode is selected.
20. HIGH VOLT: Press this key to begin setup of a High Volt treatment (the Hi Volt LED is
lighted). The HV output channel is automatically selected (the LED for the channel
selected is GREEN). Connect the patient lead wire to the HV output jack.
HIGH VOLT POLARITY: To select or change the polarity of a high volt treatment, use
the HI VOLT key together with the FUNCTION key. Press and continue holding the
FUNCTION key while pressing the Hi Volt key one or more times to select positive only
(the “+” LED is lighted), negative polarity only (the “-” LED is lighted), or both (both
LEDs are lighted).
21. TREATMENT MODE TOGGLE (for biphasic and Russian treatments): Press this key
one or more times to select Normal, Co-Contraction, or Reciprocal contraction. The
output channel is automatically selected. When Normal is selected, one output jack only
is selected (1 or 2). When Co-contraction or Reciprocal is selected, channels 1 and 2 are
selected. Connect the patient lead wire(s) to the output jack(s) for the channel(s)
selected.
22. CONTRACTION/REST CYCLE TOGGLE (for Russian, biphasic, and high volt
treatments): Press this key one or more times to select the desired contraction/rest
(on/off) cycle . Available cycles include 10/10, 10/30, 10/50, and Continuous. The first
value indicates the on-time in seconds, and the second value indicates the off-time. For
example; 10/30 indicates the current is on (muscle is contracting) for 10 seconds, and
current is off (muscle is relaxed) for 30 seconds. With Continuous mode, current is
applied continuously with no off cycle. The continuous duty cycle is not recommended
for electrical muscle stimulation, but may be used for settings that are intended to effect
other results than a muscle contraction.
23. RAMP TIME TOGGLE (for Russian, biphasic, and high volt treatments): This key is
pressed to select the ramp time . The ramp time is applied before and after the “On”
segment of the cycle (it provides both a ramp up and a ramp down). Available ramp
times are .5, 1, 1.5, and 2 seconds.
NOTE: The High/Low option described previously under the Interferential/Premod options,
is also used for High Volt treatments (see item 13 above).
11
Installation & Features
Dynatron® 850plus & 550plus
Microcurrent Selections:
24. MICRO: Press this key to begin setup of a microcurrent treatment. This key is also used
to turn the conductance tone off and on after a microcurrent treatment is started. When
MICRO is selected, Channel 1 is automatically selected for the default electrodes
treatment and the LED for that channel is lighted. For a microcurrent treatment setup
with electrodes, connect a patient lead wire to Channel 1 output jack. For a microcurrent
probes treatment, press the channel toggle key (CH) to select the MICRO channel
selection (Channel 1 LED will become yellow and MICRO will become green to indicate
the probes option is selected). Connect the microcurrent probe to the microcurrent output
jack on the left side panel of the device.
NOTE: Channel 1 is committed to the microcurrent output during a probes treatment as
well as during a treatment with electrodes, and is not available for use by any other
modality while any microcurrent treatment is in progress.
MICROCURRENT POLARITY: To select or change the polarity of a microcurrent
treatment, use the MICRO key together with the FUNCTION key. Press and continue
holding the FUNCTION key while pressing the MICRO key one or more times to select
positive only (the “+” LED is lighted), negative polarity only (the “-” LED is lighted), or
dual polarity (both LEDs are lighted).
25. OUTPUT JACK CHANNELS 1 and 2: These are the output jacks for delivering
interferential, premodulated, Russian, biphasic, and microcurrent treatments. Consult
treatment instructions later in this manual for appropriate channels and treatment setup
for each modality.
26. HIGH VOLT OUTPUT JACK HV. This is the output jack for delivering high volt
treatments. Consult treatment instructions later in this manual for high volt treatment
setup.
Dynatron 850plus and
550plus Back Panel:
27. POWER ON/OFF
28. Electrical Stim to Ultrasound
Input Jack (only on D850plus)
29. Jack for Optional Battery Pack
SWITCH: Located on the
back of the unit this switch
is labeled “1” and “0”. Set
the switch to “1” for ON;
set the switch to “0” for
OFF.
