you get the UFX flow meter up and running quickly.
Refer to the detailed instructions if you require additional
information.
• The remaining chapters provide a detailed description of
all software settings and hardware installation guidance.
MPORTANTI
Read this manual carefully before attempting any installation or
operation. Keep the manual accessible for future reference.
UNPACKING AND INSPECTION
Upon opening the shipping container, visually inspect the
product and applicable accessories for any physical damage
such as scratches, loose or broken parts, or any other sign of
damage that may have occurred during shipment.
OTE:NIf damage is found, request an inspection by the
carrier’s agent within 48 hours of delivery and file
a claim with the carrier. A claim for equipment
damage in transit is the sole responsibility of
the purchaser.
QUICKSTART
OPERATING OVERVIEW
This manual contains detailed operating instructions for
the UFX flow meter. The following condensed instructions
are provided to assist an experienced operator in basic
operation of the instrument. If the operator is unfamiliar with
this type of instrument, refer to the detailed explanations in
this document.
1. Select a transducer mounting location at least 10 pipe
diameters downstream and 5 diameters upstream of ow
disturbances (such as elbows, tees and valves).
10 dia.5 dia.
Figure 1: Pipe diameters
2. On horizontal pipe, choose a transducer mounting
location approximately 90 degrees from the top of
the pipe.
No
YesYes
SAFETY
Terminology and Symbols
Indicates a hazardous situation, which,
if not avoided, could result in severe
personal injury or death.
Indicates a hazardous situation, which, if
not avoided, is estimated to be capable
of causing minor or moderate personal
injury or damage to property.
No
Figure 2: Transducer location on the pipe
3. Remove rust, scale and paint from the transducer
mounting location. Clean to bare metal. Plastic pipes do
not require preparation.
4. Plug the transducer plug into the transducer jack.
Figure 3: Transducer plug
Page 3 August 2018DPP-UM-01613-EN-03
Introduction and Operating Theory
5. Apply approximately 1/8 inch (3 mm) of silicone grease
to the transducer face. Place the transducer face on the
prepared area of the pipe. Hold the transducer parallel to
the pipe with the cable pointing downstream of the
ow direction.
Flow
Figure 4: Operation
6. Press the ON/OFF key. Wait at least 30 seconds before
recording a reading. The UFX ow meter automatically
turns o after 3 minutes.
7. To change units of measure, press the FT/SEC M/SEC key.
INTRODUCTION AND
OPERATING THEORY
CONTROLS AND TERMINOLOGY
Transducer Cable
Display
Keypad
(On/O key
and units of
measure key)
Transducer Plug
Transducer Jack
The UFX Doppler Flow Meter provides flow velocity readings
in closed piping systems with suspended solids or entrained
gases in liquids. The UFX flow meter uses a non-invasive
transducer, which is handheld or strapped to the outside of
a pipe. The meter is designed to operate on metal, plastic or
rubber pipes containing liquids with greater than 100 ppm of
suspended solids or entrained gases that act as reflectors.
Doppler flow meters use two piezoelectric crystals contained
within one transducer to transmit ultrasonic sound energy
into the fluid stream and receive reflected sound from
reflectors (suspended solids or entrained gases) within the
liquid. See Figure 4. When the liquid is moving and carrying
the reflectors with it, and ultrasonic waves are transmitted
on the moving reflector, the reflected wave’s frequency will
be altered with respect to the transmitted frequency. The
magnitude of frequency change is directly proportional to
the velocity of the reflector.
Standard Pipe (over 1 in.) Transducer
Figure 5: Controls
OTE:NThe UFX flow meter’s battery compartment is on the
back of the enclosure. Remove the two upper screws
to gain access to the battery compartment.
FLOW
Top view of pipe
Figure 6: Small pipe (1/4…1 in.) transducer
Bottom view of transducer
Page 4 August 2018DPP-UM-01613-EN-03
Transducer Location
TRANSDUCER LOCATION
Selecting the proper location for taking a flow measurement
is the single most critical step in the operational procedure.
The transducer used by the UFX flow meter contains
two piezoelectric crystals for transmitting and receiving
ultrasonic signals through the wall of a pipe. Select a
transducer location with adequate straight runs (without
flow disturbances) of pipe, both upstream and downstream,
to achieve stable and accurate readings. Examples of
common piping configurations and the recommended
minimum upstream and downstream pipe lengths are
included in Table 1.
Piping Configuration and
Transducer Position
Flow
Flow
Flow
Upstream
Dimension
Pipe
Diameters
** *
93
143
244
Downstream
Dimension
Pipe
Diameters
MOUNTING LOCATIONS
ON THE PIPE
If the transducer is applied to horizontal pipe, choose a
mounting position at approximately 90 degrees from the
top of the pipe. These positions typically provide optimum
acoustic penetration into the moving liquid. As shown in
Figure 7, placement at the top or bottom of the pipe can
result in poor sound penetration due to air pockets (on the
top of the pipe).
If the transducer is applied to vertical pipe, orientation does
not matter.
No
YesYes
No
Figure 7: Transducer location on the pipe
PIPE PREPARATION
Before the transducer face can be coupled to a pipe
surface, an area slightly larger than the flat surface of the
transducer must be cleaned to bare metal on the pipe.