1
0
Batt. Input
28. STIM INPUT JACK FOR
COMBINATION
TREATMENTS: The special
lead wire for combination
treatments is plugged into this
jack for a combination
27. Power On/Off Switch
Dynatron 550plus and 850plus Back Panel
30. Patient Remote Stop Jack
treatment setup providing stim output through the ultrasound head. The special lead wire is
also plugged into the jack on the front of the device which has been selected for the specific
combo treatment. See combination treatment instructions later in this manual for detailed
information regarding combination treatment setup.
Installation & Features
12
Dynatron® 850plus & 550plus
9. BATTERY: This jack may be used to supply power to the device using an optional
battery pack. More information about the optional battery operation is provided later in
this manual.
30. REMOTE STOP: A cord with a remote stop button is inserted in this jack. The remote
stop is controlled by the patient during unattended therapy to allow the patient to stop the
treatment at any time. When the button on the remote stop cable is pressed, output for all
stim modalities is stopped and the tone sounds briefly. During Combo treatments, both
sound and stim outputs are stopped.
Right side of device (D850plus only)
UltrasoundMicro Probe
Left side of device
31. Ultrasound Ouput Jack
Side Panels
Dynatron 850plus and 550plus Side Panels:
31. ULTRASOUND OUTPUT JACK (DYNATRON 850plus ONLY—not available on the
Dynatron 550plus): The applicator soundhead plugs into this jack for ultrasound therapy.
32. MICROCURRENT OUTPUT JACK: The microcurrent probe plugs into this jack for
microcurrent probe therapy.
Instructions for Using Toggle Keys
Toggle keys are used to make selections from two or more options in a given area. Toggle
keys are pressed one or more times to make a desired selection. A GREEN light (LED) next
to the toggle key shows the option that has been selected. Pressing the toggle key one or
more times allows you to scan through the available options.
Each toggle key has unique capabilities. Most toggle keys allow only one selection. For
example, the Ramp toggle key requires you to select just one of the four ramp times
available. However, some toggle keys allow you to select two options. For example, in
interferential you can press the High/Low toggle key once to select High, press again to select
Low, and press again to select both High and Low.
The following is a list of all the toggle keys available with each modality:
32. Ouput Jack for Microcurrent Probe
13
Installation & Features
Dynatron® 850plus & 550plus
IFC and Premod
• Target/Sweep (IFC only)
• High/Low Frequency Ranges
Russian and Biphasic Stim
• Treatment mode
• Contraction/Rest cycle
• Ramp Time
High Volt Stim
• Contraction/Rest cycle
• Ramp Time
• Polarity
• High/Low Pulse Rate Ranges
Microcurrent
• Microcurrent Polarity
• Channel toggle to select Micro or Channel 1 (during setup only)
All Modalities
• Time/Frequency/Rate/Width
• Channel Toggle (to view individual channels one at a time while treatments are in
progress
Ultrasound (Dynatron 850plus only)
• Ultrasound Frequency
• Ultrasound Duty Cycle
Stim Intensity Display
The intensity display is intended as an
incremental value only for the convenience
of the practitioner.
POWER-INTENSITY
Ultrasound Power Display
Ultrasound power is displayed, with the value
shown either in WATTS or W/cm2.
POWER-INTENSITY
Increasing the intensity does increase
current delivery. However, intensity and
current are not a 1:1 comparison. Actual
current delivered to the patient depends
upon the intensity setting and the current
density (affected by the size and
condition of the electrodes).
Microcurrent Intensity Display
Microcurrent intensity is displayed in
microamperes
POWER-INTENSITY
Examples of Power-Intensity displays for the Dynatron 850plus and 550plus.
A blinking decimal indicates the power is displayed
as WATTS. A steady (non-blinking) decimal
indicates WATTS/cm2. Pressing and holding the
PAUSE key changes the display from WATTS to
W/cm2 or reverse.
High Volt Intensity Display
High volt intensity is displayed in volts
POWER-INTENSITY
Installation & Features
14
Time and Power Displays
The “Time” display and the “Power-Intensity” display can show the settings for only one
channel at a time. The Time and Power-Intensity settings displayed are for the channel with
the GREEN light only. Any other channel in use at that time will have a YELLOW light to
show it is active but its parameters are not currently displayed.
To view the settings for another
channel or output, press the channel
toggle key one or more times until
the light for the desired channel
becomes GREEN. The GREEN
light appears next to a different
channel or output each time you
press the toggle key, and the Time
and Power-Intensity displays change
to show the parameters currently in
effect for that channel.
The green and yellow channel lights
each will also appear solid (nonflashing) or flashing. A solid light
means current is being delivered to
the channel at this time (for example, during the ON cycle of the Russian stimulation
treatment). A flashing light means current is not being delivered to the channel at this time
(for example, during the OFF cycle of the Russian stimulation treatment).
Dynatron® 850plus & 550plus
When a channel is selected its parameters are displayed in
the Time and Power-Intensity displays, and treatment
parameters for that treatment may be modified.
Press the
channel toggle
key to select any
channel that is
currently in use.
The LED for the channel selected becomes GREEN and the
LEDs for all other active channels become YELLOW.
Channel Toggle Key
CH
TIME
HV 1
1
MICRO
2
SOUND
TIME
FREQ
RATE
WIDTH
CHANNEL / OUTPUT INDICATOR LIGHTS
GREEN Solid • You CAN see this channel’s parameters displayed on Time
and Power-Intensity displays.
• The channel IS delivering current. GREEN Flashing • You CAN see this channel’s parameters displayed on Time
and Power-Intensity displays.
• The channel IS NOT delivering current. YELLOW Solid • You CANNOT see this channel’s parameters displayed on
Time and Power-Intensity displays.
• The channel IS delivering current. YELLOW Flashing • You CANNOT see this channel’s parameters displayed on
Time and Power-Intensity displays.
• The channel IS NOT delivering current.
Ultrasound Heads (Dynatron 850plus only)
The Dynatron 850plus features the new Dynatronics SmartHead™ ultrasound applicator
soundhead that has all calibration information self-contained. SmartHeads may be used
interchangeably with any 50 Series Plus model or the Dynatron 125 model with no need for
entering calibration numbers into the device.
15
Installation & Features
Dynatron® 850plus & 550plus
Soundheads for the Dynatron 850plus come in 1 cm2, 2 cm2, 5 cm2, and 10 cm2 sizes. The
2
, 5 cm2, and 10 cm2 soundhead operate at 1, 2, and 3 MHz; while the 1 cm2 soundhead
2 cm
operates at 2 and 3 MHz. The soundheads are waterproof allowing ultrasound (not
combination) therapy to be administered in water, if desired.
Do not drop the soundhead nor allow it to strike a hard surface. Do not place soundheads
into ice water. Do not allow soundheads to repeatedly reach maximum head temperature.
All of these conditions can cause damage to the crystal. Any such damage is not covered by
the warranty.
For new or replacement soundheads, contact Dynatronics or your authorized Dynatronics
representative.
Microcurrent Probes (optional accessory)
The Dynatron 850plus and Dynatron 550plus devices offer optional probes for delivering
hands-on microcurrent treatment. The microcurrent probes can only be used for microcurrent
therapy. It is not possible to select the probe output nor to use the microcurrent probe to
deliver a probe treatment for interferential, premodulated, Russian, biphasic, or high volt
modes.
The setup includes an active probe along with either a ground probe or a ground electrode.
Use the end of a cotton swab (such as a Q-Tip
probe and the ground probes. Cut the end of the swab to a short length; the cotton must touch
the probe’s metal ring. Use water or a conductive electrolyte spray to wet the cotton swab
before treating.
Active Probe: To deliver the current to the patient through the active probe, press and
release the button on the probe. Press and release the button again to stop the current. The
active probe should touch the patient’s skin at the treatment site, and the ground probe should
touch the patient’s skin elsewhere. This completes the circuit and delivers current to the
patient. You do not need to hold the button down. Once you have pressed and released the
button, the current is delivered until you press and release the button again to stop the current.