Remove all scale rust and paint. Thoroughly dry and clean the
mounting surface.
OTE:NFor plastic pipes, such as PVC or PVDF, pipe
preparation is typically not required.
Flow
Flow
Flow
83
83
244
Table 1: Transducer location
CONNECTING THE TRANSDUCER
Connect the transducer plug into the transducer jack.
The connection is polarized, so alignment of the keyway
is necessary.
OTE:NThe splash-resistant environmental seal is provided
only when the transducer plug is secured into
the transducer jack. It is advisable to make this
connection before entering an area where the
UFX flow meter may be dropped into or splashed
with liquid.
APPLYING THE TRANSDUCER
To provide an acoustically conductive path between the
transducer face and the prepared pipe surface, use the
coupling compound (Dow Corning 111 silicone-based
grease) supplied with the UFX flow meter. This grease is
adequate for the majority of installations. If an alternate
grease is used, the grease must be specified not to flow at the
temperature of the pipe surface or the ambient conditions.
Page 5 August 2018DPP-UM-01613-EN-03
Obtaining a Reading
Apply an even layer of grease, approximately 1/8 inch
(3 mm) thick to the flat surface of the transducer. Place
the transducer on the prepared area of the pipe, with the
cable pointing downstream. Align the transducer with the
pipe so it is parallel with the pipe wall. Apply only enough
pressure to hold the transducer in place. If no reading is
shown, perform a “rub test” by rubbing your thumb across
the dry transducer head. The screen should display a low flow
reading. See the “Troubleshooting Guide” on page7 for
further recommendations.
Transducer Cable
Flow
Figure 8: Operation
CHANGING UNITS OF MEASURE
Toggle between FT/S (feet per second) and M/S (meters per
second) by pressing the FT/SEC M/SEC key.
CONVERSION FROM VELOCITY
TO VOLUME
The velocity readings taken from the UFX flow meter can
be readily converted to volumetric flow rate measurements
such as GPM (gallons per minute) or LPM (liters per minute).
See the “Liquid Velocity to Volume Conversion Chart” on
page8 for conversions of many popular schedule 40
pipe sizes. If the pipe size is not located in this chart, use the
following equations:
• For conversion to GPM (pipe I.D. in inches):
GPM = FT/SEC × 2.448 x I.D.
• For conversion to LPM (pipe I.D. in millimeters):
LPM = M/SEC × 0.047 x I.D.
2
2
LOW BATTERY INDICATION
OBTAINING A READING
Press the ON/OFF key on the keypad. On power up, the
UFX flow meter performs internal diagnostics and starts
the ultrasonic transmitter. The display will read Hold for
approximately 5 seconds during this time. If the transducer is
properly mounted to a pipe containing flowing liquid with at
least 100 ppm of 100 micron or larger particles or entrained
gases, the display will begin to indicate velocity. The default
units of measure are FT/S (feet per second). The response
time of the UFX flow meter is approximately 10 seconds to
obtain a reading. Adequate signal strength is obtained when
the OK icon appears in the upper right corner of the display.
See Figure 9. Readings taken without this icon present may
be erroneous. It is recommended that survey readings of at
least 30 seconds be taken to provide reading stability. The
UFX flow meter can be turned off by pressing the ON/OFF
key on the keypad, or the unit will automatically turn off after
approximately 3 minutes of operation.
When the power supply batteries become drained to a level
of 4 volts, the LOW BAT icon will illuminate on the right side
of the display. Typical alkaline batteries will provide about 2
hours of operation after the LOW BAT icon turns on. The flow
meter will no longer operate when battery voltage is less
than 3.5 volts. At 3.5 volts and lower, an Err1 will be displayed
and then unit will shut off.
Replace the batteries by removing the two screws located
on the upper portion of the enclosure back. Install four fresh
AA alkaline batteries following the polarity indicated within
the battery compartment and replace the battery cover.
A fresh set of alkaline batteries will provide approximately
30 hours of service. Use of carbon-based batteries is
not recommended.
DO NOT ALLOW DISCHARGED BATTERIES TO REMAIN IN
THE UFX FLOW METER DURING STORAGE. DISCHARGED
BAT TERIES CAN LEAK AND MAY CAUSE SEVERE DAMAGE
TO THE INTERNAL CIRCUITS OF THE UFX FLOW METER.
DAMAGE CAUSED BY LEAKING BAT TERIES WILL NOT BE
COVERED UNDER THE MANUFACTURER’S WARRANTY.
Figure 9: Reading
Page 6 August 2018DPP-UM-01613-EN-03
DISPLAY TEST
The UFX flow meter contains software to verify operation of
the individual LCD segments. To run the display test:
1. Turn o the UFX ow meter.
2. Press and hold the FT/SEC M/SEC key, then press the
ON/OFF key.
3. Release both keys and verify that all segments illustrated
in Figure 9 illuminate.
The UFX firmware version number (FX.XX) is displayed at the
end of the test.
TROUBLESHOOTING GUIDE
Display Test
Unit does not turn
“ON” when ON/OFF
key is pressed
“Err1” is indicatedThe batteries must be replaced.
No display readings
are obtained and no
“OK” icon is observed
Dashes appear on
display
Readings are
obtained, but the “OK”
icon does not turn on