While you are delivering
current, the treatment
timer counts up in
seconds from zero. When
the current is stopped, the
timer returns to zero.
Press the button again to
commence the next time
sequence. Continue in
this way until treatment is
completed.
Ground Probe: The
ground probe is attached
directly to the banana pin
connector on the probe
cable. This probe is used to complete the circuit to allow flow of current through patient
tissue. The ground probe should touch the patient’s skin at any location away from the
treatment point. As an alternative, you may also use a ground electrode as explained below,
Press and release button
to start and to stop
current delivery during
treatment.
Cut the end off a cotton swab and insert it into
the tip of the cone. Saturate the cotton with
water or an electrolyte spray before treating.
Microcurrent Active Probe (optional accessory)
®) inserted into the ends of both the active
Installation & Features
16
Dynatron® 850plus & 550plus
in place of the ground probe. This is particularly convenient when treating in several
different places around one point.
Ground Electrode: You can use an electrode in place of the ground probe. Just unplug the
ground probe from its cable, attach a banana-to-pin adapter to the cable, then attach an
electrode to the pin. Place the electrode on the patient at a site where it will not interfere with
placement of the active probe during treatment. The banana-to-pin adapter is an optional
accessory available from Dynatronics.
High Volt Probe (optional accessory)
An optional probe may be purchased which
allows you to deliver an attended high volt
treatment. The high volt probe is connected to
the dedicated high volt jack (HV) on the front
of the machine. The probe is used in
conjunction with a large dispersive electrode.
The high volt probe is used only for delivering
high volt therapy.
Complete setup and use instructions are
provided in this manual under the treatment
setup instructions for high volt.
High Volt Probe (optional accessory)
17
Installation & Features
Dynatron® 850plus & 550plus
Quick Reference of Special Key Presses
The following is a brief list of special key presses available with this device. These options are explained in
detail where they apply in the treatment instructions later in this manual. NOTE: Where two keys are
required (ie, FUNCTION-STOP) you must press both keys simultaneously to achieve the result.
KEY PRESS RESULT DESCRIPTION
FUNCTION-STOP
FUNCTION-HI VOLT
FUNCTION-TIME CONDUCTANCE BAR GRAPH
FUNCTION-SOUND
SOUND-TARGET
SOUND-HI/LO keys
FUNCTION DISPLAY WATTS or W/cm2
FUNCTIONTARGET/SWEEP key
FUNCTION-MICRO
MICRO AUDIO TONE ON/OFF
FUNCTION-INTENSITY
UP & DOWN ARROW
KEYS
START SAVE NEW DEFAULT SETTINGS
STOP TREATMENT Stop one treatment
CHANGE HIGH VOLT POLARITY Change high volt polarity. Repeat to change again.
Press during IFC, premod or mcrocurrent treatment to view
conductance bar graph. Press TIME to return to normal display.
HEAD WARMING ON/OFF
ULTRASOUND COUPLING BAR
GRAPH
SOUNDHEAD TEMPERATURE BAR
GRAPH
TEST LEADS
CHANGE MICROCURRENT
POLARITY
CHANGE VOLUME
Set head warming feature to on (HD 1) or off (HD 0)
During an ultrasound treatment only, press to view the ultrasound
coupling bar graph. Press TIME to return to normal display.
During an ultrasound treatment, press to view the ultrasound head
temperature bar graph. Press TIME to return to normal display.
During ultrasound treatment hold for 2 seconds to change
ultrasound power display from w/cm2 to Watts or reverse.
Enter the Lead Test Function. Then press START to begin lead
test, and press STOP to exit the lead test function. NEVER DO THIS WHILE ELECTRODES ARE ATTACHED TO PATIENT.
Change microcurrent polarity. Repeat to change again.
During a microcurrent treatment, turn audio tone on and off.
Change loudness of tone (during microcurrent treatment only)
Hold for 2 seconds to save defaults for current treatment.
The following key presses are applied only when powering up the device:
AT POWER ON, HOLD DOWN THE FOLLOWING KEYS WHILE THE DEVICE STARTS UP:
TARGET/SWEEP LEAD TEST MODE
START RESTORE FACTORY SETTINGS
FUNCTION ENTER CALIBRATION MODE
ULTRASOUND DUTY
CYCLE KEY
FUNCTION-TIME MODE
KEY
(Rev 1.02 and later)
FUNCTION-INTENSITY
DOWN ARROW
(Rev 1.02 and later)
ENTER SOUNDHEAD
PARAMETERS
DISABLE CONDUCTANCE BAR
GRAPH
DISABLE ULTRASOUND COUPLING
DETECTION
Press and hold on power up to enter Lead Test feature.
Press and hold on power up and wait for beep to restore factory
default settings.
Press and hold on power up to enter soundhead calibration mode
(for technicians only)
Press and hold on power up to enter soundhead parameter entry
mode (for technicians only)
Press and hold these keys at the same timeto disable the
conductance bar graph for treatment setup (the graph will still be
available for viewing after a treatment has started). After the
device starts up, select 1 (in the intensity screen) to turn the option
on and 0 to turn the option off.
Press and hold these keys to disable the ultrasound coupling
detection feature. The Time display will show the current setting
briefly: CP1=feature on, CP0=feature off. Repeat this step to
revert to the prior setting.
Installation & Features
18
Dynatron® 850plus & 550plus
Section II
Operation and Treatment
Instructions
19
Operation & Treatment Instructions
Dynatron® 850plus & 550plus
General Operating Instructions
Power On/Off
Press the power switch located on the rear panel of the unit. The self-diagnostic and calibration
function automatically occurs each time the unit is turned on. It is recommended that the unit
remain on throughout the day if it is used on a regular basis.
CAUTION: Never turn power on or off while the unit is connected to the patient.
Set intensity to minimum power before pressing START.
Basic Treatment Setup
Treatment setup is simple and fast with the Dynatron 850plus and 550plus. Each modality
has default settings that are automatically selected when you press a modality key. The basic
steps for setting up any treatment are:
1. Press a modality key once or twice to choose the desired modality (IFC, Premod,
Russian, Biphasic, High Volt, Micro, Sound, or Combo). The lighted LED on the key
indicates the modality chosen. The first available output channel(s) and default settings
for that modality are automatically selected. For ultrasound and combination treatments,
the soundhead must be connected to the device before selecting Sound or Combo.
2. Plug lead wires or probe cable into the appropriate jack(s) on the device for the
modality to be delivered. Observe which channels the device selected for this treatment—channels are selected automatically by the device. Attach the electrodes
to the leads, and affix the electrodes to the patient. For unattended stim treatments, attach
the Remote Stop cable, and give the cable to the patient
3. Change any options desired using the toggle keys available for that modality.
4. Increase power or intensity. Current delivery to the patient starts when you increase
intensity. Only do this when you are ready to proceed with the treatment.
5. • For Target treatments only, locate the treatment site using the Target pad.
• For Russian / biphasic two-channel treatments, press Start or the channel toggle key after
setting the intensity for the first channel, then set the intensity for the second channel.
• For Combo treatments, set the parameters for the stim modality (or ultrasound), press
Start, then set parameters for ultrasound (or stim), then press Start again. You can select
the desired stim modality in this step by pressing the desired modality key after pressing
Combo.
6. Press Start. The treatment timer begins.
Operation & Treatment Instructions
20
7. Administer treatment until the treatment time expires, or press STOP to stop all
treatments in progress at any time.
Each of the therapies you can deliver with the Dynatron 850plus and 550plus offers its own
default settings. To make your treatment setup even more convenient, the Dynatron 850plus
and 550plus allow you to change the defaults for each modality to your own most common
treatment selections. See “Setting Defaults” later in this manual.
The following pages provide more detailed setup instructions for each individual modality.
Before administering a treatment to a patient, be sure to read all treatment instructions
as well as contraindications, warnings, and precautions for that modality provided later
in this manual. Also, consult published literature for recommended application of these
modalities.
Conductance Bar Graph
This feature for models having software greater than Rev. 1.01 only (see Page 2).
During setup of an interferential or premodulated treatment, a conductance graph is
automatically displayed in the Time display while you set the treatment intensity. The graph
allows you to see whether conductance is good for that treatment setup and if current is being
delivered efficiently through the electrodes. The display reverts to the Time display when
you press any other key after setting the intensity. You may again view this conductance
graph at any time during an interferential, premodulated, or microcurrent treatment by
pressing FUNCTION-TIME. The conductance bar graph is not available for combination
treatments.
The bar graph uses the eight segments in the top half of the display, with eight lighted
segments indicating best conductance, and one lighted segment indicating poorest
conductance. In the illustration below, eight of the eight segments are lighted indicating good
conductance.
Eight lighted segments indicate conductance is good. If only
two or three segments are lighted, conductance is poor.
Dynatron® 850plus & 550plus
Conductance Bar Graph
CH
MICRO
3
SOUND
4
HV 1
1
HV 2
2
Conductance Bar Graph.
TIME
START
STOP
PAUSE
TIME
FREQ
RATE
WIDTH
During an IFC, Premodulated, or Microcurrent treatment, press
the FUNCTION and the TIME keys simultaneously to see the
conductance graph. Press TIME to return to the treatment
time display.
The graph is automatically displayed when the intensity is
increased during treatment setup for IFC and Premod.
FUNCTION
POWER-INTENSITY
Operation & Treatment Instructions
21
Dynatron® 850plus & 550plus
If the graph indicates poor conductance, check to be sure leads and electrodes are in good
condition, that leads are fully plugged in and the entire surface of the electrode is making
contact with patient’s skin. Extremely dry skin can also be a cause of poor conductance. If
leads or electrodes are found to be worn, they should be replaced. Leads may be tested using
the Lead Test function described in the operator’s manual for this device. Guidelines for
replacing leads and electrodes are also provided in the operator’s manual.
When you no longer wish to view the conductance graph, press the TIME key to redisplay
the treatment time. NOTE: The treatment time counter does not stop, but continues counting
down when the graph is displayed.
Disable the Conductance Bar Graph. If you do not wish to see the conductance bar graph
automatically during the treatment setup, you can disable this function. Press and hold the
FUNCTION and TIME mode keys while powering the device on (NEVER TURN THE DEVICE ON OR OFF WHILE LEADS ARE CONNECTED TO A PATIENT). When
the device completes starting up, the word “bar” appears in the TIME display, and “1”
appears in the Power/Intensity display. Use the intensity up/down arrow keys to change from
“1” to “0” (select 1 to turn this option on, select 0 to turn the option off). Press START after
making your selection. Once you have disabled the automatic portion of this option, the
device will default to the disabled mode each time the device is powered on. However, you
may still view the conductance graph at any time after the treatment has started as described
above. If you wish to reset the default to re-enable the bar graph to appear automatically
during treatment setup, again press and hold the FUNCTION and TIME mode keys while
powering the device on and select the desired setting.
“LEAD” Warning - No Patient Current
This feature for models having software greater than Rev. 1.01 only (see Page 2).
This added safety feature warns you if the device detects an “open” or incomplete circuit
during setup of an interferential treatment. An open can be due to a lead that is not plugged
in, an electrode that has fallen off or otherwise is not making contact with the patient, or
another reason that would cause an incomplete circuit. When this occurs, current is not being
delivered to the patient, and the condition must be corrected before continuing with treatment
setup.
"LEAD" ERROR - No Patient Current
No Patient Current. During an interferential treatment setup, if an open is detected, the “Lead” warning
is displayed along with the channel number that triggered the error.
When the device detects this condition during the treatment setup, it will beep and display
“LEAD” in the Time display, and will indicate in the Intensity display the number of the
channel that has triggered the error (for example, CH1), and you will be unable to increase
the intensity setting. If this error occurs, reduce the intensity for the channel indicated,
correct the condition, then return the intensity to the desired treatment level. In adjusting
Operation & Treatment Instructions
TIME
22
START
STOP
PAUSE
FUNCTION
POWER-INTENSITY
Loading...
+ 123 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.